Published clinical literature (meta-analyses); FDA Updated Assessment of the Use of Laparoscopic Power Morcellators to Treat Uterine Fibroids (December 2017)
The sponsor used published clinical literature and meta-analyses of real-world patient outcomes to refine the patient population (age-based risk stratification) for the device's indications and contraindications.
Literature review; Risk stratification; Occult cancer risk; Uterine fibroids
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Seidhoff et al. (2017) meta-analysis; Meta-analysis / Decision analysis
Women undergoing hysterectomy for presumed fibroids
Not applicable for this study
Relationship between age, menopausal status, and risk of leiomyosarcoma (LMS)
AAGL Tissue Extraction Task Force paper; Literature review / Clinical update
Women with presumed fibroids
Not applicable for this study
Risk of malignancy in patients with presumed fibroids
Indications for Use
The PneumoLiner device is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue, considered benign, resected during single-port or multi- site laparoscopic surgery during power morcellation and removal. When used in women with fibroids, the PneumoLiner is for women who are pre-menopausal and under age 50. The PneumoLiner is compatible with bipolar or electromechanical laparoscopic power morcellators that are between 15mm and 18mm in shaft outer diameter and 135mm and 180mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation.
Device Story
PneumoLiner is a sterile, single-use, inflated containment system for gynecological tissue, cells, and fluids during laparoscopic power morcellation. System components include a retractor introducer, boot assembly, and containment bag. Used in hospital ORs by surgeons. Procedure: abdominal wall incision retracted; device deployed under pneumoperitoneum; tissue sample placed inside bag; bag inflated to retract viscera and create working space; morcellation performed under laparoscopic visualization within the contained space; bag and debris removed. Grid pattern on bag aids visual differentiation between target tissue and external viscera. Prevents spillage of potentially malignant cells during morcellation. Benefits include containment of benign tissue during removal, reducing risk of occult cancer dissemination.
Clinical Evidence
No clinical data generated for this submission. Rationale for labeling changes based on published clinical literature, including a meta-analysis (Seidhoff et al.) and FDA assessments (2017) regarding the risk of occult leiomyosarcoma (LMS) in patients with presumed fibroids. Data supports age 50 as a significant marker for risk stratification.
Technological Characteristics
Materials: polyurethane, nylon, PVC, silicone, polycarbonate, HDPE, stainless steel. Principle: pneumatic containment and visceral retraction. Energy: none (mechanical). Connectivity: none. Sterilization: gamma irradiation. Shelf life: 3 years. Form factor: multi-component system including retractor, introducer, and containment bag.
Indications for Use
Indicated for women undergoing minimally invasive gynecologic laparoscopic surgery for benign tissue resection. When used for fibroids, indicated for pre-menopausal women under age 50. Contraindicated for post-menopausal women or those over 50 years of age, and patients requiring en-bloc tissue removal via vagina or mini-laparotomy.
Regulatory Classification
Identification
A gynecologic laparoscopic power morcellation containment system is a prescription device consisting of an instrument port and tissue containment method that creates a working space allowing for direct visualization during a power morcellation procedure following a laparoscopic procedure for the excision of benign gynecologic tissue that is not suspected to contain malignancy.
Special Controls
In combination with the general controls of the FD&C Act, the Gynecologic Laparoscopic Power Morcellation Containment System is subject to the following special controls:
- 1. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 2. Device components that are labeled sterile must be validated to a sterility assurance level of 10-°.
- 3. Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components, package integrity, and device functionality over the intended shelf life.
- 4. Non-clinical performance data must demonstrate that the device meets all design specifications and performance requirements. The following performance characteristics must be tested:
- a. Demonstration of the device impermeability to tissue, cells and fluids.
- b. Demonstration that the device allows for the insertion/withdrawal of laparoscopic instruments while maintaining pneumoperitoneum.
- c. Demonstration that the containment system provides adequate space to perform morcellation and adequate visualization of the laparoscopic instruments and tissue specimen relative to the external viscera.
- d. Demonstration that intended laparoscopic instruments and morcellators do not compromise the integrity of the containment system.
- Demonstration that intended users can adequately deploy the device, morcellate a e. specimen without compromising the integrity of the device and remove the device without spillage of contents.
- 5. Training must be developed and validated to ensure users can follow the instructions for use.
- 6. Labeling must include:
- . Contraindication for use in gynecologic surgery in which the tissue to be morcellated is known or suspected to contain malignancy.
- Unless clinical performance data demonstrates that it can be removed or modified, a contraindication for removal of uterine tissue containing suspected fibroids in patients who are: peri- or post-menopausal; or candidates for en bloc tissue removal, for example, through the vagina or via a mini-laparotomy incision.
