Synthetic Polyisoprene Lubricated Male Condom - 5 senses
K182438 · Sxwell USA, LLC · MOL · Dec 6, 2018 · Obstetrics/Gynecology
Device Facts
Record ID
K182438
Device Name
Synthetic Polyisoprene Lubricated Male Condom - 5 senses
Applicant
Sxwell USA, LLC
Product Code
MOL · Obstetrics/Gynecology
Decision Date
Dec 6, 2018
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
Synthetic Polyisoprene Lubricated Male Condom - 5 Senses is used for contraception and for prophylactic purposes to help reduce the risk of pregnancy and the transmission of sexually transmitted infections (STI's).
Device Story
Synthetic polyisoprene male condom; fitted sheath with integral ring and reservoir tip. Lubricated with silicone gel containing sensate additives (cooling, tingling, warming). Used by individuals for contraception and STI risk reduction. Device acts as physical barrier to prevent exchange of bodily fluids. No electronic or software components.
Clinical Evidence
Bench testing only. Evaluated per ISO 23409:2011 and FDA guidance for non-latex condoms. Biocompatibility testing included cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), irritation (ISO 10993-10), and acute systemic toxicity (ISO 10993-11). Mechanical burst properties and shelf-life (3-year) were verified post-aging.
Technological Characteristics
Material: Synthetic polyisoprene rubber. Dimensions: 190±10mm length, 53±2mm width, 0.065-0.075mm thickness. Lubricant: Silicone oil with sensate additives. Dusting agent: Corn starch. Design: Straight wall with reservoir tip. Conforms to ISO 23409:2011. Non-electronic, no software.
Indications for Use
Indicated for individuals seeking contraception and protection against sexually transmitted infections (STIs).
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
Synthetic Polyisoprene Male Condom - Flavored (K171172)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
December 6, 2018
SXWELL USA LLC Robert Mahler Director, QA/RA North America and European Union 111 Wood Avenue South Iselin, New Jersey 08830
Re: K182438
Trade/Device Name: Synthetic Polyisoprene Lubricated Male Condom - 5 Senses Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: MOL Dated: August 14, 2018 Received: September 7, 2018
Dear Robert Mahler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon M. Andrews -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182438
Device Name
Synthetic Polyisoprene Lubricated Male Condom - 5 Senses
Indications for Use (Describe)
Synthetic Polyisoprene Lubricated Male Condom - 5 Senses is used for contraception and for prophylactic purposes to help reduce the risk of pregnancy and the transmission of sexually transmitted infections (STI's).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510k Summary K182438
#### Submitter:
SXWELL USA LLC. 111 Wood Avenue South Iselin, NJ 08830, USA
#### Contact Person:
Robert S. Mahler Director of Quality Assurance & Regulatory Affairs for North America and the European Union Phone: 732-345-2174 Rob.mahler@lifestyles.com
### Date Prepared:
December 4, 2018
### Device Name:
| Proprietary Name: | Synthetic Polyisoprene Lubricated Male Condom – 5 Senses |
|----------------------------|----------------------------------------------------------|
| Common Name: | Condom, Synthetic |
| Classification Name: | Condom |
| Classification Regulation: | 21 CFR 884.5300 |
| Device Class: | Class II |
| Product Code: | MOL (Condom, Synthetic) |
### Predicate Device:
K171172 – Synthetic Polyisoprene Male Condom - Flavored
The predicate device has not been subject to a design-related recall.
### Device Description:
The Synthetic Polyisoprene Lubricated Male Condom – 5 Senses is a male contraceptive and prophylactic device made from synthetic rubber polyisoprene latex with a lubricant coating containing silicone gel with flavor and sensate. The subject condom is provided in a variety pack which includes three variants; Cooling, Tingling, and Warming.
The condom is a fitted sheath with an integral ring at the open end and a reservoir (nipple end) at the closed end to contain semen. The condom dimensions are length 190±10mm, width 53±2mm, and thickness 0.065-0.075mm. The condom is designed to conform to the requirements of ISO 23409:2011. This product has a 3-year shelf-life.
{4}------------------------------------------------
## Indications for Use:
The Synthetic Polyisoprene Lubricated Male Condom – 5 Senses is used for contraception and for prophylactic purposes to help reduce the risk of pregnancy and the transmission of sexually transmitted infections (STI's).
## Technological Characteristics:
The comparison of technological characteristics between the Synthetic Polyisoprene Lubricated Male Condom - 5 Senses and the predicate device is shown in the table below:
| Device & Predicate Device(s): | K182438 (subject<br>device) | K171172 (predicate<br>device) | Comparison |
|-------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------|------------|
| Condom Material | Polyisoprene | Polyisoprene | Same |
| Dimensions | Length 190±10 mm<br>Nominal width 53±2<br>mm<br>Thickness 0.065-0.075<br>mm | Length 190±10 mm<br>Nominal width 53±2<br>mm<br>Thickness 0.065-0.075<br>mm | Same |
| Shape | straight wall with a<br>reservoir tip | straight wall with a<br>reservoir tip | Same |
| Texture | Smooth | Smooth | Same |
| Lubricant | Silicone oil | Silicone oil | Same |
| Color Additives | None | None | Same |
| Fragrance Additives | None | Yes | Different |
| Sensate Additives | Yes | None | Different |
| Dusting Agent | Corn starch | Corn starch (different<br>amount per device) | Different |
The subject and predicate devices have similar technological characteristics. The primary difference is the use of fragrance and sensate additives in the subject device. These differences in technological characteristics do not raise different questions of safety or effectiveness as compared to the predicate device.
## Performance Data:
Performance testing to assess the addition of new fragrant and sensate chemicals to the subject device was conducted in accordance with ISO 23409:2011, "Male Condoms -Requirements and test methods for condoms made from synthetic materials," and the FDA Guidance Document, "Testing guidance for Male Condoms Made from New Material (Non-Latex)." Results of testing showed that the mechanical properties of the subject device were equivalent to the predicate device.
Shelf-life testing was also conducted to demonstrate the subject device met its three-year shelf-life specification. The shelf life testing included an assessment of the mechanical burst properties of the subject device following aging.
{5}------------------------------------------------
In addition, biocompatibility testing was conducted in accordance with ISO 10993-1 to assess the addition of fragrant and sensate additives to the condom lubricant. The following biocompatibility testing was conducted:
- Cytotoxicity (Extraction Method, ISO 10993-5: 2009)
- Sensitization (Guinea Pig Maximization Sensitization, ISO 10993-10:2010)
- Irritation (Dermal, Vaginal, and Penile Irritation Studies in Rabbits, ISO 10993-10:2010) ●
- Acute Systemic Toxicity (Tests for Systemic Toxicity, ISO 10993-11: 2006) ●
The remainder of the performance testing outlined in the guidance document was deemed not needed to support the addition of fragrant and sensate chemicals to the condom lubricant.
### Conclusion:
The subject and predicate devices have the same intended use and similar fundamental technological characteristics. The performance data demonstrate that the subject device is substantially equivalent to the predicate device in terms of safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.