HISTOBRUSH
Device Facts
| Record ID | K111681 |
|---|---|
| Device Name | HISTOBRUSH |
| Applicant | Puritan Medical Products Company, LLC |
| Product Code | HHT · Obstetrics/Gynecology |
| Decision Date | Jan 19, 2012 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.4530 |
| Device Class | Class 2 |
Indications for Use
To obtain endocervical cells for Pap smear cytology test. The Puritan Histobrush is used to obtain cervical samples for transfer to a microscope slide or specimen preservation vial.
Device Story
The Puritan Histobrush Model 2188 is a manual, single-use, non-sterile cervical cytology brush. It consists of nylon bristles secured by a stainless steel 304 wire to a polystyrene handle. The device is used by clinicians to collect endocervical cell samples for Pap smear cytology. The collected samples are transferred to a microscope slide or specimen preservation vial for diagnostic evaluation. The device functions as a mechanical tool for cell collection.
Clinical Evidence
Bench testing only. Testing included biocompatibility, tensile strength of brush fibers, force to remove fibers from wire, and force to remove wire from handle.
Technological Characteristics
Manual cervical cytology brush. Materials: Nylon bristles, stainless steel 304 wire, polystyrene handle. Dimensions: 0.750 inch tip length, 0.250/0.200 inch tip diameter, 8 inch overall length. Single-use, non-sterile.
Indications for Use
Indicated for obtaining endocervical cells for Pap smear cytology testing.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.