KOH COLPOTOMIZER SYSTEM

K954311 · CooperSurgical, Inc. · HEW · Nov 19, 1996 · Obstetrics/Gynecology

Device Facts

Record IDK954311
Device NameKOH COLPOTOMIZER SYSTEM
ApplicantCooperSurgical, Inc.
Product CodeHEW · Obstetrics/Gynecology
Decision DateNov 19, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1640
Device ClassClass 2

Indications for Use

The Koh Colpotomizer™ System, comprised of the Koh Cup™ Vaginal Fornices Delineator and Colpo-Pneumo Occluder, are accessories to the RUMI uterine manipulator, K932115. These accessories are designed to attach to the RUMI uterine manipulator device and are intended to be used by the surgeon as an anatomical marker to locate the position of the vaginal fornices as viewed through uterine tissue from within the abdominal cavity (intracorporeal), and also to maintain pneumoperitoneum during laparoscopic surgical procedures in which a colpotomy or culdotomy incision is made. The Koh Colpotomizer™ System is used in conjunction with the RUMI uterine manipulator device.

Device Story

System consists of Koh Cup (vaginal fornices delineator) and Colpo-Pneumo Occluder; accessories attach to RUMI uterine manipulator. Koh Cup: plastic cup-shaped device; fits over RUMI tip; stretches vaginal fornices to provide visual anatomical marker for surgeon during intracorporeal viewing. Colpo-Pneumo Occluder: silicone sleeve with inflatable balloon; inflated with saline via external catheter; creates pressure barrier in vagina to prevent pneumoperitoneum leakage during colpotomy/culdotomy. Used in OR by surgeons; provides visual guidance and maintains surgical field pressure; benefits include improved anatomical identification and prevention of gas loss during laparoscopic procedures.

Clinical Evidence

Bench testing only. Static burst testing performed on the Colpo-Pneumo Occluder; balloons inflated with 250 cc of water (2.5x typical surgical volume) maintained integrity for 12 hours without bursting, meeting the pass criteria of 30 minutes without loss.

Technological Characteristics

Koh Cup: plastic cup-shaped delineator (3.0, 3.5, 4.5 cm diameters). Colpo-Pneumo Occluder: silicone sleeve with inflatable balloon; saline-filled. Attaches to RUMI uterine manipulator. Non-powered, manual operation.

Indications for Use

Indicated for use during laparoscopic surgical procedures requiring colpotomy or culdotomy, specifically to delineate vaginal fornices as an anatomical marker and to maintain pneumoperitoneum.

