K153513 · W.O.M. World of Medicine GmbH · HIF · Mar 4, 2016 · Obstetrics/Gynecology
Device Facts
Record ID
K153513
Device Name
Insufflator 50L FM134
Applicant
W.O.M. World of Medicine GmbH
Product Code
HIF · Obstetrics/Gynecology
Decision Date
Mar 4, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1730
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device Insufflator 50L FM134 is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The Standard, Pediatric and Bariatric operating modes of the device are indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is specifically indicated for pediatric laparoscopic procedures. The Vessel Harvesting operating mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery.
Device Story
Microprocessor-controlled CO2 insufflator; inputs include gas supply and user settings via GUI touch screen; transforms inputs by regulating gas flow and pressure to distend body cavities; outputs include controlled CO2 delivery to surgical site. Used in OR/clinic settings by physicians; provides real-time pressure/flow monitoring to maintain cavity distension during laparoscopic or vessel harvesting procedures. Benefits include stable surgical field visualization and access. Device is non-sterile; uses sterile single-use tube sets (with/without heating wire/filter). Healthcare providers use GUI to monitor insufflation parameters and adjust flow/pressure to ensure patient safety and procedural efficacy.
Clinical Evidence
Bench testing only. Comparative performance testing against predicate F114 demonstrated equivalent ability to reach set pressure, compensate for leakages, and manage overpressure. Testing confirmed 50 L/min flow rate performance. Electrical safety/EMC testing per IEC 60601-1 and IEC 60601-1-2. Biocompatibility of tube sets per ISO 10993 series. Sterilization validation per ISO 11135-1, ISO 14937, and ISO 10993-7.
Technological Characteristics
Microprocessor-controlled CO2 insufflator. Features: pressure reducers, venting system, redundant pressure measurement, fluid sensor, gas heater, GUI touch screen. Connectivity: ISO 15/22 mm tube set connectors. Power: World power supply. Software: IEC 62304 compliant. Sterilization: Single-use tube sets ETO sterilized per ISO 11135-1/ISO 14937.
Indications for Use
Indicated for patients undergoing diagnostic/therapeutic endoscopic procedures requiring cavity distension via CO2 gas. Includes Standard, Pediatric, and Bariatric laparoscopic procedures for peritoneal cavity distension, and endoscopic vessel harvesting procedures for saphenous vein or radial artery access.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 4, 2016
W.O.M World of Medicine Gmbh Susanne Raab Regulatory Consultant 1480 Cambridge Street Cambridge, MA 02139
Re: K153513 Trade/Device Name: Insufflator 50L FM134 Regulation Number: 21 CFR 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: Class II Product Code: HIF, OSV Dated: November 25, 2015 Received: December 7, 2015
Dear Susanne Raab,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153513
Device Name Insufflator 50L FM134
#### Indications for Use (Describe)
The device Insufflator 50L FM134 is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The Standard, Pediatric operating modes of the device are indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is specifically indicated for pediatric laparoscopic procedures. The Vessel Harvesting mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for WOM, which is a medical device company. The logo consists of a circle with a stylized "W" inside, and the text "WOM" is written below the logo. Below the logo and name is the text "Insufflator 50L FM134" and "510(k) Premarket Notification".
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
# 1) General Information:
Product Code:
| Submitter: | W.O.M. WORLD OF MEDICINE GmbH<br>Salzufer 8<br>10587 Berlin<br>Germany<br>Phone: +49 30 39981 594<br>Fax: +49 30 39981 593 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Registration Number: | 3001556604 |
| Contact Person: | Susanne Raab<br>Regulatory Consultant<br>1480 Cambridge Street<br>Cambridge, MA 02139<br>Phone: (617) 547-0628<br>Fax: (617) 547-0628<br>e-mail: sbraab@comcast.net |
| Date Prepared: | November 24, 2015 |
| 2) Proposed Device | |
| Trade Name:<br>Common Name:<br>Classification Name:<br>Regulation Number and Name:<br>Regulatory Class:<br>Product Code: | Insufflator 50L FM134<br>Carbon Dioxide Insufflator for Laparoscopic<br>and Vessel Harvesting<br>Insufflator, Laparoscopic<br>21 CFR 884.1730, Laparoscopic Insufflator<br>II<br>HIF, OSV |
| 3) Predicate Devices: | |
| Primary Predicate Device:<br>Trade Name:<br>510(k) Number:<br>Classification Name:<br>Regulation Number and Name:<br>Regulatory Class:<br>Product Code: | 45L Core Insufflator F114<br>K063367<br>Insufflator, Laparoscopic<br>21 CFR 884.1730, Laparoscopic Insufflator<br>II<br>HIF |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the WOM logo, which includes a circle with a stylized figure inside and a red triangle at the bottom. Below the logo, the text "WOM" is displayed. Further down, the text "Insufflator 50L FM134" and "510(k) Premarket Notification" are written.
