Schoelly Cystoscopes/Hysteroscopes and Accessories
Applicant
Schoelly Fiberoptic GmbH
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Mar 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Schoelly Cystoscopes Hysteroscopes and Accessories are indicated to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples for the devices include the visualization and manipulation of anatomy as the surgeon deems appropriate. The Schoelly Cystoscopes and Accessories are intended to be used in general urological and gynecological surgery through a minimally invasive approach by utilizing natural orifices to access the surgical site.
Device Story
System comprises rigid reusable endoscopes, sheaths, obturators, bridges, and grasping forceps. Endoscopes transmit light via fiber optics from external source to surgical site; rigid lens system transfers images back to eyepiece for connection to camera/monitor/printer. Sheaths provide access channel for instruments and irrigation ports. Obturators facilitate sheath introduction. Bridges connect endoscope to sheath; some include accessory ports. Grasping forceps (flexible, 7Fr) allow tissue/foreign body retrieval. Used in clinic/OR by surgeons for minimally invasive visualization and manipulation of anatomy. Output is visual image for surgeon decision-making; benefits include minimally invasive access to urological/gynecological sites.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993), thermal safety (IEC 60601-2-18), optical performance (ISO 8600-1), packaging (ISO 2248), and reprocessing validation (AAMI TIR12/TIR30, ISO 17664, ISO 17665-1, ASTM E 1837). No clinical data presented.
Technological Characteristics
Rigid reusable endoscopes (2.9mm-4mm diameter, 300mm-365mm length). Materials: Phynox cobalt-nickel-chromium stainless steel alloy, 304 stainless steel, glass, glass fibers, adhesive, brazing alloy. Optical: 0°-70° direction of view, 70°-85° field of view. Fiber optic light transmission, rigid lens image transmission. Reprocessing: Manual/automated cleaning, high-level disinfection, steam sterilization. Compliant with IEC 60601-2-18.
Indications for Use
Indicated for patients undergoing endoscopic diagnostic and therapeutic surgical procedures in general urology and gynecology via natural orifices.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Stryker Urology and Gynecology Hardware System (K040390)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
JUL 2 7 2015
Schoelly Fiberoptic GmbH % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street NW Buffalo, MN 55313
Re: K150158
Trade/Device Name: Schoelly Cystoscopies/Hysteroscope and Accessories Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: HIH, FAJ Dated (Date on orig SE Itr): February 20, 2015 Received (Date on orig SE ltr): February 23, 2015
Dear Mark Job,
This letter corrects our substantially equivalent letter of March 9, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K150158
#### Device Name
Schoelly Cystoscopes/Hysteroscopes and Accessories
#### Indications for Use (Describe)
The Schoelly Cystoscopes Hysteroscopes and Accessories are indicated to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples for the devices include the visualization and manipulation of anatomy as the surgeon deems appropriate. The Schoelly Cystoscopes and Accessories are intended to be used in general urological and gynecological surgery through a minimally invasive approach by utilizing natural orifices to access the surgical site.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments reqarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## General Information
.
