Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Light Absorbency; Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Regular & Super; Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Super Plus; Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Super Absorbency
K151470 · First Quality Hygienic, Inc. · HEB · Nov 2, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K151470
Device Name
Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Light Absorbency; Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Regular & Super; Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Super Plus; Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Super Absorbency
Applicant
First Quality Hygienic, Inc.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Nov 2, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 884.5470
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Unscented tampon for women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the placement of the tampon correctly into the vagina.
Device Story
Compact applicator menstrual tampon; consists of viscose fiber strip with inserted cord, pressed into shape; covered with non-woven fiber layer; housed in polyethylene (PE) applicator tubes; wrapped in polypropylene (PP). Used by women for personal hygiene; facilitates intra-vaginal absorption of menstrual/vaginal discharge. Plastic applicator eases correct vaginal placement. Device is non-sterile, over-the-counter product. Benefits include hygienic management of menstrual flow.
Clinical Evidence
No clinical data. Bench testing only. Absorbency tested per 21 CFR 801.430 (Syngina method) meeting 90:90 ranges. Chemical residue testing (TCDD/TCDF, pesticides) showed no detectable levels. Extractables tested per ISO 10993-12/18. Withdrawal cord strength met AS/NZS 2869:1998 (min 22.4N). Fiber shed evaluated against predicate. Biocompatibility (cytotoxicity, irritation, systemic toxicity) and microbiological testing (bioburden, S. aureus bacteriostatic effect, TSST-1 production, vaginal microflora impact) confirmed safety profile equivalent to predicate.
Technological Characteristics
Viscose fiber tampon with hydrophobic cotton cord; non-woven PET/PE cover. Applicator: Low Density Polyethylene (LDPE) with food-grade colorants, mold release, and antistatic agents (conforming to 21 CFR 177-178). Compact design with mushroom-shaped tip and straight grooves. Non-sterile. Packaging: Polypropylene wrapper.
Indications for Use
Indicated for women for personal hygiene, specifically for the intra-vaginal absorption of menstrual or other vaginal discharge.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2015
First Quality Hygienic, Inc. % Dave Yungvirt Official Correspondent 45 Rockerfeller Plaza, Suite 2000 New York, NY 10111
Re: K151470
> Trade/Device Name: Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Light Absorbency; Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Regular & Super; Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Super Plus; Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Super Absorbency Regulation Number: 21 CFR 884.5470 Regulation Name: Unscented menstrual tampon Regulatory Class: II Product Code: HEB Dated: October 14, 2015 Received: October 19, 2015
Dear Dave Yungvirt,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151470
#### Device Name
Opal by Femtex and Private Label Unscented Compact Applicator Tampon - Light Absorbency, Super Absorbency, and Super Plus Absorbency
#### Indications for Use (Describe)
Unscented tampon for women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the placement of the tampon correctly into the vagina.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 5 - 510(k) Summary
### 1. Submitter's name and address:
First Quality Hygienic, Inc. 121 North Road McElhattan, PA 17748
#### 2. Submitter's telephone number and fax number:
Tel: 516-498-3665 Fax: 516-829-4949
#### 3. Contact person:
Melanie Leibowitz Director of Regulatory Affairs
### 4. Date this 510(k) summary prepared: 04/23/2015
### 5. Regulatory Description:
| Trade name: | Opal by Femtex and Private Label Unscented Compact<br>Applicator Tampon |
|-------------------------|-------------------------------------------------------------------------|
| Common Name: | Unscented Menstrual Tampon |
| Classification Name: | Tampon, Menstrual, Unscented |
| Regulation Description: | Unscented Menstrual Tampon |
| Regulation Number: | 21 CFR 884.5470 |
| Class: | II |
| Product Code: | HEB |
### 6. Legally marketed device to which substantial equivalence is claimed:
Tosama Hygiene Tampon (Tosama Maxim Compact Plastic Applicator Tampon), K080775
### 7. Description of the device:
The Opal by Femtex and Private Label Unscented Compact Applicator Tampons is an unscented, compact applicator menstrual tampon. The tampon itself is made of a strip of viscose fibers in which a cord is inserted and pressed firmly together. The viscose tampon is covered with a thin layer of non-woven fiber material which is melted on the strip. The applicator tubes are made from polyethylene (PE). The tampon with applicator is
{4}------------------------------------------------
wrapped in printed polypropylene (PP) wrapping. The tampon is marketed in Light, Regular, Super, and Super Plus Absorbencies.
### 8. Intended use and indication for use:
For over-the-counter use: Unscented tampon for women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the placement of the tampon correctly into the vagina.