Secondary Predicate Device:
Trade Name: 510(k) Number: Classification Name: Regulation Number and Name: Regulatory Class: Product Code:
Nebulae™ I 50 LPM Insufflator K120151 Insufflator, Laparoscopic 21 CFR 884.1730, Laparoscopic Insufflator II HIF, FCX, OSV
#### 4) Device Description:
The Insufflator 50L FM134 is a microprocessor controlled CO2 insufflator that consists of the following major components and features: a casing, a world power supply, pressure reducers, a venting system, redundant pressure measurement, a fluid sensor, a gas heater, a software controlled graphical user interface (GUI) touch screen and various setting keys and display elements. The Insufflator 50L FM134 is not intended to enter the sterile field, and cannot be sterilized. The device is to be used with specially designed single-use tube sets that are delivered sterile. Specifically, the proposed device is to be used with a single-use tube set with heating wire and integrated filter or with a single-use tube set with integrated filter but without heating wire.
## 5) Intended Use:
The device Insufflator 50L FM134 is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The Standard, Pediatric and Bariatric operating modes of the device are indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is specifically indicated for pediatric laparoscopic procedures. The Vessel Harvesting operating mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with a circle and a design inside, followed by the text "WOM". Below this, the text "Insufflator 50L FM134" is displayed, followed by "510(k) Premarket Notification". The logo and text appear to be related to a medical device or product, possibly an insufflator.
#### 6) Comparison of Technological Characteristics:
The Insufflator 50L FM134 is substantially equivalent to the 45L Core Insufflator F114 (K063367). Both the proposed device and the predicate device F114 have the same intended use and are used for the same indications. Specifically, the Insufflator 50L FM134 and the predicate device 45L Core Insufflator F114 are CO2 insufflators intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. In addition, both the proposed device and the predicate device are designed with Standard, Pediatric and Bariatric operating modes that are indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. Finally, both the proposed device and the predicate device incorporate a Vessel Harvesting operating mode that is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery.
Furthermore, the Insufflator 50L FM134 and the predicate device 45L Core Insufflator F114 use the same basic operating principles and incorporate the same basic design. Both the proposed device Insufflator 50L FM134 and the predicate device 45L Core Insufflator F114 are microprocessor controlled CO2 insufflators that consist of the same major components and incorporate the same features. In addition, both the proposed device and the predicate device 45L Core Insufflator F114 are to be used with specially designed sterile, single-use tube sets. The technical differences between the proposed device and the predicate device 45L Core Insufflator F114 consist of the following:
- . The casing of the proposed device is slightly larger in dimensions and higher in weight.
- . The maximum flow rate in the Bariatric operating mode has been
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with a circle and a design inside, followed by the letters "WOM". Below the logo, the text "Insufflator 50L FM134" is displayed, followed by "510(k) Premarket Notification". The text provides information about a medical device, specifically an insufflator, and its regulatory status.
increased from 45 1/min to 50 1/min.
- The proposed device does not incorporate an optional continuous . pressure measurement feature.
- . The proposed device is not equipped with an optional Sidne control system interface.
- The proposed device is not designed with a proprietary connection for the tube sets but instead incorporates an ISO 15/22 mm connector.
- The proposed device is designed with separate connections for gas insufflation and for gas heating.
