JoyDrops® Natural Personal Lubricant Gel is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity, and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
Device Story
Water-based personal lubricant gel; packaged in 100 mL polymer bottle with pump dispenser. Applied topically by user to penis, vagina, or condom to reduce friction during sexual activity. Enhances comfort and supplements natural lubrication. Non-sterile; water-soluble; pH adjusted to ~7.0. No software or electrical components. Used in home setting by consumers.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993: in vitro cytotoxicity (no effects), systemic toxicity (no mortality/toxicity), vaginal irritation (non-irritant), and skin sensitization (no sensitization). Condom compatibility tested per ASTM D7661-10 and ARDL method; confirmed compatible with latex, polyisoprene, and polyurethane condoms.
Technological Characteristics
Water-based gel; ingredients include hydroxyethylcellulose, glycerin, mono propylene glycol, citric acid, phenoxyethanol, ethylhexylglycerin, and deionized water. pH ~7.0. Packaged in 100 mL plastic polymer bottle. Non-sterile. No electrical or software components.
Indications for Use
Indicated for penile and/or vaginal application to moisturize, lubricate, and enhance comfort during intimate sexual activity; supplements natural lubrication. Compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized design of three human profiles facing right, arranged in a cascading manner. The profiles are connected by flowing lines, creating a sense of unity and movement. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 17, 2015
Turkuaz Saglik Hizmetleri Medikal Temizlik Kimyasa % Michael Scott President ST&T Research 2237 Chestnut Street San Francisco, CA 94123
Re: K140506
Trade/Device Name: Joy Drops™ Natural Personal Lubricant Gel Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: March 13. 2015 Received: March 17, 2015
Dear Michael Scott,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number: K140506
#### Device Name: JoyDrops® Natural Personal Lubricant Gel
#### Indications For Use:
JoyDrops® Natural Personal Lubricant Gel is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity, and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
______________________________________________________________________________________________________________________________________________________________________________ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use __X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
## ST&T Research Departments:
PRODUCT DEVELOPMENT & CLINICAL RESEARCH DRUG-DEVICES-COSMETICS-SUPPLEMENTS-FOOD/BEVERAGE US AGENT FDA-FTC-EPA-CUSTOMS - & GRAS REGISTRATION LABELING & INTERNATIONAL REGULATORY CONSULTATION
PRE & POST- MARKET CONSUMER PRODUCT USE RESEARCH □INFORMATION& CONSULTING: 800-869-4636
Image /page/3/Picture/3 description: The image shows a logo for "S&T Research Intl." with the tagline "Science, Toxicology & Technology" underneath. The logo features a stylized "S" and "T" connected by a red bar at the top, with the "S" in black and the "T" partially in red. The words "RESEARCH INTL." are in a smaller font size and placed below the "S&T", followed by the tagline in an even smaller font.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS [21 CFR 807.92(c)]
## Date Revised: April 17, 2015
#### 510(k) Number: K140506
# Submitter:
Official Contact Person: Michael Scott, President. ST&T Research Intl. On Behalf of: Turkuaz Sağlik Hizmetleri Medikal Temizlik Kimyasal Urünler San. ve Tic. LTD ŞTI (Turkauz) Saadetdere Mah. 67 Sok No: 3 Esenyurt Istanbul TR-34513 Turkey Phone: 90-212-4286848 FAX: 90-212-4286853 Contacts: Nurhan Irmak
Device (807.92(a)(2): Trade or Proprietary Name: JoyDrops® Natural Personal
#### Lubricant Gel
Common or Usual Name: Personal Lubricant Classification Name: Condom Product Code: NUC Requlation Number: 21 CFR 884.5300 Device Class: Class II
Predicate Device [807.92(a)(3): Walgreens Personal Lubricating Jelly (K080978)
# Indications for Use
JoyDrops® Natural Personal Lubricant Gel is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity, and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms.
SCIENCE · TOXICOLOGY & TECHNOLOGY RESEARCH
□ SAN FRANCISCO (mailing address) • C/O P.O. BOX 470116 • SAN FRANCISCO, CALIFORNIA 94147 • (415) 441-2163
{4}------------------------------------------------
K140506 Turkuaz Medical Page 2
#### Device Description
JoyDrops® Natural Personal Lubricant Gel is a water-based personal lubricant gel that is contained in a 3.38 fluid ounces (100 mL) polymer container with a pump dispenser.