### 9. Summary of technological characteristics compared to the predicate device:
| | Opal by Femtex and Private Label Unscented Compact<br>Applicator Tampon | Tosama Hygiene Tampon<br>K080775 |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification<br>Regulation | 884.5470 | 884.5470 |
| Product Code | HEB | HEB |
| Intended Use | Unscented tampon for women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the placement of tampon correctly into the vagina. | Unscented tampon for women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the placement of tampon correctly into the vagina. |
| Sterile? | No | No |
| Design | Compact tampon with mushroom-shaped tip and straight grooves | Compact tampon with mushroom-shaped tip and straight grooves |
| Absorbency, grams | Light $\leq$ 6.0<br>Regular 6.0 - 9.0<br>Super 9.0 - 12.0<br>Super Plus 12.0 - 15.0 | N/A<br>Regular 6.0 - 9.0<br>Super 9.0 - 12.0<br>Super Plus 12.0 - 15.0 |
| Pledget Length<br>Dimensions (mm) | Light 40.0 - 45.0<br>Regular 40.0 - 45.0<br>Super 43.0 - 46.0<br>Super Plus 43.0 - 46.0 | N/A<br>Regular 45.0 - 49.0<br>Super 45.0 - 49.0<br>Super Plus 46.0 - 50.0 |
| Pledget Diameter<br>Dimensions (mm) | Light 8.7 – 9.7<br>Regular 8.7 – 9.7<br>Super 11.0 – 13.0<br>Super Plus 13.0 - 15.0 | N/A<br>Regular 11.0 - 13.0<br>Super 11.0 - 13.0<br>Super Plus 13.0 - 15.0 |
| Applicator Inner Tube<br>Length x Diameter | Light 66.0 - 68.0 x<br>10.5 - 11.5 | N/A |
{5}------------------------------------------------
| Dimensions (mm) | | Regular | 66.0 - 68.0 x<br>10.5 - 11.5 | Regular | 66.5 x 18.0 |
|--------------------------------------------------------------|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| | | Super | 64.0 - 66.0 x<br>13.5 - 14.5 | Super | 66.5 x 18.0 |
| | | Super Plus | 65.5 - 68.5 x<br>15.5 - 16.5 | Super Plus | 66.5 x 21.4 |
| Applicator OuterTube<br>Length x Diameter<br>Dimensions (mm) | | Light | 58.0 x 60.0 x<br>13.0 - 14-0 | N/A | |
| | | Regular | 58.0 x 60.0 x<br>13.0 - 14-0 | Regular | 59.5 x 18.9 |
| | | Super | 59.0 x 61.0 x<br>15.0 - 16.0 | Super | 59.5 x 18.9 |
| | | Super Plus | 60.0 x 62.0 x<br>17.0 - 18.0 | Super Plus | 59.5 x 22.7 |
| Cotton Cord | | 100% Hydrophobic Cotton | | 100% Hydrophobic Cotton | |
| Nonwoven | PET/PE - conforms with 21<br>CFR 177-178 | | PET/PE - conforms with 21<br>CFR 177-178 | | |
| | Plastic | Low Density Polyethylene | | Low Density Polyethylene | |
| Applicator<br>Tube | Colorant | Four new colorants, for Green,<br>Pink, Blue, and Pearl White<br>applicator tubes; New colorants<br>have the following<br>characteristics:<br>• Meets Toxics Substance Control Act<br>Inventory (TSCA)<br>• 21 CFR 178.3297 and/or EU<br>Resolution (89)1 compliant;<br>Extractables testing confirmed that<br>no harmful compounds are<br>bioavailable<br>• conforms with 21 CFR 177-178 | | Pearl White colorant with the<br>following characteristics:<br>• Meets Toxics Substance Control Act<br>Inventory (TSCA)<br>• 21 CFR 178.3297 and/or EU<br>Resolution (89)1 compliant;<br>Extractables testing confirmed that<br>no harmful compounds are<br>bioavailable<br>• conforms with 21 CFR 177-178 | |
| | Mold | Food grade mold release | | Food grade mold release | |
| | Release | agent—conforms with 21 CFR | | agent—conforms with 21 CFR | |
| | Agent | 177-178 | | 177-178 | |
| | Antistatic<br>Agent | Food grade antistatic agent—<br>conforms with 21 CFR 177-178 | | N/A | |
| Wrapper (Primary<br>Packaging) | | Polypropylene | | Polypropylene | |
{6}------------------------------------------------
### 10. Non-Clinical Performance Data
The following testing was performed:
### Absorbency
Absorbency testing for the new device was performed per the 21 CFR 801.430 for Syngina absorbency testing. Data for each absorbency were then analyzed for 90% population distribution at 90% confidence, as the regulation directs, and it was confirmed that 90:90 ranges of Light, Regular, Super, and Super Plus absorbencies met the absorbency range requirements defined by 21 CFR 801.430.