| Preparation date: | 3/4/2015 | | |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Owner's Name: | Schoelly Fiberoptic GmbH (Registration: 8043903) | | |
| Address: | Robert-Bosch-Str. 1 – 3 | | |
| | 79211 Denzlingen | | |
| | Germany | | |
| Telephone Number: | +49-7666-980-0 | | |
| Fax Number: | +49-7666-908-380 | | |
| Contact Person: | Dr. Sandra Baumann | | |
| Subject Device Name: | Schoelly Cystoscopes/Hysteroscopes and Accessories | | |
| Trade Name: | Schoelly Cystoscopes/Hysteroscopes and Accessories | | |
| Common/Usual Name: | Endoscope (Cystoscope, Hysteroscope), Sheath, Obturator,<br>Bridge, Grasping Forceps | | |
| Classification Name: | FAJ – Cystoscope and accessories, flexible/rigid<br>21 CFR 876.1500; Class II | | |
| | HIH- Hysteroscope (and accessories)<br>21 CFR 884.1690; Class II | | |
| | KOG- Endoscope and/or accessories<br>21 CFR 876.1500; Class II | | |
| Predicate Device Name: | Flexilux II Cystoscope and Hysteroscope | | |
| Trade Name: | Flexilux II Cystoscope and Hysteroscope | | |
| Common/Usual Name: | Endoscope (Cystoscope, Hysteroscope) | | |
| Classification Name: | FAJ – Cystoscope and accessories, flexible/rigid<br>21 CFR 876.1500; Class II | | |
| | HIH- Hysteroscope (and accessories)<br>21 CFR 884.1690; Class II | | |
| Premarket Notification: | K060899, Schoelly Imaging Inc., SE date March 2, 2007 | | |
| Predicate Device Name: | Stryker Urology and Gynecology Hardware System | | |
| Trade Name: | Stryker Urology and Gynecology Hardware System | | |
| Common/Usual Name: | Scope, Obturator, Working Element, Sheath, Bridge,<br>Electrode, Albarran Deflector, Timberlake Obturator,<br>Cutting Loop, Roller Ball, Cold Knives,<br>Dilator/Sound/Bougie, Bladder Syringe, Ellik Evacuator,<br>Forceps | | |
| Classification Name: | HIH- Hysteroscope (and accessories)<br>21 CFR 884.1690; Class II | | |
| | FAJ – Cystoscope and accessories, flexible/rigid | | |
{4}------------------------------------------------
21 CFR 876.1500: Class II FAS - Electrode, electrosurgical, active, urological 21 CFR 876.4300; Class II KOG- Endoscope and/or accessories 21 CFR 876.1500; Class II KQT - Evacuator, gastro-urology 21 CFR 876.4370; Class II
Premarket Notification: K040390, Stryker Endoscopy, SE date May 17, 2004
## Device Description
The proposed Schoelly Cystoscopes/Hysteroscopes and Accessories comprise several models of rigid endoscopes as well as endoscopic sheaths, obturators, instrument bridges and grasping forceps. The devices are indicated to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples for the use of the devices include the visualization and manipulation of anatomy as the surgeon deems appropriate. The Schoelly Cystoscopes/Hysteroscopes and Accessories are intended to be used in general urological and gynecological surgery through a minimally invasive approach by utilizing natural orifices to access the surgical site.
## Endoscopes (Cystoscopes/Hysteroscopes):
The endoscopes described in this submission are rigid reusable endoscopes for visualization of the operating site during cvstoscopic and hysteroscopic minimally invasive procedures in conjunction with a commercially available light source, video camera, monitor, and printer.
Light that is created by an external light source is transmitted from the endoscope's light guide connector through the endoscope itself to the tip via a fiber optic system. Images are transferred the other way back through a rigid lens system.
Technical parameters of the Schoelly Cystoscopes/Hysteroscopes that characterize the optical view are the Direction of View (0°-70°) and the Field of View (70°-85°). The image can be displayed by a camera/monitor system which can be connected to the endoscope eyepiece. Models of the Schoelly Cystoscopes/Hysteroscopes differ in diameter and length of the insertion tube (2.9mm; 4mm / 300mm - 365mm). None of the endoscope models have a working channel.
Like other currently marketed rigid cystoscopes and hysteroscopes, all endoscope models have outer surfaces mainly made from metal (Phynox cobalt-nickelchromium stainless steel alloy, 304 stainless steel) and incorporate fiber optics for light transmission and rigid lenses for image transmission.
Some Schoelly endoscope models have already been cleared for marketing by FDA (K060899) for the same general intended use but with a different material of the
{5}------------------------------------------------
insertion tube and a different bonding material used for the fixation of the plan glass at the endoscope's distal end. Accessories do not have a prior 510(k) clearance.
#### Sheath:
The endoscopic sheaths included in this submission are rigid reusable instruments with an inner lumen and mainly made from stainless steel; the sheath serves as the most outer part of the whole device setup in cystoscopic or hysteroscopic procedures. The proximal end of the endoscopic sheath has two irrigation ports with integral stopcocks for the introduction and the egress of irrigation media. The distal end of the sheath is cut away in a fenestration to permit the use of working instruments and endoscopes with a Direction of View of more than 0°. Opposite to the fenestration, the sheath is bevelled to facilitate its introduction into natural body orifices (transurethral/transvaginal). The outer diameter of the sheaths in this submission ranges from 17Fr - 25Fr, which are standard sizes used in cystoscopic or hysteroscopic procedures in adults. The sheaths can be used with all endoscopes models, obturators and bridges that are included in this submission.
