MAKO 7

K182975 · Boston Scientific Corporation · HIH · May 21, 2019 · Obstetrics/Gynecology

Device Facts

Record IDK182975
Device NameMAKO 7
ApplicantBoston Scientific Corporation
Product CodeHIH · Obstetrics/Gynecology
Decision DateMay 21, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1690
Device ClassClass 2

Indications for Use

The MAKO 7 is a hysteroscope accessory, placed through the working channel of a hysteroscope to obtain samples from the fallopian tube for cytological evaluation.

Device Story

MAKO 7 is a sterile, single-use hysteroscope accessory used to collect cell samples from the fallopian tube for cytological evaluation. The device consists of a handle and a catheter featuring a balloon, a stainless-steel and Nylon shaft, and a Nylon 12 sheath. A physician inserts the device through the working channel of a hysteroscope, positioning the distal tip near the fallopian tube ostium. The balloon is advanced into the fallopian tube using a drive wheel on the handle to collect cells on its surface. The device connects to a commercially available inflation device via a luer and stopcock. The procedure requires hysteroscopic visualization. The device benefits patients by enabling minimally invasive collection of fallopian tube samples for diagnostic analysis.

Clinical Evidence

No new clinical data was provided. The clinical performance data demonstrating the ability to navigate target anatomy and obtain samples sufficient for cytological evaluation, as submitted for the predicate device (K160510), is applicable to the subject device.

Technological Characteristics

Catheter-based accessory for hysteroscopes. Materials: stainless-steel tube, Nylon tube, Nylon 12 sheath, Polycarbonate sheath knob. Features a balloon for sample collection, drive wheel for actuation, and three depth markers. Sterilization: Ethylene oxide (EO). Single-use. No electronic or software components.

Indications for Use

Indicated for patients requiring fallopian tube cytological sampling via hysteroscope. Prescription use only.

Regulatory Classification

Identification

A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 21, 2019 Boston Scientific Corporation Ibhina Sherchan Principal Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, MA 01752 Re: K182975 Trade/Device Name: MAKO 7 Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope And Accessories Regulatory Class: Class II Product Code: HIH Dated: April 10, 2019 Received: April 11, 2019 Dear Ibhina Sherchan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Sharon M. Andrews Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182975 Device Name MAKO 7 Indications for Use (Describe) The MAKO 7 is a hysteroscope accessory, placed through the working channel of a hysteroscope to obtain samples from the fallopian tube for cytological evaluation. Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary for MAKO 7 #### A. Date Prepared October 24, 2018 #### B. Submitter Boston Scientific Corporation Urology and Pelvic Health Division 100 Boston Scientific Way Marlborough, MA 01752 #### C. Contact Primary Contact: Ibhina Sherchan Regulatory Affairs Specialist Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Tel: (508) 683-4215 Fax: (508) 683-5041 Email: Ibhina.Sherchan@bsci.com Secondary Contact: Lisa Sullivan Director, Regulatory Affairs Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Tel: (508) 683-4745 Fax: (508) 683-5041 Email: Lisa.Sullivan@bsci.com ## D. Device Name Trade name: Common/usual name: Classification Name: #### MAKO 7 Hysteroscope Accessory HIH - Hysteroscope and accessories Obstetrics/Gynecology 21 CFR 884.1690, Class II ## E. Predicate Device Trade name: Common/usual name: MAKO 7 Hysteroscope Accessory {4}------------------------------------------------ | Classification Name: | HIH - Hysteroscope and accessories<br>Obstetrics/Gynecology<br>21 CFR 884.1690, Class II | |----------------------|------------------------------------------------------------------------------------------| | 510(k) number: | K160510 | The predicate device has not been subject to a design related recall. #### F. Device Description The MAKO 7 is a hysteroscope accessory intended to collect cell samples from the fallopian tube. The device is comprised of a catheter and a handle. The catheter includes a balloon, a shaft (which is made up of a stainless-steel tube and a Nylon tube), a sheath (Nylon 12), and a sheath knob (Polycarbonate). The catheter shaft has 3 markers to facilitate ease of use: the distal marker at the tip of the device, the retraction marker in the middle of the shaft, and the preparation (prep) marker near the proximal end. The handle includes a drive wheel and a luer that attaches to a commercially available inflation device via a stopcock. In summary, the physician inserts the MAKO 7 into the working channel of the hysteroscope until the distal tip of the catheter is positioned immediately proximal to the ostium of the fallopian tube. Then, the balloon is advanced into the fallopian tube and cells are collected on its surface. The MAKO 7 is a sterile, single-use device. The device is terminally sterilized using ethylene oxide (EO). ## G. Intended/Indications for Use The MAKO 7 is a hysteroscope accessory, placed through the working channel of a hysteroscope to obtain samples from the fallopian tube for cytological evaluation. The subject and predicate device have identical Indications for Use. #### H. Technological Characteristics The mechanism of action and principles of operation of the subject device and the predicate device are identical. Both devices use an identical handle, catheter body, balloon length, and, most distally, a balloon that collects the sample. For both devices, this distal portion is actuated by a drive wheel on the handle and a hysteroscopic view is required. The proposed MAKO 7 device has the same technical characteristics as its predicate, the currently cleared MAKO 7 (K160510). #### I. Performance Data No performance testing is required to evaluate the update of the Instructions for Use (IFU) to modify a warning that was previously listed and make an update to the directions for use section related to the updated warning. Performance testing {5}------------------------------------------------ for bending (kink resistance), pulling (tensile testing), balloon extension length and presence of leaks submitted for currently cleared MAKO 7 (K160510) is applicable to the subject device. Clinical performance data which demonstrated ability to navigate the target anatomy and obtain a sample sufficient for cytological evaluation submitted for currently cleared MAKO 7 (K160510) is applicable to the subject device. #### J. Conclusion The information provided in this submission demonstrates that the proposed MAKO 7 device is substantially equivalent to the predicate device MAKO 7, cleared in K160510.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...