Rösch Thurmond Fallopian Tube Catheterization Set, Fallopian Tube Catheterization Set, and Fallopian Tube Catheterization Wire Guide with Platinum Tip
Applicant
Cook Incorporated
Product Code
LKF · Obstetrics/Gynecology
Decision Date
Feb 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Indications for Use
The Rosch Thurmond Fallopian Tube Catheterization Set is intended to be used for through a uterine access device for fluoroscopic selective catheterization of the proximal fallopian tube(s) and injection of contrast medium in the evaluation of tubal patency. The Fallopian Tube Catheterization Set is intended to be used through a uterine access device for fluoroscopic selective catheterization of the proximal fallopian tube(s) and injection of contrast medium in the evaluation of tubal patency. The Fallopian Tube Catheterization Wire Guide with Platinum Tip is to be used through a dimensionally compatible (>0.018 inch inner diameter) uterine access device or selective salpingography catheter to assist in the evaluation of tubal patency.
Device Story
Catheterization sets and wire guides used for fluoroscopic evaluation of fallopian tube patency; sets include introducing catheters, torque control catheters, injection catheters, and wire guides. Device operates by manual insertion through uterine access device into proximal fallopian tube; contrast medium injected through catheter to visualize patency under fluoroscopy. Used in clinical settings by physicians. Output is visual confirmation of tubal patency on fluoroscopic imaging; assists in diagnosing tubal obstruction or infertility. Benefits include minimally invasive assessment of fallopian tube status.
Clinical Evidence
Bench testing only. Performance studies included sterilization validation (ISO 11135-1:2007), biocompatibility (ISO 10993-5/10), transportation simulation (ASTM D4169-16), and package integrity (ASTM F2096-04, ASTM F88-09). Bench performance testing verified dimensional compatibility, tensile strength (JIS T 3246:2011), torque response, liquid leakage, corrosion resistance, fracture resistance, flex damage resistance, radiopacity, and tip deflection (BS EN ISO 11070:2014).
Technological Characteristics
Catheters made of Teflon, polyethylene, or nylon; wire guides made of stainless steel or nitinol. Radiopaque tips included. Sterilized via ethylene oxide. Shelf-life 3 years. Single-use. Dimensional compatibility with >0.018 inch ID uterine access devices. Manual operation.
Indications for Use
Indicated for patients undergoing evaluation of fallopian tube patency via fluoroscopic selective catheterization and contrast medium injection. Used with uterine access devices.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
Predicate Devices
Rösch-Thurmond Fallopian Tube Catheterization Set (K955508)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 15, 2018
Cook Incorporated Ian Herrman Regulatory Affairs Specialist, China Team 750 Daniels Way, P.O. Box 489 Bloomington, IN 47402
Re: K171604
> Trade/Device Name: Rösch-Thurmond Fallopian Tube Catheterization Set, Fallopian Tube Catheterization Set, and Fallopian Tube Catheterization Wire Guide with Platinum Tip Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: LKF Dated: January 22, 2018 Received: January 23, 2018
Dear Ian Herrman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and is set against a white background. The letters are clearly legible and well-defined. The name appears to be the focus of the image.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171604
#### Device Name
Rösch Thurmond Fallopian Tube Catheterization Set, Fallopian Tube Catheterization Set, and Fallopian Tube Catheterization Wire Guide with Platinum Tip
#### Indications for Use (Describe)
The Rosch Thurmond Fallopian Tube Catheterization Set is intended to be used for through a uterine access device for fluoroscopic selective catheterization of the proximal fallopian tube(s) and injection of contrast medium in the evaluation of tubal patency.
The Fallopian Tube Catheterization Set is intended to be used through a uterine access device for fluoroscopic selective catheterization of the proximal fallopian tube(s) and injection of contrast medium in the evaluation of tubal patency.
The Fallopian Tube Catheterization Wire Guide with Platinum Tip is to be used through a dimensionally compatible (>0.018 inch inner diameter) uterine access device or selective salpingography catheter to assist in the evaluation of tubal patency.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K171604
| Submitted By: | Ian Herrman |
|----------------|-------------------------------|
| | Rohini Patel |
| | Cook Incorporated |
| | 750 Daniels Way |
| | P.O. Box 489 |
| | Bloomington, IN 47402 |
| | Phone: (812) 339-2235 x104034 |
| | Fax: (812) 332-0281 |
| Date Prepared: | 05/31/2017 |
| Date Prepared: | 05/31/2017 |
|----------------------------|-------------------------------------------------------------|
| Submission: | Traditional 510(k) Premarket Notification |
| Trade/Device Name: | Rösch Thurmond Fallopian Tube Catheterization Set, |
| | Fallopian Tube Catheterization Set |
| | Fallopian Tube Catheterization Wire Guide with Platinum Tip |
| Common Name: | Salpingography Catheter |
| Classification Regulation: | 21 CFR §884.4530 |
| Product Code: | LKF, Cannula, Manipulator/Injector, Uterine |
| Classification Name: | Obstetric-gynecologic specialized manual instrument |
| Device Classification: | Class II |
| Classification Panel: | Obstetrics/Gynecology |
## Predicate Device:
The predicate device is the Rösch-Thurmond Fallopian Tube Catheterization Set cleared under 510(k) K955508. This predicate device has not been subject to any design related recall.
