K153092 · CooperSurgical, Inc. · LKF · Dec 21, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K153092
Device Name
Ally Uterine Positioning System
Applicant
CooperSurgical, Inc.
Product Code
LKF · Obstetrics/Gynecology
Decision Date
Dec 21, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.4530
Device Class
Class 2
Indications for Use
The Ally Uterine Positioning System™ (UPS) is intended to assist the surgical staff in mounting, positioning and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment. The Advincula Delineator™ Adapter Drape for the Ally Uterine Positioning System™ UPS is intended to assist the surgical staff in mounting, positioning, and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.
Device Story
The Ally Uterine Positioning System (UPS) is a mechanical, multi-segmented articulating arm that attaches to an operating room bed rail. It utilizes a sterile, single-use disposable adapter drape to secure a uterine manipulator (RUMI II, Advincula Arch, or Advincula Delineator) to the distal end of the arm. The device is operated by surgical staff in an operating room. The arm is positioned manually when unlocked; it is locked into a fixed position by releasing a foot pedal, which activates a linear actuator to apply tension to an internal cable, drawing segments together. The system provides stable positioning of the uterine manipulator during gynecological laparoscopic surgery, facilitating surgical access and visualization.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance testing, including adapter pull-off force, drape pull-off force, and handle push force testing. Sterilization validation performed per ISO 11135-1:2007 and ISO 11135-2:2008. Shelf life stability testing conducted per ASTM F1980-07.
Technological Characteristics
Multi-segmented articulating arm constructed of stainless steel and aluminum. Adapter drape made of Lexan and 2 mil polyethylene film. System uses a linear actuator (energy source) to lock segments via internal cable tension. Sterilization of adapter drape via ethylene oxide (SAL 10^-6). Complies with IEC 60601-1 and IEC 60601-1-2. Non-patient contacting.
Indications for Use
Indicated for use by trained operating room personnel to assist in mounting, positioning, and holding uterine manipulators during gynecological laparoscopic surgical procedures.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an emblem featuring a stylized depiction of three human profiles facing right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 21, 2015
CooperSurgical, Inc. Roaida Johnson Associate Director, New Product Development 95 Corporate Drive Trumbull, CT 06611
Re: K153092 Trade/Device Name: Ally Uterine Positioning System™ Regulation Number: N/A Regulation Name: N/A Regulatory Class: Not Classified Product Code: LKF Dated: December 8, 2015 Received: December 9, 2015
Dear Roaida Johnson,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Herbert P. Lerner -S**
Benjamin R. Fisher, Ph.D. for Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153092
Device Name Ally Uterine Positioning System™
# Indications for Use (Describe)
Ally Uterine Positioning System™
The Ally Uterine Positioning System™ (UPS) is intended to assist the surgical staff in mounting, positioning and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.
#### Adapter Drape for the Ally UPS System
The Advincula Delineator™ Adapter Drape for the Ally Uterine Positioning System™ UPS is intended to assist the surgical staff in mounting, positioning, and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# (Goper Surgical
# 510(k) Summary CooperSurgical Ally Uterine Positioning System™ (UPS) K153092
#### Submitter Information
| Company Name: | CooperSurgical Inc. |
|------------------|-----------------------------------------------------------------|
| Company Address: | 95 Corporate Drive<br>Trumbull, CT 06611 |
| | Telephone: 203-601-5200 |
| | Fax: 203-601-9870 |
| Contact Person: | Roaida Johnson<br>Associate Director of New Product Development |
| | Phone: 203-601-5200 Ext: 3325 |
| | Fax: 203-601-9870 |
| | roaida.johnson@coopersurgical.com |
| Date Prepared: | December 15, 2015 |
#### Device Information:
| Trade Name: | Ally Uterine Positioning System™ |
|----------------------|--------------------------------------|
| Common Name: | Holder, Manipulator, Positioner, Arm |
| Classification: | Unclassified |
| Classification Name: | Unclassified |
| Product Code: | LKF |
#### Predicate Device Information:
The CooperSurgical Ally Uterine Positioning System™ is substantially equivalent to the predicate CooperSurgical Ally Uterine Positioning System™ (K141523)
#### Device Description:
The CooperSurgical Ally Uterine Positioning System™ (UPS) consists of a single, multi- segmented, articulating arm that attaches to a standard operating room bed rail, and a separate, sterile, disposable Ally UPS Adapter Drape that is used to attach a uterine manipulator to the Ally UPS. When unlocked, the articulation of the arm allows the attached manipulator to be positioned by the user. The segmented design of the arm allows lateral/medial movement from a single point to position the uterine manipulator. The arm is then locked in the desired position by releasing a food pedal, activating a linear actuator that applies tension to an internal cable, drawing the segments together and thus locking the arm.
