The AP50/30 Insufflator is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The Standard/High Flow, Pediatric and Bariatric operating modes of the device are indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is specifically indicated for pediatric laparoscopic procedures. The Vessel Harvesting mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery. The Insuflow® Port (5 mm, 8 mm, 10mm and 12mm) devices have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.
Device Story
The Lexion AP50/30 Insufflator is a microprocessor-controlled CO2 insufflator used in hospital settings for laparoscopic and vessel harvesting procedures. It accepts CO2 gas, which it regulates, heats, and humidifies via the single-use Insuflow Port. The device features a touch-screen interface for setting pressure (1-30 mmHg) and flow rates (up to 50 lpm). It performs continuous pressure monitoring using sensors integrated into the Insuflow Port or standard sensing tubes. The system includes safety alarms for overpressure, occlusion, and malfunctions. Physicians or surgical staff operate the device to maintain cavity distension. By providing warmed, humidified gas, the device aims to reduce patient discomfort and potential complications associated with cold, dry gas insufflation. The system is designed for minimally invasive surgery, providing a controlled path of entry for instruments while managing the surgical environment.
Clinical Evidence
Bench testing only. No clinical data. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and comparative bench performance (set pressure, leakage compensation, overpressure, pressure control). Biocompatibility testing for the Insuflow Port (cytotoxicity, irritation, sensitization, acute systemic toxicity) performed per ISO 10993-5, -10, and -11.
Technological Characteristics
Microprocessor-controlled CO2 insufflator. Components: metal housing, power supply, pressure reducers, venting system, gas heater/humidifier control, touch screen. Connectivity: plug connector cable for Insuflow Port. Sterilization: ETO for single-use tubing/port. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993. Software: Major level of concern.
Indications for Use
Indicated for patients undergoing diagnostic/therapeutic endoscopic procedures requiring cavity distension via CO2 insufflation, including laparoscopic procedures (Standard/High Flow, Pediatric, Bariatric modes) and endoscopic vessel harvesting (saphenous vein or radial artery). Pediatric mode specifically for pediatric laparoscopic procedures.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
Lexion Insuflow® Synergy™ Port and Synergy™ XL Port (K140263)
Submission Summary (Full Text)
{1}------------------------------------------------
November 14, 2017
LEXION Medical LLC % Bernard Horwath Regulatory Consultant HRG 4486 Timberline Ct Vadnais Heights, MN 55127
Re: K170799
Trade/Device Name: Lexion AP50/30 Insufflator with Insuflow Port Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF. GCJ. OSV Dated: October 13, 2017 Received: October 16, 2017
Dear Bernard Horwath:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the text "Benjamin R. Fisher -S". The text is written in a simple, sans-serif font and is horizontally aligned. The background of the image is a light blue color.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170799
Device Name Lexion AP50/30 Insufflator with Insuflow Port
#### Indications for Use (Describe)
The AP50/30 Insufflator is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The Standard/High Flow, Pediatric operating modes of the device are indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is specifically indicated for pediatric laparoscopic procedures. The Vessel Harvesting mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery.
The Insuflow® Port (5 mm, 8 mm, 10mm and 12mm) devices have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
## Lexion AP50/30 Insufflator with Insuflow® Port
| Date Prepared: | November 7, 2017 |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Lexion Medical, LLC<br>545 Atwater Circle<br>St. Paul, MN 55103<br>Telephone: 651-855-1447<br>Fax: 651-636-1671 |
| Contact: | Mr. Bernard Horwath<br>Regulatory Affairs Consultant<br>4486 Timberline Ct<br>Vadnais Heights, MN 55127<br>Telephone: 651- 231-1761 |
Device Name: Lexion AP50/30 Insufflator with Insuflow® Port
Common/Usual Name: Carbon dioxide insufflator for laparoscopic insufflation and endoscopic vessel harvesting (AP50/30), and gas conditioner path of entry device (Insuflow Port)
Classification Name/Code: Class II per regulations 884.1730, Laparoscopic Insufflator Product Code: HIF=Insufflator, laparoscopic Class II per regulations 884.1730, Laparoscopic Insufflator Product Code: OSV=Insufflator, endoscopic vessel harvesting Class II per regulations 876.1500, Endoscope and Accessories Product Code: GCJ=Laparoscope, general and plastic surgery
# Predicate Devices:
LEXION Medical believes that the AP50/30 Insufflator with Insuflow® Port devices is substantially equivalent to the following devices:
- Primary Predicate: Stryker/WOM 45L CORE Insufflator F114, K063367
- WOM Insufflator 50L FM134, K153513
- Lexion Insuflow® Synergy™ Port and Synergy™ XL Port, K140263 ●
None of the cited predicate devices have been subject to a design-related recall.
