K212659 · Agency For Medical Innovations GmbH · PMU · Nov 10, 2021 · Obstetrics/Gynecology
Device Facts
Record ID
K212659
Device Name
More-Cell-System
Applicant
Agency For Medical Innovations GmbH
Product Code
PMU · Obstetrics/Gynecology
Decision Date
Nov 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4050
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K212659 · Nov 10, 2021
More-Cell-System
Agency For Medical Innovations GmbH
Observational clinical case series
An observational study of 26 clinical cases was conducted to evaluate the usability of the More-Cell-System in women undergoing laparoscopic gynecologic morcellation in routine clinical practice.
Usability; Observational study; Clinical cases
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Observational usability study; Observational study
Women undergoing laparoscopic gynecologic morcellation; Sample Size: 26 clinical cases; Number of Sites: 22 sites
Not applicable for this study
Ease of bag insertion/removal, operating time, ease of instrument use, ability to maintain distension/inflation, visualization, and device integrity (leak testing)
Indications for Use
More-Cell-System is intended for use as a tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue considered benign, resected during multi-site laparoscopic surgery for power morcellation and removal. When used in women with fibroids, the More-Cell System is for women who are pre-menopausal and under age 50. More-Cell-System is compatible with electromechanical laparoscopic power morcellators that are between 12 mm and 20 mm in shaft outer diameter and between 95 mm and 170 mm in shaft working length.
Device Story
More-Cell-System is a tissue containment system for gynecologic laparoscopic surgery. It consists of a single-use, insufflatable polyurethane bag (2.5L capacity) and a Visi-Shield (polycarbonate window/metal sleeve) for laparoscope protection. During multi-site surgery, the bag is inserted into the abdomen, filled with resected tissue, and insufflated with CO2. The Visi-Shield protects the camera while allowing visualization. A morcellator is inserted through the bag's large opening to morcellate tissue under visual control. After morcellation, the bag is closed by tying knots in the neck and removed. Used by surgeons in clinical settings to isolate tissue and prevent potential spread of unsuspected malignancy during power morcellation. Benefits include containment of tissue and fluids during extraction.
Clinical Evidence
Evidence includes an in vivo porcine study (n=16) comparing the device to conventional surgery; results showed no disseminated cells in the MCS group vs. 5/8 positive cytology in controls (p=0.0256). Usability studies included an observational study (n=26 cases) and a U.S. simulated-use study (n=11 surgeons) on pelvic trainers. No use errors or leaks were reported. Bench testing confirmed impermeability to E. coli and mechanical integrity after 2-year accelerated aging.
Technological Characteristics
Materials: 50 µm polyurethane film, polyester thread, stainless steel, polycarbonate, silicone. Sensing/Actuation: Mechanical containment. Energy: None (passive). Connectivity: None. Sterilization: EtO. Shelf life: 2 years. Design: Two-opening bag (16 cm and 16 mm) with Visi-Shield optic sleeve.
Indications for Use
Indicated for women undergoing minimally invasive gynecologic laparoscopic surgery for benign tissue resection and power morcellation. For fibroid cases, restricted to pre-menopausal women under age 50. Contraindicated for patients with known/suspected malignancy, post-menopausal patients, patients over 50, or candidates for en-bloc removal.
Regulatory Classification
Identification
A gynecologic laparoscopic power morcellation containment system is a prescription device consisting of an instrument port and tissue containment method that creates a working space allowing for direct visualization during a power morcellation procedure following a laparoscopic procedure for the excision of benign gynecologic tissue that is not suspected to contain malignancy.
Special Controls
In combination with the general controls of the FD&C Act, the Gynecologic Laparoscopic Power Morcellation Containment System is subject to the following special controls:
- 1. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 2. Device components that are labeled sterile must be validated to a sterility assurance level of 10-°.
- 3. Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components, package integrity, and device functionality over the intended shelf life.
- 4. Non-clinical performance data must demonstrate that the device meets all design specifications and performance requirements. The following performance characteristics must be tested:
- a. Demonstration of the device impermeability to tissue, cells and fluids.
- b. Demonstration that the device allows for the insertion/withdrawal of laparoscopic instruments while maintaining pneumoperitoneum.
- c. Demonstration that the containment system provides adequate space to perform morcellation and adequate visualization of the laparoscopic instruments and tissue specimen relative to the external viscera.
