K100767 · Church & Dwight Co., Inc. · MOL · Apr 5, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K100767
Device Name
TROJAN SUPRA LUBRICATED POLYURETHANE MALE CONDOM
Applicant
Church & Dwight Co., Inc.
Product Code
MOL · Obstetrics/Gynecology
Decision Date
Apr 5, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
The TROJAN SUPRA® Lubricated Polyurethane Male Condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).
Device Story
TROJAN SUPRA® is a straight-walled, non-textured, nipple-end male condom made of polyurethane resin with a lubricated coating. It is an over-the-counter device intended for use by individuals to prevent pregnancy and STIs. The device functions as a physical barrier. It is manufactured in Japan. The device is not compatible with all lubricant types, and labeling reflects this limitation. Clinical performance was evaluated via a slippage and breakage study comparing the polyurethane condom to a standard latex condom control; results showed no statistically significant difference in clinical breakage or slippage rates. The device provides a barrier against viral penetration.
Clinical Evidence
Clinical slippage and breakage study conducted per FDA guidance with >290 couples. Subject polyurethane condom compared to standard latex control. Clinical breakage rate: 1.1% (polyurethane) vs 0.9% (control). Clinical slippage rate: 2.0% (polyurethane) vs 1.0% (control). Results indicated no statistically significant difference between the subject and control devices.
Technological Characteristics
Material: Polyurethane resin. Design: Straight-walled, non-textured, nipple-end, integral formed ring. Dimensions: 190 mm length, 58 mm flat-width. Standards: ASTM D 6324-08 (tensile, air burst, seal integrity), ISO 4074 (water leakage), ASTM D 624-00 (tear strength).
Indications for Use
Indicated for use by individuals for contraception and to help reduce the risk of pregnancy and sexually transmitted infections (STIs).
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
TROJAN-ENZ® Brand Lubricated Latex Male Condom (Pre-1976)
Submission Summary (Full Text)
{0}------------------------------------------------
Page 1 of 3
K100767
APR - 5 2011
April 4, 2011
## 510(k) Summary (revised)
| Submitted by: | Church & Dwight, Co. Inc.<br>469 North Harrison Street<br>Princeton, NJ 08543 |
|----------------------|-------------------------------------------------------------------------------|
| Contact Person: | Joseph Ciccone<br>Manager, Regulatory Affairs<br>(609) 497-7251 |
| Date Prepared: | March 17, 2010 |
| Proprietary Name: | TROJAN SUPRA® Lubricated Polyurethane Male Condom |
| Common Name: | Polyurethane Condom |
| Classification Name: | Condom [21 CFR §884.5300] |
TROJAN-ENZ® Brand Lubricated Latex Male Condom Predicate Device: Pre-1976
Description of Device: The TROJAN SUPRA® Lubricated Polyurethane Condom is a male condom consisting of a sheath of polyurethane with a lubricated coating. The condom is a straight-walled, nontextured, nipple-end condom with a nominal length of 190 mm and an nominal flat-width of 58 mm.
Intended Use of the Device: The 510(k)-subject condom has the same intended use as the predicate condom. The condom is used for contraception and for prophylactic purposes (intended to prevent pregnancy, HIV/AIDS, and other sexually transmitted infections, STIs).
Technological Characteristics: There is no difference in the basic technological characteristics of the 510(k)-subject condom and the predicate condom. Both the 510(k)-subject condom and the predicate condom are of the same basic design, both are straight-walled, nipple-ended, lubricated condoms with an integral formed ring at the open-end. The 510(k)-subject condom differs from the predicate condom in the material used to manufacture each condom: the predicate condom is made from natural rubber latex, the 510(k)-subject condom made from a polyurethane resin. With minor exceptions, the labeling for the 510(k)-subject device is the same as the predicate device. Exceptions to the predicate device labeling include: country of origin, the 510(k)-subject condom will be made in Japan; and absence of natural rubber latex warning for latex sensitive users, unlike the predicate device, the 510(k) subject device does not contain natural rubber.
Image /page/0/Picture/10 description: The image shows the logo for Trojan Brand Condoms. The word "TROJAN" is written in bold, sans-serif font. Below the word "TROJAN" is the phrase "BRAND CONDOMS" in a smaller font. To the right of the word "TROJAN" is a profile of a Trojan warrior's helmet.
Church & Dwight Co., Inc.
{1}------------------------------------------------
#### K100767
### April 4, 2011
lubricant type for 60 minutes at 37°C. Compatibility was assessed at three different aging conditions. Changes in airburst and tensile properties were evaluated compared to a control (subject condom conditioned but with no additional lubricant). The results of testing demonstrated that the subject condom is not compatible with any lubricant type. The subject condom is labeled accordingly.
Image /page/1/Picture/4 description: The image shows the logo for Trojan brand condoms. The logo features the word "TROJAN" in bold, black letters. Above the word is an image of a Trojan helmet. Below the word "TROJAN" are the words "BRAND CONDOMS".