# Description:
The Lexion AP50/30 Insufflator is a microprocessor controlled CO2 (carbon dioxide) insufflator with multiple operating modes. The insufflator instrumentation is intended for hospital use for Standard/High Flow, Pediatric, and Bariatric laparoscopic procedures used in conjunction with a laparoscope to fill and distend a peritoneal cavity with gas, and for Vessel Harvesting procedures used to create a cavity along the saphenous vein and/or the radial artery during an endoscopic vessel harvesting procedure.
The device incorporates the following major components and features: a metal housing, a
{5}------------------------------------------------
world power supply, pressure reducers, a venting system, a gas heater control and a touch screen user interface with various settings and display elements. The software operation includes system checks, user interface, setting adjustments, warning/error messages and service info. The device is equipped with a continuous pressure measurement mode that conformity of the actual pressure in the peritoneal or extraperitoneal cavity with the pre-set nominal pressure. The AP50/30 Insufflator is designed with several alarms/warnings to inform the operator in case of an overpressure or other malfunctions. The device is to be used with specially designed single-use tubing sets, the Insuflow Port, in order to utilize the full capabilities of continuous pressure measurement and gas heating and humidification. (The insufflator can also accept other tubing sets for gas deliver only.)
The Insuflow® Port (5, 8, 10, 12 mm) devices are gas conditioning/access port devices that attach to the outlet of the AP50/30 Insufflator and are designed to warm and humidify the CO2 gas stream prior to insufflation via an integral path of entry device during minimally invasive surgery. The Insuflow Port consists of an ethylene oxide sterilized, disposable single use tubing set and a path of entry access port device which contains the pressure sensors, a filter, and gas heater/humidifier. The access port device materials are intended for patient contact of less than 24 hours. The Insuflow® Port is connected to the AP50/30 Insufflator via a plug connector cable, which controls the pressure sensing, gas heating and safety circuits for the system.
The technology for gas conditioning and path of entry for the modified port devices is the same as the predicate Insuflow® Synergy™ Port marketed devices. Regulated CO2 gas from the AP50/30 Insufflator flows into the Insuflow® Port, through the in-line filter, continues along the tubing to enter the path of entry access device that contains the heating element and humidification media, and through the path of entry access device lumen for delivery into the patient's surgical cavity.
## Indications for Use:
The AP50/30 Insufflator is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The Standard/High Flow, Pediatric and Bariatric operating modes of the device are indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is specifically indicated for pediatric laparoscopic procedures. The Vessel Harvesting mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery.
The Insuflow® Port (5 mm, 8 mm, 10mm and 12mm) devices have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.
## Technological characteristics, comparison to predicate devices.
## AP50/30
The Lexion AP50/30 Insufflator is substantially equivalent to the Stryker/WOM 45L Core Insufflator (K063367) in terms of indications for use, operating mode, basic features, and custom tubing sets. Both insufflators have the same intended use and are used for the same indications. The AP50/30 Insufflator and the predicate device 45L Core Insufflator are CO2 insufflators intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. In addition, both the proposed device and the predicate device are designed with Standard/High Flow, Pediatric and Bariatric operating modes that are indicated to fill and distend a peritoneal cavity with
{6}------------------------------------------------
gas during a laparoscopic procedure. Both insufflators also incorporate a Vessel Harvesting operating mode that is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery.
The AP50/30 Insufflator and the predicate 45L Core Insufflator use the same basic operating principles and incorporate the same basic design technology. Both are microprocessor controlled CO2 insufflators that consist of similar major components and incorporate the same type of features, including continuous pressure measurement. In addition, both the proposed device and the predicate device 45L Core Insufflator are to be used with specially designed sterile, single-use tube sets. The following table provides a comparison of key performance specs for the two insufflators.
