K201526 · Karl Storz Endoscopy America · GCJ · Nov 20, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K201526
Device Name
TIPCAM1 Rubina Video Endoscope System
Applicant
Karl Storz Endoscopy America
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Nov 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Intended Use
The TIPCAM®1 Rubina Video Endoscope System is intended to be used together with the camera control unit during diagnostic and/or surgical procedures when endoscopic video assistance is required. For use in all endoscopy and endoscopic surgery within the peritoneal and thoracic cavity, including gynecological anatomy.
Device Story
Integrated video endoscope system; includes camera and endoscope. Illumination light from external source travels via optical fiber bundle to distal end; objective lens receives reflected light from patient lumen. Dual CMOS sensors convert light to electrical signals; transmitted to camera control unit (Image1 S Connect II + 4U-Link) for processing and display. Used in OR by physicians for minimally invasive surgery. Features 2D auto-leveling (autorotation) to maintain fixed horizon and 2D auto-switch mode to transition from 3D to 2D when rotated outside 3D range. Provides real-time visualization of surgical field; assists surgeon in navigation and procedure performance; benefits patient through minimally invasive access.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical/mechanical/thermal safety (IEC 60601-1, 60601-2-18), EMC (IEC 60601-1-2), photobiological safety (IEC 62471), and performance verification (color reproduction, distortion, depth of field, SNR, latency, spatial resolution, FOV per ISO 8600-3). Biocompatibility and sterilization validation leveraged from predicate K150525.
Technological Characteristics
Integrated video endoscope with dual CMOS sensors. Features 0° and 30° direction-of-view models. Connectivity via cable to camera control unit. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62471, ISO 8600-3. Software-controlled image processing for auto-leveling and 3D-to-2D switching.
Indications for Use
Indicated for use in diagnostic and/or surgical endoscopic procedures within the peritoneal and thoracic cavity, including gynecological and urological anatomy, requiring endoscopic video assistance.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K220895 — 1688 4K Camera Control Unit with Advanced Imaging Modality (1688010000), 1688 4K Pendulum Camera Head (1688310130) · Stryker · Apr 22, 2022
K063457 — TRUEHD (1080P) DIGITAL CAMERA SYSTEM · Conmed Linvatec · Dec 5, 2006
K152619 — Neoscope 3D - 3D Digital Video Endoscopic System (Rigid/Flexible) · Prosurg, Inc. · Apr 29, 2016
K150525 — SPIES 3D Tip Cam, Image 1 SPIES CCU Image Connect, Image 1 SPIES CCU D-3 Link · KARL STORZ Endoscopy-America, Inc. · Nov 20, 2015
K970369 — VISTA MINI CAMERA SYSTEM · Oktas · Apr 14, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 20, 2020
Karl Storz Endoscopy-America, Inc. Mario Trujillo Associate Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, CA 90245
K201526 Re:
> Trade/Device Name: TIPCAM®1 Rubina Video Endoscope System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: GCJ, FGB, HET Dated: October 21, 2020 Received: October 23, 2020
Dear Mario Trujillo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201526
Device Name TIPCAM®1 Rubina Video Endoscope System
#### Indications for Use (Describe)
The TIPCAM®1 Rubina Video Endoscope System is intended to be used together with the camera control unit during diagnostic and/or surgical procedures when endoscopic video assistance is required. For use in all endoscopy and endoscopic surgery within the peritoneal and thoracic cavity, including gynecological anatomy.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">✘</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">❍</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Karl Storz Endoskope. The word "STORZ" is in large, bold, blue letters. Below that, the words "KARL STORZ - ENDOSKOPE" are in smaller, blue letters.
K201526
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KARL STORZ Endoscopy-America's knowledge.