#### Directions for Product Use
Remove the cap of the JoyDrops® Natural Personal Lubricant Gel, and then twist the pump to open. Gently squeeze the tube to obtain the desired amount of lubricant. A 1 to 2 inch strip should be sufficient. May be applied directly onto the condom or onto the penis or into the vagina.
#### Product Performance and Condom Compatibility
JoyDrops® Natural Personal Lubricant Gel has been tested for performance by subjecting it to Condom Compatibility testing. Condom Compatibility was tested using the procedure described in the ASTM D7 6661-10 method and the ARDL method. JoyDrops® Natural Personal Lubricant Gel was found to be compatible with latex condoms, and polyisoprene and polyurethane condoms. This is the same as the statement made by Walgeens Personal Lubricant Jelly, which is the "substantially equivalent" comparison product.
## Summary of test data that confirm that JoyDrops® Natural Personal Lubricant Gel is safe for use. 807.92(b)(3)1
JoyDrops® Natural Personal Lubricant Gel has been tested for safety by subjecting it to cytotoxicity testing, systemic toxicity testing, vaginal irritation testing, and sensitization testing.
1. The cytotoxicity tests were conducted using the ISO 10993 methodology. No cellular toxicity was found.
2. The systemic toxicity testing was conducted using a standard protocol developed by NAMSA, and approved by FDA upon submission. No systemic toxicity was found.
3. Vaqinal irritation was studied using the International Organization for Standardization 10993-10, Biological Evaluation of Medical Devices. The test article was considered a nonirritant to vaginal tissue.
4. Skin sensitization patch tests in humans and skin sensitization tests in animals were conducted according to standard practices, including ISO 10993-10/AI. No skin sensitization was reported.
These tests are consistent with the expectations of this class of device, and are substantially equivalent to the predicate product.
# Biocompatibility
| Test | Result |
|--------------------------------------------------------|---------------------------------------|
| In Vitro Cytotoxicity | No Cytotoxic Effects Seen |
| Skin Irritation (Human Patch Test)<br>30 minutes | No Skin Irritation was Observed |
| Skin Irritation (Human Patch Test)<br>24 hours | No Skin Irritation was Observed |
| Skin Sensitization - Animals<br>(Guinea Pigs); 3 Weeks | No Skin Sensitization was<br>Observed |
| Vaginal Irritation<br>(Rabbits) | No Irritation Observed |
| Systemic Toxicity | No Mortality or Evidence of |
{5}------------------------------------------------
| (Mice); 7 Days | Systemic Toxicity |
|----------------|-------------------|
|----------------|-------------------|
The test sample extract was prepared according to ISO 10993-12. Table 1 "Standard surface areas and extract liquid volumes". Since the test sample is irregularly shaped solid devices (powder, pellets, foam, non-absorbent molded items) 0.2 g/ml extraction ratio was used. In order to use diluted extract of sample, 0.1 g/ml extraction ratio was also used. According to ISO 10993-12; 10.3.5 Extraction using both polar and non-polar extraction vehicles were performed.
The supplementation of basal culture media with animal serum of different origins is essential for cell growth, metabolism, and to stimulate proliferation ("mitogenic effect"). The major functions of serum in culture media are to provide (i) hormonal factors stimulating cell growth and proliferation and promoting differentiated functions, (ii) transport proteins carrying hormones (e.g. transcortin), minerals and trace elements (e.g. transferrin) and lipids (e.g. lipoproteins), (iii) attachment and spreading factors (Brunner et al., 2010, Altex 27; 1/10).
In order to support cell growth, 5% FBS was used until it was approximately 50% confluent. By adding extract of test materials the FBS concentration was decreased to 1% in order to continue cell growth and avoid cells to be overexposed to hormonal and growth factors according to the basic cell culture knowledge, since the percentage of the FBS has not been mentioned in the standard. According to our procedure, the extract condition includes 1% FBS.