| Key<br>Performance<br>Specs | AP50/30 Insufflator | 45L CORE Insufflator | Comparison |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Insufflator<br>Indications for<br>Use | The AP50/30 Insufflator is a<br>CO2 insufflator intended for<br>use during diagnostic and/or<br>therapeutic endoscopic<br>procedures to distend a cavity<br>by filling it with gas. The<br>Standard/High Flow, Pediatric<br>and Bariatric operating modes<br>of the device are indicated to<br>fill and distend a peritoneal<br>cavity with gas during a<br>laparoscopic procedure. The<br>Pediatric operating mode is<br>specifically indicated for<br>pediatric laparoscopic<br>procedures. The Vessel<br>Harvesting operating mode is<br>indicated for use during<br>endoscopic vessel harvesting<br>procedures to create a cavity<br>along the saphenous vein or<br>radial artery. | The 45L CORE Insufflator<br>F114 is a CO2 insufflator<br>intended for use in diagnostic<br>and/or therapeutic endoscopic<br>procedures to distend a cavity<br>by filling it with gas. The high<br>flow application, the low flow<br>application and the bariatric<br>application of the device are<br>each indicated for use in<br>facilitating the use of a<br>laparoscope by filling the<br>peritoneal cavity with gas to<br>distend it. The low flow<br>application of the device is<br>indicated for pediatric use. The<br>vessel harvesting application of<br>the 45L CORE Insufflator F<br>114 is indicated for use during<br>endoscopic vessel harvesting<br>procedures to create a cavity<br>along the saphenous vein and/or<br>radial artery during endoscopic<br>vessel harvesting procedures. | Different; the<br>differences between<br>the subject and<br>predicate indications<br>statements are minor<br>and do not alter the<br>intended use |
| Operating Modes | Standard/High Flow, Pediatric,<br>Bariatric and Vessel<br>Harvesting | Standard/High Flow, Pediatric,<br>Bariatric and Vessel Harvesting | Same |
| Max Gas Flow | 20 lpm (Pediatric)<br>45 lpm (High Flow)<br>50 lpm (Bariatric)<br>10 lpm (Vessel Harvest) | 20 lpm (Pediatric)<br>40 lpm (High Flow)<br>45 lpm (Bariatric)<br>10 lpm (Vessel Harvest) | Different; the flow<br>rate of the subject<br>device in bariatric<br>and high flow is<br>higher than the<br>predicate; these<br>differences do not<br>raise different |
| Key<br>Performance<br>Specs | AP50/30 Insufflator | 45L CORE Insufflator | Comparison |
| Pressure Range | 1-20mmHg (Pediatric/Vessel<br>Harvest)<br>1-30mmHg (High<br>Flow/Bariatric) | 1-20mmHg (Pediatric/Vessel<br>Harvest)<br>1-30mmHg (High<br>Flow/Bariatric) | Same |
| Intra-abdominal<br>Max Pressure | 30 mmHg (High<br>Flow/Bariatric)<br>20 mmHg (Pediatric/Vessel<br>Harvest) | 30 mmHg (High<br>Flow/Bariatric)<br>20 mmHg (Pediatric/Vessel<br>Harvest) | Same |
| Maximum Supply<br>Pressure | 65 mm Hg in Bariatric, High<br>Flow, Pediatric and Vessel<br>Harvest modes and 60 mm<br>Hg in Veress mode | 70 mm Hg for Bariatric<br>mode, 65 mm Hg for High<br>Flow mode and 60 mm Hg in<br>Veress mode | Different; the<br>subject device<br>maximum supply<br>pressure is lower<br>for bariatric setting.<br>This difference<br>does not raise<br>different questions<br>of safety and<br>effectiveness. |
| Adjustable<br>Values | Pressure: 1-30 mmHg<br>Flow: 1-50 lpm | Pressure: 1-30 mmHg<br>Flow: 1-45 lpm | Different; the flow<br>rate of the subject<br>device is higher than<br>the predicate; this<br>differences does not<br>raise different<br>questions of safety<br>and effectiveness |
| Pneumo Pressure<br>Control | Constantly held based upon<br>pressure sensors in Insuflow<br>Port or standard insufflation<br>using pressure sensing in the<br>insufflator | Constantly held based upon real<br>time pressure sensing tube or<br>standard insufflation using<br>pressure sensing in the<br>insufflator | Same |
| User Interface | Touch screen | Touch screen | Same |
| Power | 100-240 V | 100-240 V | Same |
