K071512 · ETHICON, Inc. · OTP · May 15, 2008 · Obstetrics/Gynecology
Device Facts
Record ID
K071512
Device Name
GYNECARE PROLIFT +M* PELVIC FLOOR REPAIR SYSTEMS
Applicant
ETHICON, Inc.
Product Code
OTP · Obstetrics/Gynecology
Decision Date
May 15, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5980
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
The GYNECARE PROLIFT* Total, Anterior, and Posterior Pelvic Floor Repair Systems, through the placement of GYNECARE GYNEMESH PS* PROLENE Soft Mesh, are indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended, either as mechanical support or bridging material for the fascial defect. The GYNECARE PROLIFT+M* Total, Anterior, and Posterior Pelvic Floor Repair Systems, through the placement of GYNECARE GYNEMESH M* Partially Absorbable Mesh, are indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended, either as mechanical support or bridging material for the fascial defect.
Device Story
System comprises pre-cut mesh implants and specialized surgical instruments (guides, retrieval devices, cannulas) for pelvic floor repair. Prolift systems utilize non-absorbable mesh; Prolift+M systems utilize partially absorbable mesh. Used by surgeons in clinical settings to treat vaginal wall prolapse. Instruments facilitate precise placement of mesh to provide mechanical support or bridge fascial defects. Mesh acts as permanent or partially absorbable reinforcement to stabilize pelvic floor structures. Benefit includes long-lasting stabilization of fascial tissues.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing and cadaveric procedural performance evaluation. Biological reactivity assessed per ISO 10993-1.
Technological Characteristics
Mesh implants made of non-absorbable or partially absorbable polymers identical to predicate materials. System includes manual surgical instruments (guides, retrieval devices, cannulas). Biocompatibility tested per ISO 10993-1. Non-electronic, mechanical device.
Indications for Use
Indicated for tissue reinforcement and stabilization of pelvic floor fascial structures in patients with vaginal wall prolapse requiring surgical intervention.
Regulatory Classification
Identification
Surgical mesh for transvaginal pelvic organ prolapse repair is a prescription device intended to reinforce soft tissue in the pelvic floor. This device is a porous implant that is made of synthetic material, non-synthetic material, or a combination of synthetic and non-synthetic materials. This device does not include surgical mesh for other intended uses (§ 878.3300 of this chapter).
{0}------------------------------------------------
K071512
pg 1 of 2
# Summary of Safety and Effectiveness
Information supporting claims of substantial equivalence, as defined under the Federal Food, Drug and Cosmetic Act, in respect to safety and effectiveness is summarized below.
### Submitted by:
Bryan A. Lisa Regulatory Affairs Project Manager ETHICON, Inc., A Johnson & Johnson Company Route 22 West, PO Box 151 Somerville, NJ 08876
MAY 1 5 2008
#### Name/Classification of Device:
Class II in 21 CFR § 878.3300, Surgical Mesh (0TP) Class I in 21 CFR § 878.4800, Manual surgical instrument for general use
#### Trade Name:
GYNECARE PROLIFT Total, Anterior, and Posterior Pelvic Floor Repair Systems GYNECARE PROLIFT+M* Total, Anterior, and Posterior Pelvic Floor Repair Systems
#### Predicate Devices:
- GYNECARE GYNEMESH PS* PROLENE* Soft Mesh (ETHICON, Inc.) K013718 .
- ULTRAPRO* Mesh (ETHICON, Inc.) K033337
- AMS APOGEE Vault Suspension System (American Medical Systems, Inc.) K040537
- AMS PERIGEE System (American Medical Systems, Inc.) K040623 .
Statement of Intended Use:
The GYNECARE PROLIFT* Total, Anterior, and Posterior Pelvic Floor Repair Systems, through the placement of GYNECARE GYNEMESH PS* PROLENE Soft Mesh, are indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended, either as mechanical support or bridging material for the fascial defect.
The GYNECARE PROLIFT+M* Total, Anterior, and Posterior Pelvic Floor Repair Systems, through the placement of GYNECARE GYNEMESH M* Partially Absorbable Mesh, are indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended, either as mechanical support or bridging material for the fascial defect.
