The Elvie Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Elvie Pump is intended for a single user.
Device Story
Elvie Pump is a battery-powered, portable, electric breast pump for lactating women. It integrates a pump motor, breast shield, and milk collection bottle into a single unit worn inside a nursing bra. The device uses piezoelectric technology to generate negative pressure, expressing milk into the collection bottle. It operates hands-free without external cords or tubes. Users control the pump via physical buttons or a mobile companion app. The app provides real-time monitoring of milk volume, battery life, pumping time, and history. The device benefits patients by enabling discreet, hands-free milk expression in home settings. Healthcare providers do not operate the device; it is for patient self-use.
Clinical Evidence
No clinical data. Evidence consists of bench testing: biocompatibility (ISO 10993-5, -10), electrical safety (IEC 60601-1, -11), electromagnetic compatibility (IEC 60601-1-2), usability testing (IEC 62366-1), cleaning validation, and performance testing (vacuum/backflow). Software validation included cybersecurity assessment and validation of the milk measurement algorithm.
Technological Characteristics
Piezoelectric pump mechanism; rechargeable Li-Polymer battery; food-grade materials (21 CFR 174-179); portable form factor for bra-integration. Connectivity via mobile app. Software-controlled via microcontroller. Sterilization/cleaning: reusable components with validated cleaning instructions. Standards: ISO 10993, IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 62366-1.
Indications for Use
Indicated for lactating women to express and collect breast milk. Intended for single-user, home-use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 10, 2018
Chiaro Technology Limited % Marc C. Sanchez, Esq. Regulatory Attorney Contract In-House Counsel and Consultants, LLC 53516 Bickett Chapel Hill, NC 27517
Re: K181863 Trade/Device Name: Elvie Pump Regulation Number: 21 CFR& 884.5160 Regulation Name: Powered breast pump Regulatory Class: II Product Code: HGX Dated: July 11, 2018 Received: July 12, 2018
Dear Marc C. Sanchez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael T. Bailey -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181863
Device Name Elvie Pump
Indications for Use (Describe)
The Elvie Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Elvie Pump is intended for a single user.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510k Submission Elvie Pump
### 510(K) Summary - K181863
The following information is provided as required by 21 CFR 807.87 for the Elvie Pump 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence is based.
| Sponsor: | Chiaro Technology Limited<br>63-66 Hatton Garden<br>London EC1N 8LE United Kingdom<br>Establishment Registration: 3012098706 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Jabil Circuit (Shanghai) LTD.<br>600 Tian Lin Road, Shanghain, China 200233<br>Establishment Registration: 3005210579 |
| Contact: | Marc C. Sanchez, Esq.<br>Contract In-House Counsel and Consultants, LLC<br>53516 Bickett Chapel Hill NC 27517<br>Ph: 202.765.4491<br>E-mail: msanchez@fdaatty.com |
| Date Prepared: | October 9, 2018 |
| Proprietary Name: | Elvie Pump |
| Common Name: | Powered breast pump |
| Regulation Number: | 21 CFR 884.5160, Powered breast pump |
| Regulatory Class: | Class II |
| Product Code: | HGX (pump, breast, powered) |
| Predicate Device(s): | Medela Freestyle (K150499) |
The predicate device has not been subject to a design related recall.
### Device Description:
The Elvie Pump is an electric breast pump system comprised of a single or double portable unit that integrates the pump body and milk collection bottle. It consists of a pump (motor unit) and includes a breast shield, bottle, seal, valve, spout and bra adjustor. All components (minus the pump) are reusable and may be manually cleaned. The kit components may be purchased separately and are available in different sizes and configurations.
{4}------------------------------------------------
Traditional 510k Submission Elvie Pump
The Elvie Pump includes piezoelectric pump technology which generates negative pressure on the nipple to express milk, which is collected in the integrated milk collection bottle. It is designed to work in the user's nursing bra and has a rechargeable battery so it can be used hands-free without external power cords or milk collection tubes.
The Elvie Pump is a battery-powered electro-mechanical device that contains software. It can be controlled through the physical interface on the device or through a mobile companion app, which also provides real-time milk monitoring, pump battery life, pumping time elapsed and pumping history information.
All milk contacting components are constructed out of food grade materials that are compliant with 21 CFR 174-179.
### Indications for Use:
The Elvie Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Elvie Pump is intended for a single user.
The subject and predicate device have the same intended use.
### Comparison of Technological Characteristics:
The Elvie Pump uses the same fundamental technology and mode of action as the predicate device, the Medela Freestyle powered electric breast pump (K150499). The table below summarizes the key specifications of the Elvie Pump and the predicate.
| Attribute | Predicate Device<br>Medela Freestyle | Subject Device<br>Elvie Pump |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K150499 | K181863 |
| Product code | HGX | HGX |
| Classification Name | Pump, Breast, Powered | Pump, Breast, Powered |
| Indications for Use | The Freestyle is a powered breast pump<br>to be used by lactating women to<br>express and collect milk from their<br>breasts. The Freestyle is intended for a<br>single user. | The Elvie Pump is a powered breast<br>pump to be used by lactating women<br>to express and collect milk from their<br>breasts. The Elvie Pump is intended<br>for a single user. |
| Intended Use | Express milk from breast | Express milk from breast |
{5}------------------------------------------------
Traditional 510k Submission
| | Yes | Elvie Pump |
|------------------------------|----------------------------------------------|---------------------------------|
| Single User Device | Yes | Yes |
| Single / Double<br>Pumping | Both | Both |
| Patient Population | Lactating women | Lactating women |
| Environment of<br>Use | Home | Home |
| Over the Counter<br>(OTC) | Yes | Yes |
| Power Source | AC adaptor or Rechargeable Li-Ion<br>Battery | Rechargeable Li-Polymer Battery |
| Adjustable Suction<br>Levels | Yes | Yes |
| Backflow<br>Protection | Yes | Yes |
| Suction Level<br>(mmHg) | 40-245 | 40-220 |
| Cycles per<br>Second | 1.7-1.93 | 1.1-2.0 |
| Cycles per<br>Second | 0.83-1.36 | 0.6-1.0 |
| Two-phase<br>Expression | Yes | Yes |
| User Interface | Pump body | Pump body and/or mobile app |
| Software | Yes | Yes |
| Control<br>Mechanism | Microcontroller | Microcontroller |
| Portable | Yes | Yes |
| Milk Quantity<br>Measurement | No | Yes |
As evidenced by the table above, there are differences in technological characteristics between the subject and predicate device. These differences in technological characteristics do not raise different questions of safety or effectiveness.
{6}------------------------------------------------
#### Summary of Non-Clinical Test Reports:
The following performance tests were conducted on the Elvie Pump device:
- . Biocompatibility testing demonstrating that the patient contacting device components are biocompatibility, including the following specific tests:
- Cytotoxicity per ISO 10993-5 O
- Irritation per ISO 10993-10 O
- o Sensitization per ISO 10993-10
- . Software validation for a minor level of concern per the FDA guidance document "Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005, including cybersecurity assessment and validation of the milk measurement algorithm
- Simulated use testing (usability study) per IEC 62366-1 Application of Usability Engineering to Medical Devices
- . Use life testing to support the proposed use life
- Electrical safety testing per IEC 60601-1 and IEC 60601-1-11 ●
- Electromagnetic compatibility testing per and IEC 60601-1-2
- Cleaning validation on the reprocessing instructions included in the labeling
- . Device performance testing including evaluation of vacuum performance and backflow demonstrating that the subject device meets its specifications
### Conclusion:
The Elvie Pump is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.