- . The following boxed warning: "Warning: Information regarding the potential risks of a procedure with this device should be shared with patients. Uterine tissue may contain unsuspected cancer. The use of laparoscopic power morcellators during fibroid surgery may spread cancer. The use of this containment system has not been clinically demonstrated to reduce this risk."
- Statement limiting use of device to physicians who have completed the training ● program.
- An expiration date or shelf life. ●
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible;
(2) Device components that are labeled sterile must be validated to a sterility assurance level of 10
−6 ;(3) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components, package integrity, and device functionality over the intended shelf life;
(4) Non-clinical performance data must demonstrate that the device meets all design specifications and performance requirements. The following performance characteristics must be tested:
(i) Demonstration of the device impermeability to tissue, cells, and fluids;
(ii) Demonstration that the device allows for the insertion and withdrawal of laparoscopic instruments while maintaining pneumoperitoneum;
(iii) Demonstration that the containment system provides adequate space to perform morcellation and adequate visualization of the laparoscopic instruments and tissue specimen relative to the external viscera;
(iv) Demonstration that intended laparoscopic instruments and morcellators do not compromise the integrity of the containment system; and
(v) Demonstration that intended users can adequately deploy the device, morcellate a specimen without compromising the integrity of the device, and remove the device without spillage of contents;
(5) Training must be developed and validated to ensure users can follow the instructions for use; and
(6) Labeling must include the following:
(i) A contraindication for use in gynecologic surgery in which the tissue to be morcellated is known or suspected to contain malignancy;
(ii) Unless clinical performance data demonstrates that it can be removed or modified, a contraindication for removal of uterine tissue containing suspected fibroids in patients who are: Peri- or postmenopausal, or candidates for en bloc tissue removal, for example, through the vagina or via a mini-laparotomy incision;
(iii) The following boxed warning: “Warning: Information regarding the potential risks of a procedure with this device should be shared with patients. Uterine tissue may contain unsuspected cancer. The use of laparoscopic power morcellators during fibroid surgery may spread cancer. The use of this containment system has not been clinically demonstrated to reduce this risk.”
(iv) A statement limiting use of device to physicians who have completed the training program; and
(v) An expiration date or shelf life.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 25, 2020
Advanced Surgical Concepts % Jonathan Kahan Partner Hogan Lovells US LLP Columbia Square, 555 Thirteenth Street, NW Washington, DC 20004
Re: K192898 Trade/Device Name: PneumoLiner Regulation Number: 21 CFR 884.4050 Regulation Name: Gynecologic Laparoscopic Power Morcellation Containment System Regulatory Class: II Product Code: PMU Dated: October 11, 2019 Received: October 11, 2019
Dear Jonathan Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Sharon M. Andrews for Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192898
Device Name PneumoLiner
Indications for Use (Describe)
The PneumoLiner device is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue, considered benign, resected during single-port or multi- site laparoscopic surgery during power morcellation and removal. When used in women with fibroids, the PneumoLiner is for women who are pre-menopausal and under age 50. The PneumoLiner is compatible with bipolar or electromechanical laparoscopic power morcellators that are between 15mm in shaft outer diameter and 135mm and 180mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
Advanced Surgical Concepts' PneumoLiner
#### Submitter
Advanced Surgical Concepts Ltd. Unit 4 Sunnybank Center Upper Dargle Road Bray County Wicklow A98 E339 Ireland
Phone: +353 1 286 4777 Facsimile: +353 1 286 4776
Contact Person: Edward Hyland Date Prepared: February 7, 2020
Name of Device: PneumoLiner
Common or Usual Name: Gynecologic Laparoscopic Power Morcellation Containment System
Classification Number: 21 C.F.R. §884.4050
Classification Name: Gynecologic Laparoscopic Power Morcellation Containment System
Regulatory Class: Class II
Product Code: PMU
#### Predicate Device
Advanced Surgical Concepts Ltd. PneumoLiner DEN150028
Classification Name: Gynecologic Laparoscopic Power Morcellation Containment System Regulatory Class: Class II Product Code: PMU
This device has never been the subject of a recall, design related or otherwise.
{4}------------------------------------------------
# Device Description
The ASC PneumoLiner is an inflated morcellation containment system that allows for containment of gynecological tissue, cells and fluids during power morcellation and removal. It is a sterile single use device and requires the surgeon to successfully complete a validated training program before use. The ASC PneumoLiner is for clinical use in a hospital or surgical center operating room.
