K052715 · Lms Medical Systems (Canada), Ltd. · HGM · Dec 22, 2005 · Obstetrics/Gynecology
Device Facts
Record ID
K052715
Device Name
CALM CURVE, VERSION 01.00
Applicant
Lms Medical Systems (Canada), Ltd.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Dec 22, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.2740
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
CALM Curve is intended for use as a data analysis system to assess labor progress in singleton, cephalic presentation pregnancies when the gestational age is > 35 weeks.
Device Story
CALM Curve is a computerized labor progress monitoring system. It accepts user-entered clinical data (parity, dilatation, effacement, station, epidural use) and contraction frequency data (either manually entered or via fetal monitor uterine activity signal). The system processes these inputs using established algorithms to calculate labor progress, displaying results graphically and as percentile values compared to a reference database of vaginal deliveries. Used in clinical settings, it assists obstetricians and clinicians in evaluating labor progression. The output provides a visual labor curve and percentile metrics, aiding clinical decision-making regarding labor management. Benefits include standardized assessment of labor progress and flexibility in data entry, allowing monitoring even when patients are mobile and disconnected from fetal monitors.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design control activities, including Failure Mode and Effects Analysis (FMEA) and verification/validation of software modifications.
Technological Characteristics
Computerized data analysis system; operates as standalone PC, networked client/server, or web-based application. Inputs: manual clinical data and fetal monitor uterine activity signals. Outputs: graphical labor progress curves and percentile values. Algorithm: same as predicate K980719. No specific hardware materials or sterilization requirements specified as it is a software-based system.
Indications for Use
Indicated for assessment of labor progress in singleton, cephalic presentation pregnancies with gestational age ≥ 35 weeks.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
Predicate Devices
Labor Progress Monitoring (LPM) module of LMS Medical Systems CALM (Computer Assisted Labor Monitoring) software (K980719)
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 2 2 2005
# 510(k) Summary for CALM Curve - K052715
Revised - 22 November 2005
### Submitted by
LMS Medical Systems, Ltd. 5252 de Maisonneuve Boulevard West Suite 314 Montreal, Quebec Canada H4A 3S5
FDA facility registration number: 9615396
Telephone: (514) 488-3461 Fax: (514) 488-1880
Contact Person: Pamela J. Haswell Director, QA/RA
#### Device Information
Trade Name: CALM Curve
Common Name: Computer-based labor progress monitoring system for obstetrical care.
Device Class. Perinatal monitoring systems and accessories were reviewed by the Device Oldoo. Formatornient and are classified in Class II per 21 CFR 884.2740.
### Predicate Device Information
The predicate device for CALM Curve is the Labor Progress Monitoring module (LPM me producte devilos for Sriem Sarvet CALM (Computer Assisted Labor Monitoring) software
K980719, clearance date 31 July 1998
#### Labeling and Intended Use
A draft User Guide for CALM Curve is provided as Appendix 5. The User Guide has been updated to reflect the modifications described in this submission.
A draft promotional piece for CALM Curve is provided as Appendix 6.
#### Intended Use
CALM Curve is intended for use as a data analysis system to assess labor progress in singleton, cephalic presentation pregnancies when the gestational age is > 35 weeks. This intended use is the same as for the Labor Progress module of CALM.
The Indications for Use statement is provided as Appendix 2.
{1}------------------------------------------------
#### Device Description
CALM Curve is a computerized method to assess progress during labor. Its algorithms calculate the patient's progress based on user-entered values for parity, dilatation, effacement, station, contraction frequency, and epidural use. Results of these calculations are displayed graphically and as percentile values derived from comparison to a database of women with similar characteristics who delivered vaginally. CALM Curve is intended for use as a data analysis system to assess labor progress in singleton, cephalic presentation pregnancies when the gestational age is > 35 weeks.
CALM Curve can function in a networked environment, as a standalone workstation, or as a web-based application. When connected to a fetal monitor, CALM Curve can use the monitor's uterine activity signal in lieu of user-entered values to calculate contraction counts.
A more detailed description of the architecture of the proposed changes is provided as Appendix 3.
### Summary of Design Control Activities - Risk Analysis
The risk analysis method used to assess the impact of the modifications was a Failure Mode and Effects Analysis (FMEA). The results of that analysis are provided as Appendix 4.
# Summary of Design Control Activities -- Verification and Validation
The risk analysis method used to assess the impact of the modifications was a Failure Mode and Effects Analysis (FMEA). The results of that analysis are provided as Appendix 5.
#### Comparison to the Predicate Device
There are only three differences between CALM Curve and the LPM module of the CALM software:
- The LPM module runs as a module within the CALM software suite. CALM Curve can t also be used as a stand-alone product.
