K241595 · Veol Medical Technologies Pvt , Ltd. · FBK · Feb 27, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K241595
Device Name
EzVu Visual Vasopressor injector (EV-19)
Applicant
Veol Medical Technologies Pvt , Ltd.
Product Code
FBK · Gastroenterology, Urology
Decision Date
Feb 27, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
EzVu Visual Vasopressor Injector (EV-19) is intended to inject vasopressor solution into the uterine musculature during laparoscopic surgery.
Device Story
Laparoscopic instrument (330 mm length, 5 mm diameter) used by gynecological surgeons during laparoscopic myomectomy. Device features distal stainless steel needle, transparent plastic window, and proximal luer-connected stopcock. Input: manual aspiration of tissue fluid/blood via needle. Operation: surgeon inserts device through trocar; aspirates to confirm needle placement via flashback visualization in transparent window; injects vasopressor solution into uterine musculature. Output: visual confirmation of aspiration and delivery of therapeutic agent. Benefits: enables precise drug delivery to uterine tissue, reducing potential for systemic absorption or incorrect placement. Used in clinical OR settings.
Clinical Evidence
Multi-center, prospective observational study of 13 women (ages 30-40, BMI 18-40) undergoing laparoscopic myomectomy. 24 injections performed by 5 surgeons. Primary endpoints: clear visibility of blood stain in transparent window upon aspiration; no device leakage or breakage. Results: 100% success in structural integrity; blood aspirate detected in 4/25 cases; no adverse events reported. Usability ratings averaged ≥4.6/5.0.
Technological Characteristics
Materials: stainless steel needle/tube, transparent plastic shaft, heat shrink. Dimensions: 330 mm length, 5 mm diameter, 19G needle. Principle: manual injection/aspiration. Connectivity: none. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 10993-1 (biocompatibility), ISO 11135 (sterilization), ISO 7864/9626 (needle specs), ASTM F1929/F88 (packaging).
Indications for Use
Indicated for injection of vasopressor solution into the uterine musculature during laparoscopic surgery in adult women.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 27, 2025
Veol Medical Technologies Pvt Ltd. % Tatiana Jabor Botura Regulatory Affairs Consultant Passarini Edifício Pórtico Sul - R. José Jaime Delibo, 160 - Nova Alia Ribeirão Preto, SP 14026-563 BRAZIL
Re: K241595
Trade/Device Name: EzVu Visual Vasopressor Injector (EV-19) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FBK Dated: January 24, 2025 Received: January 28, 2025
Dear Tatiana Jabor Botura:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Reginald K. Avery -S
for Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241595
Device Name
EzVu Visual Vasopressor Injector (EV-19)
Indications for Use (Describe)
EzVu Visual Vasopressor Injector (EV-19) is intended to inject vasopressor solution into the uterine musculature during laparoscopic surgery.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
{4}------------------------------------------------
# EzVu Visual Vasopressor Injector (EV-19)
### 510(k) Summary
### ADMINISTRATIVE INFORMATION
| Sponsor | Veol Medical Technologies PVT. LTD.<br>Plot - A 747, Near Pavan Bus Stop,<br>Pawane MIDC, TTC Industrial Estate,<br>Koparkhairane,<br>Navi Mumbai<br>Mumbai<br>Maharashtra 400705 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person and Preparer | Tatiana Jabor Botura<br>Regulatory Affairs Specialist<br>Passarini Regulatory Affairs<br>e-mail: tatiana@passarini.com.br<br>Telephone: +55 (16) 3421 8488 |
| Data Prepared | February 26, 2025 |
### DEVICE NAME AND CLASSIFICATION
| Trade Name | EzVu Visual Vasopressor Injector (EV-19) |
|----------------------|-------------------------------------------------------------|
| Common Name | Endoscope and accessories |
| Regulation Number | 21 CFR 876.1500 |
| Regulatory Class | II |
| Product Code | FBK (endoscopic injection needle, gastroenterology-urology) |
| Classification Panel | Gastroenterology/Urology |
### PREDICATE DEVICE INFORMATION
K220292, Disposable Endoscope Injection Needle by Yangzhou Fartley Medical Instrument Technology Co., Ltd has been identified as the predicate device for the 510(k) submission for the EzVu Visual Vasopressor Injector (EV-19) by Veol Medical Technologies Pvt Ltd to support the substantial equivalence comparison.
