AVETA SYSTEM: Hysteroscopy: The Aveta System is intended for intrauterine use by trained gynecologists to permit viewing of the cervical canal and the uterine cavity, provide liquid distension of the uterus and monitor the volume differential between the irrigation fluid flowing into and out of the uterus during diagnostic and surgical procedures to resect and remove tissue such as submucous myomas, endometrial polyps and retained products of conception. Cystoscopy: The Aveta System is intended for use in endoscopic access to and examination of the lower urinary tract, including the bladder. When combined with accessory instruments, the system allows the user to perform various diagnostic and therapeutic procedures. AVETA DISPOSABLE HYSTEROSCOPE (Pearl/Opal/Coral): The Aveta Disposable Hysteroscope (Pearl/Opal/Coral) is intended to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. AVETA DISPOSABLE CYSTOSCOPE (Coral): The Aveta Disposable Cystoscope (Coral) is intended for use in endoscopic access to and examination of the lower urinary tract, including the bladder. When combined with accessory instruments, the Cystoscope allows the user to perform various diagnostic and therapeutic procedures.
Device Story
Integrated endoscopic system for visualization and mechanical tissue resection; includes controller, disposable scopes (hysteroscope/cystoscope), and resecting devices. Controller uses dual peristaltic pumps for continuous saline irrigation/aspiration; monitors intrauterine/urinary tract pressure and fluid deficit. Resecting devices (mechanical) oscillate to remove tissue; some models feature integrated motors in the handset. Used in clinical settings by physicians. Output displayed on external monitor for real-time visualization and procedural monitoring. System aids in diagnostic and therapeutic procedures by maintaining cavity distension and enabling tissue removal, potentially reducing procedure time and improving patient outcomes.
Clinical Evidence
Bench testing only. Performance data included software verification/validation (IEC 62304), system integrity, pressure/flow accuracy, fluid deficit monitoring, optical testing, LED tip temperature, motor oscillation speed, biocompatibility (ISO 10993-1), sterilization (ISO 11135/11137), packaging (ASTM D4169), accelerated aging (ASTM F1980), and electrical safety/EMC (IEC 60601 series). Simulated use testing confirmed tissue resection and imaging capabilities.
Technological Characteristics
Integrated system with CMOS sensor, LED illumination, and dual peristaltic pump fluid management. Components include disposable scopes and mechanical resecting devices (stainless steel blades). Connectivity via external monitor. Sterilization per ISO 11135/11137. Software-controlled pressure regulation and fluid deficit monitoring. Electrical safety per IEC 60601-1, EMC per IEC 60601-1-2, and endoscopic safety per IEC 60601-2-18.
Indications for Use
Indicated for trained gynecologists performing diagnostic and surgical hysteroscopy (resection of myomas, polyps, retained products) and for endoscopic access/examination of the lower urinary tract (including bladder) in patients requiring diagnostic or therapeutic procedures.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it.
May 26, 2022
Meditrina, Inc. Csaba Truckai President & CEO 1190 Saratoga Avenue, Suite 180 San Jose, CA 95129
Re: K213171
Trade/Device Name: Aveta System, Aveta Disposable Hysteroscope (Pearl/Opal/Coral), Aveta Disposable Cystoscope (Coral) Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH, HIG, FAJ Dated: April 21, 2022 Received: April 25, 2022
Dear Csaba Truckai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213171
#### Device Name
Aveta System, Aveta Disposable Hysteroscope (Pearl/Opal/Coral), Aveta Disposable Cystoscope (Coral)
## Indications for Use (Describe)
## AVETA SYSTEM:
#### Hysteroscopy:
The Aveta System is intended for intrauterine use by trained gynecologists to permit viewing of the cervical canal and the uterine cavity, provide liquid distension of the uterus and monitor the volume differential between the irrigation fluid flowing into and out of the uterus during diagnostic and surgical procedures to resect and remove tissue such as submucous myomas, endometrial polyps and retained products of conception.
#### Cystoscopy:
The Aveta System is intended for use in endoscopic access to and examination of the lower urinary tract, including the bladder. When combined with accessory instruments, the system allows the user to perform various diagnostic and therapeutic procedures.