The Ally UPS is not intended for patient contact. The Adapter Drape for the Ally UPS is single- use and sterilized by ethylene oxide. The Ally UPS Adapter Drape secures either the Cooper Surgical RUMI II
{4}------------------------------------------------
Uterine Manipulator Handle, or the Advincula Arch Uterine Manipulator Handle, to the distal end of the Ally UPS arm. The purpose of this submission is to gain clearance to add the new Advincula Delineator™ Adapter Drape to the system, which allows the Ally UPS to be used with the CooperSurgical Advincula Delineator™ Uterine Manipulator.
#### Indications for Use:
## Ally Uterine Positioning System™ (UPS)
The Ally Uterine Positioning System™ (UPS) is intended to assist the surgical staff in mounting, positioning and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.
## Adapter Drape for the Ally UPS System
The Advincula Delineator™ Adapter Drape for the Ally Uterine Positioning System™ UPS is intended to assist the surgical staff in mounting, and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.
| Attribute | Subject<br>Ally Uterine Positioning System™ | Predicate<br>Ally Uterine Positioning<br>System - K141523 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Indications for Use: Ally<br>UPS | The Ally UPS is intended to assist the<br>surgical staff in mounting, positioning, and<br>holding uterine manipulators during<br>gynecological laparoscopic surgical<br>procedures. It is intended for use by trained<br>operating room personnel in an operating<br>room environment. | Same |
| Indications for Use:<br>Adapter Drape | The Advincula Delineator™ Adapter Drape<br>for the Ally UPS System is intended to assist<br>the surgical staff in mounting, positioning,<br>and holding uterine manipulators during<br>gynecological laparoscopic surgical<br>procedures. It is intended for use by trained<br>operating room personnel and in an operating<br>room environment. | Same |
| Patient Contact Type | Non-Patient Contacting | Same |
| Ally UPS Material | Stainless steel, aluminum | Same |
| Adapter Drape Material | Lexan, 2 mil polyethylene film | Same |
| Packaging | Tyvek 1073B/polyester/polyethylene heat<br>sealed chevron pouch. The pouches are<br>packaged in SBS paperboard cartons and are<br>placed into corrugated shipping cases for<br>transport and storage. | Same |
| Sterilization | The Delineator Adapter Drape is sterilized<br>by ethylene oxide to a Sterility Assurance<br>Level (SAL) of 10-6. | Same |
#### Substantial Equivalence Analysis
The substantial equivalence of the subject Ally UPS System with the Advincula Delineator™ Adapter Drape to the predicate is shown by similarity of intended use, indications for use, materials, design, function, sterilization, and performance.
{5}------------------------------------------------
## Non-Clinical Performance Testing
The modified Ally UPS with the Advincula Delineator™ Adapter Drape was compared to the predicate for the following performance characteristics:
- . Adapter Pull-off Force
- Drape Pull-off Force
- Handle Push Force .
The Advincula Delineator™ Adapter Drape maintained its specifications for the duration of its shelf life as demonstrated by stability testing in accordance with ASTM F1980-07.
Sterilization validation performed per ISO 11135-1:2007- Sterilization of healthcare products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices, and ISO 11135-2:2008 - Sterilization of health care products - Ethylene oxide - Part 2: Guidance on the application of ISO 11135-1 demonstrated that the Advincula Delineator™ Adapter Drape meets the appropriate ethylene oxide and ethylene chlorohydrin residual levels per the requirements of ANSVAAMI/ISO 10993-7:2008(R)2012 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals.
The Ally UPS is in compliance with the following standards:
- IEC 60601-1 CORR 1 & 2 2007 Medical Electrical Equipment Part 1: General Requirements . for Basic Safety and Essential Performance
- IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements for Safety ●
#### Clinical Performance Testing
Clinical testing was not required to support the conclusion of substantial equivalence.
#### Conclusion
Based on the indications for use, technological characteristics, performance testing, and comparison to the predicate, the modified Ally Uterine Positioning System™ (UPS) has been shown to be substantially equivalent to the predicate device identified, and does not present any new issues of safety or effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.