K192100 · Meditrina, Inc. · HIH · Dec 19, 2019 · Obstetrics/Gynecology
Device Facts
Record ID
K192100
Device Name
Aveta Disposable Hysteroscope
Applicant
Meditrina, Inc.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Dec 19, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aveta Disposable Hysteroscope is used to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.
Device Story
Disposable hysteroscope used by physicians in office or OR settings; includes integrated camera and LED light. Connects to Aveta Controller for image processing, storage, and display. Operates with gravity-fed saline or external fluid management system; compatible with resecting devices for tissue removal. Handle buttons control image capture and transfer; controller software ignores other inputs. Provides real-time visualization to assist clinicians in diagnostic and surgical decision-making; benefits include sterile, single-use design reducing cross-contamination risk.
Clinical Evidence
Bench testing only. Software verification and simulated use testing in a model uterine system demonstrated performance of cavity pressure regulation, flow rates, and tissue resection, confirming specifications were met.
Technological Characteristics
Materials: Stainless steel, ABS, silicone elastomer, PVC, polycarbonate. Optics: Fixed focus lens with CMOS sensor (90° FOV, 0° direction of view). Connectivity: Wired connection to Aveta Controller. Sterilization: Ethylene oxide (SAL 10^-6). Form factor: 5.5mm outer diameter, 224mm working length, 3.5mm working channel.
Indications for Use
Indicated for diagnostic and surgical procedures requiring visualization of the cervical canal and uterine cavity in patients undergoing hysteroscopy.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 19, 2019
Meditrina, Inc. Csaba Truckai President & CEO 1190 Saratoga Avenue, Suite 180 San Jose. CA 95129
Re: K192100
Trade/Device Name: Aveta Disposable Hysteroscope Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope And Accessories Regulatory Class: II Product Code: HIH Dated: November 18, 2019 Received: November 21, 2019
Dear Csaba Truckai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Sharon M. Andrews Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192100
Device Name Aveta Disposable Hysteroscope
Indications for Use (Describe)
The Aveta Disposable Hysteroscope is used to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K192100: 510(k) Summary
# I. Submitter Information
| Submitter name: | Meditrina, Inc.<br>1190 Saratoga Avenue, Suite 180<br>San Jose, CA 95129 | |
|-----------------|----------------------------------------------------------------------------------------------------------------------|---------------------|
| Contact person: | Csaba Truckai<br>President & CEO<br>Mobile: (415) 215-7233<br>Office: (408) 471-4877<br>csabat@hermesinnovations.com | Fax: (408) 418-4815 |
| Date Prepared: | 18 December 2019 | |
## II. Product Classification
| Device Name: | Aveta Disposable Hysteroscope | |
|------------------|-------------------------------|----------------|
| Common Name: | Hysteroscope | Subject Device |
| Regulation: | 21 CFR 884.1690 | |
| Regulation Name: | Hysteroscope and accessories | |
| Class: | II | |
| Product Code: | HIH | |
## III. Predicate Device Information
| Predicate Devices | Manufacturer | Predicate Device Names | 510(k)# | Clearance Date |
|-------------------|-----------------|------------------------|---------|----------------|
| Predicate Device | Meditrina, Inc. | Aveta System | K190372 | May 16, 2019 |
Predicate has not been the subject of a design related recall.
## IV. Device Description
The Aveta Disposable Hysteroscope is used by physicians in an office or operating room setting. The subject device includes the Aveta Disposable Hysteroscope and Aveta Controller. Like the Aveta Disposable Hysteroscope cleared under K190372, the modified Aveta Disposable Hysteroscope includes a camera and LED light. The Aveta Disposable Hysteroscope is connected to the Aveta Controller to provide the camera and light functions and to process and store the image obtained from the Aveta Disposable Hysteroscope.
For fluid management, the Aveta Disposable Hysteroscope can be used with gravity fed pressurized saline bag or the fluid management system cleared under K190372. For therapeutic procedures, the Aveta Disposable Hysteroscope is used with the cleared Aveta Disposable Resecting Device (K190372). The only functional buttons on the Aveta Disposable Hysteroscope handle are the image capture and the image transfer buttons. For the modified device, when other buttons on the handle are pressed, the Aveta Controller does not take any action.
Other than minor change to the Aveta Controller software to disable fluid management related buttons on the Aveta Disposable Hysteroscope, there is no change to the components of the cleared Aveta
{4}------------------------------------------------
System (K190372). Provided in Table 1 are additional details on the modified Aveta Disposable Hysteroscope.
| Aveta Disposable<br>Hysteroscope and<br>Accessory | Device<br>Characteristics &<br>Packaging | Materials; Patient<br>Contact and Contact<br>Duration | Functions Performed |
|---------------------------------------------------|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Aveta Disposable<br>Hysteroscope | Sterile (EO), single<br>use, used in the sterile<br>field. Tyvek/Film<br>Pouch | Polymers (PVC, ABS);<br>Stainless Steel, direct<br>patient contact for<br>limited duration. | • Visualization of uterine cavity.<br>• Provides conduits/lumens for inflow,<br>outflow.<br>• Provides conduit for mechanical Resecting<br>Device for operative hysteroscopy.<br>• Provides controls to the user to record<br>images.<br>• Provides controls to the user to perform<br>image transfer |
| Aveta Controller<br>(accessory) | Non-sterile, reusable,<br>used outside the<br>sterile field | No patient contact | • Camera, light, image processing and image<br>storing. |
Table 1. Modified Aveta Disposable Hysteroscope and accessory
### V. Indications for Use
The indications for use of the modified Aveta System is a subset of the indications for use of the predicate Aveta System.
| Device | Indications For Use |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Aveta Disposable Hysteroscope<br>(Subject Device) | The Aveta Disposable Hysteroscope is used to permit viewing of the<br>cervical canal and the uterine cavity for the purpose of performing<br>diagnostic and surgical procedures. |
| Aveta System<br>(Predicate)<br>(K190372) | The Aveta System is intended for intrauterine use by trained gynecologists<br>to permit viewing of the cervical canal and the uterine cavity, provide<br>liquid distension of the uterus and monitor the volume differential between<br>the irrigation fluid flowing into and out of the uterus during diagnostic and<br>surgical procedures to resect and remove tissue such as submucous<br>myomas. endometrial polyps and retained products of conception. |
#### Comparison of Indications for Use
This difference does not represent a new intended use.
