K993749 · Gyrus Medical , Ltd. · HIH · Dec 3, 1999 · Obstetrics/Gynecology
Device Facts
Record ID
K993749
Device Name
GYRUS HYSTEROSCOPIC RESECTOSCOPE
Applicant
Gyrus Medical , Ltd.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Dec 3, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Gyrus Hysteroscopic Resectoscope is intended to enable the viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. The surgical applications include tissue cutting, removal, and dissection as required or encountered in gynecologic hysteroscopic electrosurgical procedures for the treatment of intrauterine myomas, polyps, adhesions, and septa.
Device Story
Gyrus Hysteroscopic Resectoscope; electrosurgical instrument for gynecologic procedures. Device comprises working element, rotating outer sheath, and visual obturator. Surgeon inserts device through cervical canal; visualizes uterine cavity via telescope housed in working element. High-frequency current transmitted to electrode for tissue cutting, removal, dissection, or coagulation. Features 'quick connection' for working element attachment and rotating collar for 360-degree fluid inlet rotation. Visual obturator facilitates insertion while maintaining visualization. Used in clinical/surgical settings by physicians. Output is direct visualization and electrosurgical tissue modification; assists in treating myomas, polyps, adhesions, septa, and endometrial tissue.
Clinical Evidence
Bench testing only. Preclinical and bench-top testing performed to verify performance requirements and substantiate efficacy for hysteroscopic usage.
Technological Characteristics
Materials: stainless steel, chrome-plated stainless steel, plastic, chrome-plated brass. Principle: electrosurgical resection/coagulation via high-frequency current. Form factor: hysteroscopic resectoscope with rotating outer sheath and visual obturator. Connectivity: none. Sterilization: not specified.
Indications for Use
Indicated for direct viewing of cervical canal and uterine cavity for diagnostic and surgical procedures. Specific applications include excision of intrauterine myomas, polyps, lysis of adhesions, incision of uterine septa, and endometrial ablation when used with G-VAP electrode.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}------------------------------------------------
DEC - 3 1999
GYRUS MEDICAL DOCUMENT No. 813740
K993749 page 1
Gyrus Medical Limited
Cardiff CF33 0LT, UK
510k Premarket Notification
November, 1999
| | 510(k) Summary of Safety and Effectiveness |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Statement | Information supporting claims of substantial equivalence, as defined under the Federal Food, Drug, and Cosmetic Act, respecting safety and effectiveness is summarized below. For the convenience of the Reviewer, this summary is formatted in accordance with the Agency's final rule "...510(k) Summaries and 510(k) Statements..." (21 CFR 807) and can be used to provide a substantial equivalence summary to anyone requesting it from the Agency. |
| | MODIFIED DEVICE NAME: Gyrus Hysteroscopic Resectoscope<br>Accessories Visual Obturator and Rotating<br>Outer Sheath |
| | PREDICATE DEVICE NAME: COMEG Endoscopy Resectoscope |
| Device<br>Description | The Gyrus Hysteroscopic Resectoscope that we intend to market employs an electrode specific resectoscope-working element. All other sheaths, etc., are standard COMEG components. A "Quick Connection" feature is used to attach the working element to the sheath. |
| | Endoscopic electrosurgical instruments and accessories are described under "Endoscope and Accessories" in 21 CFR 876.1500, and Hysteroscopes and described in 21 CFR 884.1690. |
| | The Hysteroscopic Resectoscope is intended to enable the viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. |
| | The surgeon performs the examination through the cervical canal. The working elements are the devices that house and control various electrodes and lasers used to remove, cut, coagulate, and transect tissue. The surgeon controls the back and forth movement of the electrode using finger controls. The working elements also house a telescope for visualization. Current is transmitted to the electrode through a high frequency cable. |
| | Continued on next page |
17
{1}------------------------------------------------
K993749
GYRUS MEDICAL DOCUMENT No. 813740
Gyrus Medical Limited
Cardiff CF3 0LT, UK
510k Premarket Notification
November, 1999
