K190639 · CooperSurgical, Inc. · HIH · May 15, 2019 · Obstetrics/Gynecology
Device Facts
Record ID
K190639
Device Name
Endosee System
Applicant
CooperSurgical, Inc.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
May 15, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Endosee System is used to permit viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and therapeutic procedures. The types of procedures where the Endosee System could offer visualization include: - Abnormal bleeding - Infertility and pregnancy wastage - Evaluation of abnormal hysterosalpingogram - Intrauterine foreign body - Amenorrhea - Pelvic pain - Directed biopsy - Removal of fibroids and polyps - Transection of intrauterine adhesions - Transection of intrauterine septa The Endosee System can also be used to permit viewing of the adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic and therapeutic procedures.
Device Story
Handheld, battery-operated, portable endoscope; consists of reusable LCD touchscreen display module and sterile, single-use cannula. Cannula features distal light source and camera; captures image/video signals. Signals transferred electronically to display module for real-time visualization. Used in OR, outpatient, or office settings by clinicians. Docking station recharges battery and enables image/video download to external computer. Device facilitates diagnostic and therapeutic hysteroscopy and cystoscopy; provides direct visualization for procedures like biopsy, polyp removal, and adhesion/septa transection. Enables minimally invasive access via natural orifices; improves clinical decision-making through immediate visual feedback.
Clinical Evidence
Bench testing only. Performance verified through design validation and functional testing on 6-month aged samples, including tip retention, instrument access, and EMC/electrical safety re-evaluation per IEC 60601-2-18.
Technological Characteristics
Handheld, battery-operated endoscope. Components: reusable LCD display module, sterile single-use flexible cannula (276mm length, 4.5mm diameter). Light source: LED. Materials: PEEK, ink, glass, adhesive, lens coating, surface black coating. Connectivity: Wi-Fi, USB for data download. Sterilization: Ethylene oxide (SAL 10^-6).
Indications for Use
Indicated for adult patients requiring visualization of the cervical canal, uterine cavity, or urinary bladder for diagnostic and therapeutic procedures, including management of abnormal bleeding, infertility, foreign bodies, amenorrhea, pelvic pain, and tissue removal/biopsy.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 15, 2019
CooperSurgical, Inc. Christine Kupchick Regulatory Affairs Associate 95 Corporate Drive Trumbull, CT 06611
Re: K190639
Trade/Device Name: Endosee® System Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope And Accessories Regulatory Class: Class II Product Code: HIH, FAJ Dated: April 12, 2019 Received: April 15, 2019
Dear Christine Kupchick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon Andrews Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190639
Device Name Endosee® System
Indications for Use (Describe)
The Endosee System is used to permit viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and therapeutic procedures. The types of procedures where the Endosee System could offer visualization include:
- · Abnormal bleeding
- · Infertility and pregnancy wastage
- · Evaluation of abnormal hysterosalpingogram
- · Intrauterine foreign body
- · Amenorrhea
- · Pelvic pain
- · Directed biopsy
- · Removal of fibroids and polyps
- · Transection of intrauterine adhesions
- · Transection of intrauterine septa
The Endosee System can also be used to permit viewing of the adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic and therapeutic procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for CooperSurgical. The logo is in blue and consists of the words "CooperSurgical" in a stylized font. Below the logo is the tagline "Healthy women, babies, and families™" also in blue.
# 510(k) SUMMARY K190639
#### Submitter Information
| Company Name: | CooperSurgical Inc. |
|------------------|------------------------------------------|
| Company Address: | 95 Corporate Drive<br>Trumbull, CT 06611 |
| | Telephone: 203-601-5200 Ext. 3370 |
| | Fax: 203-601-9870 |
- Christine Kupchick Contact Person:
- Date Prepared: May 14, 2019
## Device Information
| Trade Names: | Endosee® System |
|--------------------|------------------------------|
| Common Name: | Hysteroscope; Cystoscope |
| Regulation Number: | 21 CFR 884.1690 |
| Regulation Name: | Hysteroscope and Accessories |
| Product Code: | HIH; FAJ |
| Regulatory Class: | Class II |
#### Predicate Device Information
Endosee System (K183020).
The predicate device has not been subject to a design-related recall.
#### Device Description
The Endosee System is a handheld, battery-operated, portable endoscope that consists of a reusable Display Module with an LCD touchscreen monitor and a sterile, single-use cannula. It is intended for use in viewing the adult cervical canal, uterine cavity, or female urinary tract, including the bladder, to perform diagnostic and therapeutic hysteroscopy or cystoscopy procedures in an operating room, outpatient, or office setting. The Cannula has a light source and camera at the distal end, which are used for visualization and to capture image and video of the diagnostic area. The image and video signals are electronically transferred from the Cannula to the Display module via an electrical connector and cable, and the LCD monitor is used for viewing. The Endosee is provided with a Docking Station that recharges the battery in the Display Module and allows the user to download images/video from the internal device memory to an external computer.
