The proposed LITE mesh kits are intended for tissue reinforcement and stabilization of fascial structures of the pelvic floor for vaginal wall prolapse, where surgical treatment is intended, either as mechanical support or bridging material for the fascial defect.
Device Story
LITE Pelvic Floor Repair Kits are sterile, single-use surgical mesh assemblies designed for transvaginal repair of anterior, apical, and posterior prolapse. Device consists of macroporous polypropylene knitted mesh with integrated legs protected by disposable polymer sleeves; includes lead with needle for use with Capio Open Access Suture Capturing Device. Used by surgeons in clinical/OR settings to provide mechanical support or bridging material for fascial defects. Mesh is lightweight, low-density version of previously cleared Pinnacle and Uphold systems. Facilitates passage through bodily tissues for placement; provides structural stabilization to pelvic floor.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included mechanical evaluation (thickness, pore size, density, tensile strength, stiffness, burst strength, suture pullout strength, flexibility) and biocompatibility testing (cytotoxicity, sensitization, irritation, acute systemic toxicity, sub-acute toxicity, genotoxicity, implantation).
Technological Characteristics
Macroporous polypropylene knitted mesh; lightweight, low-density configuration. Same fiber material and diameter as predicates but with larger pore size. Includes disposable polymer sleeves, needle, and lead assembly. Sterile, single-use.
Indications for Use
Indicated for patients requiring surgical treatment for vaginal wall prolapse, specifically for tissue reinforcement and stabilization of pelvic floor fascial structures.
Regulatory Classification
Identification
Surgical mesh for transvaginal pelvic organ prolapse repair is a prescription device intended to reinforce soft tissue in the pelvic floor. This device is a porous implant that is made of synthetic material, non-synthetic material, or a combination of synthetic and non-synthetic materials. This device does not include surgical mesh for other intended uses (§ 878.3300 of this chapter).
{0}------------------------------------------------
K103426
Page 1 of 2.
### 510(k) Summary for the LITE Pelvic Floor Repair Kits
### Date Prepared: September 13, 2011
SEP 1 4 2011
#### A. Sponsor
Boston Scientific Corporation Urology and Gynecology Division 100 Boston Scientific Way Marlborough, MA 01756
### B. Contact
Michelle Berry Senior, Regulatory Affairs Specialist 508-683-4941 or Donna Gardner
Director, Regulatory Affairs 508-683-4398
### C. Device Name
LITE Pelvic Floor Repair Kits Tradename: Common/usual name: Surgical Mesh Classification Name: OTP .- Mesh, Surgical, Polymeric and OTP - Mesh, Surgical, gynecologic, for pelvic organ prolapse, transvaginally placed 21 CFR 878.3300, Class II
# D. Predicate Device(s)
Tradename: Common/usual name: Classification Name:
Premarket Notification:
Tradename: Common/usual name: Classification Name:
Premarket Notification:
Tradename: Common/usual name: Classification Name:
Premarket Notification:
Pinnacle and Uphold Pelvic Floor Repair Kits Surgical Mesh Oil - Mesh, Surgical, Polymeric 21 CFR 878.3300, Class II Boston Scientific, K071957, K081048, K091131
Novasilk Mesh Surgical Mesh OTP - Mesh, Surgical, Polymeric 21 CFR 878.3300, Class II K053414
Restorelle Polypropylene Mesh Surgical Mesh OTY oldMesh, Surgical, Polymeric 21 CFR 878.3300, Class II K092207
{1}------------------------------------------------
# K103426 Page 2 of 2
# E. Device Description
The proposed devices were submitted to introduce lightweight versions of previously cleared pelvic floor repair kits: Pinnacle Anterior/Apical, Pinnacle Posterior, Pinnacle Duet and Uphold Vaginal Support System. The proposed mesh configurations were designed for performing transvaginal vaginal wall repair to facilitate treatment of anterior, apical and posterior prolapse repairs. These devices are sterile, single use devices, consisting of one or two light-weight synthetic mesh assemblies and a needle holder.
The mesh assemblies consist of a macroporous polypropylene knitted mesh body with integrated legs that are protected by disposable polymer sleeves. At the distal end of the disposable polymer sleeve is a lead with needle designed for use with the Capio The Open Access Suture Capturing Device. The disposable lead was designed to facilitate the passage of the proposed mesh through bodily tissues for placement.
### F. Intended Use
The proposed LITE mesh kits are intended for tissue reinforcement and stabilization of fascial structures of the pelvic floor for vaginal wall prolapse, where surgical treatment is intended, either as mechanical support or bridging material for the fascial defect.
### G. Technological Characteristics
The proposed device has the same mesh fiber materials and fiber diameter as the predicates but is knit with a larger pore size to create a lighter, lower density mesh.
# H. Substantial Equivalence/ Performance Testing (Bench)
Utilizing FDA's Guidance for Industry and FDA Staff "Format for Traditional and Abbreviated 510(k)s" and "Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh", a direct comparison of key characteristics demonstrates that the proposed mesh is substantially equivalent to the predicate mesh in terms of intended use, technological characteristics, and performance characteristics tested.
Boston Scientific conducted mechanical performance and biocompatibility testing to demonstrate the equivalence of the LITE mesh to the predicate devices. The mesh mechanical performance testing included mesh thickness, pore size, density, tensile strength, stiffness, burst strength, suture pullout strength, and mesh flexibility. In addition, Boston Scientific evaluated the tensile strengths of the mesh assembly including the components of the leg assemblies (needle, lead, dilator, protective sleeve and leader loop). Boston Scientific also evaluated the force to remove the leg assemblies from the mesh.
Biocompatibility testing on the LITE kits' materials of construction included cytotoxicity, sensitization, irritation, acute systemic toxicity, sub acute toxicity, genotoxicity and implantation.
The results of the performance testing demonstrate equivalence of the LITE Pelvic Floor Repair kits to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, with its wings forming a flowing, abstract shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Michelle Berry Sr. Regulatory Affairs Specialist Boston Scientific Corporation 100 Boston Scientific Way MARLBOROUGH MA 01752
SEP 2 8 2012
Re: K103426
Trade/Device Name: LITE Pelvic Floor Repair Kits Regulation Number: 21 CFR§ 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTP Dated: September 2, 2011 Received: September 6, 2011
Dear Ms. Berry:
This letter corrects our substantially equivalent letter of September 14, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 -
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin R. Perkins
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
510(k) Number (if Known): _ K 1034 2 6
Device Name: LITE Pelvic Floor Repair Kits
Indications For Use:
The proposed LITE Pelvic Floor Repair Kits are indicated for tissue reinforcement and stabilization of fascial structures of the pelvic floor for vaginal wall prolapse, where surgical treatment is intended, either as mechanical support or bridging material for the fascial defect.
Prescription Use X (21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hl. Leun
ro-Renal, and
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.