GLISSANT Intimate Lubricant Fragrance Free/Sea Salt & Caramel is a personal lubricant, for penile and/or vaginal application, intended to moisturize and to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
Device Story
GLISSANT Intimate Lubricant is a water-based, non-sterile, colorless, viscous personal lubricant; intended for OTC use to supplement natural lubrication during sexual activity. Device is packaged in a polyethylene bottle with a lockdown pump. It does not contain contraceptives or spermicides. Users apply the lubricant topically to the penis or vagina. The lubricant enhances comfort and ease of sexual activity. It is compatible with natural rubber latex, polyisoprene, and polyurethane condoms. The device benefits patients by reducing friction and providing moisture during intimate activity.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing per ISO 10993-1 (cytotoxicity, systemic toxicity, sensitization, vaginal irritation) showed non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic results. Condom compatibility confirmed per ASTM D7661-10. Shelf life of 12 months validated via accelerated aging (ASTM F1980-16) with successful testing of physical/chemical properties (pH, viscosity, osmolarity) and microbial limits (USP <51>, <61>, <62>).
Technological Characteristics
Water-based, non-sterile, viscous liquid. Ingredients include organic aloe barbadensis leaf juice, potassium sorbate, hydroxyethylcellulose, stevia, flavor. pH 3.9–4.1; viscosity 250–500 cps; osmolarity 160–222 mOsm/kg. Packaged in polyethylene bottle with lockdown pump. Tested per ASTM D7661-10 for condom compatibility and ASTM F1980-16 for shelf life. Microbial testing per USP <51>, <61>, <62>.
Indications for Use
Indicated for individuals requiring a personal lubricant for penile and/or vaginal application to moisturize, lubricate, and enhance comfort during intimate sexual activity. Compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
Aloe Cadabra® Personal Lubricant and Aloe Cadabra® Flavored/Scented Personal Lubricants (K124044)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 8, 2019
Vjuvenate, LLC Karyn Eilber President. CEO 3435 Santa Monica Blvd. Suite 107 Santa Monica, CA 90405
Re: K183050 Trade/Device Name: GLISSANT Intimate Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: May 31, 2019 Received: June 7, 2019
#### Dear Karyn Eilber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon Andrews Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183050
Device Name GLISSANT Intimate Lubricant
#### Indications for Use (Describe)
GLISSANT Intimate Lubricant Fragrance Free/Sea Salt & Caramel is a personal lubricant, for penile and/or vaginal application, intended to moisturize and to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
Type of Use (Select one or both, as applicable)
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K183050 - Traditional 510(k) Premarket Notification GLISSANT Intimate Lubricant
## GLISSANT Intimate Lubricant – K183050
## 510(k) Summary
| Submitter | VJUVENATE, LLC<br>3435 Ocean Park Boulevard, Suite 107<br>Santa Monica, CA 90405<br>(424) 835-4372 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person | Karyn S. Eilber, MD<br>dreilber@vjuvenateyou.com<br>(424) 835-4372 Office<br>(310) 909-4967 Mobile |
| Date prepared | July 5, 2019 |
| Device trade names | GLISSANT Intimate Lubricant<br>GLISSANT Intimate Lubricant - Fragrance Free<br>GLISSANT Intimate Lubricant - Sea Salt & Caramel |
| Device common name | Personal lubricant |
| Regulation number | 21 CFR 884.5300 |
| Regulation name | Condom |
| Regulatory class | Class II |
| Product code | NUC |
| Classification panel | Obstetrics/gynecology |
| Predicate device | Aloe CadabraⓇ Personal Lubricant and Aloe CadabraⓇ<br>Flavored/Scented Personal Lubricants (K124044)<br><br>The predicate device has not been subject to a design related recall. |
| Indications for use | GLISSANT Intimate Lubricant Fragrance Free/Sea Salt & Caramel<br>is a personal lubricant, for penile and/or vaginal application,<br>intended to moisturize and lubricate, and to enhance the ease and<br>comfort of intimate sexual activity and supplement the body's<br>natural lubrication. This product is compatible with natural rubber<br>latex, polyisoprene, and polyurethane condoms. |
| Device description | GLISSANT Intimate Lubricant is a water-based, non-sterile,<br>colorless, viscous personal lubricant It does not contain a |
| | contraceptive or spermicidal component. The device is packaged in<br>a polyethylene bottle with a lockdown pump. The primary<br>packaging is plastic placed inside of a display box (dimensions<br>1.375" width x 1.375" depth x 4.75" height). |
{5}------------------------------------------------
The device specifications are listed in the table below:
| Parameter | Specification acceptance<br>criteria |
|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| Appearance | Clear, viscous |
| Color | Colorless |
| Odor | Odorless |
| pH | 3.9 – 4.1 |
| Specific gravity | 1.003 g/mL |
| Viscosity | 250 – 500 cps |
| Osmolarity | 160 – 222 mOsm/kg |
| Antimicrobial effectiveness | Per USP <51> (Category 2) |
| Total microbial count | Per USP <61> (<100 cfu/g) |
| Fungal/yeast/mold limits | Per USP <61> (<10 cfu/g) |
| Absence of pathogenic<br>organisms (e. coli, c.<br>salmonella sp., pseudomonas<br>aeruginosa, s. aureus,<br>clostridia sp., c. albicans) | Per USP <62> (absent) |
## SUBSTANTIAL EQUIVALENCE DISCUSSION
GLISSANT Intimate Lubricant has the same intended use when compared to the predicate device (Aloe Cadabra® Personal Lubricant and Aloe Cadabra® Flavored/Scented Personal Lubricants (K124044). The GLISSANT Intimate Lubricant has different technological characteristics compared to the predicate device; however, these differences do not raise different questions of safety or effectiveness. The following table compares GLISSANT™ Intimate Lubricant to the predicate device.
