pjur AQUA Baseline, pjur AQUA Guarana, pjur AQUA ProVitamin B5, pjur BACK DOOR anal Provitamin B5
K200731 · Pjur Group Luxembourg S.A. · NUC · Jan 21, 2021 · Obstetrics/Gynecology
Device Facts
Record ID
K200731
Device Name
pjur AQUA Baseline, pjur AQUA Guarana, pjur AQUA ProVitamin B5, pjur BACK DOOR anal Provitamin B5
Applicant
Pjur Group Luxembourg S.A.
Product Code
NUC · Obstetrics/Gynecology
Decision Date
Jan 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
pjur® AQUA Baseline for penile, vaginal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms. pjur® AQUA Guarana for penile, vaginal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms. pjur® BACK DOOR anal Provitamin B5 for penile, vaginal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms. pjur® AQUA ProVitamin B5 for penile, vaginal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
Device Story
Water-based personal lubricants; intended for OTC use to moisturize and lubricate during intimate sexual activity. Applied topically to penile, vaginal, or anal areas. Formulated with ingredients including glycerin, hydroxyethylcellulose, and various extracts/provitamins. Provided non-sterile in polyethylene bottles. Not compatible with natural rubber latex, polyisoprene, or polyurethane condoms. Benefits include supplementation of natural lubrication and enhanced comfort during sexual activity.
Clinical Evidence
Bench testing only. Includes biocompatibility (cytotoxicity, sensitization, vaginal irritation, systemic toxicity per ISO 10993), condom compatibility (ASTM D7661-10), and shelf-life stability testing (accelerated and real-time).
Technological Characteristics
Water-based personal lubricant; non-sterile; polyethylene bottle packaging. Formulations include glycerin, ethoxydiglycol, phenoxyethanol, hydroxyethylcellulose, and various botanical extracts/provitamins. Specifications include pH 4.1-4.7, viscosity 690-1250 mPa*s, and osmolality 500-625 mOsmol/kg. Tested per USP <791>, <912>, <785>, <61>, <62>, and <51>.
Indications for Use
Indicated for individuals requiring personal lubrication for penile, vaginal, and/or anal application to enhance comfort during intimate sexual activity. Not compatible with natural rubber latex, polyisoprene, or polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 21, 2021
pjur group Luxembourg SA % Candace Cederman Consultant Candace F. Cederman 722 Arjean Drive Wilmington, NC 28411
Re: K200731
> Trade/Device Name: piur® AOUA Baseline, piur® AOUA Guarana, piur® BACK DOOR anal Provitamin B5, pjur® AQUA ProVitamin B5 Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: December 21, 2020 Received: December 22, 2020
Dear Candace Cederman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200731
Device Name
pjur® AQUA Baseline, pjur® AQUA Guarana, pjur® BACK DOOR anal Provitamin B5, pjur® AQUA ProVitamin B5
#### Indications for Use (Describe)
pjur® AQUA Baseline for penile, vaginal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
pjur® AQUA Guarana for penile, vaginal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
pjur® BACK DOOR anal Provitamin B5 for penile, vaginal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
pjur® AQUA ProVitamin B5 for penile, vaginal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the "pjur group" with the word "pjur" in large, bold, dark gray letters with a yellow circle above the "j". Below "pjur" is the word "group" in a smaller, bold, light gray font. Below the logo is the text "510(k) Summary" followed by "K200731".
| DATE PREPARED: | 19 January 2021 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER INFORMATION: | pjur group Luxembourg SA<br>87 esplanade de la Moselle<br>L-6637 Wasserbillig, Luxembourg<br>Telephone:<br>+352 74-8989<br>Fax:<br>+352 74-8990<br>Contact: Andrea Giebel, Quality Management Representative |
| DEVICE INFORMATION: | Trade name:<br>• pjur® AQUA Baseline<br>• pjur® AQUA Guarana<br>• pjur® BACK DOOR anal Provitamin B5<br>• pjur® AQUA ProVitamin B5<br>Common name: Personal Lubricant<br>Regulation Number: 21 CFR 884.5300<br>Regulation Name: Condom<br>Product Code: NUC (Lubricant, Personal)<br>Class: II |
| PREDICATE DEVICE: | Astroglide® Organix® Liquid, K200114<br>The predicate device has not been subject to a design-<br>related recall. |
## DEVICE DESCRIPTION:
The piur® AQUA Baseline, pjur® AQUA Guarana, pjur® BACK DOOR anal Provitamin B5, and pjur® AQUA ProVitamin B5 are water-based personal lubricants provided non-sterile. These personal lubricants are for over-the-counter use. They are not compatible with natural rubber latex, polyurethane, and polyisoprene condoms. The products are provided in 30 mL and 100 mL polyethylene bottles.
