For rapid transcervical aspiration of the uterine cavity.
Device Story
Berkeley VC-10 Vacuum Curettage System provides suction for uterine content removal. System comprises electrically powered vacuum pump console, collection canisters, tubing, and sterile single-use vacurettes. Clinician operates system via foot actuator pedal to control vacuum supply. Vacuum pressure draws uterine contents through vacurette into collection container; tissue traps or gauze sacks collect tissue samples. Device used in clinical settings. Modification involves upgrading vacuum pump motor from 1/3 HP to 1/2 HP to maintain performance. Output is physical aspiration of uterine cavity, facilitating clinical procedures. Benefits include efficient, rapid uterine evacuation.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included maximum vacuum pressure, free flow rate, time to develop/reach full vacuum, and evacuation time for 1L using 6mm and 7mm curettes. All tests met pre-specified acceptance criteria.
Technological Characteristics
System includes stainless steel cabinet, double diaphragm pump, collection canisters, and PVC tubing. Electrical power: 115VAC/60Hz. Vacuum performance: 71 cm Hg max vacuum. Standards: AAMI/ANSI ES 60601-1:2005/(R) 2012 and A1:2012 (Electrical Safety); IEC 60601-1-2 Edition 3: 2007-03 (EMC). Vacurettes are sterile, single-use disposable. Console is multi-use, non-sterile.
Indications for Use
Indicated for rapid transcervical aspiration of the uterine cavity. No specific patient population, age, or contraindications are defined.
Regulatory Classification
Identification
A vacuum abortion system is a device designed to aspirate transcervically the products of conception or menstruation from the uterus by using a cannula connected to a suction source. This device is used for pregnancy termination or menstrual regulation. This type of device may include aspiration cannula, vacuum source, and vacuum controller.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle with three human profiles incorporated into its design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 29, 2017
Gyrus ACMI, Inc. Christina Flores Senior Specialist, Regulatory Affairs 136 Turnpike Road Southborough, MA 01772
Re: K171440
Trade/Device Name: Berkeley VC-10 Vacuum Curettage System Regulation Number: 21 CFR& 884.5070 Regulation Name: Vacuum Abortion System Regulatory Class: II Product Code: HHI, HHK Dated: August 31, 2017 Received: September 1, 2017
Dear Christina Flores:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Charles Viviano -S
For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171440
Device Name Berkeley VC-10 Vacuum Curettage System
Indications for Use (Describe) For rapid transcervical aspiration of the uterine cavity.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K171440
# 510(k) Summary Gyrus ACMI, Inc. Berkeley VC-10 Vacuum Curettage System
## General Information
| 510(k) Submitter | Gyrus ACMI, Inc. 136 Turnpike Rd Southborough, MA 01772 Phone: 508-804-2776 |
|------------------------------------|-------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 3003790304 |
| Contact Person: | Christina Flores Senior Specialist, Regulatory Affairs 508-804-2776 Christina.flores@olympus-osta.com |
| Date Prepared: | September 27, 2017 |
**Device Identification**
| Classification Name: | Vacuum Abortion System |
|----------------------|--------------------------------------------------------------------------------------|
| Regulation number | 21 CFR 884.5070 |
| Product codes | HHI (System, Abortion, Vacuum) HHK (Curette, Suction, Endometrial (And Accessories)) |
| Regulatory Class | Class II |
| Trade Name: | Berkeley VC-10 Vacuum Curettage System |
| Generic/Common Name: | Vacuum abortion system |
**Predicate Device**
| K030935 | Gyrus ACMI VC-10 |
|---------|------------------|
|---------|------------------|
The predicate device has not been subject to a design-related recall.
# Product Description
The Berkeley VC-10 Vacuum Curettage System uses aspiration to remove uterine contents through the cervix. The system includes the vacuum pump, collection containers, connection tubing, and patient contacting vacurettes. The vacuum pump
{4}------------------------------------------------
is electrically powered and generates suction through vacuum pressure. The vacuum pump console is not provided sterile and is a multi-use device. The patient contacting vacurettes of the VC-10 system are provided as sterile, single use disposable devices.