| Device Description (Cont'd) | These devices are composed of stainless steel, stainless steel chrome plated, plastic and brass chrome plated parts. The Gyrus Hysteroscopic Resectoscope as cleared in K982771, is identical to the COMEG Resectoscope cleared via K971882, with the working element modified to fit the Gyrus Medical Ltd. electrode. Endoscopic electrosurgical instruments and accessories are described in endoscope and accessories, 21 CFR 876.1500, and endoscope ESU units and accessories are described in 21 CFR 876.4300. Hysteroscopes are described in 21 CFR 884.1690. The Rotating Outer Sheath connection for the working element is of a "snap-on" construction, identical to the cleared device. The principal difference is in the rotating collar which allows 360 degree rotation of the fluid inlet. The Visual Obturator provides the same functionality as the standard obturator facilitating insertion to the uterus. In addition, the Visual Obturator allows use of the hysteroscope (telescope) during insertion, allowing visualization. |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Gyrus Hysteroscopic Resectoscope is intended to enable the viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. The surgical applications include tissue cutting, removal, and dissection as required or encountered in gynecologic hysteroscopic electrosurgical procedures for the treatment of intrauterine myomas, polyps, adhesions, and septa. |
| Indications Statement | The Gyrus Hysteroscopic Resectoscope is used to permit the direct Viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. |
| Technological Characteristics | The modified device has the same technological characteristics as the predicate devices. The form, fit and function is similar. |
| Performance Data | Preclinical as well as bench top testing has been performed to verify that The product meets the performance requirements described. This substantiates the efficacy of the resectoscope for the hysteroscopic usage. |
Continued on next page
{2}------------------------------------------------
K993749 page 3/3
GYRUS MEDICAL DOCUMENT No. 813740
Gyrus Medical Limited
Cardiff CF33 0LT, UK
510k Premarket Notification
November, 1999
| Conclusion | Based upon the 510(k) summaries and 510(k) statements (21 CFR 807)<br>and the information provided herein, we conclude that the new device is<br>substantially equivalent to the predicate devices under the Federal Food,<br>Drug, and Cosmetic Act. |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | David Kay<br>Director<br>Regulatory Affairs and Quality Assurance<br>Gyrus Medical Ltd.<br>Fortran Road<br>St. Mellons<br>Cardiff, CF3 OLT<br>United Kingdom |
| Date | November 4, 1999 |
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
- 3 1999
Mr. David Kay Director, Regulatory Affairs & Quality Assurance Gyrus Medical Limited Fortran Road St Mellons Cardiff CF3 OLT nk
Re: K993749
Hysteroscopic Rotating Sheath and Visual Obturator Dated: November 4, 1999 Receive: November 5, 1999 Regulatory Class: II 21 CFR §884.1698/Procode: 85HIH
Dear Mr. Kay:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regult in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations,
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitto diagnostic devices), piease contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97): Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## GYRUS MEDICAL DOCUMENT No. 813740
Gyrus Medical Limited
Cardiff CF33 0LT, UK
510k Premarket Notification
November, 1999
## INDICATIONS FOR USE
| 510(k) Number (if known): | K993749 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Gyrus Hysteroscopic Resectoscope |
| Indications for Use: | Used to permit direct viewing of the cervical canal and the uterine<br>Cavity for the purpose of performing diagnostic and surgical<br>procedures. As used with the Gynecare Versapoint System, the<br>indications for use include:<br>Excision on intrauterine myomas<br>Excision of intrauterine polyps<br>Lysis of intrauterine adhesions<br>Incision of uterine septa<br><br>When used for resection of endometrial tissue with the G-VAP<br>Electrode, the device is indicated for Endometrial Ablation. |
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINE
ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | OR | Over-The-Counter Use |
|------------------|----|----------------------|
|------------------|----|----------------------|
(Optional Format 1-2-96)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number: K993749
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.