K181672 · Sure, Inc. · NKA · Jun 28, 2019 · Obstetrics/Gynecology
Device Facts
Record ID
K181672
Device Name
SureTouch Mobile Pressure Mapping System
Applicant
Sure, Inc.
Product Code
NKA · Obstetrics/Gynecology
Decision Date
Jun 28, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2990
Device Class
Class 2
Indications for Use
The SureTouch Mobile Pressure Mapping System is intended to produce a surface pressure map of the breast as an aid in documenting palpable breast lesions identified during a clinical breast examination. The SureTouch Mobile Pressure Mapping System is intended for use by a qualified healthcare professional trained in its use and is not for home use.
Device Story
SureTouch Mobile Pressure Mapping System aids documentation of palpable breast lesions. Device consists of rechargeable, battery-powered handheld wand with 30x40 mm array of 192 pressure-sensing elements; tablet display; calibration/training phantom; scale. Healthcare professional moves wand across breast; sensor collects tactile data; data wirelessly transferred to Android-based tablet. System generates tactile images; provides information on lesion size, shape, hardness. User manually inputs lesion location on display. Used in clinical settings by trained professionals. Benefits include standardized documentation of palpable findings. Includes disposable polyurethane sheath and lotion. Calibration performed using provided phantom and scale.
Clinical Evidence
Bench testing only. No clinical data. Testing included intra- and inter-observer and inter-system accuracy and reproducibility per FDA Class II Special Controls Guidance. Force gauge validation, algorithm output sensitivity to calibration errors, and phantom testing (sensor accuracy under uniform force and pad aging) met all acceptance criteria.
Technological Characteristics
Handheld wand with 192-element (30x40 mm) pressure sensor array. Android-based touchscreen tablet. Wireless data transfer. Battery-powered. Uses disposable polyurethane sheaths and lotion. Calibration via external phantom and scale. Electrical safety per ANSI/AAMI EN60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for qualified healthcare professionals to document palpable breast lesions (5mm to 25x35 mm) identified during clinical breast examinations. Not for home use.
Regulatory Classification
Identification
A breast lesion documentation system is a device for use in producing a surface map of the breast as an aid to document palpable breast lesions identified during a clinical breast examination.
Special Controls
*Classification.* Class II (special controls). The device, when it is a breast examination recording sheet, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9. The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Breast Lesion Documentation System.” See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it.
June 28, 2019
Sure, Inc. Laurence Harvey Senior Quality Engineer Certified Compliance Solutions 11665 Avena Place, Suite 203 San Diego, CA 92128
Re: K181672
> Trade/Device Name: SureTouch Mobile Pressure Mapping System Regulation Number: 21 CFR§ 884.2990 Regulation Name: Breast Lesion Documentation System Regulatory Class: II Product Code: NKA Dated: May 17, 2019 Received: May 28, 2019
Dear Laurence Harvey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon M. Andrews Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181672
Device Name SureTouch Mobile Pressure Mapping System
#### Indications for Use (Describe)
The SureTouch Mobile Pressure Mapping System is intended to produce a surface pressure map of the breast as an aid in documenting palpable breast lesions identified during a clinical breast examination. The SureTouch Mobile Pressure Mapping System is intended for use by a qualified healthcare professional trained in its use and is not for home use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary K181672 - SureTouch Mobile Pressure Mapping System
| Submitter: | Sure, Inc |
|--------------------|----------------------------------------------------------|
| Address: | 1404 Granvia Altamira<br>Palos Verdes Estates, CA. 90274 |
| Phone number: | (833) 787-3642 |
| Contact person: | David Ables, CTO |
| Phone number: | (833) 787-3642 x 701 |
| Date prepared: | June 25, 2019 |
| Trade name: | SureTouch Mobile Pressure Mapping System |
| Common Name: | Documentation System for Breast Lesions |
| Regulation Number: | 21 CFR 884.2990 |
| Regulation Name: | Breast Lesion Documentation System |
| Product Code: | NKA (System, Documentation, Breast Lesion) |
| Regulatory Class: | Class II |
Predicate Device: BreastView Visual Mapping System (DEN020001). This predicate device has not been subject to any design related recalls.
### Device Description:
The SureTouch Mobile Pressure Mapping System ("SureTouch") is a computer-based device that produces a pressure map, called a tactile image, of specific areas of the breast as an aid to document lesions detected during a clinical breast exam. SureTouch utilizes a rechargeable, battery-powered hand-held wand (sensor unit) that incorporates a 30 x 40 mm array of pressure sensing elements to collect tactile data as the device is moved across the breast. Data collected using the wand are wirelessly transferred to the tablet display where they are used to generate tactile images and provide information on a lesion's size, shape and hardness. The final report includes a tactile image of each lesion along with its user inputted location. The SureTouch System also includes a calibration and training phantom, a scale to ensure correct force applied during calibration procedures, and a holder for the wand.
### Indications for Use:
The SureTouch Mobile Pressure Mapping System is intended to produce a surface pressure map of the breast as an aid in documenting palpable breast lesions identified during a clinical breast examination. The SureTouch Mobile Pressure Mapping System is intended for use by a qualified healthcare professional trained in its use and is not for home use.