The differences in the technological characteristics of both the proposed device Insufflator 50L FM134 and the predicate device F114 do not raise new questions of safety and effectiveness. Moreover, with regards to the increase of the maximum flow rate to 50 l/min, the Insufflator 50L FM134 is substantially equivalent to the predicate device Nebulae™ I 50 LPM Insufflator (K120151).
## 7) Performance Data:
Electrical safety and electromagnetic compatibility testing was performed by independent laboratories in accordance with the following standards:
- IEC 60601-1:2005 Medical Electrical Equipment Part 1: General ● Requirements for Basic Safety and Essential Performance; and
- IEC 60601-1-2:2007 Medical Electrical Equipment Part 1 -2: General ● Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
Test results demonstrate that the proposed device conforms to the above standards.
The software was developed, tested and verified in accordance with the FDA guidance document Software Contained in Medical Devices" and in accordance with the following standard:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the WOM logo, which is a circle with a design inside and a red triangle at the bottom. Below the logo, the text "WOM" is written in a bold font. The text "Insufflator 50L FM134 510(k) Premarket Notification" is written below the logo.
- IEC 62304:2006 Medical Device Software Software Life Cycle . Processes.
Design verification testing of the Insufflator 50L FM134 demonstrates that the device performs as intended and that the performance does not raise new questions of safety and effectiveness.
In addition, comparative bench testing was performed to demonstrate that the performance of the Insufflator 50L FM134 in the Bariatric operating mode is substantially equivalent to that of the predicate device 45L Core Insufflator F114. The comparative bench test included comparison of both devices with regards to reaching the set pressure, compensation of small and large leakages, overpressure scenarios and increase of the maximum flow rate from 45 1/min to 50 1/min. The test results demonstrate that the performance of the proposed device related to reaching the set pressure, compensation of small and large leakages and overpressure scenarios are comparable to the performance of the predicate device 45L Core Insufflator F114. In addition, the test results demonstrate that a maximum flow rate of 50 1/min leads on average to a slightly better insufflation performance of the proposed device compared to the predicate device 45L Core Insufflator.
Biocompatibility testing was performed on the insufflation tube sets of the proposed device Insufflator 50L FM134 in accordance with:
- ISO 10993-1:2009 Biological evaluation of medical devices- Evaluation and testing within a risk management system;
- · ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity;
- · ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization; and
- · ISO 10993-11:2006 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo with a circle and a design inside, followed by the text "WOM". Below this, the text "Insufflator 50L FM134" is displayed, followed by "510(k) Premarket Notification". The logo and text appear to be related to a medical device or product, possibly an insufflator.
In addition, the insufflation tube set with heating wire of the proposed device FM134 was tested in accordance with ISO 10993 - 18:2005, Chemical characterization of materials.
ETO sterilization validation on the tube sets was performed in accordance with the below standards:
- . ISO 11135-1:2007 - Sterilization of health care products - Ethylene oxide -Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices;
- ISO 14937:2009 Sterilization of health care products General criteria for ● characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices; and
- ISO 10993-7:2008 - Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals.
- AAMI TIR 28:2009, Product adoption and process equivalency for ethylene ● oxide sterilization
Residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) data shows that the limit of EO < 4 mg and ECH < 5 mg after 10 days of aeration (gas release) that remains on the tube set will not be exceeded. The sterility assurance level (SAL) was < 10 . Package and product integrity of the tube sets were tested in accordance with ISO 11607-1 - Packaging for terminally sterilized medical devices and ASTM-F-1980:2002 - Standard for accelerated aging of sterile medical device packages.
## 8) Conclusion:
Based on the same intended use, the same basic technological characteristics and performance testing, the Insufflator 50L FM134 is substantially equivalent to the predicate devices 45L Core Insufflator F114 and to the Nebulae™ I 50 LPM Insufflator. The differences between the proposed device and the predicate devices
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the WOM logo, which is a stylized symbol above the letters "WOM". Below the logo is the text "Insufflator 50L FM134" followed by "510(k) Premarket Notification". The text appears to be advertising or providing information about a medical device.
do not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.