- d. Demonstration that intended laparoscopic instruments and morcellators do not compromise the integrity of the containment system.
- Demonstration that intended users can adequately deploy the device, morcellate a e. specimen without compromising the integrity of the device and remove the device without spillage of contents.
- 5. Training must be developed and validated to ensure users can follow the instructions for use.
- 6. Labeling must include:
- . Contraindication for use in gynecologic surgery in which the tissue to be morcellated is known or suspected to contain malignancy.
- Unless clinical performance data demonstrates that it can be removed or modified, a contraindication for removal of uterine tissue containing suspected fibroids in patients who are: peri- or post-menopausal; or candidates for en bloc tissue removal, for example, through the vagina or via a mini-laparotomy incision.
- . The following boxed warning: "Warning: Information regarding the potential risks of a procedure with this device should be shared with patients. Uterine tissue may contain unsuspected cancer. The use of laparoscopic power morcellators during fibroid surgery may spread cancer. The use of this containment system has not been clinically demonstrated to reduce this risk."
- Statement limiting use of device to physicians who have completed the training ● program.
- An expiration date or shelf life. ●
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible;
(2) Device components that are labeled sterile must be validated to a sterility assurance level of 10
−6 ;(3) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components, package integrity, and device functionality over the intended shelf life;
(4) Non-clinical performance data must demonstrate that the device meets all design specifications and performance requirements. The following performance characteristics must be tested:
(i) Demonstration of the device impermeability to tissue, cells, and fluids;
(ii) Demonstration that the device allows for the insertion and withdrawal of laparoscopic instruments while maintaining pneumoperitoneum;
(iii) Demonstration that the containment system provides adequate space to perform morcellation and adequate visualization of the laparoscopic instruments and tissue specimen relative to the external viscera;
(iv) Demonstration that intended laparoscopic instruments and morcellators do not compromise the integrity of the containment system; and
(v) Demonstration that intended users can adequately deploy the device, morcellate a specimen without compromising the integrity of the device, and remove the device without spillage of contents;
(5) Training must be developed and validated to ensure users can follow the instructions for use; and
(6) Labeling must include the following:
(i) A contraindication for use in gynecologic surgery in which the tissue to be morcellated is known or suspected to contain malignancy;
(ii) Unless clinical performance data demonstrates that it can be removed or modified, a contraindication for removal of uterine tissue containing suspected fibroids in patients who are: Peri- or postmenopausal, or candidates for en bloc tissue removal, for example, through the vagina or via a mini-laparotomy incision;
(iii) The following boxed warning: “Warning: Information regarding the potential risks of a procedure with this device should be shared with patients. Uterine tissue may contain unsuspected cancer. The use of laparoscopic power morcellators during fibroid surgery may spread cancer. The use of this containment system has not been clinically demonstrated to reduce this risk.”
(iv) A statement limiting use of device to physicians who have completed the training program; and
(v) An expiration date or shelf life.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
November 10, 2021
Agency for Medical Innovations GmbH % Allison C. Komiyama, Ph.D., R.A.C. Principal Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Avenue, Suite B-257 San Diego, CA 92110
Re: K212659
> Trade/Device Name: More-Cell-System Regulation Number: 21 CFR§ 884.4050 Regulation Name: Gynecologic Laparoscopic Power Morcellation Containment System Regulatory Class: II Product Code: PMU Dated: August 16, 2021 Received: August 23, 2021
Dear Allison C. Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212659
Device Name More-Cell-System
#### Indications for Use (Describe)
More-Cell-System is intended for use as a tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue considered benign, resected during multi-site laparoscopic surgery for power morcellation and removal.
When used in women with fibroids, the More-Cell System is for women who are pre-menopausal and under age 50. More-Cell-System is compatible with electromechanical laparoscopic power morcellators that are between 12 mm and 20 mm in shaft outer diameter and between 95 mm and 170 mm in shaft working length.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for A.M.I. The logo is in blue and consists of the letters A, M, and I. There is a circle below each letter. The registered trademark symbol is located to the right of the last I.