{4}------------------------------------------------
| Parameter | Subject Device<br>SureTouch System<br>K181672 | Predicate Device<br>BreastView Visual<br>Mapping System<br>(DEN020001) | Comparison |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The SureTouch<br>Mobile Pressure<br>Mapping System is<br>intended to produce a<br>surface pressure map<br>of the breast as an<br>aid in documenting<br>palpable breast<br>lesions identified<br>during a clinical<br>breast examination.<br>The SureTouch<br>Mobile Pressure<br>Mapping System is<br>intended for use by a<br>qualified healthcare<br>professional trained<br>in its use and is not<br>for home use. | The Breast View<br>Visual Mapping<br>System is intended<br>for use in producing<br>a surface map of the<br>breast as an aid to<br>document palpable<br>breast lesions<br>identified during a<br>clinical breast<br>examination. | The subject and<br>predicate devices<br>have the same<br>intended use. |
| Sensor | Handheld sensor unit<br>containing a 30 x 40<br>mm rectangular array<br>of 192 sensors. | Handheld sensor unit<br>containing a 25 x 40<br>mm array of 416<br>sensors. | Different – The<br>predicate device has a<br>greater number of<br>sensors than the<br>subject device.<br>Differences in sensor<br>numbers do not raise<br>different questions of<br>Safety and<br>Effectiveness (S&E) |
| Display | Android-based<br>touchscreen tablet<br>that controls device<br>functions and<br>displays exam<br>information. | Windows-based PC<br>that controls device<br>functions and<br>displays exam<br>information. | Different - The<br>subject and predicate<br>devices use different<br>display types. This<br>difference does not<br>raise different<br>questions of S&E. |
| Calibration/Training<br>Pad and Scale | Yes – pad includes<br>simulated lesions for<br>training and is also | No | Different - The<br>subject device is<br>provided with a |
| | used for device<br>calibration in<br>conjunction with the<br>provided scale. | | standalone pad and<br>scale that are used for<br>calibrating the device.<br>These differences do<br>not raise different<br>questions of S&E. |
| Lesion<br>Documentation | Lesion size,<br>hardness, shape and<br>user-inputted<br>location | Lesion size, stiffness,<br>shape, and location. | Different – The<br>predicate device<br>includes an automated<br>lesion location<br>position system that is<br>not part of the subject<br>device that requires<br>the user to manually<br>identify the location<br>of a specific lesion on<br>the display screen.<br>This difference does<br>not raise different<br>questions of S&E. |
| Detectable Lesion<br>Size | 5mm to 25x 35 mm | 5-40 mm | Similar |
| Disposable Sheath<br>and Lotion Use | Device to be used<br>with a polyurethane<br>sheath and lotion.<br>Specifications for<br>suitable,<br>commercially-<br>available sheaths and<br>lotion provided in<br>labeling | Device to be used<br>with a polyurethane<br>sheath and lotion. | Same |
# Comparison of Subject and Predicate Devices:
{5}------------------------------------------------
As shown above, the indications for use of the subject and predicate devices are not identical; however, their intended uses are the same (i.e., documenting palpable breast lesions detected during a clinical breast exam).
Regarding technological characteristics, the subject and predicate devices have similarities in their general designs. However, differences do exist as described in the table above (e.g., sensor number, display types, etc.). The differences identified do not raise different questions of safety and effectiveness as compared to the predicate device as stated in the table.
## Performance Testing:
The following tests have been conducted to ensure reliable design and performance under the specified testing parameters. These tests include:
{6}------------------------------------------------
- . Software information that met the recommendations for a minor level of concern software per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
- Cybersecurity information that met the recommendations in the FDA guidance document, . "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices."
- Electrical safety testing per ANSI/AAMI EN60601-1:2006+A11:2013+A12:2014 .
- Electromagnetic compatibility testing per IEC 60601-1-2:2007 (3rd edition) .
- Wireless technology information that met the recommendations in the FDA guidance ● document "Radio Frequency Wireless Technology in Medical Device."
- . Intra and inter-observer, and inter-system accuracy and reproducibility testing as described in the FDA "Class II Special Controls Guidance Document: Breast Lesion Documentation System." Testing met all acceptance criteria.
- Force gauge validation: Testing to assess the accuracy and reproducibility of the force gauge ● shown on the device display that is used for device calibration and to assess force applied during an exam. Testing met all acceptance criteria.
- Algorithm output sensitivity: Testing to assess sensitivity of the device algorithms to ● calibration errors. Testing met all acceptance criteria.
- Phantom testing:
- Testing to assess accuracy of sensor measurement when placed under a uniform O calibration force as compared to calibration force applied using the phantom and scale. Testing met all acceptance criteria.
- Testing to demonstrate that aging of the calibration and training pad did not impact o device calibration results. Testing met all acceptance criteria.
### Conclusion:
The results of the performance testing described above demonstrates that the subject device is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.