## Obturator:
The obturators included in this submission are rigid reusable instruments with (visual obturators) or without (blind obturators) an inner lumen and mainly made from stainless steel. During application, the obturator can be attached to the endoscopic sheath; it fills the space inside the sheath to provide a smooth surface. The visual obturator has an inner channel that accommodates the endoscope and allows the sheath to be introduced under direct vision. The overall length of the obturator used in conjunction with the sheath complies with the working length of the endoscope.
#### Bridges:
This submission includes standard endoscope bridges, which are rigid reusable instruments with an inner lumen and mainly made from stainless steel. The evaluation bridges without any accessory port allow the connection of the endoscope to the endoscopic sheath; the single horn bridges include an accessory port with stopcock to allow the insertion of instruments through the inner lumen of the sheath. Both types of bridges are available in a range of lengths for use with the full range of endoscope working lengths.
#### Grasping forceps:
As with standard endoscopic grasping forceps, the one described in this submission is a flexible forceps solely made from stainless steel and designed for grasping tissue and/or retrieving foreign bodies under endoscopic visualization. It consists of a flexible shaft and a manual proximal control handle. Operation of the proximal control handle actuates the distal tip grasping jaws. The outer diameter and the working length of the grasping forceps comprised in this submission are 7Fr and 400 mm, respectively. The grasping forceps can be introduced into the instrument bridge and moved forward through the inner lumen of the sheath towards the tip.
{6}------------------------------------------------
The Schoelly Cystoscopes/Hysteroscopes and Accessories are delivered in a nonsterile condition and will have CE mark.
## Indications for Use
The Schoelly Cystoscopes/Hysteroscopes and Accessories are indicated to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples for the use of the devices include the visualization and manipulation of anatomy as the surgeon deems appropriate. The Schoelly Cystoscopes/Hysteroscopes and Accessories are intended to be used in general urological and gynecological surgery through a minimally invasive approach by utilizing natural orifices to access the surgical site.
## Predicate Device Comparison
The Indications for Use for the proposed Schoelly Cystoscopes/Hysteroscopes and Accessories comprises a subset of the Indications for Use of the predicate Schoelly Flexilux II Cystoscope and Hysteroscope (K060899) and the predicate Stryker Urology and Gynecology Hardware System cleared in K040390.
The predicate Schoelly Flexilux II Cystoscope and Hysteroscope (K060899), which consists only of an endoscope with no accessories, has a very general indications statement referring to use of the instruments for visualization of and therapeutic access to the organs and anatomical structures of interest (hollow organs such as the bladder, uterus, cervical canal). Therefore, the indications statement for the proposed Schoelly Cystoscopes/Hysteroscopes and Accessories was created as an exact subset of the indications for use of the predicate Stryker Urology and Gynecology Hardware System cleared in K040390.
Because the proposed Schoelly Cystoscopes/Hysteroscopes and Accessories system does not include accessories such as knives, scissors, electrodes, etc., the specific examples of uses involving these accessory devices that are present in the Stryker predicate indications statement are not included in the indications statement for this submission.