| Applicant: | KARL STORZ Endoscopy-America, Inc.<br>2151 E. Grand Avenue<br>El Segundo, CA 90245 |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Mario Trujillo<br>Associate Regulatory Affairs Specialist<br>Tel.: (424) 218-8481<br>Email: Mario.Trujillo@karlstorz.com |
| Date of<br>Preparation: | November 19, 2020 |
| Type of 510(k)<br>Submission: | Traditional |
| Device<br>Identification: | Trade Name: TIPCAM®1 Rubina Video Endoscope System<br>Regulation Name: Endoscope and Accessories |
| Regulatory Class: | 2 |
| Product Code: | GCJ, FGB, HET |
| Regulation<br>Number: | 21 CFR 876.1500 |
| Predicate<br>Device(s): | SPIES 3D System (K150525)<br>This predicate device has not been subject to a design-related recall. |
| Device<br>Description: | This predicate device has not been subject to a design-related recall.<br>The TIPCAM®1 Rubina video endoscope is an integrated unit that<br>includes a camera and an endoscope. The endoscope receives the<br>illumination light from the light source by the light guide connector<br>connected to the light source device. The illumination light is<br>transferred to the distal end through the optical fiber bundle inside of<br>the endoscope and illuminates the inside of the patient body through<br>the illumination lens at the distal end. The endoscope receives the<br>reflected light from the inner lumen of a patient by the objective lens at<br>the distal end. The built-in dual CMOS sensors convert the light to an<br>electrical signal, and the signal is sent to the camera control unit<br>(Image1 S Connect II + 4U-Link) via the attached cable for processing<br>and display. |
| Intended Use: | TIPCAM®1 Rubina videoscopes and corresponding accessories<br>(camera control unit, light sources, monitor) are used for endoscopic<br>imaging of the surgical field during diagnostic and/or surgical<br>procedures in the abdominal and thoracic cavity.<br>TIPCAM®1 Rubina videoscopes are designed for short-term use in<br>surgically invasive procedures |
| Indications For<br>Use: | The TIPCAM®1 Rubina Video Endoscope System is intended to be<br>used together with the camera control unit during diagnostic and/or<br>surgical procedures when endoscopic video assistance is required.<br>For use in all endoscopy and endoscopic surgery within the peritoneal<br>and thoracic cavity, including gynecological and urological anatomy. |
| Technological<br>Characteristics: | The predicate and subject devices are both camera systems that are<br>used for observation purposes in general endoscopic surgery within<br>the thoracic and peritoneal cavity, including gynecological and<br>urological anatomy. There are some minor differences in the<br>technological characteristics. These differences are:<br>The subject device has two image viewing features: A 2D<br>auto-leveling (autorotation) mode which allows the user to<br>maintain fixed horizon while rotating the endoscope and a 2D<br>Auto-switch mode that automatically switches the image from<br>3D mode to 2D Auto-leveling mode when the scope is rotated<br>outside the 3D viewing range to prevent the user from losing<br>the perspective of the image. The subject device is also available in a 30° direction of view<br>model in addition to the 0° model. The total length of the subject endoscope is 5mm longer<br>than the predicate These differences do not raise different questions of safety and<br>effectiveness and can be evaluated through performance testing. |
| Non-Clinical<br>Performance Data: | The TIPCAM®1 Rubina System was tested for<br>electrical/mechanical/thermal safety and EMC according to the<br>following standards:<br>IEC 60601-1 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1<br>(2012) IEC 60601-1-2:2014 (4th Edition) IEC 60601-2-18:2009 IEC 62471:2009 The following additional bench testing for performance verification<br>and validation purposes was performed: Photobiological Safety per IEC 62471:2006 Color Reproduction and Color Contrast Enhancement Distortion Depth of Field Detection Linearity Illumination Detection Uniformity Signal-to-Noise Ratio (SNR) & Sensitivity Latency Spatial Resolution Dynamic Range Direction of View per ISO 8600-3:2019 |
| | • Field of View per ISO 8600-3:2019<br>• System Horizontal Parallax<br>The subject device software was validated per the FDA guidance<br>"Guidance for the Content of Premarket Submissions for Software<br>Contained in Medical Devices" as a moderate level of concern. |
| | The subject device materials and manufacturing are identical to that of<br>the predicate, therefore, biocompatibility testing was leveraged from<br>K150525 to support the biocompatibility of the subject device. |
| | Cleaning and sterilization validation of the subject device was<br>leveraged from the predicate device K150525, as the subject device is<br>substantially similar in specifications to that of the predicate, and the<br>sterilization process remains identical to the predicate. |
| | The bench testing noted above confirmed that the TIPCAM®1 Rubina<br>system has met all its design specification and is as safe and effective<br>as the predicate device, SPIES 3D System. |
| Clinical<br>Performance Data: | Clinical testing was not required to demonstrate the substantial<br>equivalence to the predicate device. Non-clinical bench testing was<br>sufficient to assess safety and effectiveness and to establish the<br>substantial equivalence of the modifications. |
| Conclusion: | The subject device has the same intended use as the predicate. The<br>technological differences between the subject and predicate do not<br>raise different questions of safety and effectiveness. The conclusions<br>drawn from the non-clinical performance data demonstrated that the<br>subject device is as safe as and as effective as the predicate device.<br>Therefore, the subject device is substantially equivalent to the<br>predicate. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in a smaller font.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for KARL STORZ. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.