LAPARO CO2-PNEU (AND WITH VIDEO MONITORING)

K981334 · Richard Wolf Medical Instruments Corp. · HIF · Sep 10, 1998 · Obstetrics/Gynecology

Device Facts

Record IDK981334
Device NameLAPARO CO2-PNEU (AND WITH VIDEO MONITORING)
ApplicantRichard Wolf Medical Instruments Corp.
Product CodeHIF · Obstetrics/Gynecology
Decision DateSep 10, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1730
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Laparo CO2-Pneu generates and maintains pneumoperitoneum CO2 gas and can be used in diagnostic and operative laparoscopy.

Device Story

The Laparo CO2-Pneu 2232 is an automatic laparoscopic insufflator used to dilate the abdomen for diagnostic and operative laparoscopy. It connects to CO2 gas cylinders or central gas supplies to deliver gas into the patient's abdomen. The device features microprocessor control to regulate and maintain intra-abdominal pressure and flow rates. Operators can preselect pressure (5-25 mmHg) and flow (1-20 l/min) settings. The system includes independent safety mechanisms to ensure safe insufflation and automatically reduces pressure if the intra-abdominal pressure exceeds the preselected limit. Real-time data regarding pressure, flow, and gas consumption are displayed on-screen via video equipment for clinician monitoring during surgical interventions. The device is intended for use by healthcare professionals in clinical settings.

Clinical Evidence

No clinical tests performed. Bench testing only.

Technological Characteristics

Microprocessor-controlled laparoscopic insufflator. Features stepless pressure preselection (5-25 mmHg) and flow rate control (1-20 l/min). Provides on-screen display of pressure, flow, and gas consumption via video integration. Includes independent safety devices for pressure regulation.

Indications for Use

Indicated for patients undergoing diagnostic or operative laparoscopic procedures requiring pneumoperitoneum via CO2 gas insufflation.

Regulatory Classification

Identification

A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 353 Corporate Woods Parkway Vernon Hills, Illinois 60061 Phone: 847.913.1113 Fax: 847.913.1488 SEP 1 0 1998 RICHARD V MEDICAL INSTRUMENTS CO K981334 # .510(k) Summary of Safety and Effectiveness | Company / Institution name: | Richard Wolf Medical Instruments Corp. | Date of Preparation: | April 8, 1998 | |--------------------------------|----------------------------------------|--------------------------------------|----------------| | Division name (if applicable): | N.A. | FDA establishment regulation number: | 14 184 79 | | Street address: | 353 Corporate Woods Parkway | Phone number (include area code): | (847) 913-1113 | | City: | Vernon Hills | FAX number (include area code): | (847) 913-0924 | | State/Province: | Illinois | ZIP/Postal Code: | 60061 | | Country: | USA | | | | Contact name: | Mr. Robert L. Casarsa | | | | Contact title: | Quality Assurance Manager | | | Product Information: | Trade name: | Laparo CO₂ Pneu | Model number: | 2232.601, .611 | |--------------|----------------------------------------|----------------------|--------------------------| | Common name: | Automatic Laparoscopic CO₂ Insufflator | Classification Name: | Laparoscopic Insufflator | Information on devices to which substantial equivalence is claimed: | 510(k) Number | Trade or proprietary or model name | Manufacturer | |---------------|------------------------------------|----------------| | 1 K934884 | 1 2231 Laparoscopic Insufflator | 1 Richard Wolf | | 2 | 2 System 4000 | 2 CABOT | | 3 | 3 Medicam 900 | 3 MP Video | | 4 | 4 26012C | 4 Karl Storz | #### 1.0 Description The Laparo CO2-Pneu 2232 supplies CO2 gas from cylinders or central gas supplies to dilate the abdomen by CO2 gas insufflation. #### 2.0 Intended Use The Laparo CO2-Pneu generates and maintains pneumoperitoneum CO2 gas and can be used in diagnostic and operative laparoscopy. {1}------------------------------------------------ #### Technological Characteristics 3.0 The pressure and flow values are checked by microprocessor control. A series of independent safety devices guarantee safe and problem-free insufflation. The intra-abdominal pressure can be steplessly preselected between 5 mmHg and 25 mmHg. The pressure is reduced if the intra-abdominal pressure is higher than the preselected value. The flow rate can be preselected between 1 1/min and 20 1/min in steps of 1 1/min. The measured values can be displayed on-screen in order to document the pressure, flow and gas consumption values during intervention via video equipment. #### Substantial Equivalence 4.0 The Laparo CO,-Pneu 2232 is substantially equivalent to existing 510(k) devices sold by Richard Wolf and to various existing laparoscopic insufflators from Cabot, MP Video, and Karl Storz. The differences do not affect the safety or effectiveness of the device. #### 5.0 Performance Data Performance data not generated. #### 6.0 Clinical Tests No clinical tests performed. #### 7.0 Conclusions Drawn Laparo CO2-Pneu 2232 was designed and tested to guarantee the safety and effectiveness furing the expected lifetime of the device if it is used according to the instruction manual. By: Robert J. Casana Robert L. Casarsa Quality Assurance Manager Date: Ape 1, 1998 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo of the U.S. Department of Health and Human Services (HHS). The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH" is arranged in a circular fashion around the left side of the emblem. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 0 1998 Mr. Robert L. Casarsa Manager of Quality Assurance RICHARD WOLF MEDICAL INSTRUMENTS CORP. 353 Corporate Woods Parkway Vernon Hills, IL 60061 Re: K981334 Laparo CO2 - Pneu 2232 (Laparoscopic Insufflator) Dated: June 26, 1998 Received: June 27, 1998 Regulatory Class: II 21 CFR 884.1730/Procode: 85 HIF Dear Mr. Casarsa: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to ievices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, garket the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html". Sincerely yours Lillian Yin, Ph.D. Director, Division of Reproducti Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ RICHARD WOLF QA DEPT. ਾ ਸ਼ 1. The Bank of the States of the States of the States of the States of the States of the States of the States of the States of the States of the States of the States of the S ## Indications for Use | 510(k) Number (if known): | K9813.34 | |---------------------------|----------------------| | Device Name: | Laparo CO, Pneu 2232 | ### Intended Use: The Laparo CO2-Pneu generates and maintains pneumoperitoneum CO3 gas and can be used in diagnostic and operative laparoscopy ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (ODE) Vimit L. Sayan (Division Sign-Off) (Division Sign-Ort) Division of Reproductive, Abdominal, ENT, and Radiological Devic 510(k) Number Prescription Use Per 21 CFR 801.109 ✓ ( )દર Over-The Counter_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...