### Chemical residues
2.3.7.8-tetrachlorodibenzo-p-dioxin (TCDD) and 2.3.7.8-tetrachlorofuran dioxin (TCDF) testing was performed on the new device per European Community (EC) Regulation 589/2014 (for food) and EC Regulation 709/2014 (for feed). No TCDD or TCDF was detectable by this method.
Pesticide (including herbicides) testing was performed on the new device per method ASU L00.00-34. No pesticides (including herbicides) were detected.
Extractables testing was performed on the new device according to ISO 10993-12: 2012, Biological evaluation of medical devices – Part 12: Sample preparation and reference materials, and ISO 10993-18: 2005, Biological evaluation of medical devices - Part 18: Chemical characterization of materials. No extractable chemicals were detectable in the aqueous extraction, which most closely simulated physiological conditions. Very low levels of non-toxic extractables were obtained in harsh, non-physiologically relevant organic extraction conditions.
### Withdrawal cord attachment test
Withdrawal cord attachment strength testing for the new device was performed per Australian Standard AS/NZS 2869:1998, Tampons-Menstrual. Each tampon tested for every absorbency level met the Standard's minimum attachment strength requirement of 22.4 Newtons (N). Each absorbency data set met an average value of ≥ 28N, which is the Standard's requirement for each production lot.
### Fiber Shed Testing
{7}------------------------------------------------
Fiber shed testing for the new device was evaluated using an internal fiber shed test protocol, since no recognized standard is available. The new device does not exhibit the potential to shed more fiber than the predicate device.
# Biocompatibility Testing
Biocompatibility assessment for the new device was evaluated in accordance with the May 1, 1995 FDA Blue Book Memorandum #G95-1, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, ' and ISO 10993-1:2003, Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing Within a Risk Management Process. Biocompatibility requirements were assessed for the tampon as having "limited contact" (i.e., < 24 hrs) and as a surface device, contacting the skin and mucous membrane. The battery of testing included the following tests:
- . Cytotoxicity
- Intracutaneous irritation ●
- . Acute systemic toxicity
# Pre-clinical microbiological testing
Preclinical microbiology evaluation was performed for the new device in as recommended by the July 27, 2005 Guidance, Guidance for Industry and FDA Staff, Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s). The following pre-clinical microbiology tests were considered:
- . Bioburden
Bioburden was evaluated for the new device per European Pharmacopeia 5.1.4, Microbiological Quality of Pharmaceutical Preparations, Category 2. No more than 29 colony forming units (CFU) of aerobic bacteria were identified on any sample tested. Furthermore, zero yeasts and no more than 5 mold CFUs were found on any sample tested. Finally, no Pseudomonas aeruginosa, Staphylococcus aureus, and enterobacteria or other gram negative bacteria were detected.
- Bacteriostatic effect on Staphylococcus aureus
Effect of the new device on growth of Staphylococcus aureus was evaluated using adaption of method United States Pharmacopeia (USP) 30, Chapter 51 (2007): Antimicrobial Effectiveness Testing. The test article did not demonstrate the potential to enhance S. aureus growth.
{8}------------------------------------------------
- Effect on normal vaginal microflora growth
Effect of the new device on the growth of normal vaginal microflora was evaluated per USP 31, Chapter 71(2008): Sterility Tests. The new device did not demonstrate the potential to inhibit the growth of normal vaginal microflora.
- Effect on production of S. aureus Toxic Shock Syndrome Toxin-1 (TSST-1)
The new device was evaluated to determine its impact on the production of TSST-1 by S. aureus. Evaluation was performed using a Tampon Sac Method with TSST-1 production detected by enzyme-linked immunosorbant assay (ELISA) and subsequent plate counting to enumerate CFUs. The new device does not demonstrate the potential to augment TSST-1 relative to current legally marketed tampons.
# 11. Conclusion
The Opal by Femtex and Private Label Unscented Compact Applicator Tampon characteristics and comparisons detailed above demonstrate that the new device has the same fundamental design, Indications for Use, technology, and performance specifications as the predicate device. Non-Clinical Performance Testing demonstrates that the new device meets the absorbency requirements defined by 21 CFR 801.430, which confirms the device's effectiveness for the Indications for Use. Biocompatibility testing, Pre-Clinical Microbiology testing, and Withdrawal Cord Attachment Strength and Fiber Shed testing demonstrate that the new device has a safe biocompatibility profile, does not increase the potential to promote TSS relative to the predicate device, and has a tampon integrity profile that is equivalent to the predicate device. Furthermore, Extractables testing demonstrates that the applicator and tampons do not have the potential to expose the user to harmful chemicals. Altogether, then, the data support the conclusion that the Opal by Femtex and Private Label Unscented Compact Applicator Tampon is substantially equivalent to the predicate device and is safe and effective for its Indications for Use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.