The pjur® AQUA Baseline consists of water, glycerin, ethoxydiglycol, phenoxyethanol, hydroxypropyl guar hydroxypropyltrimonium chloride, hydroxyethylcellulose, and citric acid. The pjur® AQUA Guarana, pjur® BACK DOOR anal Provitamin B5, and pjur® AQUA ProVitamin B5 have the same formulation and consist of water, glycerin, ethoxydiglycol, phenoxyethanol, hydroxypropyltrimonium chloride, propylene glycol, hydroxyethylcellulose, panthenol, polysorbate 20, citric acid, chamomilla recrutita (matricaria) flower extract, paullinia cupana seed extract, cananga odorata flower oil, alcohol, 1,2hexanediol, decylene glycol, lactic acid, sodium benzoate, and potassium sorbate.
{5}------------------------------------------------
| Parameter | Method | Specifications for<br>pjur® AQUA Baseline | Specifications for<br>pjur® AQUA Guarana, pjur®<br>BACK DOOR anal<br>Provitamin B5, and pjur®<br>AQUA ProVitamin B5 |
|--------------------------------------------|-----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Appearance | | Clear fluid | Clear fluid |
| Odor | | Product typical | Product typical |
| pH | USP<791> | 4.1-4.7 | 4.1-4.7 |
| Viscosity (mPa*s) | USP<912> | 750 - 1250 | 690 - 1100 |
| Osmolality<br>(mOsmol/kg) | USP <785> | 500-625 (diluted 1:3.9) | 500-625 (diluted 1:4.1) |
| Total Yeast and Mold<br>Count (TYMC) | USP<61> | <10 cfu/g | <10 cfu/g |
| Total Aerobic<br>Microbial Count<br>(TAMC) | USP<61> | <100 cfu/g | <100 cfu/g |
| Presence of<br>Pathogenic<br>Organisms | USP<62> | <i>Pseudomonas<br/>aeruginosa</i><br>Absent<br><i>Staphylococcus<br/>aureus</i><br>Absent<br><i>Candida albicans</i><br>Absent | <i>Pseudomonas<br/>aeruginosa</i><br>Absent<br><i>Staphylococcus<br/>aureus</i><br>Absent<br><i>Candida albicans</i><br>Absent |
| Antimicrobial<br>Preservation | USP<51> | Category 2 - Topical, bacteria<br>should show not less than 2.0<br>log reduction at 14 days and<br>no increase from 14-day count<br>at the 28-day count. Yeast<br>and molds should show no<br>increase from the initial<br>calculated count at 14 and 28<br>days. | Category 2 - Topical, bacteria<br>should show not less than 2.0<br>log reduction at 14 days and<br>no increase from 14-day count<br>at the 28-day count. Yeast<br>and molds should show no<br>increase from the initial<br>calculated count at 14 and 28<br>days. |
#### Table 1: Device Specifications
### INDICATIONS FOR USE:
pjur® AQUA Baseline for penile, vaginal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
pjur® AQUA Guarana for penile, vaqinal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
pjur® BACK DOOR anal Provitamin B5 for penile, vaginal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
pjur® AQUA ProVitamin B5 for penile, vaginal and/or anal application, intended to moisturize and lubricate, enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
{6}------------------------------------------------
## COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT AND PREDICATE DEVICES
| | pjur® AQUA Baseline, pjur®<br>AQUA Guarana, pjur® BACK<br>DOOR anal Provitamin B5, and<br>pjur® AQUA ProVitamin B5<br>K200731<br>Subject Devices | Astroglide® Organix® Liquid<br>K200114<br>Predicate Device | Comparison |