#### AVETA DISPOSABLE HYSTEROSCOPE (Pearl/Opal/Coral):
The Aveta Disposable Hysteroscope (Pearl/Opal/Coral) is intended to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.
#### AVETA DISPOSABLE CYSTOSCOPE (Coral):
The Aveta Disposable Cystoscope (Coral) is intended for use in endoscopic access to and examination of the lower urinary tract, including the bladder. When combined with accessory instruments, the Cystoscope allows the user to perform various diagnostic and therapeutic procedures.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K213171 510(k) Summary
### I. Submitter Information
| Submitter name: | Meditrina, Inc.<br>1190 Saratoga Avenue, Suite 180<br>San Jose, CA 95129 |
|-----------------|---------------------------------------------------------------------------------------------------------------------|
| Contact person: | Csaba Truckai<br>President & CEO<br>Email: csabat@hermesinnovations.com<br>Phone: 415-215-7233<br>Fax: 408-418-4815 |
| Date Prepared: | 25 May 2022 |
### II. Product Classification
| Device Name: | Aveta System, Aveta Disposable Hysteroscope (Pearl/Opal/Coral),<br>Aveta Disposable Cystoscope (Coral) | |
|---------------------------|--------------------------------------------------------------------------------------------------------|----------------|
| Common Name: | Hysteroscope | |
| Regulation: | 21 CFR 884.1690 | |
| Regulation Name: | Hysteroscope and accessories;<br>Hysteroscopic insufflators<br>Endoscope and accessories | Subject Device |
| Class: | II | |
| Product Code: | HIH | |
| Additional Product Codes: | HIG | |
| Additional Product Codes: | FAJ | |
#### III. Predicate Device Information
| Predicate Devices | Manufacturer | Predicate Device<br>Names | 510(k)# | Clearance Date |
|---------------------|------------------|--------------------------------------|---------|-----------------|
| PREDICATE<br>DEVICE | Meditrina, Inc. | Aveta System | K190372 | May 16, 2019 |
| Reference Device | UVision360, Inc. | Luminelle DTx<br>Hysteroscopy System | K181909 | August 16, 2018 |
Predicate has not been a subject of a design related recall.
#### IV. Device Description
The Aveta System is an integrated system which allows for visualization of the intended cavity for the purpose of performing diagnostic and operative procedures (hysteroscopy and cystoscopy). The Aveta System consists of the components listed in Table 1. The system includes a Controller with integrated fluid management which incorporates a dual peristaltic pump design to control the continuous inflow and outflow of saline to provide fluid distention of the cavity. The Controller
{4}------------------------------------------------
Aveta System, Aveta Hysteroscope, Aveta Cystoscope Traditional 510(k) Premarket Notification
provides continuous monitoring of the cavity pressure to the set pressure. For hysteroscopy, it also monitors the volume differential between saline inflow from the uterus (fluid deficit). The Controller connects to a sterile, single use disposable Scope (available in various configurations, see below) that allows visualization of the cavity and displays the images obtained from the Scope on a standard monitor. For operative hysteroscopy procedures, the Aveta System includes sterile, mechanical Disposable Resecting Device (available in various configurations, see below) powered by an integrated motor in the device handset or by use of an external Resecting Handset (available in two configurations, as a disposable device or a reusable device). The resecting device is inserted through the working channel of the sterile hysteroscope to resect the target tissue/pathology. For cystoscopy, when combined with accessory instruments the cystoscope is used for diagnostic and therapeutic procedures.
