Sperm Maintenance Medium with Glycerol is intended for use in assisted reproductive procedures that involve the manipulation of gametes. Specifically, Sperm Maintenance Medium with Glycerol is intended to be used as a cryopreservation medium for human sperm.
Device Story
Sperm Maintenance Medium with Glycerol is a synthetic, defined medium used for cryopreservation of human sperm; protects sperm cells from damage during freezing, storage, and thawing. Composition includes salts, physiologically compatible substances, glycerol (cryoprotectant), and human serum albumin (protein supplement). Used in clinical settings for sperm banking prior to assisted reproductive procedures. Product is analyzed via sperm cryopreservation and motility recovery assay, endotoxin testing, and sterility testing prior to market release. Healthcare providers use the medium to prepare sperm for long-term storage; ensures viability for future use in fertility treatments.
Clinical Evidence
No clinical trials were conducted. Evidence relies on historical information from professional literature, a history of satisfactory clinical use, and bench testing. Bench testing includes a sperm cryopreservation and motility recovery assay, endotoxin testing, and sterility testing performed as a condition of lot release to ensure functionality and absence of toxic components.
Technological Characteristics
Synthetic, defined medium containing salts, glycerol (cryoprotectant), and human serum albumin (protein supplement). Supplied in 100 mL volumes. Functionality is verified via sperm cryopreservation and motility recovery assay, endotoxin testing, and sterility testing.
Indications for Use
Indicated for the cryopreservation of human sperm in assisted reproductive procedures involving gamete manipulation.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
Notice of Final Rule, 63 FR 48428, Docket number 97N-0335
Submission Summary (Full Text)
{0}------------------------------------------------
Irvine Scientific
JIII - 8 1999
K991337
Image /page/0/Picture/4 description: The image is a black and white graphic of an abstract design. The design features a combination of straight lines and geometric shapes, creating an open, modern look. The contrast between the black and white areas is sharp, highlighting the structure of the design. The image is simple, yet visually striking.
IRVINE SCIENTIFIC
# IRVINE SCIENTIFIC 510(k) SUMMARY AS REQUIRED BY SECTION 807.92(c)
Submitted by:
Irvine Scientific Sales Co., Inc. 2511 Daimler Street Santa Ana, CA 92705-5588
Telephone: (800) 437-5706 Facsimile: (949) 261-6522
Contact: Roberta L. Johnson
Date Submitted: April 16, 1999
#### Device Identification:
| Trade Name: | Sperm Maintenance Medium with Glycerol |
|----------------------|----------------------------------------|
| Common Name: | Sperm Cryopreservation Medium |
| Classification Name: | Reproductive Media (21 CFR, 886.6180) |
#### Predicate Device:
Notice of Final Rule, 63 FR 48428, Docket number 97N-0335
### Description:
Sperm Maintenance Medium with Glycerol is a synthetic, defined medium composed of a mixture of salts and other physiologically compatible substances, with glycerol as a cryoprotectant and human serum albumin as a protein supplement. It is supplied in fill volumes of 100 mL.
Sperm Maintenance Medium with Glycerol
Page 24 of 29
{1}------------------------------------------------
#### Intended Use:
Sperm Maintenance Medium with Glycerol is intended for the cryopreservation of human sperm prior to assisted reproductive procedures.
#### Technological Characteristics:
Sperm Maintenance Medium with Glycerol is used to protect human sperm during cryopreservation for future assisted reproductive procedures. Sperm is routinely saved, or "banked" for a variety of reasons. Products with ingredients such as glycerol protect the cells from damage during the initial freezing process, the storage and the thawing prior to use.
#### Performance Data:
Sperm Maintenance Medium with Glycerol will be analyzed by a sperm cryopreservation and motility recovery assay prior to release to market. This assay assures that the product is both functional for its intended use, the cryopreservation of human sperm, and that no toxic components are present in the formulation. Sperm Maintenance Medium has been used in a variety of clinical settings, for its intended use, for a number of years. In that time, the product has become one of the standard media used for the cryopreservation of human sperm.
### Additional Information:
The sperm cryopreservation and motility recovery assay will be performed as a condition of release for this product, as well as endotoxin and sterility testing. Results of all release assays performed will be reported on a lot-specific certificate of analysis, and will be indicated on the labeling.
### Conclusion:
The conclusion from performance testing, as well as a review of the historical information contained in professional literature and a history of satisfactory use,
{2}------------------------------------------------
shows that Sperm Maintenance Medium with Glycerol is suitable for its intended use, and meet the criteria outlined in the Notice of Final Rule, 63 FR 48428, Docket number 97N-0335.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird in flight, with three curved lines representing the wings and body. The bird is positioned to the right of the text, which is arranged in a circular pattern around the image. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 8 1999
Ms. Roberta L. Johnson Manager, Regulatory Affairs Irvine Scientific, Inc. 2511 Daimler Street Santa Ana, CA 92705
Re: K991337 Sperm Maintenance Medium with Glycerol Dated: April 16, 1999 Received: April 19, 1999 Requlatory Class: Il 21 CFR §884.6180/Procode: 85 MQL
Dear Ms. Johnson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compilance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Irvine Scientific
#### (page 1 of 1) INDICATIONS FOR USE STATEMENT
K991337 510(K) Number:_
Device Name: Sperm Maintenance Medium with Glycerol
Indications for Use:
Sperm Maintenance Medium with Glycerol is intended for use in assisted reproductive procedures that involve the manipulation of gametes. Specifically, Sperm Maintenance Medium with Glycerol is intended to be used as a cryopreservation medium for human sperm.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Dev
510(k) Number K941321
Prescription Use
(Per 21 CFR 801.109)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.