Similar to the predicate device VC-10, the modified Berkeley VC-10 includes a high capacity. double diaphragm pump and motor, enclosed in a stainless steel cabinet. Collection canisters sit on top of the cabinet, and a large vacuum gauge and vacuum control knob are easily accessed on the top of the cabinet. A vacurette (available in various shapes and sizes) is attached to a collection set (includes a plastic swivel handle with a slip ring to control the vacuum and PVC tubing) and is attached to the patient port of the first collection container. Two accessories, which can be used to collect tissue, include a tissue trap, which is connected to the collection tubing and the collection container, or a gauze sack, which is placed inside the collection container. The modified VC-10 offers the same foot actuator pedal for vacuum supply.
## Indications for Use
For rapid transcervical aspiration of the uterine cavity.
The indications for use and intended use of the modified/unmodified Berkeley VC-10 Vacuum Curettage System has not changed as a result of the vacuum pump motor modification in this submission. The indications for use of the subject and predicate devices are identical.
# Comparison to Predicate Device:
The modified Berkeley VC-10 has been compared to the predicate VC-10 device with respect to intended use, design, performance, and fundamental scientific technology.
The Berkeley VC-10 was originally cleared in K030935. The modification that is the subject of this submission is to the vacuum pump and motor in the console of the Berkeley VC-10 Vacuum Curettage System. The horsepower rating for the predicate and new vacuum pump is changing from 1/3 horsepower to 1/2 horsepower. All other components of the system are identical to those cleared in K030935. The system's interface with the disposable components has not changed with this modification.
No changes were made to patient contacting material or packaging to the modified VC-10 since the predicate system was cleared in K030935.
A comparison of the pump specifications for the modified Berkeley VC-10 system and the predicate Berkeley VC-10 system is shown in the table below:
{5}------------------------------------------------
| | VC-10 (K030935) | VC-10 (this submission) |
|-------------------|-------------------|-------------------------|
| Motor part number | 72R547-V116-D303X | 87R647-101-N470X |
| Voltage/Frequency | 115VAC/60 Hz | Same |
| Power/Amps | 600VA/5A | 500VA/T6.3AH |
| Horsepower | 1/3 HP | 1/2 HP |
| Maximum Vacuum | 73 cm Hg | 71 cm Hg |
Except for the replacement of a the vacuum pump motor and the adjustments made to the internal enclosure to accommodate the new vacuum pump motor, the modified Berkeley VC-10 system and unmodified Berkeley VC-10 are identical. These minor differences do not raise different questions of safety or effectiveness. As demonstrated in the non-clinical testing (e.g., electrical safety, EMC, and functional testing), the different technological characteristics do not affect the safety and effectiveness of the subject devices. The comparisons and summary of testing results presented in this Special 510(k) Notification show this device to be substantially equivalent to the predicate VC-10 device.
## Summary of Biocompatibility, Sterilization and Shelf Life Discussion
The subject of the modification is the pump motor in the non-patient contacting console of the VC-10 system. The sterile, single use disposable devices that were cleared with the system under K030935 have not been modified since receiving FDA clearance. Therefore, the current submission relies on the biocompatibility, sterilization and shelf life information provided in K030935.
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical Safety was tested according to AAMI/ANSI ES 60601-1:2005/(R) 2012 and A1:2012.
EMC was tested according to IEC 60601-1-2 Edition 3: 2007-03.
# Summary of Bench. Performance Testing
The following bench performance tests were conducted to assess the functional performance of the modified device:
- Maximum Measured Vacuum Pressure ●
- Free Flow Rate
- Time to Develop Vacuum
- Time to Reach Full Vacuum ●
{6}------------------------------------------------
- Time to Evacuate 1L w/ 7mm Curette .
- . Time to Evacuate 1L w/ 6mm Curette
All performance testing passed or met the pre-specified acceptance criteria.
# Conclusion
The modified Berkeley VC-10 has the same intended use, fundamental technology and similar design as its predicate VC-10 device. The technological differences do not raise different questions of safety or effectiveness and the performance testing demonstrates substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.