#### Device Description:
The GYNECARE PROLIFT Total, Anterior, and Posterior Pelvic Floor Repair Systems consist of pre-cut non-absorbable mesh implants and a set of instruments to facilitate mesh implant placement. The following table summarizes the instruments included with each system:
#### 510(k) Summary
{1}------------------------------------------------
| REPAIR SYSTEM | COMPONENTS | | | |
|---------------|--------------|-------|-------------------|----------|
| | Mesh Implant | Guide | Retrieval Devices | Cannulas |
| Total | 1 Total | 1 | 6 | 6 |
| Anterior | 1 Anterior | 1 | 4 | 4 |
| Posterior | 1 Posterior | 1 | 2 | 2 |
The GYNECARE PROLIFT+M Total, Anterior, and Posterior Pelvic Floor Repair Systems consist of pre-cut partially absorbable mesh implants and a set of instruments to facilitate mesh implant placement. The following table summarizes the instruments included with each system:
1071512
2022
| REPAIR SYSTEM | COMPONENTS | | | |
|---------------|--------------|-------|-------------------|----------|
| | Mesh Implant | Guide | Retrieval Devices | Cannulas |
| Total | 1 Total | 1 | 6 | 6 |
| Anterior | 1 Anterior | 1 | 4 | 4 |
| Posterior | 1 Posterior | 1 | 2 | 2 |
#### Summary of Technological Characteristics of New Device to Predicate Devices:
The modified devices have similar technological characteristics as the predicate devices.
GYNECARE PROLIFT: Like currently marketed devices, the implantable component is a sterile, mesh implant intended for the repair of pelvic floor defects. The mesh implant component of the proposed device is made of nonabsorbable polymers, which are identical to those found in GYNECARE GYNEMESH PS, currently marketed by ETHICON, Inc. ....
GYNECARE PROLIFI+M: Like currently marketed devices, the implantable component is a sterile, mesh implant intended for the repair of pelvic floor defects. The mesh implant component of the proposed device is made of nonabsorbable and absorbable polymers, which are identical to those found in ULTRAPRO Mesh, currently marketed by ETHICON, Inc.
#### Performance Data:
Biological reactivity of the materials has been assessed using methods specified in ISO 10993-1. and the materials were found to be acceptable for their intended uses. Results of functional performance testing (bench and cadaver testing) indicate that the proposed device meets or exceeds all functional requirements, based on FDA's Guidance Document "Guidance for the Preparation of a Premarket Notification Arplication for a Surgical Mesh". Preclinical modelling was used to evaluate procedural performance of the systems.
#### Conclusions:
Based on the similarities to the predicate devices identified in this submission, we conclude that the modified device is substantially equivalent to the predicate devices under the Federal Food, Drug, and Cosmetic Act.
* Trademark of ETHICON, Inc.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings outstretched, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 2 8 2012
Ethicon, Inc. % Mr. Bryan A. Lisa Route 22 West P.O. Box 151 SOMERVILLE NJ 08876
Re: K071512
Trade/Device Name: Gynecare Prolifi™ Total, Anterior, and Posterior Pelvic Floor Repair Systems; Gynecare Prolift +MTM Total, Anterior, and Posterior Pelvic Floor Repair Systems
Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTP Dated: February 22, 2008 Received: February 25, 2008
Dear Mr. Lisa:
This letter corrects our substantially equivalent letter of May 15, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Benjamin R. Perkins
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
BAL-2007-002
## Indications for Use
## 510(k) Number (if known): K071512
## Device Name: GYNECARE PROLIFT* and GYNECARE PROLIFT+M* Total, Anterior, and Posterior Pelvic Floor Repair Systems
Indications for Use:
The GYNECARE PROLIFT* Total, Anterior, and Posterior Pelvic Floor Repair Systems are indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended, either as mechanical support or bridging material for the fascial defect.
The GYNECARE PROLIFT+M* Total, Anterior, and Posterior Pelvic Floor Repair Systems, through the placement of GYNECARE GYNEMESH M* Partially Absorbable Mesh, are indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended, either as mechanical support or bridging material for the fascial defect.
*Trademark.
(21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
NILREOL for reims
Division of General, Restorative, and Neurological Devices
510(k) Number K071512
INDICATIONS FOR USE
ETHICON CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.