# 510(k) Summary
DATE PREPARED
November 8, 2021
#### MANUFACTURER AND 510(k) OWNER
Agency for Medical Innovations GmbH Im Letten 1, 6800 Feldkirch, Vorarlberg, Austria Telephone: +43 5522 90505-4006 Fax: Official Contact: Martin Hohlrieder, Ph.D. (CTO)
#### REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C.; Lucie Dalet, Ph.D. AcKnowledge Regulatory Strategies, LLC Telephone: +1 (619) 458-9547 akomiyama@acknowledge-rs.com; Idalet@acknowledge-rs.com Email:
#### PROPRIETARY NAME OF SUBJECT DEVICE
More-Cell-System
### COMMON NAME
Containment system, laparoscopic power morcellation, with instrument port
#### DEVICE CLASSIFICATION
Gynecologic laparoscopic power morcellation containment system (21 CFR 884.4050, Product Code PMU, Class II)
#### PREDICATE DEVICE IDENTIFICATION
| 510(k) Number | Predicate Device Name / Manufacturer |
|---------------|------------------------------------------|
| K192898 | PneumoLiner / Advanced Surgical Concepts |
The predicate device has not been subject to a design-related recall.
### DEVICE DESCRIPTION
The More-Cell-System device is used as a receptacle for collection and extraction of tissue during laparoscopic surgical procedures (e.g., hysterectomy and myomectomy). The key components of the device include the following:
- A. More-Cell Bag: A single use, insufflatable, transparent, EO sterilized bag made of 50 µm thick polyurethane film that is biocompatible and cell-tight. The bag includes printed lines to aid in visualization during surgery. There are two openings to the bag: one
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for A.M.I. The letters are in blue and are large and bold. There are three gray dots underneath the letters, one under each letter. There is a registered trademark symbol next to the I.
opening that is 16 cm in diameter where a morcellator can be inserted (and where tissue can be removed), and a second opening that is 16 mm in diameter x 190 mm in length where a laparoscope can be inserted to visualize the surgical procedure. The bag capacity is 2.5 L (340 x 250 mm).
- B. Visi-Shield: A metal sleeve with a polycarbonate window for lens protection (comes in 0° or 30° angle) that is cell-tight to protect the laparoscopic camera during a surgical procedure. Using the Visi-Shield the surgical procedure can be visualized without compromising the integrity of the bag, laparoscopic view, or safety of the laparoscopic camera.
The More-Cell-System functions in the following manner:
- 1. A first incision (suprapubic) is made and a surgical port is inserted using a port obturator. The port obturator is then removed.
- 2. The More-Cell Bag is introduced through the port into the abdomen.
- 3. The bag is unfolded inside the abdomen and is filled with the tissue that is to undergo morcellation.
- 4. The port is removed and the large opening of the bag is pulled out through the first incision where it had been introduced. The small opening of the bag is pushed through a second incision (umbilical).
- 5. A trocar port is inserted into the small opening of the bag towards the encased tissue. Through this port, the More-Cell Bag is insufflated with CO2.
- 6. The optic that is encased in the Visi-Shield is inserted into the umbilical trocar port.
- 7. Through the large opening of the bag, a morcellator equipped with a blunt obturator is inserted towards the encased tissue. The tissue undergoes morcellation under visual control of the optic.
- 8. Following completion of morcellation, the morcellator, optic (including Visi-Shield), and umbilical trocar port are removed. The small opening is closed tightly by tying two (2) knots in the neck of the bag.
- 9. The More-Cell Bag is removed from the abdomen by grasping the bag at the large opening and withdrawing it through the first incision.
## INDICATIONS FOR USE
More-Cell-System is intended for use as a tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue considered benign, resected during multi-site laparoscopic surgery for power morcellation and removal. When used in women with fibroids, the More-Cell System is for women who are premenopausal and under age 50.
More-Cell-System is compatible with electromechanical laparoscopic power morcellators that are between 12 mm and 20 mm in shaft outer diameter and between 95 mm and 170 mm in shaft working length.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for A.M.I. The logo is in blue and consists of the letters A, M, and I. There are gray dots below each of the letters. There is a registered trademark symbol next to the I.
### COMPARISON OF INDICATIONS FOR USE
The predicate device includes the following indications for use:
The PneumoLiner device is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue, considered benign, resected during single-port or multiport laparoscopic surgery during power morcellation and removal. When used in women with fibroids, the PneumoLiner is for women who are pre-menopausal and under age 50. The PneumoLiner is compatible with bipolar or electromechanical laparoscopic power morcellators that are between 15mm and 18 mm in shaft outer diameter and 135mm and 180mm in shaft working length and which have an external component that allows for the proper orienation of the laparoscope to perform a contained morcellation.
