K122605 · Gyrus Acmi, Inc. · HIN · Dec 5, 2012 · Obstetrics/Gynecology
Device Facts
Record ID
K122605
Device Name
PKS BILL
Applicant
Gyrus Acmi, Inc.
Product Code
HIN · Obstetrics/Gynecology
Decision Date
Dec 5, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.4150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PKS™ (Plasmakinetic System) BiLL™ (Bipolar Laparoscopic Loop) instrument is a 5mm bipolar electrosurgical device. The device is intended to be used for the amputation of the mobilised uterus during Laparoscopic Supracervical (Subtotal) Hysterectomy and the resection of devascularized subserosal pedunculated myomas. It is used in conjunction with the Gyrus ACMI G400 Workstation Generator.
Device Story
5mm bipolar electrosurgical instrument; used for uterine amputation during laparoscopic supracervical hysterectomy and resection of pedunculated myomas. Operates with Gyrus ACMI G400 Workstation Generator; utilizes Bipolar PK (Plasma Kinetic) technology. Device includes identification capacitor in connector cable to communicate optimal power parameters to generator. Single-use, sterile, gamma-irradiated. Operated by physicians in surgical settings. Provides electrosurgical cutting/coagulation via high-frequency RF energy. Benefits include efficient tissue resection and hemostasis during minimally invasive gynecological procedures.
Clinical Evidence
No clinical data. Pre-clinical evaluation performed using extirpated human uterine tissue with three physicians conducting mock procedures. Qualitative feedback confirmed device met performance expectations and user needs. Bench testing included design verification, validation, and electrical safety testing per IEC 60601. Biocompatibility testing passed per ISO-10993-1.
Technological Characteristics
5mm bipolar electrosurgical loop; high-frequency RF energy source. Uses identification capacitor for generator communication. Compliant with IEC 60601 (electrical/thermal safety, EMI). Biocompatible materials per ISO-10993. Sterile, single-use, gamma-irradiated.
Indications for Use
Indicated for patients undergoing Laparoscopic Supracervical (Subtotal) Hysterectomy for uterine amputation and resection of devascularized subserosal pedunculated myomas.
Regulatory Classification
Identification
A bipolar endoscopic coagulator-cutter is a device used to perform female sterilization and other operative procedures under endoscopic observation. It destroys tissue with high temperatures by directing a high frequency electrical current through tissue between two electrical contacts of a probe. This generic type of device may include the following accessories: an electrical generator, probes, and electrical cables.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance 2/12/90 (K-90),” and
(iii) “Guidance (‘Guidelines’) for Evaluation of Laproscopic Bipolar and Thermal Coagulators (and Accessories),”
(2) International Electrotechnical Commission's IEC 60601-1-AM2 (1995-03), Amendment 2, “Medical Electrical Equipment—Part 1: General Requirements for Safety,”
(3) American National Standards Institute/American Association for Medical Instrumentation's HF-18, 1993, “Electrosurgical Devices,”
(4) Labeling:
(i) Indication: For female tubal sterilization, and
(ii) Instructions for use:
(A) Destroy at least 2 centimeters of the fallopian tubes,
(B) Use a cut or undampened sinusoidal waveform,
(C) Use a minimum power of 25 watts, and
(D) For devices with ammeters: continue electrode activation for 5 seconds after the visual endpoint (tissue blanching) is reached or current flow ceases indicating adequate tissue destruction.
{0}------------------------------------------------
Olympus Surgical Technologies of America · Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772
| 510(K) SUMMARY |
|----------------|
|----------------|
PKS" BiLL""
General Information
Manufacturer:
510(k) Submitter:
Contact Person:
Date Prepared:
Classification Name:
Trade Name:
Generic/Common Name:
Bipolar Loop / Bipolar Laparoscopic Loop
Predicate Device
Gyrus ACMI PKS™ BiLLTM
K11059
Indications for Use
The PKS™ (Plasmakinetic System) BiLL™ (Bipolar Laparoscopic Loop) instrument is a 5mm bipolar electrosurgical device. The device is intended to be used for the amputation of the mobilised uterus during Laparoscopic Supracervical (Subtotal) Hysterectomy and the resection of devascularized subserosal pedunculated myomas. It is used in conjunction with the Gyrus ACMI G400 Workstation Generator.
Product Description
The Bipolar Laparoscopic Loop is a single use disposable high frequency RF bipolar accessory to be used in conjunction with the G400 generator. It is available in an 88mm x 227mm loop size. The device is sterile for single use sterilized by gamma irradiation to an SAL of 10-6.