| Touch Screen<br>Menus (Menu<br>Options) | Configuration Menus:<br>- First Nominal Pressure<br>15mmHg (high flow)<br>15mmHg (bariatric)<br>10mmHg (vessel harvesting)<br>8mmHg (pediatric)<br>- Venting Valve Status<br>with Veress insufflation on<br>or off, Venting system on or<br>off<br>- Venting Pressure Limit: set at | Configuration Menus:<br>- First Nominal Pressure<br>15mmHg (high flow)<br>15mmHg (bariatric)<br>10mmHg (vessel harvesting)<br>8mmHg (pediatric)<br>- Venting Valve Status<br>with Veress insufflation on<br>or off, Venting system on or<br>off<br>- Venting Pressure Limit: set | Different; Pressure<br>limits for venting are<br>preset in the subject<br>device, while they<br>can be adjusted in the<br>predicate. Gas flow<br>rates differ to reflect<br>differences in flow<br>range. These<br>differences do not<br>raise different |
| Key<br>Performance<br>Specs | AP50/30 Insufflator | 45L CORE Insufflator | Comparison |
| | 3mmHg (cannot be changed)<br>- Venting Response time:<br>set at 3 sec (cannot be<br>changed)<br>- Gas Supply: House or Bottle<br>- Alarm Volume: set between<br>Level 1-3<br>- Gas Flow Rates: Quick set<br>rate 1-3<br>3, 20, 45 LPM (high flow)<br>3, 25, 50 LPM<br>(bariatric)<br>1.0, 4.0, 10.0 LPM (vessel<br>harvesting)<br>0.1 LPM (pediatric)<br>- Maximum Nominal Pressure:<br>setting range 5-30 mmHg for<br>bariatric and high flow<br>setting range of 5-20 mmHg<br>for vessel harvesting and<br>pediatric<br>- Flow Safety Limit: Limit On<br>or Off<br>- Warning Signal Occlusion:<br>Signal On or Off<br>- Language: English, Spanish | between 2-5 mmHg<br>- Venting Response time:<br>between 2-5 sec<br>- Gas Supply: House or Bottle<br>- Alarm Volume: set between<br>Level 1-3<br>- Gas Flow Rates: Quick set<br>rate 1-3<br>3, 20, 40 LPM (high flow)<br>5, 25, 45 LPM<br>(bariatric)<br>1.0, 4.0, 10.0 LPM (vessel<br>harvesting)<br>0.1 LPM (pediatric)<br>- Maximum Nominal Pressure:<br>setting range 5-30 mmHg for<br>bariatric and high flow<br>setting range of 5-20 mmHg<br>for vessel harvesting and<br>pediatric<br>- Flow Safety Limit: Limit On<br>or Off<br>- Warning Signal Occlusion:<br>Signal On or Off<br>- Language: English, Spanish | questions of safety<br>and effectiveness. |
| Error Warnings | Check gas supply, low supply<br>gas pressure, overpressure,<br>venting system active,<br>overpressure/venting system<br>active, occlusion,<br>contamination,<br>contamination/call for service,<br>gas heater/call for service, error<br>message/call for service,<br>device temperature error/turn<br>off device, venting valve non-<br>functioning, continuous<br>pressure sensing<br>deactivated/call for service,<br>flow safety limit, safety limit,<br>valve non-functioning/call for<br>service | Check gas supply, low supply<br>gas pressure, overpressure,<br>venting system active,<br>overpressure/venting system<br>active, occlusion,<br>contamination,<br>contamination/call for service,<br>gas heating defective/call for<br>service, gas temperature >42C,<br>error message/call for service,<br>device temperature error/turn<br>off device, venting valve<br>defective, RTP defective | Different; the minor<br>differences do not<br>raise different<br>questions of safety<br>and effectiveness |
| Gas Conditioning | Heat and humidify | Heat | Different; the subject<br>device includes<br>humidification with<br>the use of the<br>Insuflow port; this |
| Key<br>Performance<br>Specs | AP50/30 Insufflator | 45L CORE Insufflator | Comparison |
| | | | difference does not<br>raise different<br>questions of safety<br>and effectiveness |
| Tubing Sets | Custom, sterile, single use<br><i>Insuflow</i> Port | Custom, sterile, single use | Same |
# Comparison Table: Key Performance Specs
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
The technical differences between the proposed Lexion AP50/30 Insufflator and the predicate Stryker/WOM 45L Core Insufflator include the following:
- The housing of the proposed AP50/30 Insufflator is slightly smaller in dimensions and lighter in weight.
- The maximum flow rate in the Bariatric operating mode has been increased from 45 1/min to 50 l/min in the AP50/30 Insufflator.