{5}------------------------------------------------
The predicate device has not been subject to any design-related recall.
# SUBJECT DEVICE INFORMATION
### Indications For Use
EzVu Visual Vasopressor Injector (EV-19) is intended to inject vasopressor solution into the uterine musculature during laparoscopic surgery.
### Subject Device Description
EzVu Visual Vasopressor Injector (EV-19) is a 330 mm long and 5 mm diameter hollow bore laparoscopic instrument, used to enable laparoscopic flashback visualization before injecting Vasopressor drug into the uterine musculature during laparoscopic surgery. It consists of a stainless steel needle at the distal end, securely connected to a hub, an elongated shaft with transparent window at the distal end and stopcock with luer connection at the proximal end onto which a syringe can be attached. The shaft is made of transparent plastic tube, surrounded by a stainless steel tube along the proximal portion of the injector to provide stability. The stainless steel tube around the transparent plastic tube is wrapped in a heat shrink. The injector is available in one needle size (19 gauge).
### Technological characteristics
{6}------------------------------------------------
# K241595
| Comparison parameters | Subject Device | Predicate Device |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | EzVu Visual Vasopressor<br>Injector (EV-19) | Disposable Endoscope Injection<br>Needle |
| 510(k) number | K241595 | K220292 |
| Product Code | FBK | FBK |
| Regulation No. | 21 CFR 876.1500 | 21 CFR 876.1500 |
| Regulation Name | Endoscope and<br>Accessories | Endoscope and Accessories |
| Class | Class II | Class II |
| Indication for Use | EzVu Visual Vasopressor<br>Injector (EV-19) is<br>intended to inject<br>vasopressor solution into<br>the uterine musculature<br>during laparoscopic<br>surgery. | The Disposable Endoscope<br>Injection Needle is used for<br>endoscopic injection into<br>gastrointestinal mucosa and<br>submucosa<br>to:<br>* introduce a sclerosing agent,<br>vasoconstrictor, or other<br>solutions into selected sites to<br>control actual or potential<br>bleeding lesions in the digestive<br>system;<br>* aid in Endoscopic Mucosal<br>Resection (EMR), Endoscopic<br>Submucosal Dissection (ESD), or<br>polypectomy procedures;<br>* control non-variceal<br>hemorrhage |
| Configuration | EzVu is a 330 mm long<br>and 5 mm diameter<br>hollow bore laparoscopic<br>instrument with a<br>metallic needle at its<br>distal end. | The disposable endoscope<br>Injection Needle consists of<br>Sheath, Handle buckle, handle,<br>Booster tube, handle end cap,<br>Outer tube, Inner tube, Metal<br>end cap and injection needle tip |
| Single Use | Single Use | Single Use |
| Needle Gauge | 19G | 19G, 22G,23G, 25G |
| Outer Diameter (mm) | 5 | 1.8, 2.4 |
| Working length (mm) | 330 | 600,1200,1600,1800,2300,<br>2500 |
| Biocompatibility | Comply with ISO 10993-1 | Comply with ISO 10993-1 |
| Method | EO Sterilization | EO Sterilization |
| SAL | 10-6 | 10-6 |
{7}------------------------------------------------
### K241595 Page 4 of EzVu Visual Vasopressor Injector (EV-19)
### Differences between Subject Device & Predicate Device
• Indications for Use: The indications for use for the subject device slightly differs from that of the predicate due to variations in the targeted application area. While the subject device is designed for laparoscopic use, the predicate device is intended for use in conjunction with an endoscope. However, the intended use is the same – delivery of liquid agents into the tissue or mucosa.
- Device Name: Each device bears a distinct name.
• Configuration: Although the fundamental configuration of all devices remains similar comprising a tube with an attached needle for drug delivery—the subject and predicate device have different working lengths, diameters, and needle gauges to address the anatomical environment. However, the different dimensions do not introduce different questions of safety and effectiveness.
Overall, the differences do not result in a new intended use or raise different questions of safety and effectiveness.
