K964276 · Richard Wolf Medical Instruments Corp. · HEW · Nov 25, 1997 · Obstetrics/Gynecology
Device Facts
Record ID
K964276
Device Name
ENDOSCOPIC HYSTERECTOMY AND UTERUS MANIPULATOR
Applicant
Richard Wolf Medical Instruments Corp.
Product Code
HEW · Obstetrics/Gynecology
Decision Date
Nov 25, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The instrument set consisting of vaginal tube, obturator (mandrin) and instrumentation inserts in combination with the uterus manipulator and uterus probe is used to support LAVH (laparoscopically assisted vaginal hysterectomy).
Device Story
Endoscopic Hysterectomy and Uterus Manipulator set supports LAVH procedures. System includes vaginal tube, obturator (mandrin), instrumentation inserts, and uterus manipulator with probe. Vaginal tube provides visualization of cervical portio, serves as surgical guide, and maintains pneumoperitoneum. Obturator facilitates tube introduction. Instrumentation inserts provide gas-tight ports for surgical tools. Uterus manipulator features deflectible distal tip to fix and position uterus during surgery. Operated by surgeons in clinical/OR settings. Device enables uterine manipulation and access during laparoscopic hysterectomy, potentially facilitating minimally invasive surgical approach.
Clinical Evidence
No clinical tests performed. Evidence based on mechanical bench testing, including stress testing of the vaginal tube locking system (100,000 cycles) and mechanical resistance testing of the uterus manipulator hinge. Steam sterilization compatibility was verified using the fractional method.
Technological Characteristics
Autoclavable, gas-tight vaginal tube with instrumental ports, specifically shaped tubal tip, and deflectible uterus manipulator probe. Materials are medical-grade for steam sterilization. System is manual/mechanical; no software or electronic components.
Indications for Use
Indicated for patients undergoing laparoscopically assisted vaginal hysterectomy (LAVH) for conditions including uterus myomatosis and abnormal uterine bleeding. Contraindicated in patients where vaginal hysterectomy is possible, patients with pelvic masses >8x5 inches, cesarean hysterectomy, placenta accreta, uterine atony, unspecified uterine bleeding, uterine rupture, or stage III ovarian cancer requiring large abdominal incision. Medical status (anemia, diabetes, lung/cardiac disease, bleeding diathesis) must be evaluated; surgeon inexperience is a contraindication.
Regulatory Classification
Identification
A culdoscope is a device designed to permit direct viewing of the organs within the peritoneum by a telescopic system introduced into the pelvic cavity through the posterior vaginal fornix. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include trocar and cannula, instruments used through an operating channel, scope preheaters, light source and cables, and component parts.
Dr. Reich, "The Role of Laparoscopy in Hysterectomy", Advances in Obstetrics and Gynecology, vol.1 1994
Submission Summary (Full Text)
{0}
46-1276 1/5
Richard Wolf Medical Instruments Corporation
NOV 25 1997
353 Corporate Woods Parkway
Vernon Hills, Illinois 60061
Telephone: 847.913.1113
Fax: 847.913.1488
| 510(k) Summary of Safety and Effectiveness | | | | |
| --- | --- | --- | --- | --- |
| Submitter: | | | Date of Preparation: October 25, 1996 | |
| Company / Institution name: RICHARD WOLF MEDICAL INSTRUMENTS CORP. | | | FDA establishment registration number: 14 184 79 | |
| Division name (if applicable): N.A. | | | Phone number (include area code): (847) 913-1113 | |
| Street address: 353 Corporate Woods Parkway | | | FAX number (include area code): (847) 913-0924 | |
| City: Vernon Hills | State/Province: Illinois | Country: USA | | ZIP / Postal Code: 60061 |
| Contact name: Mr. Robert L. Casarsa | | | | |
| Contact title: Quality Assurance Manager | | | | |
| Product Information: | | | | |
| Trade name: Endoscopic Hysterectomy and Uterus manipulator | | Model number: 8370.501/.511/.512/.513, 8983.011/.111/.201 | | |
| Common name: | | Classification name: Endoscope and accessories | | |
| Information on devices to which substantial equivalence is claimed: | | | | |
| 510(k) Number | Trade or proprietary or model name | | Manufacturer | |
| 1 | 1 valtchev manipulator | | 1 European Patent Applicant 94103496.9 | |
| 2 | 2 RUMI | | 2 Cooper Surgical | |
| 3 pre-enactment | 3 8370.00 | | 3 Richard Wolf Medical Instruments | |
| 4 | 4 26201A | | 4 Storz | |
| 5 | 5 24945A | | 5 Storz | |
| 6 pre-enactment | 6 8991.02 | | 6 Richard Wolf Medical Instruments | |
| 7 pre-enactment | 7 4837 C | | 7 Richard Wolf Medical Instruments | |
| 8 | 8 24945AO | | 8 Storz | |
| 9 | 9 24945E | | 9 Storz | |
## 1.0 Description
The Endoscopic Hysterectomy instruments and Uterus manipulator are designed to remove the uterus through the vagina. The obturator can be introduced into the vagina through the tube. The vaginal tube is gas tight.
The changeable distal tip of the uterus manipulator is deflectible so that the uterus can be moved in optimal operation position. The jaw with the teeth fixes the cervix part on the manipulator and can be locked and released with the locking device (17). When the turned handle is moved clockwise, the distal tip goes in the deflected angle. When the luer sealing cap is removed the instrument can be cleaned by rinsing.
