The Aqueduct 200 Cervical Dilation Balloon Catheter is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal, Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.
Device Story
Aqueduct 200 is a balloon cervical dilation catheter; enables simultaneous dilation of cervical canal. Input: visual data from integrated forward-looking CMOS camera and LED light source. Operation: physician inserts catheter through vagina/cervix into uterus; uses visual feedback to position dilation balloon at internal/external cervical os; inflates balloon with 12.5 ml saline for mechanical dilation. Output: visual feed to external monitor via HDMI; mechanical dilation of cervix. Used in clinical settings by physicians. Benefits: provides visual guidance for accurate placement and controlled mechanical dilation of cervix, facilitating various gynecological procedures. Procedure duration 6-7 minutes.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-5, -10), sterilization validation (ISO 11135), electrical safety/EMC (IEC 60601-1, -1-2, -2-18), optical performance, and mechanical bench performance (ISO 10555-1, -4). Accelerated aging studies confirmed shelf-life performance.
Technological Characteristics
Materials: Pebax balloon, polycarbonate catheter, polyvinyl chloride hub, stainless steel. Sensing: CMOS camera with LED light source. Energy: Electrical (camera/LED). Dimensions: 25 cm length, 3.3 mm (deflated) to 12 mm (inflated) diameter. Connectivity: HDMI output to monitor. Sterilization: Ethylene oxide (SAL 10-6). Standards: ISO 10993, ISO 11135, IEC 60601-1, ISO 10555.
Indications for Use
Indicated for patients requiring cervical softening and dilation for procedures including cervical stenosis treatment, IUD placement/removal, Radium placement, uterine cavity drainage, endometrial biopsy, uterine curettage, suction cannula aspiration, and operative hysteroscopy.
Regulatory Classification
Identification
A hygroscopic Laminaria cervical dilator is a device designed to dilate (stretch open) the cervical os by cervical insertion of a conical and expansible material made from the root of a seaweed (Laminaria digitata or Laminaria japonica). The device is used to induce abortion.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
October 15, 2021
GTIMD LLC Eran Levit CEO 6 Columbia Drive Amherst, NH 03031
Re: K202433
> Trade/Device Name: Aqueduct 200 Cervical Dilation Balloon Catheter Regulation Number: 21 CFR§ 884.4260 Regulation Name: Hygroscopic Laminaria Cervical Dilator Regulatory Class: II Product Code: PON Dated: September 9, 2021 Received: September 13, 2021
Dear Eran Levit:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202433
Device Name
Aqueduct 200 Cervical Dilation Balloon Catheter
Indications for Use (Describe)
The Aqueduct 200 Cervical Dilation Balloon Catheter is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal, Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | X |
|----------------------------------------------|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "GTIMD CATHETER SOLUTIONS". The text "GTIMD" is in a large, sans-serif font and is colored in a light blue. Below this, the text "CATHETER SOLUTIONS" is in a smaller, sans-serif font and is colored in gray. The logo is simple and modern, and the colors are clean and professional.
K202433 Page 1 of 7
# 510(K) SUMMARY - K202433
| I. | Submitter | |
|----|----------------------|-------------------------------------|
| | Submitter's Name: | GTIMD LLC |
| | Address: | 6 Columbia Drive, Amherst, NH 03031 |
| | Phone: | (603) 809-3089 |
| | Contact Person: | Eran Levit |
| | Contact Email: | elevit@gtimd.com |
| | Date of Preparation: | October 14, 2021 |
- II. Device
| Name of Device: | Aqueduct 200 Cervical Dilation Balloon Catheter |
|--------------------|---------------------------------------------------------|
| Common Name: | Catheter, Balloon, Dilation of the Cervical Canal |
| Regulation Number: | 21 CFR 884.4260 |
| Regulation Name: | Hygroscopic Laminaria Cervical Dilator |
| Regulatory Class: | II |
| Product Codes: | PON (Catheter, Balloon, Dilation of the Cervical Canal) |
- III. Predicate and Reference Devices
### Predicate Device:
Aqueduct 100 Cervical Dilation Balloon Catheter (K160664) manufactured by GTIMD LLC
The predicate device has not been subject to a design-related recall.
### Reference Device:
Endosee Model 8000 U-scope (K123151) manufactured by Endosee Corp
- IV. Device Description
The Aqueduct 200 Cervical Dilation Balloon Catheter is a balloon cervical dilation catheter which enables the simultaneous dilation of both sides of the cervical canal. The subject device is an updated version of the Aqueduct 100 Cervical Dilation Balloon Catheter cleared under K160664 that merges two separate dilation balloons into one balloon, removes the anchor balloon, replaces the four-way hub with a three-way hub, and removes the stylet from the shaft.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "GTIMD CATHETER SOLUTIONS". The text "GTIMD" is in a large, sans-serif font and is colored in a light blue. Below this, the text "CATHETER SOLUTIONS" is in a smaller, sans-serif font and is colored in gray.