#### Device Description:
This submission contains two catheter sets: the Rösch-Thurmond Fallopian Tube Catheterization Set and the Fallopian Tube Catheterization Set. The submission also includes the Fallopian Tube Catheterization Wire Guide with Platinum Tip, which is included in the Fallopian Tube Catheterization Set, and will also be sold separately.
## Rösch-Thurmond Fallopian Tube Catheterization Set
This set contains a 31 cm, 9 Fr, Teflon introducing catheter; a 50 cm, 5.5 Fr, polyethylene, radiopaque, torque control catheter; and a 65 cm, 3 Fr, Teflon, radiopaque, injection catheter. The introducing catheter is designed with a check flow valve at the user end while the two other catheters have a standard luer lock adapter. The set also comes with three 90 cm long, stainless steel, accessory wire guides. Two of the wire guides are 0.035 inches in diameter and the other is 0.015 inches in diameter.
{4}------------------------------------------------
## Fallopian Tube Catheterization Set
This set contains a 50 cm, 5.5 Fr, polyethylene, radiopaque torque control catheter, a 65 cm, 3.0 Fr nylon injection catheter with radiopaque tip, and two wire guides. The larger wire guide is stainless steel and has a diameter of 0.035 inches. The set also comes with a smaller wire guide which can be purchased in two variations; either the 0.015 inch diameter, 90 cm long, stainless steel Cope mandril wire guide, or the 0.018 inch diameter, 90 cm long, nitinol Cope mandril wire guide.
# Fallopian Tube Catheterization Wire Guide with Platinum Tip
This wire guide is identical to the Nitinol Cope Mandril Wire Guide included in Fallopian Tube Catheterization Set, but can be purchased individually.
Both the Rösch-Thurmond Fallopian Tube Catheterization Set, the Fallopian Tube Catheterization Set, and the Fallopian Tube Catheterization Wire Guide with Platinum Tip are sterilized using ethylene oxide, packaged in Tyvek® polyethylene peel-open pouches, and have a shelf-life of three years. The devices are intended for one time use.
# Indications for Use:
## Rösch-Thurmond Fallopian Tube Catheterization Set:
The Rösch-Thurmond Fallopian Tube Catheterization Set is intended to be used through a uterine access device for fluoroscopic selective catheterization of the proximal fallopian tube(s) and injection of contrast medium in the evaluation of tubal patency.
# Fallopian Tube Catheterization Set:
The Fallopian Tube Catheterization Set is intended to be used through a uterine access device for fluoroscopic selective catheterization of the proximal fallopian tube(s) and injection of contrast medium in the evaluation of tubal patency.
# Fallopian Tube Catheterization Wire Guide with Platinum Tip:
The Fallopian Tube Catheterization Wire Guide with platinum tip is to be used through a dimensionally compatible (> 0.018 inch inner diameter) uterine access device or selective salpingography catheter to assist in the evaluation of tubal patency.
{5}------------------------------------------------
# Comparison to Predicate Device:
The predicate device is the Rösch-Thurmond Fallopian Tube Catheterization Set (K955508). The intended use for the predicate device is as follows:
The Rösch-Thurmond Fallopian Tube Catheterization Set is a fallopian tube catheterization set intended to be used through a Thurmond-Rösch Hysterocath® (K884291) or other uterine access device for fluoroscopic selective catheterization and cannulation of the proximal fallopian tube(s), followed by the introduction of contrast medium, in the evaluation of the tubal patency.
This indication for use is identical to the indication for use in the subject device sets except it includes reference to a specific uterine access device and indicates for cannulation in addition to catheterization of the proximal fallopian tubes. The Fallopian Tube Catheterization Wire Guide with Platinum Tip is indicated to assist in the evaluation of tubal patency as indicated by both the predicate and subject device sets.
The differences between the predicate and the subject devices are listed below:
## Rösch-Thurmond Fallopian Tube Catheter Set
(1) Wording in the indications for use
## Fallopian Tube Catheterization (FTC) Set
(1) Wording in the indications for use
(2) The injection catheter for the FTC Set is made from Nylon instead of Teflon and has a radiopaque tip
(3) The wire guides included in the FTC Set vary slightly from those included in the predicate
(4) The FTC Set is not sold with an introduction catheter
As noted above the differences in the indications for use of the predicate and subject device sets do not raise any different questions of safety and effectiveness because subject device indications are within the intended use of the predicate device.
Though the nylon injection catheter included in the FTC Set is a different material and has a radiopaque tip, the material serves the same function as the Teflon in the predicate submission. Testing has been conducted to assure the nylon material performs as intended. The radiopaque tip provides visibility under fluoroscopy but as the Teflon
{6}------------------------------------------------
material was also visible under fluoroscopy there is no change to the procedure or safety and effectiveness.