{4}------------------------------------------------
#### Indications for Use
The Endosee System is used to permit viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and therapeutic procedures. The types of procedures where the Endosee System could offer visualization include:
- Abnormal bleeding
- Infertility and pregnancy wastage
- Evaluation of abnormal hysterosalpingogram
- Intrauterine foreign body
- Amenorrhea
- Pelvic pain
- Directed biopsy
- Removal of fibroids and polyps
- Transection of intrauterine adhesions
- o Transection of intrauterine septa
The Endosee System can also be used to permit viewing of the adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic and therapeutic procedures.
#### Substantial Equivalence Discussion
TABLE 1 provides a comparison of the subject and predicate devices.
| Attribute | Subject Endosee System | Predicate Endosee System (K183020) |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | CooperSurgical, Inc. | Same |
| | The Endosee System is used to permit viewing<br>of the cervical canal and uterine cavity for the<br>purpose of performing diagnostic procedures.<br>Generally recognized indications for diagnostic<br>hysteroscopy include: | The Endosee System is used to permit viewing<br>of the adult cervical canal and uterine cavity for<br>the purpose of performing diagnostic and<br>therapeutic procedures. The types of<br>procedures where the Endosee System could<br>offer visualization include: |
| | • Abnormal bleeding<br>• Infertility and pregnancy wastage<br>• Evaluation of abnormal hysterosalpingogram<br>• Intrauterine foreign body<br>• Amenorrhea and pelvic pain | • Abnormal bleeding<br>• Infertility and pregnancy wastage<br>• Evaluation of abnormal hysterosalpingogram<br>• Intrauterine foreign body<br>• Amenorrhea<br>• Pelvic pain<br>• Directed biopsy |
| Indications for Use | The Endosee System can also be used to permit<br>viewing of the urinary bladder through a<br>minimally invasive approach by utilizing natural<br>orifices for the purpose of performing diagnostic<br>procedures. | • Removal of fibroids and polyps<br>• Transection of intrauterine adhesions<br>• Transection of intrauterine septa<br>The Endosee System can also be used to permit<br>viewing of the adult urinary bladder through a<br>minimally invasive approach by utilizing natural |
#### Table 1: Subject and Predicate Comparison
{5}------------------------------------------------
| | | orifices for the purpose of performing diagnosti<br>and therapeutic procedures. |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Technology | Handheld, battery-operated endoscope that<br>consists of a reusable, detachable Display<br>Module and sterile, disposable Cannula with a<br>handle; Wi-Fi capability. | Same |
| Cannula Type and<br>Dimensions | Type: Flexible<br>Working Length: 276mm Largest Outer<br>Diameter: 4.5mm | Same |
| Cannula Working<br>Channel | Working channel for fluid infusion and<br>instrument access | Working channel for fluid infusion |
| View | Field of View: 100° ± 5°<br>Direction of View: 20° ± 3° | Same |
| Light Source | LED | Same |
| Image Display | Handheld LCD display module | Same |
| Patient Contacting<br>Materials | PEEK, ink, glass, adhesive, lens coating, surface<br>black coating | Same |
| Number of Uses | Cannula: single-use, disposable<br>Display Module: reusable | Same |
| Cannula Sterilization | Ethylene oxide, SAL 10-6 | Same |
| Cannula Shelf Life | 6 months | Same |
| Packaging | Cannula: Individually pouched or individually<br>pouched with a pipette Display Module: Boxed<br>with Docking Station, power cord, and<br>USB cable | Same |
The subject device and predicate device have different indications for use, however, the inclusion of additional procedures does not represent a new intended use. Therefore, the differences in indications do not raise different questions of safety or effectiveness as compared to the predicate.
The subject device and the predicate device are technologically comparable. The subject device differs from the predicate only in that the working channel in the cannula shaft can be used for both fluid infusion and instrument insertion. This difference does not raise different questions of safety or effectiveness as compared to the predicate.
## Non-Clinical Performance Testing
As part of demonstrating substantial equivalence, the following testing was performed:
- . Design Validation
- Functional Testing (6-month aged samples)
- o Tip Retention
- o Instrument Access
- o EMC/Electrical Safety reevaluated with Essential Performance per IEC 60601-2-18
## Conclusion
The results of the testing described above demonstrate that the subject Endosee System is substantially equivalent to the predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.