Astroglide® X Silicone Personal Lubricant is a personal lubricant for penile, vaginal, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex condoms. This product is not compatible with polyisoprene and polyurethane condoms.
Device Story
Astroglide® X Silicone Personal Lubricant is a non-sterile, clear, silicone-based lubricant; intended for OTC use to moisturize and lubricate during intimate sexual activity. Product is applied topically by the user. It supplements natural lubrication and enhances comfort. Device is packaged in a PETE bottle with a polypropylene flip-top cap. It is compatible with natural rubber latex condoms but incompatible with polyisoprene and polyurethane condoms. It is not a spermicide or contraceptive. Safety is established through biocompatibility and condom compatibility testing.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1:2009/(R)2013, including cytotoxicity (ISO 10993-5), sensitization/irritation (HRIPT per ASTM D6355), and acute systemic toxicity (ISO 10993-11). Condom compatibility tested per ASTM D7661-10. Shelf-life of 3 years validated via accelerated aging per ASTM F1980-16.
Technological Characteristics
Silicone-based lubricant (Cyclopentasiloxane, Dimethicone, Cyclohexasiloxane). Viscosity: 225-350 cP. Non-sterile. Primary packaging: PETE bottle with polypropylene cap. Biocompatibility per ISO 10993. Condom compatibility per ASTM D7661-10. Shelf-life testing per ASTM F1980-16.
Indications for Use
Indicated for penile, vaginal, and/or anal application to moisturize and lubricate, enhance comfort during intimate sexual activity, and supplement natural lubrication. Compatible with natural rubber latex condoms; contraindicated for use with polyisoprene and polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
pjur® Backdoor Anal Glide and pjur® Analyse Me! (K141913)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 28, 2021
BioFilm, Inc. Kevin Jamil Regulatory Affairs 3225 Executive Ridge Vista, CA 92081
Re: K210242
> Trade/Device Name: Astroglide® X Silicone Personal Lubricant Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: January 27, 2021 Received: January 29, 2021
Dear Kevin Jamil:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210242
Device Name Astroglide® X Silicone Personal Lubricant
Indications for Use (Describe)
Astroglide® X Silicone Personal Lubricant is a personal lubricant for penile, vaginal, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex condoms. This product is not compatible with polyisoprene and polyurethane condoms.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary - K210242 Astroglide® X Silicone Personal Lubricant
### i. General Information on Submitter
| Applicant: | BioFilm, Inc. |
|-----------------------------|---------------------------------------------|
| Address: | 3225 Executive Ridge<br>Vista, CA 92081 USA |
| Telephone: | 760-727-9030 |
| Fax: | 760-727-8080 |
| Contact Person: | Kevin Jamil |
| Contact Title: | Regulatory Affairs |
| Email: | kevin@biofilm.com |
| Date Prepared: | 4/27/2021 |
| Establishment Registration: | 2025771 |
### ii. General Information on Device
| Proprietary Name: | Astroglide® X Silicone Personal Lubricant |
|--------------------|-------------------------------------------|
| Common Name: | Personal Lubricant |
| Regulation Name: | Condom |
| Regulation Number: | 21 CFR 884.5300 |
| Regulatory Class: | II |
| Product Code: | NUC (Lubricant, Personal) |
### iii. Predicate Device
| Predicate Device | 510(k) Number |
|----------------------------------------------------|---------------|
| pjur® Backdoor Anal Glide and pjur®<br>Analyse Me! | K141913 |
This predicate device has not been subject to a device-related recall.
### iv. Description of Device
Astroglide® X Silicone Personal Lubricant is non-sterile, clear, odorless, and siliconebased lubricant. This product is not a spermicide or contraceptive. Astroglide® X Silicone Personal Lubricant is compatible with natural rubber latex condoms. Astroglide® X Silicone Personal Lubricant is not compatible with polyurethane, or polyisoprene condoms. This product's primary packaging is a PETE clear bottle with a screw on polypropylene flip-top cap. The bottle is packaged in a cardboard carton which constitutes the final packaging.
{4}------------------------------------------------
The specifications for Astroglide® X Silicone Personal Lubricant are described in the following table.
| Parameter | Specification (Test Method) |
|--------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Appearance | Clear |
| Odor | Characteristic |
| Viscosity | 225 to 350 centipoise |
| Total yeast/mold count (USP <61>) | <10 cfu/mL |
| Total aerobic microbial count (USP <61>) | <100 cfu/mL |
| Absence of Pathogenic Organisms,<br>Staphylococcus aureus, Pseudomonas<br>aeruginosa, and Candida albicans<br>(USP <62>) | Absent |
| Water activity (USP<1112>) | 0.3 Aw |
## Table 1. Physical Specifications of Astroglide® X Silicone Personal Lubricant
### V. Indications for Use
Astroglide® X Silicone Personal Lubricant is a personal lubricant for penile, vaginal, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex condoms. This product is not compatible with polyisoprene and polyurethane condoms.