The device consists of the following components: the Retractor Introducer, Boot Assembly, the PneumoLiner Introducer and the PneumoLiner. The first three components (Retractor, Retractor Introducer and Boot Assembly) are used to retract an incision in the abdominal wall and to allow for the introduction of laparoscopic instruments and the PneumoLiner Bag while under pneumoperitoneum and with vision from a laparoscope. The PneumoLiner Bag is introduced via the PneumoLiner introducer and the large valve on the Boot Assembly.
The PneumoLiner is intended for use in gynecological power morcellation. It is intended to form a complete containment barrier from the spillage of liquids, cells and tissue from the time the tissue or organ is excised and encapsulated in the PneumoLiner bag, throughout the morcellation procedure and withdrawal of the containment bag with accompanying liquids and debris.
The key technological characteristics of the PneumoLiner and predicate PneumoLiner are the following:
- . The PneumoLiner is designed to be deployed while the abdomen is under pneumoperitoneum. This ensures an easy deployment and correct placement of the PneumoLiner within the peritoneal cavity.
- . The PneumoLiner is designed to open once deployed in the abdomen so as to facilitate placement of the tissue sample in the device.
- . The device is designed to close and encapsulate the tissue sample prior to exteriorisation of the neck.
- . Once the PneumoLiner is inflated and pneumoperitoneum is re-established, the inserted morcellator is not intended to come into contact with the sides of the PneumoLiner during a morcellation.
- . The inflation of the PneumoLiner retracts the viscera and expands the abdominal wall to create a space for morcellation.
- . Morcellation takes place in the contained space of the PneumoLiner under vision with the aid of a laparoscope.
- . The size of the PneumoLiner is designed to create a large working space within the abdomen.
- The grid pattern allows for a surgeon to easily distinguish between the target tissue/organ inside the PneumoLiner and viscera on the outside of the PneumoLiner.
The PneumoLiner Bag is polyurethane with nylon tether/tabs and a nitinol opening ring with stainless steel crimp. The Retractor Introducer shaft is polycarbonate. The Retractor Introducer Plunger is HDPE with a blue colorant. The Retractor sleeve is polyurethane with polycarbonate, nylon and lubricant oil. The Boot Assembly component includes PVC, adhesive, polyurethane, HDPE and polycarbonate. The instrument valves are polycarbonate housings with silicone valves with blue and yellow colorants. The PneumoLiner introducer shaft and plunger are both HDPE. The PneumoLiner System includes materials and colorants that have direct patient contact for a duration of up to 6 hours.
{5}------------------------------------------------
### Intended Use / Indications for Use
The purpose of the 510(k) notice is to revise the Indications for Use and contraindications of the predicate PneumoLiner device cleared to market under DEN150028 to better define the patient population.
The indications for use for the cleared PneumoLiner (DEN150028) are:
The PneumoLiner device is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue, considered benign, resected during single-port or multi-site laparoscopic surgery during power morcellation and removal. The PneumoLiner is compatible with bipolar or electromechanical laparoscopic power morcellators that are between 15mm and 18mm in shaft outer diameter and 130mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation.
This 510(k) modifies the indications for use to add the following sentence: "When used in women with fibroids, the PneumoLiner is for women who are pre-menopausal and under age 50."
Thus, the indications for use for the subject device are:
The PneumoLiner device is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue, considered benign, resected during single-port or multi- site laparoscopic surgery during power morcellation and removal. When used in women with fibroids, the PneumoLiner is for women who are pre-menopausal and under age 50. The PneumoLiner is compatible with bipolar or electromechanical laparoscopic power morcellators that are between 15mm and 18mm in shaft outer diameter and 135mm and 180mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation.
Further, the 510(k) also modifies the contraindications for the PneumoLiner. The cleared PneumoLiner (DEN150028) contains the following contraindication:
Do not use for removal of uterine tissue containing suspected fibroids in patients who are: peri or post-menopausal; or candidates for enbloc tissue removal, for example, through the vagina or via a minilaparotomy incision.
This contraindication has been changed to read:
Do not use for removal of uterine tissue containing suspected fibroids in patients who are: post- menopausal or over 50 years of age; or candidates for en-bloc tissue removal through the vagina or via a mini-laparotomy incision.
The purpose of these changes is to better inform the physician of the intended patient population. This change to the wording of the contraindication does not alter the patient population but better defines it, removing the ambiquity of the term "peri-menopausal." The basis for these changes is the published clinical literature that is referenced in the Summary of Clinical Performance Testing below.
