The CooperSurgical Advincula Delineator™ Uterine Manipulator is indicated to provide delineation of the vaginal fornices and maintain pneumoperitoneum as a uterine manipulator during Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH) and/or Laparoscopic Supra-Cervical Hysterectomy (LSH).
Device Story
Single-use sterile uterine manipulator; used in operating room by surgeons during laparoscopic hysterectomy procedures. Device features cup-shaped sliding colpo-pneumo occluder for vaginal fornix delineation and balloon for vaginal vault occlusion to maintain pneumoperitoneum. Surgeon inserts device to provide anatomical marker for uterine resection; balloon inflation maintains surgical field visibility. Benefits include improved visualization and maintenance of pneumoperitoneum during resection. No clinical data required; substantial equivalence established via bench testing.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing comparing subject device to predicate. Bench tests included distal/proximal balloon burst volume, leakage, fill tube tensile strength, and mechanical testing (bending, compression, tension forces) of the tube and locking mechanism of the colpo-pneumo occluder.
Technological Characteristics
Materials: stainless steel, silicone, plastics, adhesives, inks. Single-use, sterile (ETO, SAL 10^-6). Features sliding colpo-pneumo occluder and inflatable balloon (60-120cc capacity). External communicating device, tissue contact ≤ 24 hours.
Indications for Use
Indicated for patients undergoing Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH), or Laparoscopic Supra-Cervical Hysterectomy (LSH) requiring uterine manipulation and vaginal fornix delineation.
Regulatory Classification
Identification
A culdoscope is a device designed to permit direct viewing of the organs within the peritoneum by a telescopic system introduced into the pelvic cavity through the posterior vaginal fornix. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include trocar and cannula, instruments used through an operating channel, scope preheaters, light source and cables, and component parts.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 21, 2015
Cooper Surgical Inc. % Tim Lohnes Senior Regulatory Consultant Orchid Design 80 Shelton Technology Center Shelton. CT 06484
Re: K143650
> Trade/Device Name: CooperSurgical Advincula Delineator™ Uterine Manipulator Regulation Number: 21 CFR 884.1640 Regulation Name: Culdoscope and accessories Regulatory Class: Class II Product Code: HEW Dated: June 16, 2015 Received: June 18, 2015
Dear Mr. Lohnes,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Benjamin R. Fisher -S
Benjamin Fisher, PhD. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143650
#### Device Name
CooperSurgical Advincula Delineator™ Uterine Manipulator
Indications for Use (Describe)
The CooperSurgical Advincula Delineator™ Uterine Manipulator is indicated to provide delineation of the vaginal fornices and maintain pneumoperitoneum as a uterine manipulator during Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH) and/or Laparoscopic Supra-Cervical Hysterectomy (LSH).
Type of Use (Select one or both, as applicable)
| <span></span> | <span></span> |
|----------------------------------------------|---------------|
| <div></div> | <div></div> |
| Prescription Use (Part 21 CFR 801 Subpart D) | <div></div> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div></div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Goper Surgical
### 07. 510(K) SUMMARY:
| 510(k) Summary of Safety and Effectiveness, (21 CFR 807.92): | |
|--------------------------------------------------------------|-------------------------------------------------------------|
| Submitter: | Cooper Surgical Inc., |
| | 95 Corporate Drive, |
| | Trumbull, CT 06611 |
| Contact Person: | Tim M. Lohnes |
| | Senior Regulatory Consultant, |
| | Orchid Design |
| | Phone: (203) 922 0105 |
| | Fax: (203) 922 0130 |
| | tim.lohnes@orchid-ortho.com |
| Date Prepared: | December 15, 2014 |
| Name of Device:: | CooperSurgical Advincula Delineator™<br>Uterine Manipulator |
| Common/Usual Name: | Cannula, manipulator/injector, uterine |
| Classification Name: | 884.1640, Culdoscope and accessories |
| Device Class: | Class II |
| Product Code: | HEW |
| Predicate Device(s): | Cooper Surgical Koh Colpotomizer<br>System, K954311 |
| Device Description; | |
The CooperSurgical Advincula Delineator™ Uterine Manipulator is a single use device, supplied sterile to the end user. It includes a cup shaped sliding colpo-pneumo occluder used to delineate the vaginal fornices, providing an anatomical marker to facilitate resection of the uterus, and a balloon to occlude the vaginal vault to maintain pneumoperitoneum during uterine resection.
Intended Use:
The CooperSurgical Advincula Delineator ™Uterine Manipulator is indicated to provide delineation of the vaginal fornices and maintain pneumoperitoneum as a uterine manipulator during Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH) and/or Laparoscopic Supra-Cervical Hysterectomy (LSH).
