K141523 · CooperSurgical, Inc. · LKF · Nov 25, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K141523
Device Name
ALLY UPS
Applicant
CooperSurgical, Inc.
Product Code
LKF · Obstetrics/Gynecology
Decision Date
Nov 25, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Indications for Use
The Ally Uterine Positioning System™ (Ally UPS) is intended to assist the surgical staff in mounting, positioning, and holding uterine manipulators during gynecological laparoscopic procedures. It is intended for use by trained operating room personnel in an operating room environment.
Device Story
The Ally Uterine Positioning System (UPS) is a mechanical surgical support device used in operating rooms by surgical staff. It consists of a multisegmented, articulated arm that attaches to an OR bed rail and a sterile distal adapter. The device holds uterine manipulators (e.g., RUMI II or Advincula Arch) in a fixed position during gynecological laparoscopic procedures. Operation involves manually positioning the arm when unlocked, then depressing a foot pedal to activate a linear actuator; this applies tension to an internal cable, drawing segments together to lock the arm in place. A sterile adapter drape covers the arm to maintain the sterile field and prevent contamination. By providing stable, hands-free positioning of the uterine manipulator, the device assists the surgical team in maintaining optimal visualization and access during surgery.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including holding (locking) force, durability (cycle) testing, handle push force, drape pull-off force, and adapter pull-out force. The device complies with IEC 60601-1 and IEC 60601-1-2 standards. Shelf-life stability was validated per ASTM F1980-07.
Technological Characteristics
Materials: stainless steel, aluminum. Sensing/Actuation: foot-pedal activated linear actuator applying tension to an internal cable for mechanical locking. Form factor: multisegmented, articulated arm for OR bed rail mounting. Sterilization: ethylene oxide for the adapter drape (SAL 10^-6). Connectivity: none (mechanical).
Indications for Use
Indicated for use by trained operating room personnel to assist in mounting, positioning, and holding uterine manipulators during gynecological laparoscopic surgical procedures.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three heads facing right, representing the department's mission to protect the health of all Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 25, 2014
Cooper Surgical Inc. c/o Tim M. Lohnes Senior Regulatory Consultant Orchid Design 80 Shelton Technology Center Shelton. CT 06484
Re: K141523
Trade/Device Name: Ally Uterine Positioning System™ Adapter Drape for the Ally Uterine Positioning System™ Regulation Number: Unclassified Regulation Name: Unclassified Regulatory Class: Unclassified Dated: October 17, 2014 Received: October 20, 2014
Dear Tim M. Lohnes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal. and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141523
Device Name Ally Uterine Positioning System TM
Indications for Use (Describe)
The Ally Uterine Positioning System™ (Ally UPS) is intended to assist the surgical staff in mounting, positioning, and holding uterine manipulators during gynecological laparoscopic procedures. It is intended for use by trained operating room personnel in an operating room environment.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141523
Device Name
Adapter Drape for the Ally Uterine Positioning System™
Indications for Use (Describe)
The Adapter Drape for the Ally Uterine Positioning System™ (Ally UPS) is intended to assist the surgical staff in mounting, positioning, and holding uterine manipulators during gynecological laparoscopic procedures. It is intended for use by trained operating room personnel in an operating room environment.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Goper Surgical
# 7. 510(K) SUMMARY:
510(k) Summary of Safety and Effectiveness, (21 CFR 807.92):
| Submitter: | Cooper Surgical Inc.,<br>75 Corporate Drive,<br>Trumbull, CT 06611 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tim M. Lohnes<br>Senior Regulatory Consultant,<br>Orchid Design<br>Phone: (203) 922 0105<br>Fax: (203) 922 0130<br>tim.lohnes@orchid-ortho.com |
| Date Prepared: | November 20, 2014 |
| Name of Device:: | Cooper Surgical ALLY Uterine Positioning<br>System™ (UPS) |
| Common/Usual Name: | Holder, Manipulator, Positioner, Arm |
| Classification Name: | Unclassified |
| Device Class: | Class II |
| Product Code: | LKF |
| Predicate Device(s): | Intuitive Surgical "Probe Holder System,<br>K071405 |
Device Description;
The Cooper Surgical ALLY Uterine Positioning System™ (UPS) consists of a single multisegmented, articulated Arm that attaches to a standard OR bed rail and a separate, sterile distal adapter to accommodate a uterine manipulator. When unlocked, the articulation of the Arm allows the attached manipulator to be positioned by the user. The Arm is then locked in the desired position by depressing a foot pedal, activating a linear actuator which applies tension to an internal cable, drawing the segments together and locking the Arm.
The segmented design of the Arm allows lateral/medial movement from a single point, allowing the System to be attached to one side of the OR table.
The ALLY Uterine Positioning System™ (UPS) is not intended for patient contact. The Adapter Drape for the ALL Y Uterine Positioning System™ (UPS) is provided sterile (sterilized by ethylene oxide).
{5}------------------------------------------------
### (Device Description, cont.)
The ALLY UPS Adapter Drape secures either the Cooper Surgical RUMI II Uterine Manipulator Handle or the Advincula Arch Uterine Manipulator Handle to the distal end of the ALLY Arm. After attaching the Adapter, the sterile Drape is deployed to cover the non-sterile ALLY UPS Arm, maintaining the sterile field and protecting the Arm from contamination during the procedure.
## Indications for Use:
## ALLY Positioning System;
The ALL Y Uterine Positioning System™ (UPS) is intended to assist the surgical staff in mounting. positioning and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.