| Table 1. Aveta System Components<br>Aveta System Component | Functions Performed |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Aveta Controller<br>and Footswitch | • Displays image/video and procedural information on external<br>monitor.<br>• Image / video processing / storing of the images.<br>• Enables visualization functions of the Hysteroscope /<br>Cystoscope.<br>• Fluid Management with irrigation and aspiration functions.<br>• Controls saline inflow and outflow for distention of the uterine<br>cavity or lower urinary tract with the bladder for visualization.<br>• Monitors and maintains intrauterine pressure or lover urinary<br>tract cavity pressure to set pressure.<br>• Monitors volume differential (fluid deficit for hysteroscopy).<br>• Provides ON/OFF function of the Resecting Device.<br>• Provides power to the Resecting Handsets, Disposable Resecting<br>Devices with Handsets for oscillation at a preset speed for<br>mechanical resection function of Disposable Resecting Device.<br>• Provides power to the Drape Pump. |
| Aveta Hysteroscope/Aveta Cystoscope<br>• Aveta Pearl Disposable Hysteroscope<br>• Aveta Coral Disposable Hysteroscope<br>• Aveta Opal Disposable Hysteroscope<br>• Aveta Coral Disposable Cystoscope<br>Collectively, they are called Aveta<br>Disposable Scope or Endoscope. | • Hysteroscope: Visualization of cervical canal and uterine cavity<br>• Cystoscope: Visualization of the lower urinary tract including<br>the bladder<br>• Provides conduits/lumens for fluid inflow and outflow<br>• Provides conduit (working channel) for operative instruments<br>for operative procedures<br>• Provides user interface for intrauterine or urethral cavity set<br>pressure and fluid deficit limit adjustments (for hysteroscopy),<br>flush, and recording of images |
| Aveta Fluid Management Accessory | • Provides membrane in fluid inflow line to enable intrauterine or<br>urethral cavity pressure monitoring/control using pressure<br>transducer in Controller<br>• Provides conduits for irrigation of saline and aspiration of waste.<br>• Provides FMA Cassette with tubing for peristaltic pump<br>functions of Aveta Controller. |
## Table 1 Aveta System Components
{5}------------------------------------------------
| Aveta System Component | Functions Performed |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Aveta Disposable Resecting Device:<br>• Aveta Wave+ Disposable Resecting<br>Device, 3.9mm (DRD-Wave+)<br>• Aveta Max Disposable Resecting<br>Device, 3.9mm (DRD-Max)<br>• Aveta Disposable Resecting Device,<br>3.9 (DRD-3.9)<br>• Aveta Flex Disposable Resecting<br>Device, 2.9mm (DRD-Flex)<br>• Aveta Smol Disposable Resecting<br>Device, 2.9mm (DRD-Smol)<br>• Aveta Disposable Resecting Device,<br>2.9 and (DRD-2.9)<br>• Aveta Auto Disposable Resecting<br>Device (AUTO) | • Mechanically resects and removes tissue under suction<br>• DRD-2.9 and DRD-3.9 require use of DRH or RRH for<br>oscillation of resection tip<br>• DRD-Wave+, DRD-Max, DRD-Flex, DRD-Smol, AUTO<br>include motor in the device handle to oscillate resection tip<br>• The motor in the device handle for AUTO also provides<br>suction by peristaltic action and is only for use with the<br>pressurized saline bag |
| Aveta Disposable Resecting Handset<br>(DRH) | • Includes motor for oscillation of resection tip (used with DRD-<br>2.9 and DRD-3.9) |
| Aveta Reusable Resecting Handset<br>(RRH) | • Includes motor for oscillation of resection tip (used with DRD-<br>2.9 and DRD-3.9) |
| Additional Aveta System Components /<br>Accessories<br>• Waste Management Accessory<br>• Waste Bag (6L)<br>• Aveta Reusable Cable, Opal<br>• Roll Stand with Drape Pump<br>• Monitor | • Collects tissue for pathology and stores the outflow fluid waste<br>and collects fluid from the patient's under-buttocks drape.<br>• Part of Waste Management Accessory.<br>• Provides connection from the hysteroscope pigtail outside the<br>sterile field to the Controller.<br>• Roll Stand mounts the monitor and Controller for the system<br>Drape Pump transfers collected waste fluid from drape to Waste<br>Bag.<br>• Commercially available surgical monitor. Displays image,<br>procedural parameters and notifications. |
{6}------------------------------------------------
### V. Indications for Use
There is no difference in the indications for use for the Aveta System (subject device) when compared to the combined indications of the predicate system components.
