← Product Code [KNW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/KNW) · K190378

# Spirotome Endo 08 Gauge, Spirotome Endo 10 Gauge, Spirotome Endo 14 Gauge, Spirotome Endo 14 Gauge with 10 mm Helix (K190378)

_Bioncise NV · KNW · May 24, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94diagnostic-devices/KNW/K190378

## Device Facts

- **Applicant:** Bioncise NV
- **Product Code:** [KNW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/KNW.md)
- **Decision Date:** May 24, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Spirotome Endo is a family of large core soft tissue biopsy medical devices to be used in humans to take out adequate samples of tissue. The Spirotome Endo is intended to provide soft tissue for histological and biomolecular detection and confirmation of imaged abnormalities of the internal female reproductive organs such as uterus, ovaries, and retroperitoneal lymph nodes.

## Device Story

Spirotome Endo is a manual, single-use, large-core soft tissue biopsy instrument. It consists of a stainless steel helical receiving needle, a stainless steel cutting cannula, and a polypropylene releasing element. The clinician inserts the receiving needle through the cutting cannula to the biopsy site; rotates the needle clockwise to engage tissue; rotates the cannula clockwise to sever the sample; and retracts the device. The releasing element and counter-clockwise rotation of the needle are used to extract the tissue sample. Used in clinical settings for gynecological procedures, the device provides tissue samples for histological and biomolecular analysis. This allows physicians to confirm imaged abnormalities, facilitating diagnosis of benign or malignant conditions in the uterus, ovaries, or retroperitoneal lymph nodes.

## Clinical Evidence

No new clinical trials were conducted. Substantial equivalence is supported by bench testing (functional, biocompatibility, sterilization, packaging) leveraged from predicate devices (K060384, K080095) and a review of peer-reviewed clinical literature. The literature reports demonstrate the safe and effective use of the Spirotome Endo for biopsy of internal female reproductive organs, confirming an acceptable safety profile without raising new safety or effectiveness concerns.

## Technological Characteristics

Manual core biopsy instrument. Materials: stainless steel (needle/cannula), polycarbonate, polypropylene. Variants: 8G and 14G needles; receiving needle length 375 mm, cutting cannula 325 mm; helical needle length 10-18 mm. Sterilization: steam. Single-use, disposable. Shelf life: 3 years.

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- Spirotome Cervicore ([K080095](/device/K080095.md))

## Reference Devices

- Spirotome SU ([K060384](/device/K060384.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

May 24, 2019

Bioncise NV Jaak Janssens CEO Bodemstraat 16 Wellen, Limburg Belgium 3830

Re: K190378

> Trade/Device Name: Spirotome Endo 08 Gauge, Spirotome Endo 10 Gauge, Spirotome Endo 14 Gauge, Spirotome Endo 14 Gauge with 10 mm Helix Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology-Biopsy Instrument Regulatory Class: Class II Product Code: QHC Dated: February 20, 2019 Received: February 25, 2019

Dear Jaak Janssens:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or (301) 796-7100).

Sincerely,

Sharon Andrews Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K190378

#### Device Name

Spirotome Endo 08 Gauge, Spirotome Endo 14 Gauge, Spirotome Endo 14 Gauge with 10 mm Helix

#### Indications for Use (Describe)

The Spirotome Endo is a family of large core soft tissue biopsy medical devices to be used in humans to take out adequate samples of tissue.

The Spirotome Endo is intended to provide soft tissue for histological and biomolecular detection and confirmation of imaged abnormalities of the internal female reproductive organs such as uterus, ovaries, and retroperitoneal lymph nodes.

| Type of Use (Select one or both, as applicable)                                                     |                                               |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K190378

#### Submitter Information:

Bioncise NV Bodemstraat 16 Wellen, Limburg Belgium 3830

#### Contact:

| Name:          | Jaak Janssens<br>CEO Bioncise NV |
|----------------|----------------------------------|
| Telephone:     | +3211271557                      |
| Fax:           | +3211283677                      |
| E-mail:        | jaak.janssens@bioncise.com       |
| Date prepared: | May 23, 2019                     |

#### Device Identification:

| Trade Name:        | Spirotome Endo 08 Gauge, Spirotome Endo 10 Gauge, Spirotome Endo 14 Gauge, Spirotome Endo 14 Gauge with 10 mm Helix |
|--------------------|---------------------------------------------------------------------------------------------------------------------|
| Common Name:       | Core Biopsy Instrument                                                                                              |
| Regulation Number: | 21 CFR 876.1075                                                                                                     |
| Regulation Name:   | Gastroenterology-Urology Biopsy Instrument                                                                          |
| Product Code:      | QHC                                                                                                                 |
| Regulatory Class:  | Class II                                                                                                            |

### Predicate Device Information

Spirotome Cervicore (K080095)

The predicate device has not been subject to a design-related recall.

### Device Description

The Spirotome Endo is a core biopsy device. It consists of a stainless steel receiving needle with a helical shape, stainless steel cutting cannula, and a polypropylene releasing element. The receiving needle is inserted though the cutting cannula prior to advancing to the biopsy site. The receiving needle is rotated clockwise into the tissue using the plastic hub on the proximal end. The cutting cannula is then rotated clockwise to separate the biopsy sample from the surrounding tissue. Following retraction from the biopsy site, the releasing element is used to remove the tissue sample from the receiving needle. The receiving needle is rotated counter-clockwise to release the tissue. The Spirotome Endo is operated manually.