### vi. Substantial Equivalence Discussion
The following table compares the intended use and key technological characteristics of the subject and predicate device:
{5}------------------------------------------------
| Characteristic /<br>Feature | Astroglide® X Silicone<br>Personal Lubricant<br>(Subject Device) | pjur® Backdoor Anal<br>Glide and pjur® Analyse<br>Me! (Predicate device) | Comparison |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | Astroglide® X Silicone<br>Personal Lubricant is a<br>personal lubricant for penile,<br>vaginal, and/or anal<br>application intended to<br>moisturize and lubricate, to<br>enhance the ease and<br>comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication.<br>This product is compatible<br>with natural rubber latex<br>condoms. This product is not<br>compatible with polyisoprene<br>and polyurethane condoms. | pjur® Backdoor Anal Glide<br>is a personal lubricant for<br>penile, vaginal, and/or anal<br>application intended to<br>moisturize and lubricate, to<br>enhance the ease and<br>comfort of intimate sexual<br>activity and supplement<br>the body's natural<br>lubrication. This product is<br>compatible with natural<br>rubber latex, synthetic<br>polyisoprene, and<br>polyurethane condoms. | The subject and predicate<br>devices have the same<br>indications for use except<br>for a difference in condom<br>compatibilities. |
| Condom<br>Compatibility | This product is compatible<br>with natural rubber latex<br>condoms. This product is not<br>compatible with polyisoprene<br>and polyurethane condoms. | This product is<br>compatible with natural<br>rubber latex, synthetic<br>polyurethane, and<br>polyisoprene condoms. | The condom compatibility<br>of the predicate device is<br>different, but it does not<br>raise different questions<br>of safety and<br>effectiveness. |
| Silicone-Based<br>Lubricant | Yes | Yes | Same |
| Over the Counter | Yes | Yes | Same |
| Appearance | Clear | Clear | Same |
| Not a<br>contraceptive or<br>Spermicide | Yes | Yes | Same |
| Non-sterile | Yes | Yes | Same |
| Primary<br>Ingredients | Cyclopentasiloxane<br>Dimethicone<br>Cyclohexasiloxane | cyclopentasiloxane,<br>dimethicone,<br>dimethiconol,<br>Simmondsia Chinensis<br>(Jojoba) Seed Oil | The ingredients of the<br>predicate device are<br>different; however,<br>differences in device<br>ingredients do not raise<br>different questions of<br>safety & effectiveness |
| Characteristic /<br>Feature | Astroglide® X Silicone<br>Personal Lubricant<br>(Subject Device) | pjur® Backdoor Anal<br>Glide and pjur® Analyse<br>Me! (Predicate device) | Comparison |
| Microbial Limits | - Total mold/yeast count<br><10 cfu/mL<br>- Total aerobic microbial<br>count <100 cfu/mL<br>- Absence of pathogens<br>(Candida albicans,<br>Pseudomonas aeruginosa,<br>Staphylococcus aureus) | - Total mold/yeast count<br><10 cfu/mL<br>- Total aerobic microbial<br>count <100 cfu/mL<br>- Absence of pathogens<br>(Candida albicans,<br>Pseudomonas aeruginosa,<br>Staphylococcus aureus) | Same |
| Viscosity | 225 - 350 centipoise | 600 - 1,000 centipoise | The viscosities of the<br>subject and predicate<br>devices are different;<br>however, differences<br>in viscosities do not<br>raise different<br>questions of safety<br>and effectiveness. |
| Water Activity | < 0.3 Aw | < 0.3 Aw | Same |
| Shelf-Life | 3 years | 3 years | Same |
{6}------------------------------------------------
As noted in the table above, the subject and predicate device have similar indications for use statements nad have the same intended use – to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. The subject and predicate device have difference technological characteristics, including different formulations, specifications, and condom compatibilities. The different technological characteristics identified do not raise different questions of safety and.
### Summary of Non-Clinical Performance Testing vii.
Biocompatibility
{7}------------------------------------------------
Biocompatibility testing on the subject lubricant was performed in accordance with ISO 10993-1: 2009/(R)2013, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process.
Table 3. Biocompatibility of Astroglide® X Silicone Personal Lubricant
| Test Performed |
|-----------------------------------------------------------------------------------------------------|
| Cytotoxicity per ISO 10993-5: 2009/(R)2014 |
| Human Repeat Insult Patch Testing (HRIPT) per ASTM D6355<br>to address sensitization and irritation |
| Acute Systemic Toxicity per ISO 10993-11: 2017 |
The results of the testing demonstrate that the subject device is non-cytotoxic, nonirritating, non-sensitizing, and not systemically toxic.
# Condom Compatibility
Astroglide® X Silicone Personal Lubricant was tested for compatibility with natural rubber latex condoms using ASTM D7661-10. Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms. Results show that Astroglide® X Silicone Personal Lubricant is compatible with natural rubber latex condoms only.
# Shelf Life
Astroglide® X Silicone Personal Lubricant has a shelf-life of 3 years based on 12 months of accelerated aging testing results per ASTM F1980-16, Standard Guide for Accelerated Aqing of Sterile Barrier Systems for Medical Devices. All specifications for the subject lubricant listed in Table 1 were met throughout the shelf-life study.
### viiii. Substantial Equivalence
The results of the testing described above demonstrate that the Astroglide® X Silicone Personal Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.