## ALLY Adapter Drape:
The Adapter Drape for the ALLY Uterine Positioning System™ (UPS) is intended to assist the surgical staff in mounting, positioning and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.
| Attribute | ALLY Uterine<br>Positioning System™<br>(UPS) | Intuitive Surgical<br>"Probe Holder<br>System, K071405 | Determination |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Body Contact Type | Non-Patient Contact | Non-Patient Contact | Substantially<br>Equivalent |
| Material(s) | Stainless steel, aluminum | Stainless steel,<br>aluminum | Substantially<br>Equivalent |
| Environment | Operating room | Operating room | Substantially<br>Equivalent |
| Indications for Use,<br>ALLY Positioning<br>System | The ALLY Uterine<br>Positioning System™<br>(UPS) is intended to assist<br>the surgical staff in<br>mounting, positioning and<br>holding uterine<br>manipulators during<br>gynecological laparoscopic<br>surgical procedures. It is<br>intended for use by trained<br>operating room personnel<br>in an operating room<br>environment. | The Probe Holder<br>System is intended to<br>assist the surgical<br>staff in mounting,<br>positioning and<br>holding a uterine<br>manipulator during<br>gynecological<br>laparoscopic surgical<br>procedures. It is<br>intended for use by<br>trained operating<br>room personnel in an<br>operating room<br>environment | Substantially<br>Equivalent |
Comparison of Technological Characteristics with the Predicate Device:
{6}------------------------------------------------
| Indications for Use, | The Adapter Drape for the | (Substantially |
|----------------------|-----------------------------|----------------|
| ALLY Adapter Drape | ALLY Uterine Positioning | Equivalent) |
| | System™ (UPS) is | |
| | intended to assist the | |
| | surgical staff in mounting, | |
| | positioning and holding | |
| | uterine manipulators | |
| | during gynecological | |
| | laparoscopic surgical | |
| | procedures. It is intended | |
| | for use by trained | |
| | operating room personnel | |
| | in an operating room | |
| | environment. | |
| | | |
| | | |
| Attribute | ALLY Uterine<br>Positioning System™<br>(UPS) | Intuitive Surgical<br>“Probe Holder<br>System, K071405 | Determination |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Biocompatibility | No part of the ALLY UPS<br>is intended for patient<br>contact. The Adapter<br>portion of the Adapter<br>Drape is provided sterile as<br>it attaches to the uterine<br>manipulator within the<br>sterile field. The Drape<br>portion maintains sterility<br>in the event of inadvertent<br>patient contact, as well as<br>protects the Arm from<br>contamination during the<br>procedure. | Not included in the<br>510(k) Summary<br>Statement. | Substantially<br>Equivalent |
| Packaging | The ALLY UPS Adapter<br>Drape is heat-sealed in a<br>single sterile barrier<br>DuPont Tyvek® Pouch<br>(Uncoated 1073B Tyvek®<br>bonded to .0005”<br>polyester/0.0015/0.002<br>LDPE). The pouches are<br>labeled and packaged in<br>pressboard cartons.<br>The ALLY device and<br>non-sterile accessories are<br>shipped using packaging<br>appropriate to the design<br>and construction of the<br>instrument to prevent<br>damage during shipping | The Intuitive Surgical<br>Endoscopic<br>Instrument Control<br>System included a<br>Drape, packaged in a<br>Tyvek/polymylar peel<br>pouch and gamma<br>sterilized Reusable<br>adapters are provided<br>for Cooper Surgical<br>uterine manipulators;<br>specifically for the<br>RUMI® System<br>Uterine Manipulators<br>RUMI® UMH600 and<br>RUMI® Arch™<br>UMH700. | Substantially<br>Equivalent |
{7}------------------------------------------------
| (Packaging) | This may include foam | The Adapters are | (Substantially |
|----------------------|------------------------------|-------------------------|----------------|
| | cushioned corrugated | supplied non-sterile. | Equivalent) |
| | cardboard shippers, PETG, | | |
| | vinyl, or acrylic tubes with | | |
| | end caps, and 2 or 4 mil | | |
| | polybags for small/light | | |
| | weight components. | | |
| Stability/Shelf Life | The ALLY UPS Adapter | Sterilized by end user. | Substantially |
| | Drape is sterilized by | | Equivalent |
| | Assurance Level (SAL) of | | |
| | 10°, with an initial shelf | | |
| | life of one year. | | |
The basis of substantial equivalence of the ALL Y Uterine Positioning System™ (UPS) is the similarities in materials, design, function, performance, sterilization, and indications for use in comparison to the predicate device.
## Performance Data:
The ALL Y Uterine Positioning System™ (UPS) was tested in comparison to the predicate Intuitive Surgical "Probe Holder System, K071405 for;
1) Holding (locking) force
Testing also included; 1) Durability (cycle) testing 2) Handle push force 3) Drape pull off force 4) Adapter pull out force
The device is in compliance with the following standards: IEC 60601-1 CORR 1 & 2 2007 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements for Safety
The Adapter Drape maintained its specifications for the shelf-life as demonstrated by stability testing in accordance with ASTM F1980-07.
Clinical testing was not required to support the conclusion of substantial equivalence.
## Conclusion:
The substantial equivalence of the Cooper Surgical ALLY Uterine Positioning System (UPS) has been established by the similarities in design, materials, function, and Intended Use to the previously cleared Intuitive Surgical "Probe Holder System", K071405.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.