Alexis Contained Extraction System

K211043 · Applied Medical Resources Corporation · GCJ · Mar 31, 2022 · Gastroenterology, Urology

Device Facts

Record IDK211043
Device NameAlexis Contained Extraction System
ApplicantApplied Medical Resources Corporation
Product CodeGCJ · Gastroenterology, Urology
Decision DateMar 31, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

The Alexis® Contained Extraction System is indicated to contain and isolate tissue during, or prior to, surgical removal and/or extracorporeal manual morcellation via an abdominal or vaginal approach.

Device Story

Sterile, single-use tissue containment system; consists of polyurethane tissue bag with flexible ring, tabs, strap with snap fastener, and tether; includes guard system with coiled polyethylene guard and self-retaining polyurethane retractor. Used by surgeons in OR; inserted via abdominal incision or colpotomy. Bag collapsed for insertion; deployed in cavity to receive specimen. Guard system anchors at extraction site; provides barrier against sharp instruments. Surgeon maneuvers ring to surface using tether/tabs; ring can be flipped to shorten bag for morcellation. System maintains closed environment during extracorporeal manual morcellation. Benefits include tissue isolation and protection of extraction site from potential contamination during specimen removal.

Clinical Evidence

Bench testing only. Evaluated biocompatibility (ISO 10993-1), sterilization validation (ISO 11135-2), shelf-life, and functional performance. Functional tests included bag/tether integrity, guard system protection/anchoring, and pathogen penetration resistance. Simulated use testing performed in abdominal and vaginal bench models confirmed device performance for intended anatomical sites.

Technological Characteristics

Materials: Polyurethane (bag, retractor), polyethylene (guard). Components: Flexible ring, tether, tabs, snap fastener. Energy: None (manual). Connectivity: None. Sterilization: Ethylene Oxide (EO). Form factor: Two models (14cm/3400mL and 17cm/6500mL).