# Non-Clinical Performance Data:
### Biocompatibility
The subject device meets the requirements of ISO 10993-1:2020 – "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" to address device biocompatibility for an external communicating device with tissue, bone, or dentin contact for less than 24 hours. The following endpoints were assessed:
- Cytotoxicity: ISO 10993-5:2009
- Irritation: ISO 10993-10:2021
- Sensitization: ISO 10993-10:2021
- Acute Systemic Toxicity: ISO 10993-11:2017
{8}------------------------------------------------
- . Material-Mediated Pyrogenicity: ISO 10993-11:2017
### Sterility
The subject device meets the requirements of ISO 11135:2014 – Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation, and routine control of a sterilization process for medical devices and the 2024 FDA guidance document Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile to support device sterility. The ethylene oxide and ethylene chlorohydrin residuals were measured after sterilization to meet the residual limits identified in ISO 10993-7:2008(R)2012.
### Shelf Life and Package Integrity
The subject device meets the requirements of ISO 11607-1:2019 – Packaging for terminally sterilized medical devices — Part 1: Requirements for materials, sterile barrier systems and packaging systems; The following test methods and standards were used to confirm shelf life and package integrity:
- Blister Packaging — Visual Inspection
- Label Adhesion — Visual Inspection
- Dye Penetration – ASTM F1929-23 – Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- Ink Fastness — Internal Methods
- Positive Air Pressure Leakage – Internal Methods
- . Negative Air Pressure Leakage - Internal Methods
- Seal Strength – ASTM F88-23 – Standard Test Method for Seal Strength of Flexible Barrier Materials
- Tensile Strength — Internal Methods
- Accelerated Aging – ASTM F1980-21 – Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
### Bench Performance Testing
The following bench tests were conducted on the EzVu Visual Vasopressor Injector (EV-19):
- . Connection Strength
- Flow Rate
- Positive Air Pressure Leakage
Veol Medical Technologies PVT. LTD.
{9}------------------------------------------------
- . Positive Liquid Pressure Leakage
- . Negative Air Pressure Leakage
- Patency of Lumen
- Compression
- Device Bending
- Needle Penetration Force - ISO 7864:2016 - Sterile hypodermic needles for single use -Requirements and test methods.
- Corrosion Resistance - ISO 9626:2016 – Stainless steel needle tubing for the manufacture of medical devices – Requirements and test methods.
- . Needle Stiffness - ISO 9626:2016 – Stainless steel needle tubing for the manufacture of medical devices – Requirements and test methods.
- Resistance to Breakage - ISO 9626:2016 - Stainless steel needle tubing for the manufacture of medical devices – Requirements and test methods.
- . Needle Bonding Strength -ISO 7864:2016 - Stainless steel needle tubing for the manufacture of medical devices – Requirements and test methods.
The subject device passed all performed bench testing.
### Clinical performance Data:
An observational clinical study for the EzVu Visual Vasopressor Injector (EV-19) was conducted as a multi-center, prospective observational study to evaluate the performance, usability, and safety of the device during laparoscopic myomectomy procedures. The study did not involve any interventions by the sponsor and focused on validating the intended use, transparency of the distal window, and compatibility with laparoscopic instruments. The study included 13 women, aged 30-40 years, with BMI of 18-40. A total of 24 injections across 13 patients were used, and 5 trained laparoscopic gynecological surgeons performed these procedures. The primary and secondary endpoints were as follows:
Primary Endpoint:
- . Blood Stain should be clearly visible in the transparent window when aspirated.
- . No leakage or breakage should be observed in any samples of the EzVu device.
Secondary Endpoint:
- . The performance of EzVu and other laparoscopic instruments should not be affected when used together in the procedure.
{10}------------------------------------------------
Page 7 of 7
### Data Collected
- . Device identifier tracking (UDI recorded for each use)
- . Visual inspection of the device before and after use
- . Insertion and positioning through trocar port
- . Blood aspiration visualization through the transparent window
- Injection of vasopressor drug and any potential leakage observations
- . Feedback from surgeons via Likert scale (1-5) for 19 usability questions
- . Photographic documentation for each procedure
There were no reported adverse events in the study. Blood aspirate was detected in 4 out of 25 cases. All surgeons provided usability ratings of 4.6 or greater, indicating favorable device usability. The device maintained structural integrity throughout the study. Compatibility with laparoscopic instruments was confirmed. The study results support the safety and usability of the EzVu Visual Vasopressor Injector (EV-19) in clinical settings.
# Conclusion:
In conclusion, the results of non-clinical and clinical performance testing demonstrate that the EzVu Visual Vasopressor Injector (EV-19) by Veol Medical technologies is as safe and effective as the predicate device Disposable Endoscope Injection Needle by Yangzhou Fartley Medical Instrument Technology Co., Ltd., to support a substantial equivalence determination.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.