The product is expected to be in contact with vaginal epithelial cells for less than 24 hours.
#### Packaging
JoyDrops® Natural Personal Lubricant Gel is packaged in 100 mL plastic polymer bottles. The bottle has a dispensing nozzle which in turn is covered by a plastic cap. The plastic cap has a tamper-resistant covering which must be removed in order to apply the gel. The bottle has a wrap-around label that meets the requirements of the FDA's labeling guidelines.
#### Technological Characteristics [807.92(a)(6)]:
Both the predicate device (Walgreens Personal Lubricating Jelly) and JoyDrops® Natural Personal Lubricant Gel have the same technological characteristics. They are both water-based gels that contain similar ingredients that allow vaginal lubrication and/or lubrication of condoms to ease the friction of sexual activity. Both lubricating gels are compatible with both latex-based condoms and common polymer-based condoms.
# Substantial Equivalence [807.92(b)(1)]:
JoyDrops® Natural Personal Lubricant Gel is substantially equivalent to a marketed device: Walgreens Personal Lubricating Jelly. The intended use, active (lubricating) ingredients, presence of preservatives, and application of the proposed device are substantially equivalent to those of the predicative device. Both products are/will be sold over-the-counter and are indicated as personal lubricants.
{6}------------------------------------------------
| | JoyDrops® Natural<br>Personal Lubricant Gel | Walgreens Personal<br>Lubricating Jelly |
|----------------------------------------------------|---------------------------------------------|-----------------------------------------|
| Manufacturer | Turkuaz | McNeil-PPC Inc |
| Intended use | Personal lubricant | Personal lubricant |
| Application | Topical | Topical |
| Over the counter use | Yes | Yes |
| Labeled water soluble | Yes | Yes |
| Labeled “condom compatible” | Yes | Yes |
| Contains Glycerin (Wetting Agent) | Yes | Yes |
| Contains<br>Hydroxyethylcellulose (Gel | Yes | Yes |
| Contains Preservative and<br>Buffering Ingredients | Yes | Yes |
| Contains alcohol | No | No |
| Contains fragrances | No | No |
| Container material | Plastic Polymer | N/A |
| Sterile | No | No |
# Product Characteristics
The primary characteristics of Joy Drops® Natural Personal Lubricant Gel include:
- □Lubricating agent
□Adjusted to a pH of ≈ 7.0
■Water soluble
_Chemically preserved to insure freshness
- □Non-sticky
□Gel consistency
_ Does not contain lipids
□Does not contain formaldehyde and salt
□Does not have fragrance or odor
# Ingredients and Specifications
| Ingredients | Quantity (W/V%) | CAS Number | Supplier |
|----------------------------------------|-----------------|---------------------|---------------------------------------|
| Hydroxy Ethyl<br>Cellulose | 1.0-2.0 | 9004-62-0 | Dow Chemical |
| Glycerin | 12.0-17.0 | 9004-62-0 | So Gis Industria<br>Chimica Spa Italy |
| Mono Propylene<br>Glycol | 1.0-5.0 | 57-55-6 | Sigma Aldrich<br>Germany |
| Citric Acid | 0.01-0.1 | 77-92-9 | Sigma Aldrich<br>Germany |
| Phenoxyethanol &<br>Ethylhexylglycerin | 0,05-0.2 | 122-99-6/70445-33-9 | Sigma Aldrich<br>Germany/ |
| Deionized water | q.s 100 | 7732-18-5 | |
{7}------------------------------------------------
K140506 Turkuaz Medical Page 5
#### Packaqinq
JoyDrops® Natural Personal Lubricant Gel is packaged in 100 mL plastic bottles with a dispensing nozzle and plastic cap.
## Stability 807.92(d)]:
JoyDrops® Natural Personal Lubricant Gel was placed in temperature control cabinets set at 22°C and 60°C. Based on the data obtained by an independent laboratory, a shelf-life of three years was established. Based on the data, microbiological specifications were set at <10 cfu/mL-g.
#### Software
There isn't any computer software that is required for the application of this product.
#### EMC and Electrical Safety
The device does not require EMC and Electrical safety evaluation.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.