#### VI. Comparison of Technological Characteristics with the Predicate Device
The subject Aveta Disposable Hysteroscope and the cleared Aveta Disposable Hysteroscope (cleared as a component of Aveta System K190372) have similar technological characteristics in terms of basic operating principle and basic design features with minor differences in compatible fluid sources and software of the Aveta Controller. The different technological characteristics of the subject devices do not raise different types of safety and effectiveness questions. The table below compares the subject device with the predicate device.
{5}------------------------------------------------
| Attribute | Subject Device | PREDICATE Device |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k# | K192100 | K190372 |
| Manufacturer: | Meditrina Inc. | Meditrina Inc. |
| Device Names | Aveta Disposable Hysteroscope | Aveta System |
| Class: | II | II |
| Regulation Name: | Hysteroscope and Accessories | Hysteroscope and Accessories / Hysteroscopic Insufflator |
| Regulation Number: | 884.1690 | 884.1690 / 884.1700 |
| Product Code: | HIH | HIH / HIG |
| Intended Use: | Intended to permit viewing of the cervical canal<br>and the uterine cavity. | Intended for the distention and visualization of the cervical<br>canal and the uterine cavity and for endoscopic resection and<br>removal of tissue chips via suction while monitoring volume<br>differential between the irrigation fluid flowing into and out<br>of the uterus. |
| Indications for Use: | The Aveta Disposable Hysteroscope is used to<br>permit viewing of the cervical canal and the<br>uterine cavity for the purpose of performing<br>diagnostic and surgical procedures. | The Aveta System is intended for intrauterine use by trained<br>gynecologists to permit viewing of the cervical canal and<br>the uterine cavity, provide liquid distension of the uterus<br>and monitor the volume differential between the irrigation<br>fluid flowing into and out of the uterus during diagnostic<br>and surgical procedures to resect and remove tissue such as<br>submucous myomas, endometrial polyps and retained<br>products of conception. |
| Materials of<br>Construction: | Stainless Steel,<br>ABS,<br>Silicone elastomer,<br>PVC,<br>Polycarbonate | Stainless Steel,<br>ABS.<br>Silicone elastomer,<br>PVC,<br>Polycarbonate |
| Working Length: | 224mm | 224mm |
| Outer Diameter: | 5.5mm | 5.5mm |
| Working Channel<br>Diameter: | 3.5mm | 3.5mm |
| Working Channel<br>Closure: | Silicone valve | Silicone valve |
| Recommended Light<br>Source: | Integrated LED | Integrated LED |
| Focal Length: | 5 to 25mm | 5 to 25mm |
| Working Distance: | 10mm | 10mm |
| Field of View: | 90° | 90° |
| Direction of View: | 0° | 0° |
| Optics | Fixed focus lens with CMOS sensor | Fixed focus lens with CMOS sensor |
| Attribute | Subject Device | PREDICATE Device |
| 510k# | K192100 | K190372 |
| Manufacturer: | Meditrina Inc. | Meditrina Inc. |
| Device Names | Aveta Disposable Hysteroscope | Aveta System |
| Resolution @<br>Working Distance: | 3.4 lp/mm at 25mm Working Distance | 3.4 lp/mm at 25mm Working Distance |
| Image Processing: | Image generated by the CMOS sensor in Aveta<br>Disposable Hysteroscope is processed and sent<br>to Aveta Controller for further processing and<br>displayed on a monitor | Image generated by the CMOS sensor in Aveta Disposable<br>Hysteroscope is processed and sent to Aveta Controller for<br>further processing and displayed on a monitor |
| How provided: | Sterile, single-use | Sterile, single-use |
| Sterilization<br>Method: | Ethylene oxide gas | Ethylene oxide gas |
| SAL: | 10-6 | 10-6 |
| Fluid Management<br>System: | · Use with gravity fed pressurized saline bag.<br>· Use with The Aveta Fluid Management<br>System cleared in K190372. | Used in combination with Aveta Fluid Management System |
| Tissue Resection: | • Used in combination with Aveta Disposable<br>Resecting Device cleared in K190372 | Used in combination with Aveta Disposable Resecting<br>Device |
{6}------------------------------------------------
The differences outlined above were evaluated to demonstrate safety and effectiveness of the Aveta Disposable Hysteroscope.
#### VII. Non-Clinical Performance Testing
The following summary results of design control activities have been provided in support of the substantial equivalence determination.
- . Software verification
- . Simulated Use with gravity fed pressurized saline bag (Regulation of cavity pressure was demonstrated in a model uterine system, flow rate tests and tissue resection) was performed utilizing the Aveta Disposable Hysteroscope. Results demonstrate the subject device met specifications.
#### VIII. Conclusions
The results of the non-clinical testing described above demonstrate that the subject device is as safe and effective as the predicate device and support the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.