Church & Dwight Co., Inc.
{2}------------------------------------------------
April 4, 2011
#### Summary of Studies
Pre-Clinical Safety Studies - Safety information regarding the compound materials and finished products (e.g., chemical degradation studies) was provided and raised no safety concerns. Biocompatibility studies on the non-lubricated condom and on the lubricated finished product include in vitro cytotoxicity extract test and direct contact test: sensitization test: vaginal irritation test: acute systemic toxicity; penile irritation test; 90-day muscle implantation study; bacterial reverse mutation assay; mouse lymphoma assay; and in vitro chromosomal aberration study.
Clinical In-Use Slip/Break Study - A slippage and breakage study following a protocol prepared to meet the FDA guidance: "Clinical Testing Guidance for New Material Male Condoms," was conducted using the 510(k)-subject condom with a standard latex condom serving as control. Over 290 couples were included in the study, the clinical breakage rate for the polyurethane condom was 1.1% compared to 0.9% for the control. The clinical slippage rate was 2.0% for the polyurethane and 1.0% for the control. The polyurethane condom was statistically no different from the control latex condom in both clinical breakage and slippage rate.
Physical testing data – Followed the FDA guidance: "Clinical Testing Guidance for New Material Male Condoms" and included the testing of three (3) lots of the 510(k)-subject condom. Testing included tensile strength, force at break, and elongation and air burst volume and air burst pressure were performed according to ASTM D 6324-08, and water leakage tests were carried out for the same lots in accordance with testing procedure of ISO 4074 (AQL = 0.25). In addition, condoms were tested for tear resistance and propagation according to ASTM D 624-00 ("Standard Test Methods for Tear Strength of Conventional Vulcanized Rubber and Thermoplastic Elastomers"). Further testing to simulate in-use conditions by performing tensile strength, force at break, and elongation at 37° C, 50% Relative Humidity were completed according to ASTM D 6324-08. The results of physical testing support the performance effectiveness of the 510(k) product.
Package integrity - An evaluation of seal integrity was performed on three lots (3) of the 510(k)-subject device according to ASTM D 6324-08 with satisfactory results.
Barrier Properties/Permeability: Viral Penetration Study - Was performed on three (3) test lots of the 510(k)-subject polyurethane condom and one lot of the commercially available predicate device. Results of testing on a fourth lot of the 510(k)-subject device were submitted at the request of FDA to confirm the prior testing. The cumulative results of the studies demonstrate the barrier effectiveness of the 510(k) device to viral penetration under conditions of the in vitro study.
Shelf-life - Stability of the 510(k)-subject device was established from results of physical testing data using a protocol that followed 21 CFR 801.435 as a guide. Based on the evaluation of the results of the physical testing data, the expiring date has been set at 60 months.
Lubricant Compatibility - The subject condom was evaluated for compatibility with water-based, silicone oil-based, and glycol-based personal lubricants. Condom samples were exposed to two brands of each
(continued...)
Image /page/2/Picture/12 description: The image shows a black circle on the left side of the image. The circle is solid black and appears to be a simple shape. On the right side of the image, there is some text, but it is not clear enough to read. The text is in a dark font and is positioned near the bottom right corner of the image.
Image /page/2/Picture/13 description: The image shows the logo for Trojan Grand Condominiums. The logo features the word "TROJAN" in bold, sans-serif font, with the word "GRAND CONDOMINIUMS" in a smaller font below it. To the right of the word "TROJAN" is a stylized image of a Trojan warrior's head, complete with a helmet.
Church & Dwight Co., Inc.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular pattern around the symbol. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Joseph Ciccone Manager, Regulatory Affairs Church & Dwight Co., Inc. 469 North Harrison Street Law Department - Building 100 PRINCETON NJ 08543
APR - 5 2011
Re: K100767
> Trade Name: TROJAN SUPRA® Lubricated Polyurethane Male Condom Regulation Number: 21 CFR §884.5300 Regulation Name: Condom Regulatory Class: II Product Code: MOL Dated: March 23, 2011 Received: March 23, 2011
Dear Mr. Ciccone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Hubert Lenoir us
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
| 510(k) Number: | K100767 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | TROJAN SUPRA® Lubricated Polyurethane Male Condom |
| Indications for Use: | The TROJAN SUPRA® Lubricated Polyurethane Male Condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs). |
| Prescription Use | |
| | OR Over-the-Counter Use <span style="text-decoration: underline;">X</span> |
| (Per 21 CFR §8001.109) | |
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and Urological Devices
510(k) Number: K100767
.
Image /page/5/Picture/2 description: The image shows the logo for Trojan brand condoms. The logo features the word "TROJAN" in large, bold letters, with the words "BRAND CONDOMS" in smaller letters underneath. To the right of the text is a silhouette of a Trojan warrior's helmet. The logo is simple and recognizable, and it is often used in advertising and marketing materials for Trojan condoms.
( (
(
Church & Dwight Co., Inc.
-
-
·
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.