K970961 · Richard Wolf Medical Instruments Corp. · KNF · Apr 10, 1998 · Obstetrics/Gynecology
Device Facts
Record ID
K970961
Device Name
VAPORIZATION ELECTRODES
Applicant
Richard Wolf Medical Instruments Corp.
Product Code
KNF · Obstetrics/Gynecology
Decision Date
Apr 10, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
These specific EVAP electrodes can be used for vaporization, ablation, coagulation, and or resection of soft tissue in the urology tract and/or fibroids or myomas. The electrodes can be used for endometrial ablation and endometriosis.
Device Story
EVAP electrodes are electrosurgical instruments used for soft tissue management in urology and gynecology. Constructed from medical-grade stainless steel, teflon, new silver, and Hytrel tubing, some versions feature gold-plated rollers. The device is used by clinicians to perform vaporization, ablation, coagulation, and resection of soft tissue, including fibroids, myomas, and endometriosis. Operation involves applying high-frequency electrical energy to tissue; coagulation depth is controlled via wattage adjustment and irrigation to mitigate heat damage. The device is intended for prescription use.
Clinical Evidence
Bench testing was performed to meet ANSI/AAMI standards for high-frequency devices. Clinical evidence is limited to canine model testing, which demonstrated that coagulation depth increases with higher wattage but remains within a 2-3mm range. Irrigation was shown to reduce heat damage risk. No human clinical data provided.
Technological Characteristics
Materials: medical-grade stainless steel, teflon, new silver, Hytrel tubing, optional gold-plated rollers. Principle: high-frequency electrosurgical energy for tissue ablation/coagulation. Standards: ANSI/AAMI High Frequency Devices. Form factor: electrode/roller electrode.
Indications for Use
Indicated for vaporization, ablation, coagulation, and resection of soft tissue in the urology tract, fibroids, myomas, endometrial ablation, and endometriosis.
Regulatory Classification
Identification
A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.
{0}
APR-06-1998 14:45 RICHARD WOLF GA DEPT. 1 847 913 0924 P.07
Vernon Hills, Illinois 60061
Phone: 847.913.1113
Fax: 847.913.1488
K970961
P192
RICHARD WOLF
MEDICAL INSTRUMENTS CORPORATION
APR 10 1998
Summary of Safety and Effectiveness
| Submitter | | Date of Preparation: April 6, 1998 | |
| --- | --- | --- | --- |
| Company / Institution Name: Richard Wolf Medical Instruments Corporation | | FDA establishment registration number: 1418479 | |
| Division Name (if applicable): N.A. | | Phone Number (include area code): (847) 913-1113 | |
| Street Address: 353 Corporate Woods Parkway | | FAX number (include area code): (847) 913-0924 | |
| City: Vernon Hills | State/Province: Illinois | Country: USA | ZIP/Postal Code 60061 |
| Contact Name: Robert L. Casarsa | | | |
| Contact Title: Quality Assurance Manager | | | |
| Product Information: | | | |
| Trade name: EVAP Electrodes | | Model number: | |
| Common name: Roller Electrode
Coagulation Electrode
Electrosurgical Electrode | | Classification name: Electrode, Electrosurgical | |
| Information on devices to which substantial equivalence is claimed: | | | |
| 510(k) Number | Trade or Proprietary or Model Name | Manufacturer | |
| 1. | 1. See contents of submission | 1. | |
| 2. | 2. | 2. | |
| 3. | 3. | 3. | |
| 4. | 4. | 4. | |
| 5. | 5. | 5. | |
8 - 1
{1}
APR-06-1998 14:45 RICHARD WOLF QA DEPT. 1 847 913 0924 P.08 MEDICAL INSTRUMENTS CORPORATION
$$
\begin{array}{l}
\angle 97^\circ 0961 \\
\rho_2 \rho_2 \\
\end{array}
$$
1.0 Description
The EVAP electrode is an electrosurgical device made of medical grade stainless steel, teflon, new silver, and Hytrel tubing. Some EVAP versions have gold plated rollers.
2.0 Intended Use
These specific EVAP electrodes can be used for vaporization, ablation, coagulation, and or resection of soft tissue in the urology tract and/or fibroids or myomas. The electrodes can be used for endometrial ablation and endometriosis.
3.0 Technological Characteristics
There are no significant technological characteristics of the new device compared to the existing device.
4.0 Substantial Equivalence
The EVAP electrodes are substantially equivalent to devices sold or previously sold by Richard Wolf Medical Instruments, Circon, and ProSurg. Specifically Richard Wolf models approved by 510(k) K953983, Circon model VE-B, and proSurg model SB-24W.
5.0 Performance Data
Devices were tested to meet the appropriate sections of the ANSI/AAMI standard on High Frequency Devices.
6.0 Clinical Tests
Using canine models, it has been shown that coagulation depth increases with higher wattage, but does not exceed 2-3mm. Irrigation reduces the risk of heat damage. Long term morbidity and healing were not known.
7.0 Data Conclusions
The Richard Wolf EVAP Electrode is equivalent to existing electrodes, particularly those approved by the Richard Wolf 510(k) K953983, ProSurg, and ACMI/Circon.
By: Robert L. Casarsa
Quality Assurance Manager
Date: Apr 3, 98
8 - 2
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
APR 10 1998
Mr. Robert L. Casarsa
Quality Assurance Manager
Richard Wolf Medical Instruments
353 Corporate Woods Parkway
Vernon Hills, IL 60061
Re: K970961
EVAP Electrodes
Dated: January 16, 1998
Received: January 20, 1998
Regulatory Class: II
21 CFR 884.4160/Procode: 85 KNF
Dear Mr. Casarsa:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{3}
APR-06-1998 14:44 RICHARD WOLF QA DEPT. 1 847 913 0924 P.06
510(k) Number (if known): K970961
Device Name: Vaporization Electrodes
## Indications for Use:
These specific EVAP electrodes can be used for vaporization, ablation, coagulation, and or resection of soft tissue in the urology tract and/or fibroids or myomas. The electrodes can be used for endometrial ablation and endometriosis.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IS NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODF)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K970961
Prescription Use ☑ Per CFR 21 CFR 801.103
OR
Over-The Counter Use
1 - 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.