Indications for Use

Indicated for patients undergoing surgical tissue removal or extracorporeal manual morcellation via abdominal or vaginal approach; used to contain and isolate tissue.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 31, 2022 Applied Medical Resources Corporation Apeksha Shanbhag Regulatory Affairs Specialist 22872 Avenida Empresa Rancho Santa Margarita, CA 92688 Re: K211043 > Trade/Device Name: Alexis® Contained Extraction System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: March 3, 2022 Received: March 4, 2022 Dear Apeksha Shanbhag: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K211043 #### Device Name Alexis® Contained Extraction System Indications for Use (Describe) The Alexis® Contained Extraction System is indicated to contain and isolate tissue during, or prior to, surgical removal and/or extracorporeal manual morcellation via an abdominal or vaginal approach. Type of Use (Select one or both, as applicable) | <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |-----------------------------------------------------------------------------| | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary - K211043 | 510(k) Submitter: | Applied Medical Resources Corporation<br>22872 Avenida Empresa<br>Rancho Santa Margarita, CA 92688<br>(949) 713 - 8000 | |---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Apeksha Shanbhag<br>Regulatory Affairs Manager<br>Applied Medical Resources Corporation<br>Tel: (949) 547-4136<br>Fax: (949) 713-8205<br>Email: apeksha.shanbhag@appliedmedical.com | | Preparation Date: | March 30, 2022 | | Trade Name: | Alexis® Contained Extraction System | | Common Name: | Tissue Bag | | Classification: | Regulation: 21 CFR 876.1500, Endoscope and Accessories<br>Device Class: Class II<br>Product Code: GCJ (endoscope and accessories) | | Predicate Device: | Applied Medical Tissue Containment System<br>510(k)#: K142427<br>Product Code: GCJ<br>The predicate device has not been subject to a design related recall. | | Device Description: | The Alexis® Contained Extraction System is a sterile, single use tissue<br>bag with a guard system component. The polyurethane tissue bag consists<br>of a flexible ring, tabs, strap with a snap fastener, and an attached tether<br>It is available in two models GTB14 and GTB17 with the following<br>dimensions and volumes: | | Model | Ring<br>Diameter | Volume | |-------|------------------|----------| | GTB14 | 14 cm | 3,400 mL | | GTB17 | 17 cm | 6,500 mL | The tissue bag is used to contain and isolate specimens for surgical removal and/or manual morcellation. The strap with snap fastener maintains the folded bag and flexible ring in a collapsed state during insertion. After the device is fully inserted and deployed into the abdominal or pelvic cavity, the opening of the bag returns to its original, circular shape, facilitating placement of the specimen in the bag. When {4}------------------------------------------------ the specimen is ready for removal and/or manual morcellation, the tether, tabs, and strap are used to maneuver the ring to the surface of the extraction site. If the specimen requires manual morcellation. the ring may be repeatedly flipped to shorten the bag and consequently bring the specimen closer to the extraction site. The guard system consists of a guard and a self-retaining retractor. The guard is composed of a coiled polyethylene material that conforms to the extraction site. The self-retaining retractor consists of two polyurethane rings connected by a polyurethane sheath and a tether attached to the inner ring. It helps anchor the guard at the extraction site. The guard system provides a robust barrier between the bag and sharp instruments. The Alexis® Contained Extraction System can be inserted and retrieved either through an abdominal incision or through the vaginal canal and colpotomy. In both cases, manual morcellation is performed extracorporeally at the extraction site. - Indications for use: The Alexis® Contained Extraction System is indicated to contain and isolate tissue during, or prior to, surgical removal and/or extracorporeal manual morcellation via an abdominal or vaginal approach. #### Comparison with the Predicate Device: The subject and predicate devices are both single use tissue bags intended to retrieve and contain specimen during extracorporeal manual morcellation. | | Applied Medical Tissue<br>Containment System<br>(Predicate Device) | Alexis Contained Extraction<br>System<br>(Subject Device) | |------------------------|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended<br>use | To contain and isolate tissue during, or prior to, surgical removal and/or extracorporeal manual morcellation | Same | | Indications<br>for use | Used to contain and isolate tissue during, or prior to, surgical removal and/or extracorporeal manual morcellation. | Used to contain and isolate tissue during, or prior to, surgical removal and/or extracorporeal manual morcellation via an abdominal or vaginal approach. | | Description | Consists of a tissue bag and a guard component. | Consists of a tissue bag and a guard system component. The bag includes tabs and a strap with a snap fastener. | | Size | Min: 10 cm diameter, 2,500 mL volume<br>Max: 25 cm diameter, 20,000 mL | Min: 14 cm diameter, 3,400 mL volume<br>Max: 17 cm diameter, 6,500 mL | The following table provides a comparison with the predicate: {5}------------------------------------------------ | | volume | volume | |-------------------------|-----------------------------------------------------------------|------------------------------------------------------------------------------| | Anatomical<br>Site | Bag is inserted and retrieved through<br>an abdominal incision. | Bag is inserted and retrieved through<br>an abdominal incision or colpotomy. | | Packaging<br>material | Mylar/Tyvek Pouch | Nylon/Tyvek Pouch | | Sterilization<br>method | Radiation sterilization | Ethylene Oxide (EO) sterilization | These differences in technological characteristics do not raise different questions of safety or effectiveness. ## Discussion of Performance Data The following performance data is provided in support of the substantial equivalence determination. ## Biocompatibility Biocompatibility evaluation of the subject device was conducted in accordance with ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and 2020 FDA guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. " The subject device contacts tissue for less than 24 hours, so the following endpoints were considered, and all materials were found to be biocompatible: - Cytotoxicity ● - Intracutaneous Irritation - Sensitization - Acute Systemic Toxicity - Material Mediated Pyrogenicity ## Sterilization/Shelf Life Sterilization validation was performed in accordance with ISO 11135 - 2, Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices. Shelf life testing was performed to evaluate package integrity and device functionality following artificial aging and simulated transit conditioning. ## Functional Performance Design Verification testing was performed with the subject device. The following performance characteristics were evaluated: - Integrity of tissue bag ● - Integrity of tether to pull the tissue bag out of the extraction site - Integrity of the guard system's material to protect the bag from incidental cuts ● 510(k) Summary {6}------------------------------------------------ - Ability of the tissue bag to maintain a closed system ● - Resistance of tissue bag material to penetration by blood-borne pathogens - Ability of guard system to remain anchored at the extraction site - Integrity of tabs and strap to assist in specimen capture - Ability of snap fastener to withstand insertion and ability of the user to deploy for specimen o capture Additionally, simulated use testing was performed with the subject device in both an abdominal bench model and a vaginal bench model to ensure that the device would perform as intended. This testing demonstrated the subject device's capability to perform the intended use at the different anatomical sites of insertion and retrieval. #### Conclusion: Results of testing demonstrate that the subject Alexis® Contained Extraction System is substantially equivalent to the predicate Applied Medical Tissue Containment System, and that the subject device is as safe and effective as the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...