The KIDScore D3 tool provides decision support for prediction of likelihood of embryos developing to the blastocyst stage by scoring them according to their statistical viability. Adjunctive information provided by KIDScore D3 aids in the selection of embryo(s) for either transfer on Day 3, freezing or continued embryo development when, following morphological assessment on Day 3, there are multiple embryos deemed suitable for transfer or freezing. The KIDScore D3 tool is only to be used with the EmbryoScope and EmbryoScope+ time-lapse incubator systems.
Device Story
KIDScore D3 is a software-based decision support tool for embryologists; functions as an accessory to EmbryoViewer software within EmbryoScope/EmbryoScope+ time-lapse incubator systems. Input consists of manually annotated morphokinetic parameters (number of pronuclei, timing of cell divisions t2, t3, t4, t5, t8, and pronuclei fading tPNf). Algorithm applies a sequential elimination model to these inputs; assigns embryos a score from 1 to 5 based on developmental pace and regularity. Output is a viability score displayed in the 'Compare & Select' function of the software. Clinicians use this score as adjunctive information alongside traditional morphological assessment to prioritize embryos for transfer or cryopreservation. Benefit includes improved selection of viable embryos for IVF patients.
Clinical Evidence
Prospective, multi-center, non-interventional study (6 US sites). Analyzed 4,152 embryos from 1,338 treatments (2009-2014). Primary endpoint: association between adjunct prediction and blastocyst outcome. KIDScore D3 demonstrated an Odds Ratio of 4.13 (95% CI: 3.48-4.9, p<0.0001) for blastocyst development, compared to 2.57 (95% CI: 1.88-3.51, p<0.0001) for the predicate Eeva system. Results indicate the device is informative for blastocyst outcome when used as an adjunct to morphology.
Technological Characteristics
Software-based decision support tool; accessory to EmbryoViewer software. Operates on ES Server computer. Inputs: manual annotations of morphokinetic events. Algorithm: rule-based sequential elimination model. Connectivity: networked with EmbryoScope/EmbryoScope+ incubators. No specific materials or energy sources as it is a software-only accessory.
Indications for Use
Indicated for use in assisted reproduction procedures to provide decision support for predicting the likelihood of embryos developing to the blastocyst stage. Used by embryologists to aid in the selection of embryos for transfer, freezing, or continued development on Day 3 when multiple embryos are available.
Regulatory Classification
Identification
An Assisted Reproduction Embryo Image Assessment System is a prescription device that is designed to obtain and analyze light microscopy images of developing embryos. This device provides information to aid in the selection of embryo(s) for transfer when there are multiple embryos deemed suitable for transfer or freezing.
Special Controls
In combination with the general controls of the Food, Drug &Cosmetic Act, Assisted Reproduction Embryo Image Assessment Systems are subject to the following special controls:
*Classification.* Class II (special controls). The special control(s) for this device are:(1) Clinical performance testing must demonstrate a reasonable assurance of safety and effectiveness of the device to predict embryo development. Classification performance (sensitivity and specificity) and predictive accuracy (Positive Predictive Value and Negative Predictive Value) must be assessed at the subject and embryo levels.
(2) Software validation, verification, and hazard analysis must be provided.
(3) Non-clinical performance testing data must demonstrate the performance characteristics of the device. Testing must include the following:
(i) Total light exposure and output testing;
(ii) A safety analysis must be performed based on maximum (worst-case) light exposure to embryos, which also includes the safety of the light wavelength(s) emitted by the device;
(iii) Simulated-use testing;
(iv) Mouse Embryo Assay testing to assess whether device operation impacts growth and development of mouse embryos to the blastocyst stage;
(v) Cleaning and disinfection validation of reusable components;
(vi) Package integrity and transit testing;
(vii) Hardware fail-safe validation;
(viii) Electrical equipment safety and electromagnetic compatibility testing; and
(ix) Prediction algorithm reproducibility.
