The Microthin Natural Rubber Latex Condom is used for contraception and prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
Device Story
Male contraceptive device; natural rubber latex membrane sheath; straight-walled; smooth surface; reservoir tip; lubricated with silicone oil. Used by individuals for contraception and STI prevention. Device covers penis; provides physical barrier. Benefits include pregnancy prevention and reduced transmission risk of STIs.
Clinical Evidence
Bench testing only. Physical testing (airburst) on three lots met ASTM D3492-16 and ISO 4074:2015 standards. Shelf-life testing (5 years) met 21 CFR 801.435 requirements. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), irritation (ISO 10993-10), and acute systemic toxicity (ISO 10993-11).
Technological Characteristics
Material: Natural rubber latex. Dimensions: 170mm length, 52mm width, 0.035mm thickness. Lubricant: Silicone oil. Standards: ASTM D3492-16, ISO 4074:2015. Shelf-life: 5 years. No colorants or flavors.
Indications for Use
Indicated for contraception and prevention of pregnancy and sexually transmitted infections in individuals using male condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
April 28, 2022
Suretex Limited % Carole C. Carey Senior Consultant C3-Carey Consultants, LLC 9451 Ellsworth Court Fulton, MD 20759
K213921 Re:
> Trade/Device Name: Microthin Natural Rubber Latex Condom Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: March 21, 2022 Received: March 28, 2022
Dear Carole C. Carey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213921
Device Name
Microthin Natural Rubber Latex Condom
Indications for Use (Describe)
The Microthin Natural Rubber Latex Condom is used for contraception and prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for LifeStyles. The logo consists of a shield with a gradient of colors ranging from yellow to orange to red. To the right of the shield is the word "LifeStyles" in a dark blue, sans-serif font. Below the word "LifeStyles" is the phrase "SINCE 1905" in a smaller, sans-serif font.
# 510(k) Summary - K213921
## Submitter:
Suretex Limited 31/1 Moo 4, Suratthani-Thakuapha Road Tambon Khao Hua Kwai, Amphur Phunphin Suratthani 84130 THAILAND
# Contact Person:
KC Nguyen Vice President, R&D, QA and RA Phone: +66 77 277 400 Fax: +66 77 277 428 Email: kc.nguyen@lifestyles.com
### Date Prepared:
April 26, 2022
### Device Name:
Trade Name: Common Name: Regulation Name: Regulation Number: Device Class: Product Code:
Microthin Natural Rubber Latex Condom Male Natural Rubber Latex Condom Condom 21 CFR 884.5300 Class II HIS (Condom)
### Predicate Device:
K192669, Extremely Thin 003, ZERO ZERO THREE (Okamoto U.S.A, Inc.)
The predicate device has not been subject to a design-related recall.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for LifeStyles. The logo consists of a shield-shaped emblem on the left and the word "LifeStyles" in blue on the right. Below the word "LifeStyles" is the phrase "SINCE 1905" in a smaller font. The shield emblem has a gradient of colors, including yellow, orange, red, and pink.
# Device Description:
The Microthin Natural Rubber Latex Condom is a male contraceptive made of natural rubber latex that completely covers the penis with a closely fitted membrane sheath. The condom is provided lubricated with silicone oil. The condom is straight-walled, with a smooth surface and a reservoir (nipple end) at the closed end to contain semen. The condom dimensions are length 170mm ± 10mm, width 52 ± 2mm, and thickness 0.035mm ± 0.005mm. The condom conforms to the recognized standards ASTM D3492-16 Standard Specification for Rubber Contraceptives (male condoms) and ISO 4074:2015 Natural Rubber Latex Condoms - Requirements and test methods. The condom has a 5-year shelf-life.
