The ClearView Total is intended for use in laparoscopic procedures where it is desirable to delineate the vaginal fornices and the surgeon intends to remove or access intraperitonial tissue through the vagina by use of a colpotomy or culdotomy incision; such as laparoscopically assisted vaginal hysterectomies, total laparoscopic hysterectomies, while maintaining pneumoperitoneum by sealing the vagina while a colpotomy is performed.
Device Story
ClearView Total is a single-use, sterile uterine manipulator used during laparoscopic pelvic surgeries. Device features a mechanical handle for uterine manipulation and a plastic ColpoCup accessory, available in three sizes, which screws into the manipulator to delineate vaginal fornices. A pre-attached, inflatable saline-filled balloon occluder at the base of the cup seals the vagina to maintain pneumoperitoneum during colpotomy or culdotomy. Operated by surgeons in clinical/OR settings. Device provides visual guidance via high-contrast cup colors and maintains surgical field integrity. Benefits include improved access to intraperitoneal tissue and maintenance of abdominal pressure during minimally invasive hysterectomies.
Clinical Evidence
Bench testing only. No clinical data. Integrity testing included accelerated aging, balloon leak/burst testing, and cup security/break testing. Biocompatibility testing included cytotoxicity, intracutaneous reactivity irritation, and sensitization. All tests met acceptance criteria.
Technological Characteristics
Single-use, sterile, plastic uterine manipulator. Features include mechanical handle, screw-in ColpoCup (3.0cm, 3.5cm, 4.0cm), and inflatable saline-filled balloon occluder. Compatible with standard electrosurgical and harmonic tools. No electronic components or software.
Indications for Use
Indicated for patients undergoing laparoscopic procedures requiring vaginal fornix delineation and colpotomy or culdotomy access, including laparoscopically assisted vaginal hysterectomies and total laparoscopic hysterectomies.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
510(k) Summary (As Required by 21 CFR 807.92)
22 May 2014
This 510(k) Summary of safety and effectiveness for the Clinical Innovations ClearView Total is submitted with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(k) summary.
# Submitter Information
Company Name: Company Address:
Company Phone: Company Fax:
#### Device Identification
Device trade name: Device Common Name: Classification: Product Code: Device Class: Establishment Registration Number: Clinical Innovations 747 West 4170 South Murray, UT 84123 801-268-8200 801-266-7373
ClearView Total Cannula, Manipulator/Injector, Uterine Unclassified per Pre-Amendment LKF Unclassified 1722684
### Contact Person
Tom Haueter Director, Quality Assurance and Regulatory Affairs 801-260-6028 T.Haueter@clinicalinnovations.com
### Predicate Devices:
Clinical Innovations ClearView® (K940681)
ConMed VCARE (K071907)
The Koh Colpotomizer™ System (K954311)
### Device Description:
Clinical Innovations' ClearView Total is a single-use sterile device used for uterine manipulation. Uterine manipulation is essential for laparoscopies involving the female pelvic organs (uterus, tubes, ovaries) when a uterus is present. Uterine manipulators may be helpful when clinicians perform tubal ligations, diagnostic laparoscopies for evaluating pelvic pain and infertility, treatment of endometriosis, removal of pelvic scars (adhesions) involving the uterus, fallopian tubes and ovaries, treatment of ectopic pregnancy, removal of uterine fibroids, removal of ovarian cysts, removal of ovaries, tubal repair, laparoscopic hysterectomy, laparoscopic repair of pelvic bowel or bladder, sampling of pelvic lymph nodes, laparoscopic bladder suspension procedures for treatment of incontinence, and biopsy of pelvic masses.
{1}------------------------------------------------
The ColpoCup accessory is a plastic cup which is mechanically screwed into the uterine manipulator. The ColpoCup is compatible with typical surgical devices, including harmonics and electrosurgical tools. Three different sizes of ColpoCups will be included with the device; 3.0cm, 3.5cm. and 4.0cm. Each ColpoCup will be a high contrast color in order to provide the surgeon with clear visibility during laparoscopic dissection.
At the base of the ColpoCup, past the tip pivot point, is a pre-attached Occluder constructed of an inflatable balloon and will be included to seal off the vagina and prevent pnuemoperitoneum loss. The Occluder Balloon is connected to a separate inflation valve which is located proximally from the balloon and allows for inflation after placement.
