KSEA BIPOLAR ELECTRODE

K122983 · KARL STORZ Endoscopy-America, Inc. · HIN · Jun 26, 2013 · Obstetrics/Gynecology

Device Facts

Record IDK122983
Device NameKSEA BIPOLAR ELECTRODE
ApplicantKARL STORZ Endoscopy-America, Inc.
Product CodeHIN · Obstetrics/Gynecology
Decision DateJun 26, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.4150
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Karl Storz Bipolar Electrode is intended for use in tissue cutting, removal and desiccation as required or encountered in gynecologic hysteroscopic electrosurgical procedures for the treatment of intrauterine myomas, polyps, adhesions and septa, and benign conditions requiring endometrial ablation. Procedures include: Excision of intrauterine myomas Excision of intrauterine polyps Lysis of intrauterine adhesions Incision of uterine septa Endometrial ablation

Device Story

Single-use cutting loop electrode; used with 4 mm resectoscope (24 or 26 Fr sheath) and external high-frequency generator via high-frequency cable. Operates by delivering electrical energy to tissue for cutting, removal, and desiccation. Used in clinical settings by physicians for gynecologic hysteroscopic procedures. Provides mechanical cutting and thermal desiccation; enables treatment of intrauterine pathologies. Benefits include minimally invasive access to uterine cavity for targeted tissue management.

Clinical Evidence

Bench testing only. Studies included comparison of gas production and visual depth of thermal penetration. Device also underwent sterilization, biocompatibility testing, and non-clinical conformance testing to IEC 60601-1, IEC 60601-2-2, and IEC 60601-2-18 standards.

Technological Characteristics

Single-use bipolar cutting loop electrode. Energy source: external high-frequency generator. Connectivity: high-frequency cable. Compatibility: 4 mm resectoscope with 24 or 26 Fr sheath. Standards: IEC 60601-1 (general safety), IEC 60601-2-2 (HF surgical equipment), IEC 60601-2-18 (endoscopic equipment).

Indications for Use

Indicated for gynecologic hysteroscopic electrosurgical procedures in adult women for treatment of intrauterine myomas, polyps, adhesions, septa, and benign conditions requiring endometrial ablation. Contraindicated in patients with acute cervicitis, pregnancy, cervical/uterine malignancy, acute pelvic inflammatory disease, unaddressed adnexal pathology, or where endoscopic methods are contraindicated. Not for tubal sterilization or CNS interventions.

Regulatory Classification

Identification

A bipolar endoscopic coagulator-cutter is a device used to perform female sterilization and other operative procedures under endoscopic observation. It destroys tissue with high temperatures by directing a high frequency electrical current through tissue between two electrical contacts of a probe. This generic type of device may include the following accessories: an electrical generator, probes, and electrical cables.