- The LPM module derives its cumulative contraction count from fetal monitor uterine . activity signals. CALM Curve can also accept manually-entered contraction frequency data. In obstetrical practice, clinicians routinely estimate contraction frequency without the aid of a fetal monitor. The entry of manual contraction data has the added advantage of allowing the clinician to update contraction information even when the patient is not connected to a fetal monitor, as when the patient is up and walking.
- As a part of the CALM software suite, the LPM module runs as a full client/server . application. CALM Curve can also be run as a web-based product.
{2}------------------------------------------------
# Statement of Substantial Equivalence to the Predicate Device
CALM Curve is substantially equivalent to the LPM module of LMS's CALM software CALM Curve is substantially equivalorit to the LPM module are compared in the table below.
| Feature | LPM Module of CALM | CALM Curve |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | As a system with signal analysis and<br>display functions ... [for] monitoring of<br>labor progress, data analysis and display<br>functions.* | CALM Curve is intended for use as a<br>data analysis system to assess labor<br>progress in singleton, cephalic<br>presentation pregnancies when the<br>gestational age is ≥ 35 weeks |
| Patient population | Singleton birth<br>>35 weeks gestational age<br>Cephalic presentation | Singleton birth<br>>35 weeks gestational age<br>Cephalic presentation |
| Access to the system | Client/server application | Client/server application, stand-alone PC<br>OR as a web-based product |
| Software application | As a module within the CALM system | As a module within the CALM system OR<br>as a stand-alone application |
| Algorithm | See K980719 | Same algorithm |
| Algorithm variables | Parity, dilatation, effacement, station,<br>epidural use, contraction count | Parity, dilatation, effacement, station,<br>epidural use, contraction count |
| Contraction count | Calculated from maternal/fetal monitor<br>uterine activity signal. | Calculated from maternal/fetal monitor<br>uterine activity signal OR from data<br>entered by the user. |
| Exam data | Entered by the user | Entered by the user |
| Graphical display | Patient's labor progress curve and<br>station | Patient's labor progress curve and<br>station |
| Percentile display | Patient's value (in tabular display) | Patient's value (in tabular display) |
| | Curve of mean and outer limits (3rd / 97th,<br>5th / 95th, or 10th / 90th) of the reference<br>population added to graphical display | Curve of mean and outer limits (3rd / 97th<br>5th / 95th, or 10th / 90th) of the reference<br>population added to graphical display |
| Selection of reference<br>population | Based on parity and previous delivery<br>type (vaginal or c/section) | Based on parity and previous delivery<br>type (vaginal or c/section) |
*NOTE: This statement is taken from the intended use statement for the cleared CALM software system, which is intended for use "as a central monitoring system with signal analysis and display System, which is morned a to the Perinatal clinical environment. It interfaces with standard fetal and maternal monitors and provides fetal surveillance, monitoring of labor progress, charting, reporting, and data analysis, display and archiving functions".
### Declarations, Certifications, and Statements
The following declarations, certifications, and statements are provided as Appendix 4.
- Declaration of Conformity with Design Controls *
- 510(k) Statement for the CALM Curve .
- Certification of Truthfulness and Accuracy of the CALM Curve described in this ● submission.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 2 2005
Ms. Pamela Haswell Director, QA/RA LMS Medical Systems, Inc. 5252 de Maisonneuve West, Suite 314 Montreal, Quebec H4A 3S5 CANADA
Re: K052715
Trade/Device Name: CALM Curve Regulation Number: 21 CFR §884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM Dated: November 22, 2005 Received: November 23, 2005
Dear Ms. Haswell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to the enorologics to regioned date of the Medical Device Amendments, or to devices that have been reasy 20, 11 v, accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a prematket approval application (PMA). You may, therefore, market the do inot require to the general controls provisions of the Act. The general controls provisions of the Act do reo, sacjos to an got and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket 11 your al), it may be subject to such additional controls. Existing major regulations affecting your Applorar, the found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any I Drival statutes and regulations administered by other Federal agencies. You must comply with all the r carrates are are any ing (21 CFR Part 801); good nanufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin hanketing your ence of your device of your device to a legally premarket notheation. The PDA miding of Sassannal reports and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your derites on our of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification the Also, please note the regulation entitled, "Theoration on your responsibilities under the Act from the 807.97). You may outlar outcr general informations and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use Form 2.1
Applicant: LMS Medical Systems, Inc.
510(k) Number: K052715
- **10** %
- **15** %
Device Name: CALM Curve
# Indications for Use:
CALM Curve is intended for use as a data analysis system to assess labor progress in CALM Curve is intended for use as a duta analyers eyests the new is ≥ 35 weeks.
(Per 21 CFR 801.109)
X _ Prescription Use
Over the Counter
(Per 21 CFR 801 109)
David A. Ingram
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.