The More-Cell-System indications for use differs from the predicate with respect to multi-site versus single-port surgery and the identification of compatible laparoscopic power morcellators. These differences in indications for use are based on the differences in design of the devices and the testing provided. More-Cell-System is intended for use during multi-site laparoscopic surgeries as it includes a second opening for optic insertion. This difference does not represent a new intended use as both devices are tissue containment systems to be used during minimally invasive gynecologic laparoscopic surgery. More-Cell-System is compatible with electromechanical laparoscopic power morcellators that are between 12 mm and 20 mm in shaft outer diameter and between 95 mm and 170 mm in shaft working length. The predicate is compatible with bipolar or electromechanical laparoscopic power morcellators that are between 15 mm and 18 mm in shaft outer diameter and 135 mm and 180 mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation. These differences do not represent a new intended use.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The subject device has a similar design and dimensions and uses similar or identical materials as the device cleared in K192898. Unlike the predicate device, the More-Cell Bag has two openings to the bag: one opening where a morcellator can be inserted into the abdomen and where tissue can be removed, and a second opening where a laparoscope can be inserted to visualize the surgical procedure. A disposable laparoscopic camera sleeve called the Visi-Shield is used on the laparoscope in order to protect it against any cell contamination during the surgical procedure. The subject device has a lined pattern printed on the bag while the predicate has a grid pattern. After the tissue is removed from the abdomen, the outer part of the second opening is everted and knotted. This technique is intended to keep any residual contamination during the morcellation process from being introduced to the abdominal cavity during bag removal. The subject device is provided sterile using EO sterilization and has a shelf life of 2 years, whereas the predicate device is gamma irradiated and has a shelf life of 3 year.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for A.M.I. The letters are in blue and are large and bold. There are three gray circles below the letters, one under each letter. There is a registered trademark symbol in the upper right corner of the logo.
| General Device<br>Characteristics | Subject Device<br>K212659 | Predicate Device<br>K1928998 |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Design | Tissue retrieval pouch with two<br>openings: large opening for<br>morcellator and small side<br>opening for laparoscope | Tissue retrieval pouch with<br>single opening with boot<br>assembly for insertion of<br>morcellator and laparoscope |
| Size/Capacity | Length: 340 mm<br>Width: 250 mm<br>Volume: 2.5 L | Length: 499.50 mm<br>Width: 350 mm<br>Volume: 10.58 L |
| Opening Dimensions | 16 cm diameter (large opening) | 16 cm diameter |
| | 16 mm diameter (small<br>opening) | |
| Bag removal method | Small opening knotted, pulled<br>into the abdomen and entire<br>bag removed through umbilical<br>incision | Bag removed through umbilical<br>incision |
| Insufflation required | Yes | Yes |
| Bag Materials | Polyurethane bag with<br>polyester thread, stainless steel<br>bracket/rivet, silicone and ink | Polyurethane bag with nylon<br>tether/tabs, nitinol opening ring<br>and ink |
| Optic Sleeve | Yes (Visi-Shield) | No |
| Optic Sleeve Materials | Stainless steel, polycarbonate<br>and silicone | N/A |
| LPM Compatibility | Electromechanical LPMs of<br>specific dimensions | Bipolar and electromechanical<br>LPMs of specific dimensions |
| Sterility | EtO | Gamma Radiation |
| Packaging | Tyvek pouch | Tyvek pouch |
| Shelf Life | 2 yrs | 3 yrs |
The differences between the subject and predicate devices do not raise different questions of safety and effectiveness.
### SUMMARY OF NON-CLINICAL TESTING
The following non-clinical testing has been performed on the More-Cell-System to demonstrate substantial equivalence to the predicate devices:
- Sterilization validation per ANSI/AAMI/ISO 11135 and ISO 10993-7 .
- Shelf life validation for 2-years utilizing samples that have undergone accelerated aging, . shipping and handling including:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for A.M.I. The logo is in blue and consists of the letters A, M, and I. There is a circle below each letter. There is a registered trademark symbol to the right of the letter I.