(4) Labeling must include the following:
(i) A detailed summary of clinical performance testing, including any adverse events;
(ii) Specific instructions, warnings, precautions, and training needed for safe use of the device
(iii) Appropriate electromagnetic compatibility information;
(iv) Validated methods and instructions for cleaning and disinfection of reusable components; and
(v) Information identifying compatible cultureware and explain how they are used with the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 19, 2019
Vitrolife A/S Belinda Dueholm Regulatory Affairs Specialist Jens Juuls Vej 20 Viby J, 8260 DENMARK
Re: K182798
Trade/Device Name: KIDScore D3 Regulation Number: 21 CFR 884.6195 Regulation Name: Assisted reproduction embryo image assessment system Regulatory Class: Class II Product Code: PBH Dated: June 17, 2019 Received: June 20, 2019
Dear Belinda Dueholm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Sharon M. Andrews Assistant Division Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182798
Device Name KIDScore D3
#### Indications for Use (Describe)
The KIDScore D3 tool provides decision support for prediction of embryos developing to the blastocyst stage by scoring them according to their statistical viability.
Adjunctive information provided by KIDScore D3 aids in the selection of embryo(s) for either transfer on Day 3, freezing or continued embryo development when, following morphological assessment on Day 3, there are multiple embryos deemed suitable for transfer or freezing.
The KIDScore D3 tool is only to be used with the EmbryoScope timelapse incubator systems.
Type of Use (*Select one or both, as applicable*):
| <span> <b></b>Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|----------------------------------------------------------------------|
| <span> <b></b>Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - KIDScore D3 K182798
# 1. Submitter Information
| Submitter | Vitrolife A/S<br>Jens Juuls Vej 20<br>8260 Viby J<br>Denmark |
|-------------------------------|-----------------------------------------------------------------------|
| Contact Person:<br>Telephone: | Ms. Belinda Dueholm<br>+45 7221 7900 (main)<br>+45 2076 3707 (direct) |
- 2. Date Prepared: July 19th, 2019
# 3. Device Information
| Proprietary Name: | KIDScore D3 |
|--------------------|------------------------------------------------------------|
| Common Name: | Assisted Reproduction Embryo Image Assessment System |
| Regulation Number: | 21 CFR 884.6195 |
| Regulation Name: | Assisted Reproduction Embryo Image Assessment System |
| Product Code: | PBH (Embryo Image Assessment System, Assisted Reproduction |
| Regulatory Class | II |
# 4. Predicate Device
Eeva(TM) System, K142147, Auxogyn Inc.
The predicate device has not been subject to any design related recalls.
# 5. Device Description
The KIDScore D3 decision support tool is an adjunctive algorithm that is designed to support embryologists in their decision about which embryos are suitable for transfer. The tool is an optional accessory to the EmbryoViewer software. It is used in the "Compare & Select" function. The "D3" in the name refers to the use of the algorithm on Day 3 for aiding the embryologist in preparing for transfer of the embryo to the female patient.
KIDScore D3 utilizes the following manually annotated parameters to aid in identifying embryos that are suitable for transfer:
- Pronuclei (number of pronuclei): ●
- tPNf (time from insemination until pronuclei is fading) ●
- t2 (time from insemination to complete division to two cells) ●
{4}------------------------------------------------
- t3 (time from insemination to complete division to three cells) ●
- t4 (time for insemination to complete division to four cells) .
- t5 (time from insemination to complete division to five cells) ●
- t8 (time from insemination to complete division to eight cells) ●
The KIDScore D3 assigns scores by comparing the parameters above in embryos to the model criteria, one criterion at a time until the process stops either because the embryo did not pass one of the criteria in the sequence or because the last criterion in the model was reached. From the information available at day three of incubation, the KIDScore D3 divides embryos into five score groups (1-5, as described below):
- 0 = The embryo is not 2PN
- 1 = Initial development was too fast or the embryo displayed a direct cleavage from one to three cells
- 2 = The embryo was slow to develop
- 3 = Embryo development was irregular and the development pace increased from day two to day three
- 4 = Embryo development was irregular and the development pace slowed from day two to day three
and/or
The number of cells annotated at 66 hours was not as expected
5 = The embryo passed all of the avoidance criteria included in the model.
One or more computers running the EmbryoViewer software may be connected to the ES Server. KIDScore D3 is stored on the computer running the ES Server software. Calculations related to the model in KIDScore D3 are performed on the computer running the ES Server software.