The Spirotome Endo has four variants: one 8 gauge, and two 14 gauge needles. The receiving needle is 375 mm in length and the cutting cannula is 325 mm long. The helical needle is 18 mm long in all the variants except one 14 gauge variant, where it is 10 mm long. The device is provided sterile and is for single-use only.

{4}------------------------------------------------

### Indications for Use

The Spirotome Endo is a family of large core soft tissue biopsy medical devices to be used in humans to take out adequate samples of tissue.

The Spirotome Endo is intended to provide soft tissue for histological and biomolecular detection and confirmation of imaged abnormalities of the internal female reproductive organs such as uterus, ovaries, and retroperitoneal lymph nodes.

#### Substantial Equivalence Discussion

| Attribute              | Spirotome Endo - Subject Device                                                                                                                                                                                                                                                                                                                                                                                                   | Spirotome Cervicore -<br>Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer           | Bioncise NV                                                                                                                                                                                                                                                                                                                                                                                                                       | Medinvents                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| 510(k) Number          | K190378                                                                                                                                                                                                                                                                                                                                                                                                                           | K080095                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Indications for Use    | The Spirotome Endo is a family of<br>large core soft tissue biopsy medical<br>devices to be used in humans to take<br>out adequate samples of tissue.<br><br>The Spirotome Endo is intended to<br>provide soft tissue for histological<br>and biomolecular detection and<br>confirmation of imaged<br>abnormalities of the internal female<br>reproductive organs such as uterus,<br>ovaries, and retroperitoneal lymph<br>nodes. | The Cervicore system is a family of<br>core soft tissue biopsy medical<br>devices to be used in humans to take<br>out adequate samples of tissue from<br>the human cervix.<br><br>The Cervicore system is intended to<br>provide soft tissue for histological<br>and biomolecular detection and<br>confirmation of imaged<br>abnormalities by inspection or<br>colposcopy.<br><br>The Cervicore system is intended to<br>provide soft tissue for histological<br>and biomolecular detection and<br>confirmation of palpable<br>abnormalities.<br><br>An early detection and correct<br>diagnosis is a prerequisite for a<br>successful multimodality treatment<br>of malignant disease. Therefore, an<br>adequate sampling of human tissue<br>is necessary to make a reliable<br>diagnosis of benignancy or<br>malignancy and to characterize as<br>complete as possible the nature of<br>the lesion without unnecessary<br>discomfort to the patient.<br><br>In case of malignancy, a proper<br>multimodality treatment is necessary<br>i.e. surgery and/or<br>chemotherapy and/or hormone<br>therapy and/or biological therapy. In<br>case of benignancy, adequate follow-<br>up of the patient can be provided. |
| Biopsy Tissue<br>Types | Uterine, ovarian, and retroperitoneal<br>lymph nodes                                                                                                                                                                                                                                                                                                                                                                              | Cervical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

#### Table 1: Substantial Equivalence Comparison

Bioncise NV – Bodemstraat 16 – 3830 Wellen – Belgium www.bioncise.com – info@bioncise.com tel + 32 11 271557 - fax + 32 11 283677

{5}------------------------------------------------

| Device<br>Components     | Cutting Cannula                                  | Cutting Cannula                                  |
|--------------------------|--------------------------------------------------|--------------------------------------------------|
|                          | Receiving Needle                                 | Receiving Needle                                 |
|                          | Releasing Element                                | Releasing Element                                |
| Helical Needle<br>Length | 10, 18 mm                                        | 10, 20 mm                                        |
| Component<br>Length      | Receiving Needle: 375 mm                         | Receiving Needle: 384.5 mm                       |
|                          | Cutting Cannula: 325 mm                          | Cutting Cannula: 350 mm                          |
| Device Materials         | Stainless steel, polycarbonate,<br>polypropylene | Stainless steel, polycarbonate,<br>polypropylene |
| Sterilization            | Steam Sterilized                                 | Steam Sterilized                                 |
| Number of Uses           | Single-use; disposable                           | Single-use; disposable                           |
| Shelf Life               | Three years                                      | Three years                                      |

The Spirotome Endo and the predicate device (K080095) do not have identical Indications for Use statements. The indication for the subject device includes different biopsy tissue types and does not include statements referring to diagnosis or treatment. This change does not represent a new intended use as the intended use of this device is the same as the predicate – to provide soft tissue biopsy samples.

## Non-Clinical Performance Testing

## Bench Testing

Bench testing of the subject device, Spirotome Endo, including functional testing, biocompatibility, sterilization and package testing was leveraged from testing conducted on the Spirotome SU (K060384) and Spirotome Cervicore (K080095). The subject device has identical packaging materials, materials of construction, manufacturing, and sterilization processes as the cleared devices.

### Literature

As part of demonstrating substantial equivalence to the predicate, literature was provided demonstrating safe and effective use of the subject device for internal female reproductive organs. Clinical use of the Spirotome Endo for gynecological tissues was reported in peer reviewed journals. The reports in clinical literature supported an acceptable safety profile for the subject device during biopsy procedures. The reports did not raise additional concerns regarding the safety and effectiveness of the subject device.

### Conclusion

The results of the information and testing described above demonstrate that the Spirotome Endo is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94diagnostic-devices/KNW/K190378](https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94diagnostic-devices/KNW/K190378)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