| Device | Indications For Use |
|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Aveta System<br>(Subject Device) | AVETA SYSTEM:<br>Hysteroscopy:<br>The Aveta System is intended for intrauterine use by trained gynecologists to<br>permit viewing of the cervical canal and the uterine cavity, provide liquid<br>distension of the uterus and monitor the volume differential between the irrigation<br>fluid flowing into and out of the uterus during diagnostic and surgical procedures<br>to resect and remove tissue such as submucous myomas, endometrial polyps and<br>retained products of conception.<br>Cystoscopy:<br>The Aveta System is intended for use in endoscopic access to and examination of<br>the lower urinary tract, including the bladder. When combined with accessory<br>instruments, the system allows the user to perform various diagnostic and<br>therapeutic procedures. |
| Aveta Disposable Hysteroscope<br>(Subject Device) | AVETA DISPOSABLE HYSTEROSCOPE (Pearl/Opal/Coral):<br>The Aveta Disposable Hysteroscope (Pearl/Opal/Coral) is intended to permit<br>viewing of the cervical canal and the uterine cavity for the purpose of performing<br>diagnostic and surgical procedures. |
| Aveta Disposable Cystoscope<br>(Subject Device) | AVETA DISPOSABLE CYSTOSCOPE (Coral):<br>The Aveta Disposable Cystoscope (Coral) is intended for use in endoscopic access<br>to and examination of the lower urinary tract, including the bladder. When<br>combined with accessory instruments, the Cystoscope allows the user to perform<br>various diagnostic and therapeutic procedures. |
| Aveta System<br>(Predicate Device – K190372) | The Aveta System is intended for intrauterine use by trained gynecologists to permit<br>viewing of the cervical canal and the uterine cavity, provide liquid distension of the<br>uterus and monitor the volume differential between the irrigation fluid flowing into<br>and out of the uterus during diagnostic and surgical procedures to resect and remove<br>tissue such as submucous myomas, endometrial polyps and retained products of<br>conception. |
| UVision360 Luminelle DTx<br>Hysteroscopy System<br>(Reference Device – K181909) | Hysteroscopy: The LUMINELLE DTx Hysteroscopy System is used to permit<br>direct viewing of the cervical canal and uterine cavity for the purpose of performing<br>diagnostic and surgical procedures.<br>Note: Hysteroscopes are used as tools to access the uterine cavity and are not, in and<br>of themselves, a method of surgery.<br>Generally recognized indications for diagnostic hysteroscopy include:<br>• Abnormal bleeding<br>• Infertility and pregnancy wastage<br>• Evaluation of abnormal hysterosalpingogram<br>• Intrauterine foreign body<br>• Amenorrhea<br>• Pelvic pain<br>Generally recognized indications for operative hysteroscopy include:<br>• Directed endometrial biopsy<br>• Polypectomy<br>• Submucous myomectomy<br>• Transection of intrauterine adhesions |
| Device | Indications For Use |
| | • Endometrial ablation<br>Cystoscopy: The LUMINELLE DTx Hysteroscopy System is intended for use in<br>endoscopic access to and examination of the lower urinary tract, including the<br>bladder. When combined with accessory instruments, the System allows the user to<br>perform various diagnostic and therapeutic procedures. |
### Comparison of Indications for Use
{7}------------------------------------------------
The indications for use for the subject Aveta System is identical to the reference device and has minor difference to the cleared Aveta System.
#### VI. Comparison of Technological Characteristics with the Predicate Device
A veta System and the predicate system have the same or similar technological characteristics in terms of basic operating principle and basic design features with minor differences.
