K040480 · Edan Instruments, Inc. · KNG · May 25, 2004 · Obstetrics/Gynecology
Device Facts
Record ID
K040480
Device Name
SONOTRAX ULTRASONIC POCKET DOPPLER
Applicant
Edan Instruments, Inc.
Product Code
KNG · Obstetrics/Gynecology
Decision Date
May 25, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2660
Device Class
Class 2
Indications for Use
Detection of fetal heartbeat from early gestation thru delivery. Verify fetal heart viability after patient trauma. Display of fetal heart rate. Prescription Device: Federal Law (US) restricts this device to sale by or on order of a physician. Under the direction of a physician it may be used by other health care professionals including Registered Nurses, Practical Nurses , Midwives, and Ultrasound Technicians
Device Story
Sonotrax Ultrasonic Pocket Doppler uses Doppler shift principle to detect fetal heart and artery blood flow. Device utilizes split D piezoelectric transducer; high-frequency oscillator transmits ultrasound acoustic waves into body via coupling medium. Reflected ultrasound from fetal heart/blood is received by transducer, converted to electronic signal, amplified, and filtered. Resulting baseband audio Doppler signal output via loudspeaker; fetal heart rate calculated and shown on LCD. Used in clinical settings by physicians, nurses, midwives, or ultrasound technicians to assess fetal well-being and viability. Provides real-time auditory and visual feedback to assist clinical decision-making regarding fetal status.
Clinical Evidence
Bench testing only. Device subjected to extensive safety, performance, and validation testing to ensure compliance with functional requirements and industry safety standards.
Technological Characteristics
Handheld ultrasonic fetal monitor. Uses split D piezoelectric transducer for transmission/reception. Operates via Doppler shift principle. Includes LCD for heart rate display and loudspeaker for audio output. Requires water-based coupling medium.
Indications for Use
Indicated for detection of fetal life and heart rate from early gestation through delivery, and verification of fetal heart viability following patient trauma.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
{0}------------------------------------------------
510k Submission , SONOTRAX Ultrasonic Fetal Pocket Doppler Edan Instruments, Inc.
MAY 2 5 2004
## SUMMARY
This summary of 510k safety and effectiveness information is being submitted in Accordance with 21CFD part 807.92
- 1. Submitters name, address, phone number, contact person and preparation date:
Name: Edan Instruments, Inc 4/F, New Energy Building 2009 Nanyou Road Nanshan, Shenzhen Guangdong 518054 China Phone: 86 755 26062059 86 755 26062022 Fax; Responsible person: Xie Xicheng
Official Correspondent:
William Stern Multigon Industries, Inc. 1 Odell Plaza Yonkers, N.Y. 10701 Phone: 914 376 5200 x27 Fax: 914 376 5565
Date of Preparation: 2/19/04
- 2. Device:
| Proprietary Name: | Sontotrax Ultrasonic Pocket Doppler |
|----------------------|------------------------------------------|
| Common Name: | Ultrasonic Fetal Monitor |
| Classification Name: | Fetal Ultrasonic Monitor and Accessories |
| Product Code: | KNG |
| Manufactured By: | Edan Instruments, Inc., China |
{1}------------------------------------------------
- 3. Predicate Devices:
K010889, Baby Dopplex 3000MK 2, Huntleigh Healthcare Inc. UK. K991441, Cadence Doppler Ultrasound System, Medasonics, Inc. K942441, Imexdop CT+, Imex Medical Systems, Inc. Golden, CO. K023618, Sonicaid One Fetal Heart Detector, UK
- 4. Classification Names :
Class II as per 21CFR 884-2660, Fetal Ultrasonic Monitor and accessories.
- న. Description:
The Sonotrax fetal Doppler uses the tried and true principle of Doppler shift Of an ultrasound signal to deect the blod flow within the fetal heart and arteries.
The Sonotrax Fetal Doppler uses a split D piezoelectric transducer. A high frequency oscillator supplies a continuous high frequency voltage to one half of the split D transmitter transducer. The high frequency voltage is converted to an ultrasound acoustic wave by the transducer and is transmitted to biophysical objects thru an applied coupling water based medium and moves thru biophysical objects. The acoustic ultrasound is reflected by blodd , and moving objects such as the fetal heart. The reflected ultrasound is received by the second split D receiver transducer and is converted via the piezoelectric effect into a high frequency electronic signal. The received electronic signal is amplified and detected. The result is a base band audio Doppler shifted signal which is filtered , and converted to audio via a loud speaker. At the same time the fetal heart rate is applied to and displayed on a liquid crystal counter display.
- 6. Indications for use:
Detection of fetal life .from early gestation thru delivery and as a general indication of fetal well being. It can also be used to verify fetal heart viability following patient trauma.
- 7. Contra-indications:
None known at this time.
- 8. Comparison to Predicate Devices:
{2}------------------------------------------------
The Sonotrax Basic fetal Doppler has the same device characteristics as all the predicate approved devices in item 3 above.
- ರ. Test Data:
The Sonotrax Basic Fetal Doppler device has been subjected to extensive safety, performance testing and validations before release. Final testing of the Sonotrax basic fetal Doppler includes various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications. Safety tests have further been performed to ensure the device complies with applicable industry and safety stanmdards.
The Sonotrax Basic Fetal Doppler device labeling includes instructions for safe and effective use, warnings cautions and guidance for use.
- 10. Literature Review:
A review of the literature pertaining to the safety of Doppler Blood flowmeter Has been conducted and appropriate safeguards have been incorporated in the design of the Sonotrax Basic fetal Doppler.
- 10. Conclusions:
The conclusion drawn from these tests is that the Sonotrax Basic fetal Doppler device is equivalent in safety and efficacy to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular and contains the department's name around the perimeter. In the center of the logo is an image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAY 2 5 2004
Edan Instruments, Inc. % Mr. William Stern Official Correspondent Multigon Industries, Inc. l Odell Plaza YONKERS NY 10701
Re: K040480 Trade/Device Name: Sonotrax Ultrasonic Pocket Doppler Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: 85 KNG Dated: February 23, 2004 Received: February 25, 2004
Dear Mr. Stern:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Boyden
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): k040480
SONOTRAX Ultrasonic Pocket Doppler Device Name:
Indications For Use:
Detection of fetal heartbeat from early gestation thru delivery. Verify fetal heart viability after patient trauma. Display of fetal heart rate.
Prescription Device:
Federal Law (US) restricts this device to sale by or on order of a physician. Under the direction of a physician it may be used by other health care professionals including Registered Nurses, Practical Nurses , Midwives, and Ultrasound Technicians
Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
، ،
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. bivadon
ive, Abdominal, Division of Reproduct and Radiological D 510(k) Number
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.