| | DEVICE | PREDICATE DEVICE |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | GLISSANT | ALOE CADABRA® |
| Trade Name | Intimate Lubricant<br>Fragrance free and<br>Sea Salt & Caramel<br>(K183050) | Personal Lubricant and<br>Flavored/Scented<br>Lubricants<br>(K124044) |
| Regulation number | 21 CFR 884.5300 | |
| Regulation name | Condom | SAME |
| Regulatory class | Class II | |
| Product code | NUC | |
| Indications for use | GLISSANT Intimate<br>Lubricant Fragrance<br>Free/Sea Salt and Caramel | Aloe Cadabra®<br>Lubricant and Aloe<br>Cadabra® |
| | is a personal lubricant, for<br>penile and/or vaginal<br>application, intended to<br>moisturize and lubricate,<br>and to enhance the ease<br>and comfort of intimate<br>sexual activity and<br>supplement the body's<br>natural lubrication. This<br>product is compatible with<br>natural rubber latex,<br>polyisoprene, and<br>polyurethane condoms. | Flavored/Scented<br>Lubricants are personal<br>lubricants, for penile<br>and/or vaginal<br>application, intended to<br>moisturize and lubricate,<br>to enhance the ease and<br>comfort of intimate<br>sexual activity and<br>supplement the body's<br>natural lubrication. This<br>product is compatible<br>with natural rubber latex<br>and polyisoprene<br>condoms. This product is<br>not compatible with<br>polyurethane condoms. |
| Device description | Water-based, non-sterile,<br>unscented and<br>flavored/scented personal<br>lubricant | Aloe-based, non-sterile,<br>flavored/scented<br>personal lubricant |
| Common ingredients | Organic aloe barbadensis<br>leaf juice, potassium<br>sorbate,<br>hydroxyethylcellulose,<br>stevia, flavor | Aloe |
| Biocompatibility tests<br>per ISO-10993 | Meets acceptance<br>requirements for all<br>biocompatibility tests | SAME |
| Condom compatibility<br>per ASTM D7661-10 | Compatible with natural<br>rubber latex, polyisoprene,<br>and polyurethane | Not compatible with<br>polyurethane condoms |
| Packaging | PETG plastic bottle with<br>lockdown pump packaged<br>in a carton | HDPE plastic bottle with<br>flip top cap packaged in<br>a carton |
| Shelf life | 1 year | 2 years |
| Stability tests per USP | Meets acceptance<br>requirements for microbial | SAME |
{6}------------------------------------------------
# SUMMARY OFNON-CLINICAL PERFORMANCE DATA
# Biocompatibility
{7}------------------------------------------------
The subject device was evaluated for biocompatibility in accordance with the 2016 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" and ISO 10993-1:2009 as follows:
- Cytotoxicity per ISO 10993-5:2009
- Acute systemic toxicity per ISO 10993-11:2006
- Sensitization per ISO 10993-10:2010 ●
- Vaginal irritation per ISO 10993-10:2010
The results of this testing demonstrated that the subject lubricants are non-cytotoxic, nonirritating, non-sensitizing, and non-systemically toxic.
### Condom compatibility
Condom compatibility was performed according to the FDA recognized consensus standard ASTM D7661-10 Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms. The results of this test indicate that JO subject lubricants are compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
### Shelf life (stability) testing
The shelf life of GLISSANT™ Intimate Lubricant is 12 months based on an accelerated aging study per ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
At baseline and following aging, the device met its specifications for appearance, color, odor, pH. specific gravity, viscosity, osmolarity, antimicrobial preservative effectiveness based on USP <51>, total aerobic microbial count and total yeast and mold count based on USP <61>, and absence of pathogenic organisms based on USP <62>.
### CONCLUSION
The results of the performance testing described above demonstrate that the GLISSANT Intimate Lubricants are as safe and effective as the predicate device and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.