| | Subject Device | PRIMARY<br>Predicate<br>Device | Reference Device | | |
|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| 510k# | K213171 | K190372 | K191958 | K192100 | K181909 |
| Manufacturer: | Meditrina Inc. | Meditrina, Inc. | | | UVision360 |
| Device Names | Aveta System<br>Aveta Hysteroscope<br>Aveta Cystoscope | Aveta System | Aveta System | Aveta Disposable<br>Hysteroscope | Luminelle DTx<br>Hysteroscopy<br>System |
| CONTROLLER FUNCTIONS | | | Hysteroscope Functions | | |
| Visualization<br>and Image<br>Processing | CMOS sensor, and<br>light source in<br>Endoscope with<br>image processing by<br>the Controller | CMOS sensor,<br>and light source<br>in Hysteroscope<br>with image<br>processing by<br>the Controller | N/A<br>(Used in combination with<br>hysteroscope and accessories<br>of the commercial 3rd party<br>hysteroscope) | CMOS sensor, and<br>light source in<br>Hysteroscope with<br>image processing<br>by the Controller | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Viewing<br>Functions | Controller connects<br>to a commercially<br>available external<br>Monitor and<br>displays image from<br>the cavity, plays<br>tone, displays cavity<br>pressure, fluid<br>deficit with<br>graphical user<br>interface. | Controller<br>connects to a<br>commercially<br>available external<br>Monitor and<br>displays image<br>from the cavity,<br>plays tone,<br>displays cavity<br>pressure, fluid<br>deficit with<br>graphical user<br>interface. | Controller connects to a<br>commercially available<br>external Monitor and<br>plays tone, displays<br>cavity pressure, fluid<br>deficit with graphical<br>user interface but<br>doesn't display image<br>from cavity. | Controller<br>connects to a<br>commercially<br>available external<br>Monitor and<br>displays image<br>from the cavity<br>and plays tone. | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Fluid Management Functions | | | | | |
| Fluid<br>Distension | Continuous flow of<br>saline/fluid | Continuous flow<br>of saline/fluid | YES | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| | Subject Device | PRIMARY<br>Predicate<br>Device | Reference Device | | |
| 510k# | K213171 | K190372 | K191958 | K192100 | K181909 |
| Manufacturer: | Meditrina Inc. | Meditrina, Inc. | | | UVision360 |
| Device Names | Aveta System<br>Aveta Hysteroscope<br>Aveta Cystoscope | Aveta System | Aveta System | Aveta Disposable<br>Hysteroscope | Luminelle DTx<br>Hysteroscopy<br>System |
| Irrigation for<br>Distension | Peristaltic pump<br>with dual pressure<br>sensors for irrigation<br>of fluids | Peristaltic pump<br>with dual pressure<br>sensors for<br>irrigation of fluids | Peristaltic pump with<br>dual pressure sensors for<br>irrigation of fluids | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Aspiration of<br>bodily fluids<br>and tissue | Integrated Peristaltic<br>pump for aspiration. | Integrated<br>Peristaltic pump<br>for aspiration. | Integrated Peristaltic<br>pump for aspiration. | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Intrauterine<br>Pressure<br>Measurements | Obtains two<br>independent,<br>intrauterine pressure<br>measurement by<br>sensing pressure of<br>the irrigation tube | Obtains two<br>independent,<br>intrauterine<br>pressure<br>measurement by<br>sensing pressure<br>of the irrigation<br>tube | Obtains two<br>independent, intrauterine<br>pressure measurement<br>by sensing pressure of<br>the irrigation tube | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Set Pressure<br>Range | Hysteroscopy:<br>30-120 mmHg<br>Cystoscopy:<br>30-60mmHg | Hysteroscopy:<br>30-120 mmHg<br>Cystoscopy: N/A | Hysteroscopy:<br>30-120 mmHg<br>Cystoscopy: N/A | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Set Pressure<br>User<br>Adjustments | Allows user to<br>increase/decrease the<br>set pressure | Allows user to<br>increase/decrease<br>the set pressure | Allows user to<br>increase/decrease the<br>set pressure | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Pressure Relief<br>for<br>overpressure<br>risk mitigation | Hysteroscopy:<br>Reverse rotation of<br>irrigation peristaltic<br>pump at 150mmHg<br>Cystoscopy:<br>Reverse rotation of<br>irrigation peristaltic<br>pump at 75mmHg | Hysteroscopy:<br>Reverse rotation<br>of irrigation<br>peristaltic pump at<br>150mmHg<br>Cystoscopy: N/A | Hysteroscopy:<br>Reverse rotation of<br>irrigation peristaltic pump<br>at 150mmHg<br>Cystoscopy: N/A | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Fluid Deficit<br>Measurement | YES | YES | YES | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Flow Rate | 180-500 mL/min<br>preset fixed flow<br>rates | 180-500 mL/min<br>preset fixed flow<br>rates | 180-500 mL/min preset<br>fixed flow rates | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| | Subject Device | PRIMARY<br>Predicate<br>Device | Reference Device | | |