The catheter consists of a 2-lumen shaft. One lumen inflates a cylindrical dilation balloon. A second lumen is for infusion of saline solution and also contains a camera and camera cable. In use, the catheter is inserted through the vagina and cervical canal and into the uterus. Using visual feedback from the forward-looking camera and LED light source the physician advances the catheter through the cervical canal and identifies the end of the canal when entering the internal orifice of the uterus. When the physician identifies the orifice, the catheter is positioned inside the cervix. This aligns the cylindrical dilation balloon at the internal cervical os and external cervical os. The balloon is inflated with 12.5 ml of saline, providing gradual mechanical dilation of the cervix.
The camera cable on the catheter connects to a camera module that provides power to the camera and allows connection to an HDMI port on a monitor to assess proper device placement during a clinical procedure. After 3 minutes of dilation of the internal and external orifices of the uterus, a controlled injection of 1-2.5 ml of saline mav be made through the catheter infusion lumen which exits on the distal end of the catheter. Evidence that cervical dilation is complete can be determined once droplets of the saline injection are observed exiting through the external opening of the cervix. Optimal dilation of 8-9 mm within the cervical canal is typically achieved following 5 minutes of balloon dilation. The entire procedure from catheter insertion to removal is completed in 6-7 minutes.
The deflation of the balloon is conducted by attaching a luer lock syringe to the swabable valve and removing saline from the balloon.
The subject device is packaged in a mylar/Tyvek pouch and ethylene oxide (EO) sterilized to a SAL 10-6. The packaged device has a shelf-life of 15 months.
- V. Indications for Use
The Aqueduct 200 Cervical Dilation Balloon Catheter is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal. Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for "GTIMD CATHETER SOLUTIONS". The text "GTIMD" is in a large, bold, sans-serif font and is colored in a bright blue. Below "GTIMD" is the text "CATHETER SOLUTIONS" in a smaller, sans-serif font and is colored in gray.
- VI. Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below provides a comparison of the intended use and technological characteristics of the subject and predicate device:
| | Subject Device | Predicate Device |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Aqueduct 200 Cervical<br>Dilation Balloon Catheter<br>(K202433) | Aqueduct 100 Cervical<br>Dilation Balloon Catheter<br>(K160664) |
| Manufacturer | GTIMD LLC | GTIMD LLC |
| Regulatory<br>Information | Hygroscopic Laminaria<br>Cervical Dilator (21 CFR<br>884.4260)<br>Product Code: PON<br>Class II | Hygroscopic Laminaria<br>Cervical Dilator (21 CFR<br>884.4260)<br>Product Code: PON<br>Class II |
| Indications for<br>Use | The Aqueduct 200 Cervical<br>Dilation Balloon Catheter is<br>intended to be used<br>whenever cervical softening<br>and dilation is desired. Some<br>examples are: treatment of<br>cervical stenosis, IUD<br>placement and removal,<br>Radium placement, drainage<br>of uterine cavity, endometrial<br>biopsy, uterine curettage,<br>suction cannula aspiration,<br>operative hysteroscopy. | The Aqueduct 100 Cervical<br>Dilation Balloon Catheter is<br>intended to be used<br>whenever cervical softening<br>and dilation is desired. Some<br>examples are: treatment of<br>cervical stenosis, IUD<br>placement and removal,<br>Radium placement, drainage<br>of uterine cavity, endometrial<br>biopsy, uterine curettage,<br>suction cannula aspiration,<br>operative hysteroscopy. |
| Components | Balloon catheter with an<br>incorporated camera and<br>LED light source, and camera<br>module | Balloon catheter |
| Balloon<br>volume(s) | Dilation Balloon: 12.5 mL | Anchor Balloon: 2.5 mL<br>Dilation Balloons: 7.5 mL<br>(combined) |
| Balloon burst<br>pressure | Dilation Balloon: > 8 atm | Anchor Balloon: > 2 atm<br>Dilation Balloon: > 2 atm |
| Usable Length | 25 cm | 25 cm |
| Outer diameter | 3.3 mm deflated and 12 mm | 3.3 mm deflated and 12 mm |
| | inflated | inflated |
| Shaft type | 2-lumen polycarbonate shaft | 3 lumen Pebax shaft |
| Materials | Pebax balloon, polyvinyl<br>chloride hub, polycarbonate<br>catheter, stainless steel | Pebax balloons and Pebax<br>catheter |
| Dilation<br>Mechanism | Fixed diameter balloon that<br>expands to 12 mm | Fixed diameter balloons that<br>expand to 12 mm |
| Stylet | No | Yes, unremovable |
| Shelf Life | 15 months | 2 years |
| Dilation Balloon<br>Positioning<br>Method | Optical using a CMOS camera<br>and LED light | Mechanical using an anchor<br>balloon |
| Saline flow? | Yes, single lumen for infusion | Yes, single lumen for infusion |
| Sterile/Reusable? | Single-use sterile device and<br>reusable camera module | Single-use sterile device |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for "GTIMD CATHETER SOLUTIONS". The text "GTIMD" is in a large, bold, sans-serif font and is colored in a bright blue. Below "GTIMD" is the text "CATHETER SOLUTIONS" in a smaller, sans-serif font and is colored in gray.
