To lubricate vaginal tissues to facilitate entry of diagnostic or therapeutic devices including those used in fertility interventions. Pre~Va may be applied directly to the device or may be deposited intravaginally using the applicator, prior to the insertion of diagnostic or therapeutic devices used in fertility interventions. As a personal lubricant Pre-Va supplements the body's own natural lubricating fluids, to moisturize, relieve friction and to enhance the ease and comfort of intimate sexual activity. Pre~Va is safe for use by couples who are trying to conceive and may be applied to vaginal or penile tissues for lubrication and moisturization purposes. It is compatible with latex and polyurethane condoms.
Device Story
Pre~Va is a non-sterile, water-based personal lubricant; formulated to supplement natural lubricating fluids; pH and osmolarity balanced to fertile cervical mucus and semen; non-spermicidal. Used in clinical settings to facilitate insertion of diagnostic/therapeutic devices (e.g., fertility interventions) and in home settings as a personal lubricant. Applied directly to devices, vaginal tissues, or penile tissues. Enhances comfort; facilitates device entry; supports conception efforts. Compatible with latex and polyurethane condoms.
Clinical Evidence
No clinical data provided; substantial equivalence based on identical technological characteristics and formulation to the predicate device.
Technological Characteristics
Water-based lubricant; ingredients include Hydroxyethylcellulose (NF), Pluronic 127 (NF), Sodium Chloride (USP), Arabinogalactan, Sodium Phosphate, Carbopol 934P (NF), Methyl Paraben (USP), Sodium Hydroxide (NF), and Potassium Phosphate. Non-sterile; compatible with latex and polyurethane condoms.
Indications for Use
Indicated for individuals requiring vaginal lubrication for diagnostic/therapeutic device insertion (including fertility interventions) or as a personal lubricant to relieve friction and enhance comfort during sexual activity. Safe for use by couples attempting to conceive.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
Pre' Vaginal Lubricant
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(k) Summary Pre~Va Vaginal Lubricant
JUL 1 6 2008
### l. General Information on Submitter
| Address: | INGfertility, LLC (Subsidiary of Bio-Origyn, LLC)<br>17206 S. Spangle Creek Rd.<br>Valleyford, WA 99036 USA |
|-----------------|-------------------------------------------------------------------------------------------------------------|
| Telephone: | 509.443.0149 |
| Fax: | 509.471.9638 |
| Email: | dclifton@ingfertility.com |
| Contact Person: | G. Dennis Clifton, Pharm.D. |
| Date Prepared: | March 18, 2008 |
### II. General Information on Device
| Proprietary Name: | Pre~Va Vaginal Lubricant |
|----------------------|--------------------------------------------------------------------------------------|
| Classification Name: | lubricant, patient, vaginal, latex compatible (21 CFR<br>884.5300, Product Code NUC) |
### III. Predicate Devices
| START | AREARY |
|----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Section of Allen August of Control Commens and Commens and Call All All of Concelle<br>Carol<br>1. DE RE | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
### IV. Description of Device
This product is a non-sterile, water-based personal lubricant formulated to supplement the body's own natural lubricating fluids. Pre~Va is used to lubricate vaginal tissues to facilitate entry of diagnostic or therapeutic devices including those used in fertility interventions. It is also used as a personal lubricant to supplement the body's own natural lubricating fluids and to enhance the comfort of intimate sexual activity. The formulation does not harm sperm function and has a pH and osmolarity that are physiologic ("balanced") to that of fertile cervical mucus and semen. The product is compatible with latex and polyurethane condoms. Following is the ingredient list for Pre~Va Vaginal Lubricant:
| Ingredients |
|---------------------------|
| Water |
| Hydroxyethylcellulose, NF |
| Pluronic 127, NF |
| Sodium Chloride, USP |
| Arabinogalactan |
| Sodium Phosphate |
| Carbopol 934P, NF |
| Methyl Paraben, USP |
| Sodium Hydroxide, NF |
| Potassium Phosphate |
1
{1}------------------------------------------------
### V. Intended Use
- . To lubricate vaginal tissues to facilitate entry of diagnostic or therapeutic devices including those used in fertility interventions. Pre~Va may be applied directly to the device or may be deposited intravaginally using the applicator, prior to the insertion of diagnostic or therapeutic devices used in fertility interventions
- As a personal lubricant Pre-Va supplements the body's own natural . lubricating fluids, to moisturize, relieve friction and to enhance the ease and comfort of intimate sexual activity. Pre~Va is safe for use by couples who are trying to conceive and may be applied to vaginal or penile tissues for lubrication and moisturization purposes. It is compatible with latex and polyurethane condoms.
# VI. Technological Characteristics of Device Compared to Predicate Device
All of the technological characteristics of Pre~Va are identical to the predicate device.
### VII. Summary of Performance Data
The performance data of Pre~Va are identical to the predicate.
### VIII. Conclusion
Pre~Va Vaginal Lubricant is safe for its intended use and substantially equivalent to the predicate device Pre' Vaginal Lubricant.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized symbol that resembles an abstract caduceus or a bird-like figure, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 27, 2014
INGfertility, LLC Dennis Clifton, Pharm.D. Vice President 17206 South Spangle Creek Road Valleyford, WA 99036
Re: K072741
Trade/Device Name: Pre-Va Vaginal Lubricant Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: PEB Dated (Date on orig SE Itr): July 1, 2008 Received (Date on orig SE Itr): July 8, 2008
Dear Dennis Clifton,
This letter corrects our substantially equivalent letter of July 16, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Dennis Clifton, Pharm.D.
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K072741
Device Name: Pro-Va Vaginal Lubricant
Indications for Use:
- > To lubricate vaginal tissues to facilitate entry of diagnostic or therapeutic devices including those used in fertility interventions. Pre~Va may be applied directly to the device or may be deposited intravaginally using the applicator, prior to the insertion of diagnostic or therapeutic devices used in fertility interventions.
- > As a personal lubricant Pre-Va supplements the body's own natural Tutoricaling Thuids, To moisturize, relieve Triction and To enthance The ease and comfort of intimate sexual activity. Pre~Va is safe for use by couples who are trying to conceive and may be applied to vaginal or penile tissues for lubrication and moisturization purposes. It is compatible with latex and polyurethane condoms
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hales Turner
Division Sig Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number.
Page of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.