| 510k# | K213171 | K190372 | K191958 | K192100 | K181909 |
| Manufacturer: | Meditrina Inc. | Meditrina, Inc. | | UVision360 | |
| Device Names | Aveta System<br>Aveta Hysteroscope<br>Aveta Cystoscope | Aveta System | Aveta System | Aveta Disposable<br>Hysteroscope | Luminelle DTx<br>Hysteroscopy<br>System |
| Mechanical<br>Resecting<br>Device | Connects to the<br>Resecting Handset<br>by an electrical<br>connection to<br>provide motor<br>control with a preset<br>fixed motor<br>rotation/oscillation<br>speed. | Connects to the<br>Resecting<br>Handle by an<br>electrical<br>connection to<br>provide motor<br>control with a<br>user adjustable<br>motor<br>oscillation<br>speed. | Connects to the<br>Resecting Handle by an<br>electrical connection to<br>provide motor control<br>with a user adjustable<br>motor oscillation speed. | Connects to the<br>Resecting Handle<br>by an electrical<br>connection to<br>provide motor<br>control with a user<br>adjustable motor<br>oscillation speed. | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| SCOPE | | | | | |
| | | | Disposable Hysteroscope/Cystoscope | | |
| Irrigation and<br>Aspiration<br>Lumens | Independent sterile<br>saline irrigation and<br>aspiration lumens | Independent<br>sterile saline<br>irrigation and<br>aspiration lumens | N/A | Independent<br>sterile saline<br>irrigation and<br>aspiration lumens | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Insertion OD | Pearl Hysteroscope:<br>5.7mm<br>Coral Hysteroscope:<br>4.6mm<br>Opal Hysteroscope:<br>4.6mm<br>Coral Cystoscope:<br>4.6mm | 5.5 mm | N/A | 5.5 mm | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Working<br>Length | Pearl Hysteroscope:<br>216mm<br>Coral Hysteroscope<br>206mm<br>Opal Hysteroscope<br>206mm<br>Coral Cystoscope:<br>206mm | 224mm | N/A | 224mm | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Illumination | LEDs<br>(Light Emitting<br>Diode) | LEDs<br>(Light Emitting<br>Diode) | N/A | LEDs<br>(Light Emitting<br>Diode) | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Working<br>Channel | Pearl Hysteroscope:<br>4.0mm<br>Coral Hysteroscope:<br>3.0mm<br>Opal Hysteroscope:<br>3.0mm<br>Coral Cystoscope:<br>3.0mm | 3.5mm | N/A | 3.5mm | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| | Subject Device | PRIMARY<br>Predicate<br>Device | Reference Device | | |
| 510k# | K213171 | K190372 | K191958 | K192100 | K181909 |
| Manufacturer: | Meditrina Inc. | Meditrina, Inc. | | UVision360 | |
| Device Names | Aveta System<br>Aveta Hysteroscope<br>Aveta Cystoscope | Aveta System | Aveta System | Aveta Disposable<br>Hysteroscope | Luminelle DTx<br>Hysteroscopy<br>System |
| Camera | Digital CMOS<br>Camera | Digital CMOS<br>Camera | N/A | Digital CMOS<br>Camera | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Cutting<br>Window | DRD-Wave+: 8mm<br>DRD-Max: 11mm<br>DRD-3.9: 10mm<br>DRD-Flex: 7mm<br>DRD-Smol: 7mm<br>DRD-2.9: 7mm<br>AUTO: 8mm | 8 mm | N/A | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Blade Material | Stainless steel | Stainless steel | N/A | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Working<br>Length | DRD-Wave+: 339mm<br>DRD-Max: 339mm<br>DRD-3.9: 328mm<br>DRD-Flex: 339mm<br>DRD-Smol: 339mm<br>DRD-2.9: 328mm<br>AUTO: 318mm | 328 mm | N/A | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Insertion OD | DRD-Wave+: 3.9mm<br>DRD-Max: 3.9mm<br>DRD-3.9: 3.9mm<br>DRD-Flex: 2.9mm<br>DRD-Smol: 2.9mm<br>DRD-2.9: 2.9mm<br>AUTO: 2.9mm | 3.4 mm | N/A | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Rotational<br>Speed | 3000RPM | 2000RPM | N/A | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Suction<br>Assisted<br>Resection? | YES | YES | N/A | N/A | N/A: All technical<br>information for the<br>SE comparison will<br>be taken from the<br>primary predicate. |
| Disposable (Sterile) and Reusable (Steam Sterilizable) Resecting Handset | | | | | |
| | Subject Device | PRIMARY<br>Predicate<br>Device | Reference Device | | |
| 510k# | K213171 | K190372 | K191958 | K192100…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.