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance 2/12/90 (K-90),” and (iii) “Guidance (‘Guidelines’) for Evaluation of Laproscopic Bipolar and Thermal Coagulators (and Accessories),” (2) International Electrotechnical Commission's IEC 60601-1-AM2 (1995-03), Amendment 2, “Medical Electrical Equipment—Part 1: General Requirements for Safety,” (3) American National Standards Institute/American Association for Medical Instrumentation's HF-18, 1993, “Electrosurgical Devices,” (4) Labeling: (i) Indication: For female tubal sterilization, and (ii) Instructions for use: (A) Destroy at least 2 centimeters of the fallopian tubes, (B) Use a cut or undampened sinusoidal waveform, (C) Use a minimum power of 25 watts, and (D) For devices with ammeters: continue electrode activation for 5 seconds after the visual endpoint (tissue blanching) is reached or current flow ceases indicating adequate tissue destruction.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K122983 Page 1 of 4 JUN 2 6 2013 Image /page/0/Picture/2 description: The image shows the logo for Karl Storz Endoskope. The word "STORZ" is written in large, bold letters at the top of the logo. Below the word "STORZ" is the phrase "KARL STORZ -- ENDOSKOPE" written in smaller letters. The logo is black and white. #### 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge. | Applicant: | KARL STORZ Endoscopy - America, Inc.<br>2151 E. Grand Ave<br>El Segundo, CA 90245<br>(424) 218-8379 Tel<br>(424) 218-8519 Fax | | | |------------------------|-------------------------------------------------------------------------------------------------------------------------------|--|--| | Contact: | Winkie Wong<br>Regulatory Affairs Specialist | | | | Date: | September 14, 2012 | | | | Device Identification: | | | | | Common Name: | Bipolar Electrode | | | | Classification Name: | Coagulator-Cutter, Endoscopic, Bipolar (And Accessories) [21 CFR 884.4150, Product Code HIN] | |------------------------|----------------------------------------------------------------------------------------------| | Trade Name: (optional) | The Karl Storz Bipolar Electrode | Device Description: The Karl Storz Bipolar Electrode is a single-use.cutting loop electrode that is designed to be used in combination with a 4 mm resectoscope with a 24 or 26 Fr sheath that is connected to a generator via a high frequency cable. Indications: The Karl Storz Bipolar Electrode is intended for use in tissue cutting, removal and desiccation as required or encountered in gynecologic hysteroscopic electrosurgical procedures for the treatment of intrauterine myomas, polyps, adhesions and septa, and benign conditions requiring endometrial ablation. Procedures include: Excision of intrauterine myomas Excision of intrauterine polyps Lysis of intrauterine adhesions Incision of uterine septa Endometrial ablation {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for Karl Storz Endoskope. The logo is in black and white and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. Contraindications: The Karl Storz Bipolar Electrode must not be used if the patient's health is endangered due to general condition of the patient, or if the endoscopic method as such is contraindicated. Karl Storz Bipolar Electrodes are not intended for use in tubal sterilization procedures. The use of this device is contraindicated in patients with the following conditions: - Acute cervicitis . - . Pregnancy - Cervical or Uterine malignancy - ♥ Acute pelvic inflammatory disease - . Unaddressed adnexal pathology While fluids must always be monitored during use, exercise extreme caution and very close fluid monitoring in patients with severe cardiopulmonary disease. Karl Storz Bipolar Electrodes must not be used for interventions on the CNS. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for Karl Storz Endoskope. The logo consists of the word "STORZ" in large, bold, sans-serif font, with a stylized "O" that resembles a target. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller, sans-serif font. The logo is black and white. Predicate Devices: The KSEA Bipolar Electrode is substantially equivalent to the predicate Bipolar Electrotome (K061541) and GYNECARE VERSAPOINT Bipolar Electrosurgery System (K111751) #### Technological Characteristics: #### Bipolar Electrotome The subject KSEA Electrode is substantially equivalent to Bipolar Electrotome (K061541) with the same features, dimensions, body contact and insulation materials, and intended use. The only difference between the subject electrodes and the predicate electrodes is that the subject electrodes are indicated for gynecological procedures whereas the predicate device is indicated for urological procedures. The Safety and performance equivalence are demonstrated by the testing of IEC 60601-1, IEC 60601-2-2 and IEC 60601-2-18. #### GYNECARE VERSAPOINT II Bipolar Electrosurgery System There is no significant technological difference between the predicate and subject device. Both devices are bipolar and have the same intended use. Both uses a generator for the source of energy. The minor differences in material do not affect the safety and effectiveness of the devices. The performance equivalence of the devices can be demonstrated by the passing criteria of the IEC 60601-2-2: Particular requirements for the basic safety and essential performance of high frequency surgery equipment and high frequency surgical accessories. Both the predicate and subject device have undergone and passed IEC 60601-2-2 to demonstrate the equivalence to the essential performance of the HF equipment and accessories. For additional safety requirements equivalence, both subject and predicate devices have also undergone and pass IEC 60601-1 and IEC 60601-2-18. #### Performance and Safety Testing: Two bench studies, Comparison of Gas Product and Visual Depth of Thermal Penetration, are used to demonstrate the performance substantial equivalence of the subject and predicate devices. In addition, both predicate and subject devices have undergone sterilization, biocompatibility testing and non-clinical conformance standard testing: IEC 60601-1, IEC 60601-2-2 and IEC 60601-2-18. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Karl Storz Endoskope. The logo is black and white and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. ## Conclusions: The test results presented in the testing report of abovementioned testing have demonstrated the substantial equivalence in both safety and performance of the predicate and subject device. Thus, the minor differences between the predicate and subject devices do not raise new issues of safety and effectiveness. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three stylized human figures or birds in flight. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 #### June 26, 2013 KARL STORZ Endoscopy-America, Inc. % Ms. Winkie Wong Regulatory Affairs Specialist 2151 E. Grand Avenue EL SEGUNDO CA 90245 Re: K122983 Trade/Device Name: Karl Storz Bipolar Electrode Regulation Number: 21 CFR§ 884.4150 Regulation Name: Bipolar endoscopic coagulator-cutter and accessories Regulatory Class: II Product Code: HIN Dated: May 22, 2013 Received: May 24, 2013 Dear Ms. Wong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {5}------------------------------------------------ #### Page 2 - Ms. Winkie Wong You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Benjamin R. Fisher -S Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is black and white and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters. The logo is simple and easy to read. ### 510(k) Number (if known): K122983 Device Name: Karl Storz Bipolar Electrode Indications for Use: The Karl Storz Bipolar Electrode is intended for use in tissue cutting, removal and desiccation as required or encountered in gynecologic hysteroscopic electrosurgical procedures for the treatment of intrauterine myomas, polyps, adhesions and septa, and benign conditions requiring endometrial ablation. Procedures include: of - Excision of intrauterine myomas . - Excision of intrauterine polyps . - Lysis of intrauterine adhesions . - Incision of uterine septa . - Endometrial ablation . Prescription Use _ x (Part 21 CFR 801 Subpart D) AND/OR .............................................................................................................................................................................. Over-The-Counter Use _ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # Herbert P. Lerner -S K122983 Page 004 - 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...