The MultiCut Solo Morcellator is intended for use under endoscopic visualization for the transurethral mechanical morcellation and removal of the adenoma after enucleation of the prostate during endoscopic surgical procedures in urology.
Device Story
MultiCut Solo is an endoscopic morcellator for prostate adenoma removal. Device consists of a handpiece with an integrated DC motor driving a stainless steel blade; motor controlled by a PCB. Blade performs mechanical shredding via rotation and oscillation. Aspiration tube set connects handpiece to a peristaltic pump to transport crushed tissue to a waste container. Used in hospital/healthcare facilities by physicians. Output is the physical removal of tissue; healthcare provider monitors progress via endoscopic visualization. Benefits include efficient tissue removal during urological procedures.
Clinical Evidence
Bench testing only. Comparative bench testing performed to demonstrate performance equivalence to predicates. Biocompatibility testing conducted per ISO 10993 standards (cytotoxicity, systemic toxicity, irritation, sensitization, chemical characterization). Software verification and validation performed per IEC 62304.
Technological Characteristics
Handpiece with DC motor and PCB-based motor driver. Patient-contacting components: stainless steel. Dimensions: 35.0 or 40.0 cm length, 0.47 cm (15 Fr) diameter. Power: 100-240 V, 50/60 Hz, max 90 VA. Connectivity: peristaltic aspiration pump. Sterilization: steam sterilization (reusable). Standards: EN 60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 14971, ISO 10993.
Indications for Use
Indicated for transurethral mechanical morcellation and removal of prostate adenoma following enucleation in urological endoscopic procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA logo, which includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 18, 2023
Asclepion Laser Technologies GmbH Tom Gruender Regulatory Affairs Manager Bruesseler Strasse 10 Jena. Thuringia 07747 Germany
Re: K213597 Trade/Device Name: MultiCut Solo Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: GCJ Dated: December 14, 2022 Received: December 19, 2022
Dear Tom Gruender:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark J. Antonino -S
Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213597
Device Name MultiCut Solo
Indications for Use (Describe)
The MultiCut Solo Morcellator is intended for use under endoscopic visualization for the transurethral mechanical morcellation and removal of the adenoma after enucleation of the prostate during endoscopic surgical procedures in urology.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(K) SUMMARY
| Applicant /<br>Manufacturer<br>Name and Address: | Asclepion Laser Technologies GmbH<br>Bruesseler Strasse 10<br>07745 Jena<br>Germany |
|--------------------------------------------------|-------------------------------------------------------------------------------------------|
| 510(k) Contact Person: | Tom Gruender<br>Regulatory Affairs Manager<br>Asclepion Laser Technologies |
| | Email: tom.gruender@asclepion.com<br>Phone: +49-03641-7700 201<br>Fax: +49-03641-7700 102 |
| Date Prepared: | 13th January 2023 |
| Common Name: | Soft Tissue Morcellator and Accessories |
| Device Name: | MultiCut Solo |
| Regulatory Class: | Class II |
| Regulation Name: | Endoscope and Accessories |
| Regulation Number: | 21 CFR 876.1500 |
| Product Code: | GCJ |
| Basis for Submission: | new submission |
| Predicate Devices: | VersaCut + Morcellator (K133272)<br>MORCE SCOPE SET 8970 (K041610) |
# Performance Standards:
There are no mandatory performance standards for this device.
{4}------------------------------------------------
#### Description of the device:
The MultiCut Solo consists of a special endoscopic handpiece in which a blade is inserted which mechanically shreds the tissue. The blade is driven by a dc motor, which is integrated in the handpiece. The motor itself is driven by a motor driver PCB. The blade can rotate at different speeds and thereby change the direction of rotation periodically (so called oscillation). The morcellator handpiece is also connected by an aspiration tube set to a peristaltic aspiration pump. By this pump the crushed tissue can be transported through a hollow channel in the handpiece to a waste container.
#### Other characteristics:
The device includes a controlling software.
The device has patient contacting material made of stainless steel which has short term contact (limited contact) with body tissue (External Communicating Device).
The patient contacting components are supplied non-sterile and are intended to be cleaned, disinfected and steam sterilized before the first use and after any following re-use.
The device is intended to solely be used in healthcare facility/hospital environment.
#### Indications for Use
The MultiCut Solo Morcellator is intended for use under endoscopic visualization for the transurethral mechanical morcellation and removal of the adenoma after enucleation of the prostate during endoscopic surgical procedures in urology.