- Per ASTM F1980 and ISO 11607, package integrity included visual inspection, dye O penetration test, peel strength test and microbiological stability
- O Device functionality
- Biocompatibility testing:
The biocompatibility evaluation for the More-Cell-System was conducted in accordance with the FDA biocompatibility guidance "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," issued September 4, 2020. The More-Cell-System is categorized as an externally communicating device, contacting tissue/bone/dentin for a limited duration <24 hours. The following tests were conducted:
- Cytotoxicity per ISO 10993-5 o
- Sensitization per ISO 10993-10 O
- Intracutaneous irritation per ISO 10993-10 O
- O Acute systemic toxicity per ISO 10993-11
Testing demonstrated the device was not cytotoxic, sensitizing, irritating or acutely toxic.
- Performance testing (bench):
- Impermeability testing was conducted on 36 More-Cell bags that had previously O been exposed to 2-years of aging and simulated use conditions. Impermeability to E. coli was evaluated using an immersion test. More-Cell bags were subjected to pressures intended to represent worst-case clinical conditions). Positive and negative controls were included. There was no evidence of permeability in the test samples and the positive and negative controls performed as intended. This sample size was adequate to demonstrate a failure rate of less than 90% with a Type 1 error rate of 0.025 and power of 90%.
- o The following tests were conducted on 30 samples each that had undergone two years of simulated aging (with the exception of the rivet testing). This sample size was adequate to demonstrate that the acceptance criteria were met with a 95% confidence interval and a significance level of 0.05. Acceptance criteria were developed based on testing evaluating the forces that the device will be subjected to during clinical use.
- Puncture resistance testing ■
- Insufflation pressure (burst) testing
- Material strength testing of polyurethane bag
- Tensile testing of the sleeve with knots ●
- Weld seam in sleeve ●
- Sleeve/bag junction ●
- Tensile strength of bag opening
- Rivet connections at bag opening ●
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for A.M.I. The letters are large and blue, with a registered trademark symbol in the upper right corner. There are three gray circles below the letters, one under each letter.
- Eyelet tabs used for grasping
- Weld seam in bag
- Weld seam at neck of the bag ●
- Visi-Shield performance
- Friction force
- Bond strength
- Demonstrate the Visi-Shield component does not adversely affect the optical performance of the laparoscope or the user's visualization of the operating field
- Performance testing (animal): ●
- The More-Cell-System was tested in an in vivo porcine model. Two surgeons O performed laparoscopic supracervical hysterectomy on 16 animals randomly distributed into two groups: in group A (n = 8), power morcellation (using three different electromechanical morcellators) and tissue extraction were performed using the More-Cell-System ("MCS group"), and in group B (n = 8) the same operation was performed conventionally without a tissue retrieval bag ("control group''). Overall surgery time was prolonged in the bag group by 12.86 min (p = 0.0052; 95 % confidence interval 4.64 to 21.07). No disseminated myometrial (uterine wall) cells were found in the peritoneal washings of the abdominal cavities from the animals in group A. However, positive cytology was found in five of eight cases in the control group B (p = 0.0256). A visual examination of the 8 More-Cell-Bags after morcellation with dye and another with pressurized air while submerged in water following after removal did not show evidence of a breach.
## SUMMARY OF CLINICAL TESTING
- Usability studies:
- An observational study was conducted outside the US to evaluate usability of the More-Cell-System in women undergoing laparoscopic gynecologic morcellation. Twenty six clinical cases were performed by 26 surgeons (Austria, n=6; Germany, n=19; Switzerland, n=1) at 22 sites with various levels of experience and using a wide range of morcellators, Training was limited to a review of the Instructions for Use before the surgeon performed the procedure. Surgeons were asked to evaluate and record the ease of bag insertion and removal, the time to insert the bag and total operating time, the ease of instrument insertion and removal, the ease of placing a tissue specimen inside the bag, the ability to maintain adequate distension of the bag, the ability to maintain adequate inflation of the bag at various pressures, the ability to see inside the bag using the Visi-Shield, the size evaluation of tissue specimen, whether the morcellator blade made contact with the bag wall, whether the accompanying surgical instruments damaged the bag, whether the bag was removed laparoscopically without spilling the contents, and
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for A.M.I. The logo is in blue and consists of the letters A, M, and I. There is a gray dot under each letter. There is a registered trademark symbol in the upper right corner of the letter I.
the type of morcellator used. After the 26 cases, the bags were filled with blue fluid to evaluate if any leaks could be detected. Procedures were successfully carried out with pass-criteria satisfied and no leaks observed. No significant differences were noted based on surgeon experience, specimen size or morcellator used. With no failures noted in 26 tests, an overall acceptance criteria of 87.5% reliability is met with a 95% confidence interval.