{7}------------------------------------------------
| | Attribute | Proposed<br>Device | Predicate<br>Schoelly Flexilux II Cystoscope and<br>Hysteroscope (K060899) | Predicate Stryker<br>Stryker Urology and Gynecology Hardware System<br>(K040390) |
|--|----------------------------------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| | | | Endoscope | |
| | Light transmission | Fiber optics | Fiber optics | Fiber optics |
| | Light source | External, connected via light guide to light<br>guide connector | External, connected via light guide to light guide<br>connector | External, connected via light guide to light guide<br>connector |
| | Image transmission | Rigid lenses | Rigid lenses | Rigid lenses |
| | Direction of view | 0°-70° | 0°-70° | 0°-70° |
| | Field of view | 70°-85° | 70°; 75° | Not known |
| | Image display | Camera/monitor connected to eyepiece | Camera/monitor connected to eyepiece | Camera/monitor connected to eyepiece |
| | Insertion tube working length | 300mm - 365mm | 302mm | 302mm |
| | Insertion tube outer diameter | 2.9mm;4mm | 2.9mm; 4mm | 2.9mm; 4mm |
| | Working channel | None | None | None |
| | Reprocessing | Cleaning (manual and automated),<br>High level disinfection, Sterilization (steam) | Cleaning (manual and automated),<br>Sterilization (steam, EtO, Sterrad, Amsco V-Pro) | Cleaning,<br>Sterilization (steam) |
| | Materials | Stainless steel, stainless steel alloy, glass, glass<br>fibers, adhesive, brazing alloy | Stainless steel, stainless steel alloy, glass, glass<br>fibers, adhesive | Stainless steel, stainless steel alloy, glass, adhesive |
| | Thermal safety | IEC 60601-2-18 compliant | IEC 60601-2-18 compliant | IEC 60601-2-18 compliant |
| | Instruments | | | |
| | Sheath outer diameter/ Visual<br>Obturator Compatibility | 17Fr - 25 Fr | N/A (not scope of K060899) | 17Fr - 25Fr |
| | Visual Obturator Total Length | Compatible with sheath and endoscopes with<br>300mm - 365mm working length | N/A (not scope of K060899) | Compatible with sheath and endoscopes with 302mm<br>working length |
| | Bridg Design | evaluation an single horn bridge | N/A (not scope of K060899) | evaluation, single horn and double horn bridge |
| | Bridge Total Length | Compatible with sheath and endoscopes with<br>300mm - 365mm working length | N/A (not scope of K060899) | Compatible with sheath and endoscopes with 302mm<br>working length |
| | Forceps Design | Flexible | N/A (not scope of K060899) | Flexible, rigid |
| | Forceps Working Length | 40cm | N/A (not scope of K060899) | 40cm (flexible) |
| | Forceps Outer Diameter | 7Fr | N/A (not scope of K060899) | 5Fr, 7Fr, 9Fr |
| | Reprocessing | Cleaning (manual and automated),<br>High level disinfection, Sterilization (steam) | N/A (not scope of K060899) | Cleaning,<br>Sterilization (steam) |
| | Material | Stainless steel, Plastic | N/A (not scope of K060899) | Stainless steel, Plastic |
్లిక
이 : 00:00 | 10:00 |
# Technological Characteristics: Similarities and Differences between the Proposed Device and the Predicate Devices
: 2017-04-04 11:42:13
{8}------------------------------------------------
## Non-clinical Performance Testing
Performance data demonstrated that the Schoelly Cystoscopes/Hysteroscopes and Accessories have met pre-determined acceptance criteria and are substantially equivalent to predicate devices. The devices are as safe, as effective, and perform as well as or better than the predicate devices. The risks associated with the use of the new devices were found acceptable when evaluated in accordance with • ISO 14971:2012 Medical devices - application of risk management to medical devices. Risks and benefits of predicate devices are the same as compared to the proposed ones.
The following performance testing was conducted for the new devices:
A) Biocompatibility testing as per ISO 10993-1:2009 including:
- . Cytotoxicity as per ISO 10993-5:2009
- Sensitization as per ISO 10993-10:2002 .
- . Irritation as per ISO 10993-10:2002
- Systemic toxicity as per ISO 10993-11:2006 .
B) Temperature testing as per IEC 60601-2-18:2009
C) General performance testing as per ISO 8600-1:2013 and FDA Guidance Hysteroscopes and Gynecologic Laparoscopes - Submission Guidance for a 510(k), issued March 7, 1996 including:
- Field of view ●
- Direction of view ●
- Diopters
- Eccentricity ●
- Size of view ●
- Vignetting
- Image resolution .
- Image distortion .
D) Packaging testing as per ISO 2248:1985
E) Reprocessing testing including:
- Manual and automated cleaning efficacy validation as per AAMI . TIR12:2010 and AAMI TIR30:2011
- Steam sterilization efficacy validation as per ISO 17664:2004 and . ANSI/AAMI/ISO 17665-1:2006
- For use of devices during cystoscopy: High level disinfection efficacy . validations as per ASTM E 1837:96 (2007)
Devices that have been used for reprocessing testing had underwent multiple soiling-cleaning cycles and surface marring artificially created by scratching with metal tools to simulate end of lifetime use.
## Conclusion
The Schoelly Cystoscopes/Hysteroscopes and Accessories meet all the predetermined acceptance criteria of the testing performed to confirm substantial equivalence to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.