|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Indications for Use | pjur® AQUA Baseline for penile,<br>vaginal and/or anal application,<br>intended to moisturize and<br>lubricate, enhance the ease and<br>comfort of intimate sexual activity<br>and supplement the body's natural<br>lubrication. This product is not<br>compatible with natural rubber<br>latex, polyisoprene, and<br>polyurethane condoms.<br><br>pjur AQUA Guarana for penile,<br>vaginal and/or anal application,<br>intended to moisturize and<br>lubricate, enhance the ease and<br>comfort of intimate sexual activity<br>and supplement the body's natural<br>lubrication. This product is not<br>compatible with natural rubber<br>latex, polyisoprene, and<br>polyurethane condoms.<br><br>pjur BACK DOOR anal Provitamin<br>B5 for penile, vaginal and/or anal<br>application, intended to moisturize<br>and lubricate, enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the body's<br>natural lubrication. This product is<br>not compatible with natural rubber<br>latex, polyisoprene, and<br>polyurethane condoms.<br><br>pjur® AQUA ProVitamin B5 for<br>penile, vaginal and/or anal<br>application, intended to moisturize<br>and lubricate, enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the body's<br>natural lubrication. This product is<br>not compatible with natural rubber<br>latex, polyisoprene, and<br>polyurethane condoms. | Astroglide® Organix® Liquid is a<br>personal lubricant for penile and/or<br>vaginal application, intended to<br>moisturize and lubricate, to<br>enhance the ease and comfort of<br>intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is<br>compatible with natural rubber<br>latex and polyisoprene condoms.<br>The product is not compatible with<br>polyurethane condoms. | Same<br>intended use |
| Rx / OTC<br>Base Type | OTC<br>Water | OTC<br>Water | Same<br>Same |
| Primary Ingredients | pjur® AQUA Baseline:<br>Water, Glycerin, Ethoxydiglycol,<br>Phenoxyethanol, Hydroxypropyl<br>Guar Hydroxypropyltrimonium | Water, Glycerin, Xanthan Gum,<br>Aloe Barbabensis Powder,<br>Chamomilla Recutita (Matricaria)<br>Flower Extract, Calendula<br>Officinalis Flower Extract, | Different |
| | pjur® AQUA Baseline, pjur®<br>AQUA Guarana, pjur® BACK | Astroglide® Organix® Liquid<br>K200114 | Comparison |
| | DOOR anal Provitamin B5, and<br>pjur® AQUA ProVitamin B5 | Predicate Device | |
| | K200731<br>Subject Devices | | |
| | Chloride, Hydroxyethylcellulose,<br>and Citric acid<br>pjur® AQUA Guarana, pjur® BACK<br>DOOR anal Provitamin B5, and | Vaccinium Macrocarpon<br>(Cranberry) Fruit Extract, Passiflora<br>Incarnata Flower Extract, Sodium<br>Benzoate, Potassium Sorbate,<br>Citric Acid | |
| | pjur® AQUA ProVitamin B5:<br>Water, Glycerin, Ethoxydiglycol,<br>Phenoxyethanol, Hydroxypropyl<br>Guar Hydroxypropyltrimonium<br>Chloride, Propylene Glycol,<br>Hydroxyethylcellulose, Panthenol,<br>Polysorbate 20, Citric Acid,<br>Chamomilla Recrutita (Matricaria)<br>Flower Extract, Paullinia Cupana<br>Seed Extract, Cananga Odorata<br>Flower Oil, Alcohol, 1,2-<br>Hexanediol, Decylene Glycol,<br>Lactic Acid, Sodium Benzoate,<br>Potassium Sorbate | | |
| Appearance | Clear Liquid | Colorless to slightly yellow, hazy<br>liquid | Different |
| Odor | Product typical (characteristic) | Odorless | Different |
| Viscosity | pjur® AQUA Baseline:<br>750-1250 mPa*s<br>pjur® AQUA Guarana, pjur® BACK<br>DOOR anal Provitamin B5, and<br>pjur® AQUA ProVitamin B5:<br>690-1100 mPa*s | 250-310 cps | Different |
| pH | 4.1-4.7 (all versions) | 3.5-5.5 | Similar |
| Osmolality | pjur® AQUA Baseline:<br>500-625 mOsm/kg (diluted 1:3.9)<br>pjur® AQUA Guarana, pjur® BACK<br>DOOR anal Provitamin B5, and<br>pjur® AQUA ProVitamin B5:<br>500-625 mOsm/kg (diluted 1:4.1) | 500-800 mOsm/kg (diluted1:5) | Different |
| Microbial Limits | Total aerobic microbial count<br>(TAMC): <100 cfu/g<br>Total yeast and mold count<br>(TYMC): <10 cfu/g | TAMC: <100 cfu/g<br>TYMC: <10 cfu/g | Same |