The wire guides included with the Fallopian Tube Catheterization Set are slightly different from those included in the predicate. The 0.035 inch diameter Double Flexible Tip Safe-T-J® Wire Guide included in the FTC Set has a flexible tip on one end and a curved tip on the other. This effectively combines the functionality of both the straight safety wire guide and the curved tip 0.035 Safe-T-J® wire guide included in the predicate. The 0.035 inch diameter wire guides are made from the same materials. One variation of the FTC set comes with a 0.018 inch diameter Nitinol Cope Wire Guide which varies from the corresponding predicate wire guide in that it is nitinol instead of stainless and its diameter is 0.003 inches larger. The functionality and biocompatibility of the nitinol material has been tested and the results demonstrate that no different questions of safety and effectiveness were raised.
The FTC Sets are not sold with a 9 French introduction catheter because many physicians prefer to use a uterine access device with a smaller working channel which makes this set component unnecessary.
The subject devices have similar indications for use, methods of operation, and fundamental technological characteristics as the predicate device. Modifications to the subject devices from the predicate device are supported by testing and raise no different questions of safety and effectiveness.
# Test Data:
The following testing was performed to demonstrate that Rösch-Thurmond Fallopian Tube Catheterization Set and the Fallopian Tube Catheterization Set met applicable design and performance requirements.
- Sterilization Validation testing per ISO 11135-1:2007 .
- Biocompatibility studies, as follows: .
- o Cytotoxicity testing per 10993-5:2009
- Guinea Pig Maximization Sensitization testing per ISO 10993-10:2010 o
- o Intracutaneous Irritation testing per ISO 10993-10:2010
- Transportation Simulation study per ASTM D4169-16 .
- Package Integrity testing after real-time aging: .
{7}------------------------------------------------
- Bubble Leak test per ASTM F2096-04 O
- O Seal Strength testing per ASTM F88-09
- Visual Inspection: No package displayed damage (tears, folds, puncture holes, O etc.)
- Bench Performance studies before and after aging demonstrated that all predetermined . acceptance criteria were met in the following tests (No age testing was conducted for the wire guides, radiopacity, or tip deflection):
- . Compatibility Testing –The components in each set to be used together as stated in the instructions for use were tested for dimensional compatibility. The testing demonstrated compatibility between the components included in the sets being submitted.
- . Tensile Testing (catheters hub to shaft and tip to shaft) – Tensile testing was performed on the hub to shaft connection for each catheter and also on the tip to shaft section of the catheters that have a bonded tip. The acceptance criterion for these tensile tests was determined by Cook's internal procedures and according to Section 5.3 of JIS T 3246:2011. Testing demonstrated that that the peak load value was greater than the predetermined acceptance criterion.
- Torque Response Testing (5.5 Fr catheter) Torque control catheter was ■ placed through its corresponding 9.0 French introducing catheter and the hub was rotated and the response of the distal tip was noted. Testing demonstrated that the torque-control catheter has an appropriate torque response per the predetermined acceptance criteria.
- I Liquid Leakage Test (5.5 Fr and 3 Fr catheters) – Liquid leakage testing was done on the catheters that will be used for the injection of contrast. The acceptance criterion was that the catheter assembly will not leak when an injection pressure of 38 kPa is applied. Testing demonstrated that fluid path catheter assembly did not leak under the predetermined injection pressure.
- . Corrosion Resistance (wire guide) - Representative testing was conducted according to Annex B of BS EN ISO 11070:2014 for each wire guide included in the subject device sets. Testing demonstrated that tensile
{8}------------------------------------------------
properties did not degrade after exposure of wires to corrosive conditions. The predetermined acceptance criterion was met.
- Fracture Testing (wire guide) - Representative testing was conducted for each wire guide included in the subject device sets. The fracture testing was performed according of Annex F of BS EN ISO 11070:2014 and an approved study protocol. The Testing demonstrated that no signs of fracture were observed in the region of interest, according to the predetermined acceptance criterion.
- . Tensile Test (wire guide) - Representative testing was conducted according to Annex H of BS EN ISO 11070:2014 for each wire guide included in the subject device sets. Testing demonstrated that the peak load value was greater than or equal to the predetermined acceptance criterion.
- Resistance to Damage by Flex (wire guide) Representative testing was " conducted according to Annex G of BS EN ISO 11070:2014 for each wire guide included in the subject device sets. Testing demonstrated that there were no signs of defects or damage, when subjected to repeated flexing. The predetermined acceptance criterion was met.
- Radiopacity Evaluation Testing demonstrated that the subject device was ■ determined to be radiopaque in a simulated clinical setting.
- . Tip Deflection Testing – Testing was conducted in order to compare the distal tip flexibility of the subject device to the predicate. Tested section included both the shaft material and bonded tip material of the subject device. Testing demonstrated that tip flexibility between the subject and predicate devices was comparable.
## Conclusion:
In conclusion, the minor differences in indications for use statements between the subject and predicate devices do not alter the intended use of the subject device. The differences in technological characteristics between the subject and predicate device do not raise different questions of safety and effectiveness. The results of performance testing support that the Rösch-Thurmond Fallopian Tube Catheterization Set and the Fallopian Tube
{9}------------------------------------------------
Catheterization Set are as safe and effective as the predicate device. Therefore, the subject device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.