{4}------------------------------------------------
## Comparison of Technological Characteristics to the Predicate Device:
| Attribute | Cooper Surgical<br>Advincula<br>Delineator™ Uterine<br>Manipulator<br>(subject) | Cooper Surgical<br>Koh Colpotomizer<br>System, K954311 | Determination |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Body Contact Type | External<br>communicating, tissue, ≤<br>24 hours | External<br>communicating, tissue, ≤<br>24 hours | Substantially<br>Equivalent |
| Material(s) | Stainless steel, silicone,<br>various plastics,<br>adhesives and inks. | Stainless steel,<br>silicone, various<br>plastics, adhesives and<br>inks. | Substantially<br>Equivalent |
| Environment | Operating room | Operating room | Substantially<br>Equivalent |
| Vaginal Occlusion<br>(Pneumoperitoneum) | Occluder balloon with<br>inflation capacity of 60-<br>120cc | Occluder balloon with<br>inflation capacity of 90-<br>120cc | Different |
| Indications for Use | The CooperSurgical<br>Advincula Delineator™<br>Uterine Manipulator is<br>indicated to provide<br>delineation of the<br>vaginal fornices and<br>maintain<br>pneumoperitoneum as a<br>uterine manipulator<br>during Total<br>Laparoscopic<br>Hysterectomy (TLH),<br>Laparoscopic Assisted<br>Vaginal Hysterectomy<br>(LAVH) and/or<br>Laparoscopic Supra<br>Cervical Hysterectomy<br>(LAH). | The Koh Colpotomizer<br>System is indicated for<br>use in all laparoscopic<br>procedures where the use<br>of a uterine manipulator<br>is appropriate and the<br>surgeon intends to<br>remove or access<br>intraperitoneal tissue<br>through the vagina by<br>use of a colpotomy or<br>culdotomy incision. | Substantially<br>Equivalent |
| Biocompatibility | The patient contacting<br>components of the<br>Advincula Delineator™<br>Uterine Manipulator<br>have been evaluated for<br>biocompatibility as<br>appropriate for their<br>contact classification<br>(ext. communicating,<br>tissue/bone/dentin, ≤ 24<br>hours | The patient contacting<br>components of the<br>Koh Colpotomizer<br>System have been<br>evaluated for<br>biocompatibility as<br>appropriate for their<br>contact classification<br>(ext. communicating,<br>tissue/bone/dentin, ≤<br>24 hours | Substantially<br>Equivalent |
{5}------------------------------------------------
Cooper Surgical Inc. Advincula Delineator
| Packaging | The Advincula<br>Delineator™ Uterine<br>Manipulator<br>is packaged in a 16 Mil<br>Flexible Blister pouch<br>with a heat-sealed<br>1073B Tyvek lid,<br>packed 3 to a carton. | The Koh-Efficient<br>device is packaged in a<br>.035" PETG Blister;<br>heat-sealed 1073<br>Tyvek lid, packaged 6<br>to a chipboard carton. | Substantially<br>Equivalent |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Stability/Shelf Life | The Advincula<br>Delineator™ Uterine<br>Manipulator is<br>sterilized by ETO to a<br>Assurance Level (SAL)<br>of 10-6, with an initial<br>shelf life of 1 year. | The sterile components<br>of the Koh-Effecent<br>System device are<br>sterilized by ETO gas<br>to a Assurance Level<br>(SAL) of 10-6, with a<br>shelf life of 2 years. | Substantially<br>Equivalent |
The basis of substantial equivalence of the CooperSurgical Advincula Delineator™ Uterine Manipulator is the similarities in materials, design, function, performance, sterilization, and indications for use in comparison to the predicate device. The primary technological difference between the subject and predicate devices is the inflation capacity of the vaginal occluder balloon. However, this minor difference does not raise different questions of safety or effectiveness as the maximum inflation pressure is the same for both the subject and predicate devices. In addition, burst volume testing was conducted to demonstrate that the difference in inflation pressure does not affect the safety and effectiveness of the subject device.
### Performance Data:
The Cooper Surgical Advincula Delineator™ Uterine Manipulator was tested in comparison to the predicate Cooper Surgical Koh Colpotomizer System K954311, with regards to;
- 1) Distal & proximal balloon burst volume
- 2) Distal & proximal balloon leakage
- 3) Distal & proximal balloon fill tube tensile strength
The Cooper Surgical Advincula Delineator™ Uterine Manipulator was also tested in regards to bending, compression, and tension forces applied to the tube, as well as the security of the sliding colpo-pneumo occluder when locked.
Clinical testing was not required to support the conclusion of substantial equivalence.
### Conclusion:
The substantial equivalence of the CooperSurgical Advincula Delineator™ Uterine Manipulator has been established by demonstrating the similarities in design, materials, function, performance, and Intended Use are equivalent to the previously cleared Cooper Surgical Koh Colpotomizer System, K954311.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.