### Indications for Use:
The ClearView Uterine Manipulator Device with ColpoCup and Occluder is intended for use in laparoscopic procedures where it is desirable to delineate the vaginal fornices and the surgeon intends to remove or access intraperitoneal tissue through the vagina by use of a colpotomy or culdotomy incision; such as laparoscopically assisted vaginal hysterectomies, total laparoscopic hysterectomies, while maintaining pneumoperitoneum by sealing the vagina while a colpotomy is performed.
### Indications for Use Comparison:
| Property | Result |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Laparoscopic procedures where it is<br>desirable to delineate the vaginal<br>fornices and the surgeon intends to<br>remove or access intraperitoneal tissue<br>through the vagina by use of a<br>colpotomy or culdotomy incision | Similar to predicate device, The Koh Colpotomizer™ System<br>Similar to predicate device, Clinical Innovations® ClearView |
| Laparoscopically assisted vaginal<br>hysterectomies. | Similar to predicate device, the ConMed VCARE<br>Similar to predicate device, Clinical Innovations® ClearView |
| Total laparoscopic hysterectomies | Similar to predicate device, the ConMed VCARE |
# Technical Characteristics Comparison:
| Property | Result |
|--------------------------------------------|---------------------------------------------------------------|
| Sterile, Single-use | Similar to predicate device, ConMed VCARE |
| | Similar to predicate device, Clinical Innovations * ClearView |
| Distal holes that allow suture fixation if | Similar to predicate device, ConMed VCARE |
| desired | |
| Three different cup sizes; 3.5cm, 4.0cm, | Similar to predicate device, The Koh Colpotomizer™ System |
| 4.5cm | |
| Distal handle for uterus manipulation | Similar to predicate device, The Koh Colpotomizer™ System |
| | Similar to predicate device, Clinical Innovations ClearView |
| Occluding balloon to maintain proper | Similar to predicate device, The Koh Colpotomizer™ System |
| peumoperitoneum pressure during | |
| surgery. | |
| Inflatable (saline-filled) balloon for | Similar to predicate device, ConMed VCARE |
| insertion into uterus | Similar to predicate device. Clinical Innovations "ClearView |
{2}------------------------------------------------
### Summary of Studies
Clinical Innovations performed device integrity testing to support the claim that the device is substantially equivalent to the predicate devices. All device integrity tests for the ClearView Total met the specified requirements, which consisted of:
- Accelerated Age Testing ●
- Balloon Leak/Burst Testing .
- . Cup Security and Cup Break
Clinical Innovations conducted the following biocompatibility testing in accordance to relevant industry standards:
- Cytotoxicity .
- Intracutaneous Reactivity Irritation .
- . Sensitization
All testing met the appropriate acceptance criteria.
### Basis for Substantial Equivalence:
Clinical Innovations believes that the ClearView Total is substantially equivalent based on the similarities to the predicate devices in indications and technical properties. Further, the ClearView Total introduces no new intentions, indications, or technical properties than exist currently in the previously cleared predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized emblem of an abstract human figure with three flowing lines, representing life, knowledge, and service. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 28, 2014
Clinical Innovations, LLC Tom Haueter Director, Quality and Regulatory Affairs 747 West 4170 South Murray, UT 84123
Re: K131781
> Trade/Device Name: ClearView Total Regulation Number: None Regulation Name: None Regulatory Class: Unclassified Product Code: LKF Dated: May 1, 2014 Received: May 2, 2014
Dear Tom Haueter,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Tom Haueter
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the . Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/7 description: The image shows the name "Herbert P. Lerner - S" in a bold, sans-serif font. The letters are large and clearly legible. There is a circular emblem between the "P." and "Lerner" which appears to be a seal or logo, adding a formal touch to the presentation of the name. The overall impression is clean and professional.
for Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use 510(k) Number (if known): K131781 Device Name: ClearView Total
Indications for Use:
The ClearView Total is intended for use in laparoscopic procedures where it is desirable to delineate the vaginal fornices and the surgeon intends to remove or access intraperitonial tissue through the vagina by use of a colpotomy or culdotomy incision; such as laparoscopically assisted vaginal hysterectomies, total laparoscopic hysterectomies, while maintaining pneumoperitoneum by sealing the vagina while a colpotomy is performed.
> Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D)(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1
> > Herbert P. Lerner -S 2014.05.28 13:13:53 -04'00'
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.