#### 6. Indications for Use
The KIDScore D3 tool provides decision support for prediction of likelihood of embryos developing to the blastocyst stage by scoring them according to their statistical viability.
Adjunctive information provided by KIDScore D3 aids in the selection of embryo(s) for either transfer on Day 3, freezing or continued embryo development when, following morphological assessment on Day 3, there are multiple embryos deemed suitable for transfer or freezing.
The KIDScore D3 tool is only to be used with the EmbryoScope and EmbryoScope+ timelapse incubator systems.
{5}------------------------------------------------
# 7. Comparison of Intended Use and Technological Characteristics of the Subject and Predicate Device
## Intended Use
| Devices | Subject device (K182798) – KIDScore D3 | Predicate device (K142147) – Eeva System |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The KIDScore D3 tool provides decision support for prediction of likelihood of embryos developing to the blastocyst stage by scoring them according to their statistical viability.<br><br>Adjunctive information provided by KIDScore D3 aids in the selection of embryo(s) for either transfer on Day 3, freezing or continued embryo development when, following morphological assessment on Day 3, there are multiple embryos deemed suitable for transfer or freezing.<br><br>The KIDScore D3 tool is only to be used with the EmbryoScope and EmbryoScope+ time-lapse incubator systems. | The Eeva System is indicated to provide adjunctive information on events occurring during the first two days of development that may predict further development to the blastocyst stage on Day 5 of development. This adjunctive information aids in the selection of embryo(s) for transfer on Day 3 when, following morphological assessment on Day 3, there are multiple embryos deemed suitable for transfer or freezing. The device may also be used to collect additional time-lapse images until Day 5 of development for embryos not selected for transfer, to allow monitoring of continued embryo development. |
The intended use of the subject device is the same as the predicate device – both are intended to provide adjunctive information (decision support) to the embryologist/clinician at Day 3 of culture to aid the user in selecting embryos for transfer/freezing during assisted reproduction procedures.
#### Technological Characteristics
| | Subject device<br>(KIDScoreD3,<br>K182798) | Predicate device<br>(Eeva System,<br>K142147) | Comments |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Design | Integrates the<br>KIDScore D3 decision<br>support tool on the<br>Compare & Select<br>page of the<br>Embryoviewer software<br>present on the<br>Embryoscope and<br>Embryoscope +<br>devices. | Time-lapse system<br>used in conventional<br>incubator with analysis<br>software that<br>automatically identifies<br>embryo development<br>events for use in a<br>blastocyst prediction<br>model | Different: The subject<br>device is a software<br>accessory to be used<br>with the Embryoscope<br>imaging incubator to<br>provide adjunctive<br>information regarding<br>blastocyst quality. The<br>predicate device is a<br>time lapse imaging<br>system to be used in a<br>conventional incubator<br>that utilizes automated<br>image analysis to<br>provide adjunctive<br>blastocyst quality<br>information. These |
{6}------------------------------------------------
| | | | differences do not raise<br>different questions of<br>safety and<br>effectiveness. |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Algorithm | Software<br>User annotations of<br>embryo development<br>parameters used as<br>algorithm inputs<br>Score based on criteria<br>which are either related<br>to embryo morphology<br>or development stage | Software<br>Cell tracking and event<br>inference used as<br>algorithm inputs | Different. The subject<br>device utilizes an<br>elimination model with<br>user annotation as the<br>input, whereas the<br>predicate device<br>utilizes a predictive<br>algorithm that uses<br>automated cell tracking<br>and image analysis<br>inputs. These<br>differences do not raise<br>different questions of<br>safety and<br>effectiveness. |
| Annotation method | Manual input of<br>embryo development<br>events and timing | Image analysis<br>software automatically<br>identifies embryo<br>development events<br>and timing | Different. The subject<br>device utilizes input<br>from manual<br>annotations, whereas<br>the predicate device<br>utilizes an automated<br>image analysis system.<br>These differences do<br>not raise different<br>questions of safety and<br>effectiveness. |
The technological characteristics of the subject and predicate device are different - the subject device has different predictive algorithms for blastocyst quality, and different software and annotation functions. However, different types of safety and effectiveness questions are not raised by these differences in technological characteristics.