The indications for use (IFU) for the subject device are identical to the predicate device IFU. Therefore, the intended use is the same – to dilate the cervix.
The subject device has different technological characteristics than the predicate device. The subject and predicate device have different device design specifications, number of balloons, materials, and positioning method.
The camera is used to position the dilation balloon, similar to the anchor balloon in the predicate device. Both features are used to aid in device placement.
These differences in technological characteristics do not raise different questions of safety and effectiveness as compared to predicate device (e.g., can the device be positioned correctly, are the materials biocompatible, can device use result in an injury to cervical or uterine tissues, etc.). Non-clinical performance data was conducted to address the technological differences between the subject and predicate device to demonstrate substantial equivalence.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "GTIMD CATHETER SOLUTIONS". The text "GTIMD" is in a large, sans-serif font and is colored in a bright blue. Below "GTIMD" is the text "CATHETER SOLUTIONS" in a smaller, sans-serif font and is colored in gray.
#### VII. Performance Data
To support the substantial equivalence of the subject device, the following performance testing was performed:
### Biocompatibility Testing
Biocompatibility tests were conducted on the Aqueduct 200 Cervical Dilation Balloon Catheter according to the 2020 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" as follows:
- Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2010)
- Irritation (ISO 10993-10:2010) .
The results of this testing demonstrated that the subject device is noncytotoxic, non-irritating, and non-sensitizing.
### Sterilization Validation
The subject device is ethylene oxide sterilized. The ethylene oxide sterilization cycle was validated using the overkill method, in accordance with ANSI/AAMI/ISO 11135:2014 Sterilization of health care products - Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices. Ethylene oxide residuals were evaluated according to ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals.
Packaging integrity and shipping testing were completed after accelerated aging to demonstrate the subject device packaging was able to withstand the rigors of shipping and distribution. Testing included seal strength (ASTM F88/F88M), dye penetration (ASTM F1929), shipping testing (ASTM D4169), and a visual inspection. The subject device met all the predetermined acceptance criteria.
## Electrical Safety and Electromagnetic Compatibility (EMC)
The subject device was tested for electrical safety and EMC according to the following standards:
- . IEC 60601-1:2005 (3rd Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
{8}------------------------------------------------
K202433 Page 6 of 7
Image /page/8/Picture/1 description: The image shows the logo for "GTIMD CATHETER SOLUTIONS". The text "GTIMD" is in a large, bold, blue font. Below that, the text "CATHETER SOLUTIONS" is in a smaller, gray font. The logo is simple and modern.
6-B Columbia Drive Amherst, NH 03031 USA Phone: (603) 809-3089 www.gtimd.com
- IEC 60601-1-2:2014, Medical Electrical Equipment Part 1-2: General ● Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
- 60601-2-18: Edition 3.0 2009-08, Medical electrical equipment Part 2-● 18: Particular requirements for the basic safety and essential performance of endoscopic equipment
### Optical Performance
The camera was evaluated for the following specifications and met the predetermined acceptance criteria:
- Component specifications
- Photobiological safety testing
- Field of view
- Direction of view
- Resolution
- Depth of field and optimum working distance
- Geometric distortion
- Noise and dynamic range
- Image intensity uniformity ●
### Bench Performance Testing
Performance tests were performed on the Aqueduct 200 Cervical Dilation Balloon Catheter according to the requirements of ISO 10555-1:2013 and ISO 10555-4:2013 Intravascular catheters — Sterile and single-use catheters — Part 1: General requirements and Part 4: Balloon dilation catheters, and ISO 11737-1:2018 Sterilization of health care products —Microbiological methods Part 1: Determination of a population of microorganisms on products.
Performance testing was also conducted on the Aqueduct 200 Cervical Dilation Balloon Catheter following an accelerated aging study to simulate a 15-month shelf life, including:
- Dimensional Verification ●
- Camera Functionality ●
- Balloon Burst Pressure ●
- Balloon Fatigue
- Balloon Inflation/Deflation Time
- Catheter Bond Strength ●
- Catheter Flexibility and Pushability ●
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for "GTIMD CATHETER SOLUTIONS". The letters "GTIMD" are in a large, bold, sans-serif font and are colored in a bright blue. Below the letters, the words "CATHETER SOLUTIONS" are in a smaller, sans-serif font and are colored in gray.
- Balloon Preparation, Deployment, and Retraction ●
The subject device met the predetermined acceptance criteria for each test.
- VIII. Conclusion
Based on the comparison and analysis above, the Aqueduct 200 Cervical Dilation Balloon Catheter is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.