{6}------------------------------------------------
The proposed minor revisions to the indications for use and contraindication are to better identify the population of women indicated for use of the PneumoLiner; therefore, these changes do not represent a new intended use.
The labeling was also modified to include a new warning to reflect the increased risk of occult cancer with age. The warning states:
The risk of occult cancer, including uterine sarcoma, increases with age, particularly in women over 50 years of age. This information should be shared with patients when considering surgery with the use of these devices.
### Summary of Technological Characteristics
Containment through inflation and an impervious bag material is the technological principle for both the subject and predicate device. This is supported by retraction of viscera through the inflation of the bag and vision provided through the 5mm instrument valve by a laparoscopic endoscope to enable power morcellation of the subject gynecological tissue.
Both the subject PneumoLiner and the predicate PneumoLiner use identical technological elements. All components of the devices are the same.
| | Subject PneumoLiner | PneumoLiner<br>DEN150028 |
|-------------------|---------------------------------------------------------------------------------------------------------------|--------------------------|
| Device Components | The Retractor<br>Retractor Introducer<br>The PneumoLiner Introducer<br>PneumoLiner (bag) | Same |
| Accessories | N/A | N/A |
| Power Source | N/A | N/A |
| Materials | Polyurethane, nylon, PVC, silicone,<br>polycarbonate, adhesive, stainless<br>steel, blue and yellow colorants | Same |
| Sterilization | Gamma irradiation | Same |
| Shelf Life | 3 years | 1 year |
| Packaging | Hard blister with retainer insert and<br>Tyvek | Hard blister and Tyvek |
The changes to shelf-life and packaging do not raise different questions of safety and effectiveness.
#### Summary of Non-Clinical Performance Testing
The following performance data were provided in support of the substantial equivalence determination:
- Sterilization validation to ISO 11137-1:2006 and ISO 11137-2:2013
- Shelf life testing on samples accelerated aged which underwent simulated shipping . including:
- Sterile Barrier Integrity testing .
{7}------------------------------------------------
- Visual inspection per ASTM F1996:2009 (2013) ●
- Bubble leak testing per ASTM F2096:2011 •
- . Seal strength per ASTM F88:2009.
- Device functionality ●
- Leak testing ●
- . Bond/material strength testing.
# Summary of Clinical Performance Testing
The purpose of 510(k) K192898 is a revision to the labeling; contraindication and the indication for use statement to better define the patient population. The rationale for the proposed label changes is published clinical literature as reflected in the JMIG Special Article, "Morcellation during Uterine Tissue Extraction: An Update by the Tissue Extraction Task Force Members" published in Journal of Minimally Invasive Gvnecology, Vol 25, No 4. May/June 2018 and the recent FDA paper; FDA Updated Assessment of the Use of Laparoscopic Power Morcellators to Treat Uterine Fibroids (December 2017 1 that stratify the risk of occult leiomyosarcoma (LMS) by age which provides a better- defined method to allow surgeons to mitigate the risk to their patients.
There are no clinical data, of which the company is aware, which supports an increased risk of LMS due to menopausal status. The most complete meta-analysis is included in the paper by Seidhoff et al² which indicates a relationship between age and risk of LMS but does not find any evidence of a relationship to menopausal status in the literature.
The FDA Updated Assessment of the Use of Laparoscopic Power Morcellators to Treat Uterine Fibroids (December 2017) has summarized the risk of uterine sarcoma and LMS by age. The stratification by age has clearly defined 50 years as a significant and easily defined marker; that is, there is a notably reduced risk for women under 50 years of age for the occurrence of both uterine sarcoma and LMS.
The available data has also been reviewed as part of the AAGL Tissue Extraction Task Force paper "Morcellation During Uterine Tissue Extraction: An Update," which supports the conclusion that the older the patient with presumed fibroids, the greater the risk that the tumor is malignant.
# Conclusions
The submitted published clinical data supports the labeling changes to the PneumoLiner and the nonclinical testing on packaging and shelf-life demonstrate that the device is as safe, as effective, and performs as well as the predicate.
<sup>1</sup> https://www.fda.gov/media/109018/download
<sup>2</sup> Seidhoff et al. 2017. Laparoscopic hysterectomy with morcellation vs abdominal hysterectomy for presumed fibroids: An updated decision analysis following the 2014 FDA safety communications. Am J Obstet Gynecol. 2017 Mar;216(3):259.e1-259.e6
<sup>3</sup> Morcellation During Uterine Tissue Extraction: An Update, Journal of Minimally Invasive Gynecology (2018), https://doi.org/10.1016/j.jmig.2018.03.010.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.