#### Nonclinical Performance Data:
The following performance data were applied in support of the substantial equivalence determination:
- . EN 60601-1: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests
- IEC 62304: Medical Device Software - Software life cycle processes
- . ISO 14971: Medical devices - Applications of risk management to medical devices
- . IEC 60601-2-18 Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
- . Comparative bench testing has been performed to show the performance equivalence of the subject device to its predicate
Biocompatibility testing was also conducted for the MultiCut Blades in accordance with:
- EN/ISO 10993-1: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ISO 10993-5: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-18: Biological evaluation of medical devices - Part 18: Chemical characterization of medical device materials within a risk management process
- ISO 10993-11 Biological evaluation of medical devices Part 11: Tests for systemic toxicity ●
- ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
{5}------------------------------------------------
Software verification and validation testing was conducted, and documentation is provided in this submission, as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
The MultiCut Solo System passed all of the required testing and is in compliance with all applicable sections of the above-mentioned performance standards.
| | Primary Predicate<br>Device | Secondary Predicate<br>Device | Subject Device |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model Name | VersaCut + Morcellator | MORCE SCOPE SET 8970 | MultiCut Solo |
| Manufacturer | Lumenis LTD | RICHARD WOLF MEDICAL<br>INSTRUMENTS CORP. | Asclepion Laser<br>Technologies GmbH |
| 510(k) | K133272 | K041610 | - |
| Principal of<br>Operation | Morcellation through<br>the cutting effect of<br>the cutting window,<br>produced by the<br>relative movement of<br>the internal tube and<br>the external tube. | Morcellation through the cutting<br>effect of the cutting window,<br>produced by the relative movement of<br>the internal tube and the external<br>tube. | Morcellation through<br>the cutting effect of the<br>cutting window,<br>produced by the relative<br>movement of the<br>internal tube and the<br>external tube. |
| Intended Use | The VersaCut + Tissue<br>Morcellator is<br>intended for use under<br>direct or endoscopic<br>visualization for the<br>morcellation and<br>removal<br>of dissected tissue<br>during<br>pelviscopic,<br>laparascopic,<br>percutaneous and<br>open surgical<br>procedures<br>whenever access to<br>the surgical site is<br>limited. | Morce Scope Set 8970, in conjunction<br>with a morcellation probe, and with its<br>sheaths and obturators, is used in the<br>cutting (morcellation) and continuous<br>removal of large tissue masses. In<br>combination with the corresponding<br>auxiliary instruments it can be used as<br>a nephroscope in the disintegration<br>and removal/aspiration of kidney and<br>bladder stones and the removal of<br>tumors via percutaneous (kidney) or<br>transurethral (bladder) passages, in<br>conjunction with intracorporeal 5-<br>4ithotripters e.g. operated<br>pneumatically, by ultrasound,<br>electrohydraulically or by laser, under<br>endoscopic control. The POWER<br>CONTROL 2303 in conjunction with<br>POWER STICK M4 serves to drive<br>WOLF morcellators for the continuous<br>removal of ablated tissue in | The MultiCut Solo<br>Morcellator is intended<br>for use under<br>endoscopic visualization<br>for the transurethral<br>mechanical morcellation<br>and removal of the<br>adenoma after<br>enucleation of the<br>prostate during<br>endoscopic surgical<br>procedures in urology. |
| | Primary Predicate<br>Device | Secondary Predicate<br>Device | Subject Device |
| | | PUMP is used for aspirating irrigation<br>fluid in conjunction with a<br>resectoscope or a morcellator<br>following laser TURP. | |
| Speed<br>Regulation | Possible from the<br>vacuum<br>generator | Possible from the vacuum<br>generator | Possible from the<br>vacuum<br>generator |
| External Tube<br>Dimension | 39.5 cm length<br>0.47 cm external<br>diameter | 35 or 38 cm length<br>0.475 cm external diameter | 35.0 or 40.0 cm length<br>0.47 cm (15 Fr) external<br>diameter |
| Material in<br>Contact with<br>Tissue | Stainless steel | Stainless steel | Stainless steel |
| Power Supply | AC Power Supplied | AC Power Supplied | AC Power Supplied |
| Rotation Speed | Up to 600 rpm | Up to 6000 rpm | Up to 3000 rpm |
| Aspiration | Max 1.65 l/min | Max. 1,3 l/min | Max 1.1 l/min |
| Electrical<br>Requirements | 100-240 V , 50/60 Hz ,<br>max 200 VA | 100-240 V , 50/60 Hz , max 120 VA | ~100-240 V, 50/60 Hz,<br>max. 90 VA |
| Suction Pump | Part of System Console | Part of System Console | Part of System Console |
| Supplied Sterile | NO | NO | NO |
| Re-Usable | YES | Provided Single or PluriUse | YES |
#### Summary of the technological characteristics of the new device in comparison to the predicate
{6}------------------------------------------------
## Comparison with predicate device:
The subject and predicate devices have similar intended use and the same fundamental principles of surgical cutting and aspiration of dissected tissue. Any minor difference does not raise concern about safety and effectiveness.
## Conclusions
The non-clinical performance testing conducted supports that the device can be used safely and effectively. The differences in the indications for use and technological characteristics between the subject and predicate device do not raise new types of questions regarding safety and effectiveness, and the subject device is as safe, as effective, and performs as well as the legally marketed predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.