- Continuous pressure measurement is provided by sensors in the Insuflow® Port while the . predicate device 45L Core Insufflator utilizes a real time pressure sensing tube.
- . The AP50/30 Insufflator provides heated, humidified CO2 via the Insuflow® Port while the 45L Core Insufflator provides just heated CO2.
- There are minor differences in configuration menus and settings, highlighted above. .
The differences in the technological characteristics of the proposed AP50/30 Insufflator and the predicate 45L Core Insufflator do not raise different questions of safety and effectiveness. The the maximum flow rate to 50 1/min for the AP50/30 Insufflator is identical to the secondary predicate WOM Insufflator 50L FM134 (K153513).
# Insuflow® Port
The modified Insuflow® Port (5 mm, 8 mm, 10 mm, and 12 mm) devices have intended use as the Insuflow® Synergy™ and Synergy™ XL Port predicate devices. The intended use of the modified device, as described in the labeling, has not changed as a result of the AP50/30 modifications.
The modified Insuflow® Port (5 mm, 8 mm, 10 mm and 12 mm) devices utilize the same gas conditioning technology and path of entry access port technology as the predicate Insuflow Synergy™ Port (5 mm) and the Insuflow® Synergy™ XL Port (8 mm, 10 mm and 12 mm) cleared under 510(k) K140263. All of the modified devices have the same port configuration, design and materials as the predicates. All of the Insuflow® Ports perform the same gas heating/humidification conditioning function as the predicates with minor design changes to accommodate the pressure sensors added to the circuitry. The lumen configurations, dimensions and materials of the Insuflow Port path of entry access device are equivalent to the predicate Insuflow® Synergy™ XL Port devices.
# Discussion of performance testing.
A risk analysis and resulting performance testing have been conducted to assure that the AP50/30 Insufflator with the Insuflow® Port devices perform in accordance with specifications and applicable
{10}------------------------------------------------
standards. Design verification testing of the AP50/30 Insufflator demonstrates that the device performs as intended, and that performance of the device is comparable to that of the predicate.
Electrical safety and electromagnetic compatibility testing was successfully performed by independent laboratories in accordance with the following standards:
- IEC 60601-1:2014 Ed 3- Medical Electrical Equipment Part 1: General . Requirements for Basic Safety and Essential Performance
- IEC 60601-1-2:2014 Medical Electrical Equipment Part 1 -2: General ● Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
The software used in the AP50/30 Insufflator was determined to have a major level of concern, and was developed and successfully validated in accordance with the FDA guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," issued May 11, 2005.
Comparative bench testing was performed to demonstrate that the performance of the AP50/30 Insufflator is substantially equivalent to that of the predicate device Stryker/WOM 45L Core Insufflator. The comparative bench test included the following tests: set pressure, compensation of small and large leakages, overpressure, and pressurement/control. The test results demonstrate that the performance of the proposed AP50/30 device related to reaching the set pressure, compensation of small and large leakages, pressurement/control and overpressure are comparable to the performance of the predicate device 45L Core Insufflator.
For the Insuflow® Port devices, the subject devices utilize the same design and materials as the predicate for gas conditioning, port access and sealing. The only difference is the addition of the pressure sensors on the subject device circuitry. Per FDA's biocompatibility guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," issued June 16, 2016, testing of the Insuflow device as an external communicating device with contact duration of less than 24 hours was conducted and included cytotoxicity, irritation, sensitization potential, and acute systemic toxicity per ISO 10993-5:2009(R)2014, ISO 10993-10:2010/(R)2014, and ISO 10993-11:2006/(R)2010 respectively. Analysis and testing of the sensor materials exposed to the gas flow demonstrated acceptable biocompatibility per ISO 10993-1. Because there was no change in device dimensions, the existing test data for insertion/removal testing, seal leak integrity testing and ETO sterilization previously provided in the predicate submission can be leveraged to support the safety and effectiveness of the subject device. Since the AP50/30 Insufflator now controls the heating and humidification of the Insuflow® Port device, gas temperature and humidity characterization and flow/pressure performance testing was successfully conducted to assure compliance to specification.
# Conclusion:
The subject and predicate devices have the same intended use. As identified above, there are several technological differences between the subject and predicate devices. None of these differences alter the intended use of the subject device from that of the predicate, or raise different questions of safety and effectiveness. Performance testing conducted on the subject device demonstrate that the subject device is as safe and effective as the predicate. Therefore, the AP50/30 Insufflator with Insuflow Port is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.