| Absence of | Absent | Absent | Same |
| Pathogenic<br>Organisms per<br>USP <62> | | | |
| | pjur® AQUA Baseline, pjur®<br>AQUA Guarana, pjur® BACK<br>DOOR anal Provitamin B5, and<br>pjur® AQUA ProVitamin B5<br>K200731<br>Subject Devices | Astroglide® Organix® Liquid<br>K200114<br>Predicate Device | Comparison |
| Antimicrobial<br>Effectiveness<br>Testing (USP<51><br>Category 2) | Yes | Yes | Same |
| Sterile | No | No | Same |
| Biocompatibility<br>Tested | Yes | Yes | Same |
| Condom<br>Compatibility | Not compatible with natural rubber<br>latex, polyisoprene and<br>polyurethane condoms | Compatible with natural rubber<br>latex and polyisoprene condoms.<br>Not compatible with polyurethane<br>condoms | Different |
| Packaging | Polyethylene bottles | Polyethylene terephthalate bottles<br>and foils | Different |
| Shelf Life | 18 months | 17 months (bottles)<br>9 months (foils) | Different |
Table 2: Comparison of Intended Use and technological characteristics of the subject and predicate devices:
{7}------------------------------------------------
{8}------------------------------------------------
The subject and predicate device indications for use are not identical due to differences in application area and condom compatibility. The subject device is intended to be used for penile, vaginal and/or anal applications while the predicate device is for penile and/or vaginal application. The subject device is not compatible with natural rubber latex (NRL), polyisoprene, and polyurethane condoms while the predicate device is compatible with NRL and polyisoprene condoms. These differences do not represent a new intended use as both the subject and predicate devices are intended to provide lubrication during intimate sexual activity.
The subject and predicate devices have different technological characteristics, including different formulations, packaging, specifications, shelf-life, and condom compatibility. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness.
## NON-CLINICAL TESTING/PERFORMANCE DATA:
### CONDOM COMPATIBILITY:
Condom compatibility testing was performed in accordance with ASTM D7661-10 "Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms." Testing demonstrated that the subject devices are not compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
### BIOCOMPATIBILITY:
Biocompatibility studies were performed in accordance with the 2020 FDA guidance document "Use of International Standard ISO 10993-1. Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" as follows:
- Cytotoxicity: ISO 10993-5:2009 ●
- . Sensitization: ISO 10993-10:2010
- Vaginal Irritation: ISO 10993-10:2010 ●
- Systemic Toxicity: ISO 10993-11:2017 .
The results of the testing show that the formulations of the subject devices are biocompatible.
{9}------------------------------------------------
## SHELF-LIFE:
pjur® AQUA Baseline, pjur® AQUA Guarana, pjur® BACK DOOR anal Provitamin B5, and pjur® AQUA ProVitamin B5 lubricants demonstrated a shelf-life of 18 months in accordance with the results of accelerated and real-time aged stability studies. Results from testing demonstrated that the device can maintain its specifications as shown in Table 1 over the duration of its shelf-life.
## CONCLUSION:
The results of the non-clinical testing described above demonstrate that the pjur® AQUA Baseline, pjur® AQUA Guarana, pjur® BACK DOOR anal Provitamin B5, and pjur® AQUA ProVitamin B5 personal lubricants are as safe and effective as the predicate and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.