The FM20 is a Fetal Monitor for measuring and recording maternal contraction and fetal heart rate. Data is displayed on a front panel LCD(7") Display, recorded on a strip chart recorder and may be transmitted over RS232 port to a remote data receiver. Single/Twin fetal heart rates may be measured by means of Doppler Ultrasound. Uterine Activity is measured with an external TOCO transducer. The FM20 is intended to be used during the antepartum periods on pregnant woman.
Device Story
FM20 is a fetal monitor for antepartum use; measures fetal heart rate (FHR) and uterine contractions. Inputs: ultrasound Doppler signals from fetus; pressure signals from external TOCO transducer. Processing: circuit and digital filter extract FHR and fetal movement; calculates uterine activity. Outputs: numerical values on 7" LCD; real-time strip chart recording; audible fetal heart sounds; RS232 data transmission. Features: twin monitoring via dual ultrasound transducers; event marker for patient use; heart rate limit alarms. Used in clinical settings by healthcare providers to assess fetal status and uterine activity; aids clinical decision-making regarding fetal well-being.
Clinical Evidence
Bench testing only. Performance and safety verified via electrical, mechanical, and environmental testing per EN/IEC 60601-1 and EMC testing per EN/IEC 60601-1-2. No clinical data provided.
Indicated for pregnant women during antepartum periods for monitoring fetal heart rate (single or twin) via Doppler ultrasound and maternal uterine contractions via external TOCO transducer.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{0}------------------------------------------------
NOV - 8 2011 K110612
Page 1 of 3
Submission – EM20
Mediana Co.,Ltd.510(k)
Submission - FM
## 510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### Submitter
Mediana Co., Ltd. Dongwha Medical Instrument Complex, 1650-1, Donghwa-ri, Munmak-eup, Wonju-si, Gangwon-do, Korea Tel) +82-70-7092-9964 Fax) +82-2-542-7447 Company Contact: Amy M.H. Kim Date Summary Prepared: October 26, 2010
#### Device Name
| Trade Name: | FM20 |
|----------------------|------------------------------------------------------|
| Common Name: | System, Monitoring, Perinatal |
| Classification Name: | System, Monitoring, Perinatal (21CFR 884.2740, HGM.) |
| Classification: | Class 2 |
#### Predicate Device (Legally Marketed Devices)
The predicate devices for Fetal Doppler, Model FM20 are:
- Biosys Co., Ltd. Ultrasonic fetal Monitor, IFM-500 cleared by FDA through 510(k) No. K994008 (Decision Date - Sep. 29. 2000), and
- . Bistos Co., Ltd. Ultrasonic Fetal Monitor, BT-300 cleared by FDA through 510(k) No. K05219 (Decision Date - Oct. 4. 2005)
#### Device Description
FM20 is the fetal monitor that measures uterine contractions and fetal heart rate (FHR) which may be used to predict fetal status. To detect the Doppler frequency signal reflected from the heart of the fetus, FM20 irradiates ultrasound wave to the abdomen of a pregnant woman. FM20 analyzes this signal and displays the heart rate by LCD. Also, FM20 provides the sound from the heart of fetus.
FM20 measures the uterine contraction of a pregnant woman by pressure sensors and
{1}------------------------------------------------
Submission - FM20
displays the numerical values.
And FM20 prints the heart rate of the fetus and the values of uterine contraction.
- FM20 records the heart rate of the fetus, the uterine contraction of a pregnant woman. and basic information of the equipment with a provided thermal orinter.
- FM20 is capable of Twin Monitoring with two pulsed Ultrasound Transducers.
- FM20 has a free voltage(100 - 240VAC input) power adaptor.
- FM20 has Fetal movement detection function. To extract fetal movement signal, FM20 uses a circuit and digital filter.
- FM20 has Event marker function so that the patient can record the time of important events.
- FM20 has the capability to alert the caregiver in the event a heart rate goes above or below limit for a preset time delay.
#### Indications for Use
The FM20 is a Fetal Monitor for measuring and recording maternal contraction and fetal heart rate. Data is displayed on a front panel LCD(7") Display, recorded on a strip chart recorder. Single/Twin fetal heart rates may be measured by means of Doppler Ultrasound. Uterine Activity is measured with an external TOCO transducer. The FM20 is intended to be used during the antepartum periods on pregnant woman.
## Summary of Indications and Technical characteristics of the Device compared to the Predicate Device (Legally Marketed Devices)
The Mediana Fetal Doppler, Model FM20 is substantially equivalent to the Bistos Co.,Ltd. Model BT-300, and BiOSYS Co.,Ltd, Model IFM-500.
#### -Indications
The intended use and indications are equivalent to the Bistos Co.,Ltd. Model BiOSYS BT-300. and Co.,Ltd, Model IFM-500. The mode of operation(Ultrasound mode, Contraction), Uterine System characteristics(Portable, Strip chart recorder, Power sources) of Mediana Model FM20 is same as those of the Bistos Co.,Ltd. Model BT-300, and BiOSYS Co.,Ltd, Model IFM-500.
#### Technical characteristics
The specifications and performances(FHR monitoring, UC monitoring, Transducer type, ultrasound frequency, FHR range, FHR accuracy, Detection of fetal movement via ultrasound transducer) of Mediana Model FM20 are
{2}------------------------------------------------
K110612
Page 3 of 3
Submission - FM20
equivalent to the Bistos Co.,Ltd. Model BT-300, and BiOSYS Co.,Ltd, Model IFM-500.
#### Safety, EMC and Performance Data
Electrical, mechanical, environmental safety and performance testing according to standard EN/IEC 60601-1 was performed, and EMC testing was conducted in accordance with standard EN/IEC 60601-1-2(2001). All test results were satisfactory.
#### Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Mediana Co., Ltd. concludes that FM20 are safe and effective and substantially equivalent to predicate devices as described herein.
Mediana Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA.
End of Section
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
NOV - 8 2011
Mediana Co., Ltd. % Mr. Marc M. Mouser Engineering Manager & FDA Office Coordinator, Program Reviewer Underwriters Laboratories, Inc. 2600 NW Lake Road CAMAS WA 98607
Re: K110612
Trade/Device Name: FM20 Fetal Monitor Regulation Number: 21 CFR& 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM Dated: October 7, 2011 Received: October 24, 2011
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
### Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Hubert Leuner MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K110612
FM20 Fetal Monitor Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
The FM20 is a Fetal Monitor for measuring and recording maternal contraction and fetal heart rate. Data is displayed on a front panel LCD(7") Display, recorded on a strip chart recorder and may be transmitted over RS232 port to a remote data receiver. Single/Twin fetal heart rates may be measured by means of Doppler Ultrasound. Uterine Activity is measured with an external TOCO transducer.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
1 11 : ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 11111111 11 111 11111 111 vision Sign-Off) of Reproductive. G
{6}------------------------------------------------
# FM20 Fetal Monitor
510(k) Submission · .
Appendix·F
## Diagnostic Ultrasound Indications For Use Form
Fill out one form for each ultrasound system and each transducer.
## 1MHz PW DOPPLER FETAL PROBE – MODEL : FM20
Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as 14 1 follows: ,
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | P | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(specify) | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Pheperial | | | | | | | | | | |
| Vascular | | | | | | | | | | |
| Lapaloscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
Other (spool); }); }); }); } = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
Additional Comments :
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUT ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Deyce Evaluation(ODE)
Hulsteen
Prescription Use(Part 21 CFR 801.109)
ivision Sign-Off) n of Reproductive. Gastro-Re
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.