### Indications for Use:
The Microthin Natural Rubber Latex Condom is used for contraception and prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
#### Comparison of Intended Use and Technological Characteristics:
The table below compares the intended use and technological characteristics of the subject and predicate devices:
| | Subject Device<br>K213921 | Predicate Device<br>K192669 | Comparison |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Microthin Natural<br>Rubber Latex Condom<br>is used for<br>contraception and<br>prophylactic purposes<br>(to help prevent<br>pregnancy and the<br>transmission of<br>sexually transmitted<br>infections). | The condom is used for<br>contraception and for<br>prophylactic purposes<br>(preventing<br>transmission of<br>sexually transmitted<br>infections). | The subject and<br>predicate<br>indications for<br>use are not<br>identical, but<br>both have the<br>same intended<br>use (i.e., to<br>prevent<br>pregnancy and<br>transmission of<br>sexually<br>transmitted<br>infections). |
| Condom<br>Material | Natural Rubber Latex | Natural Rubber Latex | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Lifestyles. On the left is a shield with a gradient of yellow, orange, and red. To the right of the shield is the word "LifeStyles" in a dark blue font, with the words "SINCE 1905" in a smaller font below the word "LifeStyles".
| Nominal<br>Width | $52 \pm 2$ mm | $53.2 \pm 2$ mm | Different |
|-------------------------------|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Nominal<br>Length | $170 \pm 10$ mm | $180 \pm 10$ mm | Different |
| Nominal<br>Thickness | $0.035 \pm 0.005$ | 30 mm from closed end:<br>$0.039 \pm 0.0013$<br>90 mm from closed end:<br>$0.039 \pm 0.001$<br>150 mm from closed<br>end: $0.039 \pm 0.001$ | Different |
| Lubricant | Silicone | Silicone | Same |
| Color<br>Additives | No color | No color | Same |
| Flavor<br>Additives | No flavor | No flavor | Same |
| Shape | Straight-walled<br>Reservoir-ended | Straight-walled<br>Reservoir-ended | Same |
| Texture | Smooth Surface | Smooth Surface | Same |
| Bursting<br>Pressure<br>(kPa) | Minimum of 1 kPa | $1.42 \pm 0.10$ | Different |
| Bursting<br>Volume<br>(dm3) | Minimum of 18 L | $35.4 \pm 3.83$ | Different |
| Shelf-life | 5 years | 5 years | Same |
As shown in the table above, the subject and predicate devices do not have identical indications for use statements; however, they do have the same intended use (i.e., to prevent pregnancy and transmission of sexually transmitted infections).
The technological characteristics of the subject and predicate device are similar in that they are natural rubber latex-based, do not contain colorants or flavors, have the same shape and texture, and have the same shelf-life. Differences between the subject and predicate device include dimensions of the condoms (e.g., width, length, and thickness) and burst specifications. These differences in technological characteristics do not raise different questions of safety and effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Lifestyles. The logo consists of a shield on the left and the word "LifeStyles" on the right. The shield is red, orange, and yellow. The word "LifeStyles" is blue, and below it is the text "SINCE 1905".
## Summary of Non-Clinical Performance Testing:
The following testing has been performed to support substantial equivalence to the predicate device:
#### Physical Testing
Three lots of the subject device were tested at baseline and met the airburst specifications of ASTM D3492-16 Standard Specification for Rubber Contraceptives (male condoms) and ISO 4074:2015 Natural Rubber Latex Condoms – Requirements and test methods.
#### Shelf-Life Testing
Shelf-life testing was conducted to support a shelf-life of five years per the requirements of 21 CFR §801.435. All samples met the predefined acceptance criteria.
#### Biocompatibility Testing
Biocompatibility testing was conducted in accordance with the 2020 FDA quidance document "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process." Testing included the following assessments:
- Cytotoxicity (ISO 10993-5:2009) ●
- Sensitization (ISO 10993-10:2010) ●
- Irritation (ISO 10993-10:2010) ●
- Acute systemic toxicity (ISO 10993-11:2017) ●
The device was demonstrated to be biocompatible.
#### Conclusion:
The results of the performance testing described above demonstrate that the Microthin Natural Rubber Latex Condom is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.