The Fertiloscopy Kit is intended to be used for transvaginal hydrolaparoscopy, dye-test, salpingoscopy, and hysteroscopy during one procedure. The device is indicated for use in the diagnosis of tuboperitoneal infertility.
Device Story
Fertiloscopy Kit enables transvaginal hydrolaparoscopy, dye-test, salpingoscopy, and hysteroscopy in single procedure; used for diagnosis of tuboperitoneal infertility. Kit components include intrauterine balloon catheter, transvaginal cannula, and trocar. Physician performs procedure; provides visual assessment of uterus, tubes, ovary, and peritoneum. Output allows clinician to evaluate fertility status; facilitates clinical decision-making regarding infertility treatment.
Clinical Evidence
Clinical study of 160 patients demonstrated device provides diagnostic sequence for complete status assessment of uterus, tubes, ovary, and peritoneum in one procedure.
Technological Characteristics
Kit includes intrauterine balloon catheter, transvaginal cannula, and trocar. Principles of operation involve endoscopic visualization and access for diagnostic procedures. Sterilization performed on materials.
Indications for Use
Indicated for diagnosis of tuboperitoneal infertility in patients undergoing transvaginal hydrolaparoscopy, dye-test, salpingoscopy, and hysteroscopy.
Regulatory Classification
Identification
A culdoscope is a device designed to permit direct viewing of the organs within the peritoneum by a telescopic system introduced into the pelvic cavity through the posterior vaginal fornix. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include trocar and cannula, instruments used through an operating channel, scope preheaters, light source and cables, and component parts.
Predicate Devices
Circon Corporation's Transvaginal Hydro Laparoscopy (THL) System
{0}------------------------------------------------
# MAR 1 3 2000
### 510(k) SUMMARY
### Soprane S.A. Fertiloscopy Kit® FH 1.29 / FTO 1.40
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Thierry Loubens Président Directeur Général Soprane S.A. 102 Rue Duguesclin 69006 Lyon France Telephone: (011) (33) 4 72 74 40 75 Facsimile: (011) (33) 4 75 74 04 91
Howard M. Holstein, Esq. Hogan & Hartson L.L.P. 555 Thirteenth Street, N.W. Washington, D.C. 20004-1109 Telephone: (202) 637-5813 Facsimile: (202) 637-5910
## Date Prepared:
August 6, 1999
# Name of Device and Name/Address of Sponsor
Soprane S.A, Fertiloscopy Kit® FH 1.29 / FTO 1.40
Thierry Loubens Président Directeur Général Soprane S.A. 102 Rue Duguesclin 69006 Lyon France Telephone: (011) (33) 4 72 74 40 75 Facsimile: (011) (33) 4 75 74 04 91
#### Common or Usual Name
Intrauterine Balloon Catheter and Transvaginal Cannula
#### Classification Name
Rigid culdoscope and accessories (21 C.F.R. § 884.1640); semi-rigid hysteroscope and accessories (21 C.F.R. § 884.1690); cannula. manipulator/injector, uterine; and reposable surgical trocar (21 C.F.R. \$ 884.1720)
## Predicate Devices
- Circon Corporation's ("Circon") Transvaginal Hydro® Laparoscopy . ("THLTM") System;
{1}------------------------------------------------
- Cook OB/GYN's ("Cook") Intrauterine Balloon Access Catheter ("IBA . Catheter"):
- Imagyn Medical, Inc.'s ("Imagyn") MicroSpan Hysteroscope Sheath . ("MicroSpan"); and
- Origin Medsystems, Inc.'s ("Origin") Blunt Tip Trocar. .
#### Intended Use
The Fertiloscopy Kit is intended to be used for transvaginal hydrolaparoscopy, dye-test, salpingoscopy, and hysteroscopy during one procedure. The device is indicated for use in the diagnosis of tuboperitoneal infertility.
#### Technological Characteristics and Substantial Equivalence
The Fertiloscopy Kit has the same general intended use and similar principles of operation and technological characteristics as a combination of Circon's cleared THL, Cook's cleared IBA Catheter, Imagyn's cleared MicroSpan, and Origin's cleared Blunt Tip Trocar. The minor technological differences between the Fertiloscopy Kit and the Imagyn MicroSpan, Cook IBA Catheter, Circon THL, and Origin Blunt Tip Trocar do not raise any new questions of safety or effectiveness. Therefore, the Fertiloscopy Kit is substantially equivalent to its predicate devices.
#### Performance Data
Soprane has conducted leak testing of the balloon during inflation and sterility testing of the materials following sterilization. In a clinical study of 160 patients, Fertiloscopy was shown to provide a "diagnostic sequence giving, in a one time procedure, a complete and informative status of the uterus, tubes, ovary, and peritoneum."
#### Conclusion
The Fertiloscopy Kit's substantial equivalence is based on a narrative comparison of this device and its predicate devices' intended use, principles of operation and technological characteristics and bench testing and clinical testing.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized image of an eagle with its head turned to the left, overlaid with a human profile. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle and profile.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 3 2000
Soprane S.A. c/o Mr. Howard M. Holstein Hogan & Hartson, L.L.P. 555 Thirteenth Street, N.W. Washington, D.C. 20004
Re: K992655 Fertiloscopy Kit ® FH 1.29/FTO 1.40 Dated: January 18, 2000 Received: January 19, 2000 Requiatory Class: II 21 CFR §884.1640/Procode: 85 HEW 21 CFR §884.1690/Procode: 85 HIH 21 CFR §884.1720/Procode: 85 HET
Dear Mr. Holstein:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 26, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmelic Ad (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalion assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{3}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Soprane S.A.'s Fertiloscopy Kit® FH 1.29 / FTO 1.40 Device Name:
Indications for Use:
The Fertiloscopy Kit is intended to be used for transvaginal hydro-laparoscopy, dye-test, salpingoscopy, and hysteroscopy during one procedure. The device is indicated for use in the diagnosis of tubo-peritoneal infertility.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 C.F.R. 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Tamil U. Seymore
(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological D 510(k) Number
\\\DC - 69874/1 - 0919534.01
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.