Regulatory Classification

Identification

A culdoscope is a device designed to permit direct viewing of the organs within the peritoneum by a telescopic system introduced into the pelvic cavity through the posterior vaginal fornix. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include trocar and cannula, instruments used through an operating channel, scope preheaters, light source and cables, and component parts.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} NOV 19 1996 K954311 P 1073 # 510(k) Summary ## CooperSurgical ### Koh Colpotomizer System 1. **Sponsor/Applicant** CooperSurgical, Inc. 15 Forest Parkway Shelton, CT 06484 Telephone # (800) 645-3760 Telefax # (203) 925-0135 Contact Person: John Chapman **Date of Summary Preparation** September 14, 1995 2. **Device Name** Proprietary Name: The Koh Colpotomizer™ System Common/Usual Name: Uterine Manipulator/Injector and Accessories Classification Name: Uterine Elevator 3. **Identification of Predicate or Legally Marketed Device(s) to Which Equivalence is Being Claimed** Lubke Uterine Vacuum Cannula, SEMM Uterine Vacuum Cannula and Milex Inflato Ball 4. **Device Description** The Koh Colpotomizer™ System consists of two accessory devices: The Koh Cup™ Vaginal Fornices Delineator, and a Colpo-Pneumo Occluder, each of which are specifically designed to attach to and be used with the RUMI uterine manipulator device. The CooperSurgical RUMI device as described in K932115 is intended to be used as a uterine manipulator during pelvic laparoscopic procedures. CooperSurgical, Inc. 510(k) 9/14/95 Page F-1 {1} K95-4311 p213 The Koh Cup™ Vaginal Fornices Delineator is a plastic cup-shaped device which fits over the distal end of the RUMI disposable tip and firmly attaches to the tip base. When attached to the RUMI, the Koh Cup™ becomes an extension of the uterine manipulator tip and is thus able to articulate up or down as the uterine manipulator is operated in the normal fashion. With the RUMI and Koh Cup™ properly positioned inside the patient, the outer rim of the Koh Cup™ will surround the cervix and be located in the vaginal fornices. In this position the rim of the Koh Cup™ will extend or "stretch" the vaginal fornices, thus allowing the surgeon an anatomical marker to conveniently locate the position of the fornix when viewed from an intracorporeal position. The kit will include Koh Cups™ in three outer rim diameter sizes: 3.0 cm, 3.5 cm, and 4.5 cm. The surgeon or nurse will make the selection of the appropriate size Koh Cup™ by visual inspection of the cervix at the time of instrument insertion in the patient. The Pneumo Occluder consists of a condom-like silicone sleeve whose proximal end is folded back and attached to the main shaft of the sleeve in order to form an inflatable cavity (the "occluding balloon"). Attached to the occluding balloon is an inflation catheter that is available to the surgeon outside the patient's body for inflating the occluding balloon during surgery. The occluding balloon should be inflated until it fully contacts the perimeter wall of the vagina. This creates a pressure barrier which prevents the pneumoperitoneum from leaking out of the patient through the vagina once the colpotomy or culdotomy incision is made. This occluding property allows the surgeon to maintain proper pneumoperitoneum pressure and continue the laparoscopic procedure. In addition to the Koh Cup™ and Colpo-Pneumo Occluder, a 60 cc syringe is included in the proposed kit to be used for filling the occluding balloon with saline. Typically the occluding balloon will require inflation of approximately 100 cc of saline in order to adequately occlude the vagina. 5. Intended use The Koh Colpotomizer™ System, comprised of the Koh Cup™ Vaginal Fornices Delineator and Colpo-Pneumo Occluder, are accessories to the RUMI uterine manipulator, K932115. These accessories are designed to attach to the RUMI uterine manipulator device and are intended to be used by the surgeon as an anatomical marker to locate the position of the vaginal fornices as viewed through uterine tissue from within the abdominal cavity CooperSurgical, Inc. 510(k) 9/14/95 Page F-2 {2} K954311 p 3 of 3 (intracorporeal), and also to maintain pneumoperitoneum during laparoscopic surgical procedures in which a colpotomy or culdotomy incision is made. The Koh Colpotomizer™ System is used in conjunction with the RUMI uterine manipulator device. ## 6. Technological Characteristics The Koh Cup™ Vaginal Fornices Delineator device (of the Koh Colpotomizer™ System) and the LUBKE and SEMM uterine cannula predicate device above are identical in technological characteristics in that they all consist of a cup-shaped perimeter that are used with a uterine manipulator. Both the Koh Cup™ and predicate devices consist of a cup-shaped perimeter which fit outside the cervix in the vaginal fornices while the uterine manipulator “finger” goes up through the cervix and inside the uterus. The Colpo-Pneumo Occluder device is substantially equivalent to the Milex Inflato Ball and to several common techniques utilized during laparoscopic surgery to prevent loss of pneumoperitoneum. The common technique of inflating a surgical glove with air or to stuff it with surgical gauze, and pushing the glove into the vaginal cavity to occlude the passageway is similar in technological characteristics to the Colpo-Pneumo Occluder. The latex glove is inflated as the Pneumo-Occluder and used as the occluding device. The Pneumo-Occluder is a fixed device that is firmly held in place since it is attached to the uterine manipulator and acts as a vaginal occluding device during surgery. The technological characteristics of the Pneumo-Occluder are also similar to the MILEX Inflato Ball an inflatable pessary device since it is inflated and inserted into the vagina in order to prevent leakage of the pneumoperitoneum during laparoscopic surgery. ## 7. Performance Testing The Colpo-Pneumo Occluder was subjected to and passed a static burst test which consisted of inflating the balloons with 250 cc of water and left for 12 hours. None of the balloons burst at 10 minutes, 30 minutes, 60 minutes or 12 hour time frames. The criteria for passing the test was maintaining the 250 cc of water for 30 minutes without loss. This is a factor of 2.5 times the inflation generally needed for use in surgery and has been found in practice to be an adequate design margin. CooperSurgical, Inc. 510(k) 9/14/95 Page F-3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...