Page 10
Special 510(k) Notifi
DEC 0 5 2012
Special 510(k) Notification
(508) 804-2690 neil.kelly@Olympus-OSTA.com
Neil Kelly
Gyrus Medical, Ltd. Fortran Road, St. Mellons
Cardiff CF3 OLT ERN: 9617070
Gyrus ACMI Inc. 136 Turnpike Rd.
ERN: 3003790304
Southborough, MA 01772-2104
August 24, 2012
Coagulator-Cutter, Endoscopic, Bipolar (And Accessories) Obstetrics/Gynecology HIN 21 CFR 884.4150 Class II
PKSTM BILLTM
BiLap Loop / Bipolar Laparoscopic Loop
{1}------------------------------------------------
Olympus Surgical Technologies of America Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772
# Performance Data Technological features and Substantial Equivalence
The PK STM BiLL™ utilizes features incorporated into the following legally marketed predicate device:
- The bipolar PKS™ BiLL™ connects to the same electrosurgical generator, G400 . (K050550) as the predicate PKSTM BiLLTM (K111059).
- The PKSTM BiLL™ uses Bipolar PK (Plasma Kinetic) technology and contains . an identification capacitor embedded in the single use connector cable will be recognized by the generator to set default optimal power output parameters for the subject instrument. This remained unchanged and identical to the predicate PKS™ BiLLTM (K111059).
- . The mechanical design features of the PKS™ BiLL™ are similar to that of the predicate PKSTM BiLLTM (K111059).
The proposed PKS™ BiLL™ has the identical Indications for use as the predicate PKSTM BiLLTM.
The PKS™ BiLLTM instrument is compliant to electrical standards specifically to those applicable sections of IEC 60601 incorporating electrical, thermal safety and Electromagnetic Interference. .
The PKSTM BiLLTM instrument uses materials that are well established and used in other GYRUS ACMI FDA cleared medical devices. Biocompatibility testing on all patient contacting parts has been performed in compliance to the relevant requirements of ISO-10993.
The PKS™ BiLL™ instrument is packaged and sterilized as a sterile single use device.
#### Bench, Biocompatibility, and General Safety Summary
Design verification testing was carried out to ensure the device meets the product specifications, and design validation was carried out to ensure the product meets the user requirements. All testing was completed successfully. Electrical testing was also carried out to IEC 60601 and passed as well.
In addition the representative final product was tested to and passed ISO-10993-1. The product has been determined to be biocompatible.
#### Pre-clinical Evaluation Summary
In addition a pre-clinical evaluation was carried out using extirpated tissue (Human Uterine), and three (3) Physicians. Thirty-two (32) questions were asked of each of the three physicians after they carried out a mock procedure using the proposed device. These questions ranged from ease of understanding the IFU, to performance related questions specific to the
{2}------------------------------------------------
Olympus Surgical Technologies of Američa Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772
procedure. All three physicians were happy with the proposed device and felt it met their needs and performed to their expectations.
ﺎ
## Summary
The proposed Gyrus ACMI Inc. Bipolar Laparoscopic Loop (PKS™ BiLL™), as described in this submission, is substantially equivalent to the predicate in intended use, materials, principles of operation and fundamental scientific technology and raises no new issues of safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three curved lines that resemble a person embracing another person.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
December 5, 2012
Gyrus ACMI, Inc. % Mr. Neil Kelly Regulatory Affairs Specialist 136 Turnpike Road SOUTHBOROUGH MA 01772
Re: K122605
Trade/Device Name: Gyrus ACMI® Bipolar Laparoscopic Loop (PKSTM BILLTM) Regulation Number: 21 CFR§ 884.4150 Regulation Name: Bipolar endoscopic coagulator-cutter and accessories Regulatory Class: II Product Code: HIN Dated: November 2, 2012 Received: November 5, 2012
Dear Mr. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Mr. Neil Kelly
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
**Herbert** R. **Lerner**
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use Statement
510(k) Number:
: K122605
Device Name: Gyrus ACMI® Bipolar Laparoscopic Loop (PKS™ BiLL™)
Indications for Use:
The PKS™ (Plasmakinetic System) BiLL™ (Bipolar Laparoscopic Loop) instrument is a 5mm bipolar electrosurgical device. The device is intended to be used for the amputation of the mobilised uterus during Laparoscopic Supracervical (Subtotal) Hysterectomy and the resection of devascularized subserosal pedunculated myomas. It is used in conjunction with the Gyrus ACMI G400 Workstation Generator.
Prescription Use: X
OR Over-the-Counter Use:
(Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Herbert P. Lerner**
| (Division Sign-Off) | |
|---------------------------------------------|--------|
| Division of Reproductive, Gastro-Renal, and | |
| Urological Devices | |
| 510(k) Number | K22605 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.