RICHARD WOLF
{1}
K64276 2/3
Richard Wolf Medical Instruments Corporation
355 Corporate Woods Parkway
Vernon Hills, Illinois 60061
Telephone: 847.913.1113
Fax: 847.913.1488
## 2.0 Intended Use
The instrument set consisting of vaginal tube, obturator (mandrin) and instrumentation inserts in combination with the uterus manipulator and uterus probe is used to support LAVH (laparoscopically assisted vaginal hysterectomy).
## 2.1 Contraindication
Contraindications described in the relevant literature must be observed.
Diagnosis and treatment of a pelvic mass that is too large to fit intact into an impermeable sack (≤8 x 5 in.).
Laparoscopic Hysterectomy is not indicated when vaginal hysterectomy is possible.
The medical status of the patient may prohibit surgery. Anemia, diabetes, lung disorders, cardiac disease, and bleeding diathesis must be evaluated before surgery. Age should rarely be a deterrent. Cesarean hysterectomy is an absolute contraindication. Placenta accreta, uterine atony, unspecified uterine bleeding, and uterine rupture are relative contraindications for peripartum hysterectomy at present. Another contraindication might be stage III ovarian cancer that requires a large abdominal incision.
Inexperience of the surgeon is a contraindication to the laparoscopic approach.
## 3.0 Technological Characteristics
- autoclavable
- speziell shaped tubal tip
- gas tightness
- gas tight instrumental ports
## 4.0 Substantial Equivalence
The basic design of the tube and obturator are equivalent to pre enactment devices from R.Wolf and equivalent to existing competitive devices. There are some modifications in dimension e.g. length, diameter; contour, function and working insert for its intended use in LAVH.
The manipulator is based on pre enactment devices from R.Wolf, but is changed in design and now used through the tube for its intended use in LAVH and is equivalent to competitive devices.
RICHARD WOLF
{2}
3/3
Richard Wolf Medical Instruments Corporation
353 Corporate Woods Parkway
Vernon Hills, Illinois 60061
Telephone: 847.913.1113
Fax: 847.913.1488
## 5.0 Performance Data
Mechanical stress test of the vaginal tube shows, that the bayonet locking system is tight after 100,000 cycles of opening and closing.
Mechanical test of the uterus manipulator shows that the mechanical resistance of the hinch is strong enough for the manipulating of the uterus.
The steam sterilization in the clinical use and the tests performed by Richard Wolf shows, that the steam sterilization has no influence to the quality of the endoscopic hysterectomy instruments and uterus manipulator, when using the fractional method.
## 6.0 Clinical Tests
No clinical tests performed.
(see literature Dr. Reich, "The Role of Laparoscopy in Hysterectomy", Advances in Obstetrics and Gynecology, vol.1 1994, Mosby-Year Book, Inc.).
## 7.0 Conclusions Drawn
These devices are designed and tested to guarantee the safety and effectiveness, when used according to the instruction manual.
By: Robert L. Casarsa
Quality Assurance Manager
Date: Oct 22, 96
RICHARD WOLF
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
NOV 25 1997
Mr. Robert L. Casarsa
Quality Assurance Manager
Richard Wolf Medical Instruments Corporation
353 Corporate Woods Parkway
Vernon Hills, Illinois 60061
Re: K964276
Endoscopic Hysterectomy and Uterus Manipulator
(see attached list for model names and numbers)
Dated: August 25, 1997
Received: August 29, 1997
Regulatory class: II
21 CFR §884.4160/Product code: 85 HEW
21 CFR §884.4530/Product code: 85 HDP
Dear Mr. Casarsa:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}
NOV-24-1997 09:40 RICHARD WULF & JEFF. 1 047 012 2924 P.04
510(k) Number (if known): ________________________________
P172
Device Name: **Endoscopic Hysterectomy and Uterus Manipulator**
Indications for Use:
## Indications for Use:
For laparoscopic assisted vaginal hysterectomy.
The Endoscopic Hysterectomy instrument set consisting of vaginal tube, mandrin and instrumentation insert is used to support LAVH (laparoscopically assisted vaginal hysterectomy).
- The tube is used for visualization of the cervical portio.
The distal section of the vaginal tube can be seen during laparoscopically assisted vaginal hysterectomy, and thus serves the surgeon as a guide and support during cutting. The tube allows pneumoperitoneum to be maintained during cutting and opening of the posterior vaginal vault.
- The mandrin supports the introduction of the tube into the vagina. Prior to introduction it is inserted into the vaginal tube and locked in place.
- The instrumentation insert has two openings - instrument ports - through which surgical instruments are inserted into the vaginal tube without loss of gas.
Indication: For the use of the uterus manipulator in endoscopic hysterectomy:
- Uterus myomatosis
- Abnormal uterine bleeding
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
D. J. Anthony
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number **K964276**
Prescription Use ☑
Per CFR 21 CFR 801.109
OR 1 - 3
Over-The-Counter Use
{5}
NOV-24-1997 09:40
P272
**Indications for Use:**
The Uterus Manipulator set, consisting of uterus manipulator and uterus probe is used to support LAVH.
The uterus manipulator is an instrument with a distally inclinable probe section. The part of the instrument which is inserted in the uterus serves to fix the uterus and change or manipulate its position.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robert D. Sutliff
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964276
Prescription Use ☑
Per CFR 21 CFR 801.109
OR
1 - 4
Over-The-Counter Use
TOTAL P.05
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.