# 8. Summary of Non-Clinical Performance Testing
Devices classified under 21 CFR 884.6195 (Assisted Reproduction Embryo Image Assessment System) and product code PBH must address several non-clinical special controls, including software validation, verification, and hazard analysis, an assessment of light exposure and output, simulated use, cleaning and disinfection, package integrity and transit testing, electrical safety and electromagnetic compatibility testing, and prediction algorithm reproducibility.
As the subject device is an optional accessory to the EmbryoViewer software used in the EmbryoScope+ incubator (cleared under K173264), some of the special controls do not apply to this submission. The previously cleared version of the Embryoscope+ has met the special controls associated with the majority of the non-clinical performance testing listed above.
In the current submission, the sponsor has addressed the special controls associated with software and algorithm reproducibility. The software verification testing provided in this
{7}------------------------------------------------
submission met the requirements of the FDA Guidance Document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005. The reproducibility of the prediction algorithm is addressed in the summary of clinical testing described below.
# 9. Summary of Clinical Testing
Devices classified under 21 CFR 884.6195 (Assisted Reproduction Embryo Image Assessment System) and product code PBH must address a clinical special control to demonstrate reasonable assurance of safety and effectiveness of the device to predict embryo development. The clinical study is described below.
The safety and effectiveness of the KIDScore D3 was investigated in a prospective, single arm, multi-center clinical study conducted at six sites in the United States. This was a non-interventional clinical study where the KIDScore D3 was not used during patient treatment. The purpose of the study was to collect data to evaluate the safety and effectiveness of the KIDScore D3's ability to predict which embryos are most likely to develop to blastocyst stage. This was evaluated by using the KIDScore D3 as adjunct information to traditional morphological grading. Imaging data was collected on embryos cultured to day 5. Embryologists were masked to imaging data and evaluation was only based on morphology and KIDScore D3 scores.
# Objective
The study was aimed at analyzing the utilization of established morphology methods with adjunct outcome of an algorithm (KIDScore D3) that provides a score (1 - 5) from timings of morphokinetic events.
# Study Design
A double-blinded, multi-center study, designed to evaluate the odds ratios and other measures for outcomes of methodologies used for embryo assessment: day 3 morphology alone and day 3 morphology with KIDScore D3 results as adjunct information.
# Selection of Study Data
The data included embryos from 81 patients who underwent in vitro fertilization (IVF) treatment using their own eggs or donor eggs. These data are a subset taken from a total collection of 4152 embryos from 1338 treatments where all sibling embryos have been annotated for the morphokinetic events required by KIDScore D3. These data originate from treatments that were carried out in European IVF clinics, in the period 2009 - 2014.
# Primary endpoint
The primary endpoint of the study was to assess the association between the adjunct prediction of blastocyst outcome and the actual blastocyst outcome. The purpose was to determine if KIDScore D3 was informative for embryos graded as an A, B, or C using Day 3 morphology category assignment as compared to the predicate device (Eeva system). For those Good/Fair embryos, the blastocyst Odds Ratio (OR) for the adjunct prediction is required to be statistically significantly greater than 1.
{8}------------------------------------------------
# Secondary endpoints
The following calculations were assessed as secondary endpoints:
- Embryo level diagnostic performance measures .
- o specificity
- o sensitivity
- negative predictive value o
- positive predictive value o
- negative likelihood ratio o
- positive likelihood ratio o
- . Top 2 Embryo analysis
- Treatment level analysis
# Results
The study demonstrated that the adjunct use of KIDScore D3 improved the selection of embryos for transfer compared with morphology alone. In the primary endpoint, it was shown that the adjunct use of KIDScore D3 was informative for blastocyst outcome.
The study showed that both the KIDScore D3 and the Eeva system were predictive of blastocyst outcome, with odds ratios greater than one for both systems, as described in the figure below:
| | KIDScore D3 | Eeva |
|------------|-------------|-------------|
| Odds ratio | 4.13 | 2.57 |
| 95% CI | 3.48 - 4.9 | 1.88 - 3.51 |
| P value | <0.0001 | <0.0001 |
# 10. Conclusion
The subject and predicate devices have the same intended use and fundamental technological characteristics. The differences in technological characteristics between subject and predicate devices do not raise different questions of safety and effectiveness. The performance data and clinical testing demonstrate that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.