- A simulated use study was conducted with 11 U.S. surgeons with a range of о laparoscopic surgery experience (2.5 years to 30 years) to ensure that representative users can perform the critical tasks associated with the More-Cell-System and comprehend the procedural safety information. Participants were provided with the training materials, which includes a copy of the Instructions for Use and Surgery Quick Guide Instruction training videos. After review of these materials there was a one hour break. The participants then performed a simulated use on a pelvic trainer model. They were observed to assess adherence to critical tasks. Participants completed two questionnaires to assess knowledge tasks and adequacy of device usage, and participated in group interviews for comments and open feedback from users. No use errors were found during testing. The information provided in the questionnaires demonstrated adequate comprehension of the procedural tasks to ensure safe use of the device and positive feedback on ease of device use.
These usability studies demonstrate that no additional training beyond review of labeling and videos is necessary for proper device use.
## CONFORMANCE TO SPECIAL CONTROLS
The More-Cell-System conforms to the following special controls:
- 1. The patient-contacting components of the device have been demonstrated to be biocompatible.
- 2. Device components that are labeled sterile have be validated to a sterility assurance level of 10-6.
- 3. Performance data support shelf life by demonstrating continued sterility of the device and sterile components, package integrity, and device functionality over the intended shelf life.
- 4. Non-clinical performance data demonstrate that the device meets design specifications and performance requirements. The following performance characteristics have been tested.
- a. Demonstration of the device impermeability to tissue, cells and fluids.
- b. Demonstration that the device allows for the insertion/withdrawal of laparoscopic instruments while maintaining pneumoperitoneum.
- Demonstration that the containment system provides adequate space to perform ﻥ morcellation and adequate visualization of the laparoscopic instruments and tissue specimen relative to the external viscera.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for A.M.I. The letters are in blue and are large and bold. There are three gray circles below the letters, one below each letter. There is a registered trademark symbol in the upper right corner of the logo.
- d. Demonstration that intended laparoscopic instruments and morcellators do not compromise the integrity of the containment system.
- e. Demonstration that intended users can adequately deploy the device, morcellate a specimen without compromising the integrity of the device and remove the device without spillage of contents.
- 5. Training has been developed and validated to ensure users can follow the instructions for use.
- 6. Labeling includes:
- . Contraindication for use in gynecologic surgery in which the tissue to be morcellated is known or suspected to contain malignancy.
- Contraindication for removal of uterine tissue containing suspected fibroids in patients ● who are: peri- or post-menopausal; or candidates for en bloc tissue removal, for example, through the vagina or via a mini-laparotomy incision. Note: The lanquage of this contraindication was revised during the review of the predicate device for clarity. The labeling for the More-Cell-System is consistent with the predicate device and identifies the contraindication as follows: "Do not use for removal of uterine tissue containing suspected fibroids in patients who are: post- menopausal or over 50 years of age; or candidates for en-bloc tissue removal through the vagina or via a mini-laparotomy incision."
- The following boxed warning: "Warning: Information regarding the potential risks of a procedure with this device should be shared with patients. Uterine tissue may contain unsuspected cancer. The use of laparoscopic power morcellators during fibroid surgery may spread cancer. The use of this containment system has not been clinically demonstrated to reduce this risk."
- . Statement limiting use of device to physicians who have completed the training program.
- . An expiration date or shelf life.
# CONCLUSION
The subject device has the same intended use as the predicate device. Differences in the indications for use statement do not alter the intended use. Technological differences between the subject and predicate device do not raise different questions of safety and effectiveness. The subject device has been shown to meet the special controls associated with this product type which included labeling requirements. The testing performed, including biocompatibility, sterilization and shelf life validation, non-clinical performance testing, and clinical testing, supports that the subject device is as safe and effective as the predicate. Therefore, the subject device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.