Sperm Freeze is intended for the cryopreservation of human semen samples prior to use in assisted reproductive technology procedures. Sperm Fridge is intended to protect human semen samples during refrigerated storage prior to use in assisted reproductive technology procedures.
Device Story
Sperm Freeze and Sperm Fridge are preservation media for human semen samples used in ART. Sperm Freeze, a Tris-buffered solution containing glycerol and trehalose, facilitates long-term cryopreservation. Sperm Fridge, a TES and Tris-buffered solution containing dextrose, facilitates short-term refrigerated storage. Both products are intended for laboratory use by trained professionals. The media protect sperm viability during storage, mitigating the negative effects of freezing or refrigeration on sperm motility and survival. Healthcare providers use these media to prepare samples for IVF or other ART treatments, potentially improving patient outcomes by maintaining sample quality prior to clinical use.
Clinical Evidence
Bench testing only. Performance testing included appearance, osmolality, pH (USP <791>), endotoxin (USP <85>), sterility (USP <71>), and sperm survival rates (cryo-survival and refrigeration survival). Shelf-life testing was conducted at Time 0 and beyond 12 months to verify specifications.
Technological Characteristics
Buffered solutions (Tris/TES). Sperm Freeze contains glycerol, trehalose, and recombinant human albumin. Sperm Fridge contains dextrose and recombinant human albumin. Sterilization via aseptic filtration. pH 7.2–7.6. Endotoxin ≤ 0.25 EU/mL. Storage at -20°C.
Indications for Use
Indicated for human semen sample preservation (cryopreservation or refrigerated storage) prior to assisted reproductive technology (ART) procedures. Prescription use only.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
Freezing Medium TEST Yolk Buffer (TYB) with Glycerol, Refrigeration Medium (K991421)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 17, 2020
Kitazato Corporation % Michael A. Siano, MA Regulatory Affairs Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, TX 78746
K192845 Re:
> Trade/Device Name: Sperm Freeze, Sperm Fridge Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MQL Dated: July 6, 2020 Received: July 9, 2020
Dear Michael A. Siano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K192845
Device Name Sperm Freeze Sperm Fridge
Indications for Use (Describe)
Sperm Freeze is intended for the cryopreservation of human semen samples prior to use in assisted reproductive technology procedures.
Sperm Fridge is intended to protect human semen samples during refrigerated storage prior to use in assisted reproductive technology procedures.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
---
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## K192845 - Sperm Freeze, Sperm Fridge
### 1. Submitter Information
| Applicant: | Kitazato Corporation |
|------------|------------------------------------------------------------|
| Contact: | Mr. Futoshi Inoue<br>President and Representative Director |
| Address: | 81 Nakajima<br>Fuji Shizuoka<br>JAPAN 416-0907 |
| Phone: | +81 545 66 2202 |
#### 2. Correspondent Information
| Contact: | Michael A. Siano<br>Regulatory Affairs Consultant |
|----------|---------------------------------------------------------------------------------------------------------|
| Address: | Emergo Global Consulting, LLC<br>2500 Bee Cave Road<br>Building 1, Suite 300<br>Austin, TX 78746<br>USA |
| Phone: | (512) 327-9997 |
- 3. Date Prepared: August 17, 2020
#### Device Information 4.
| Device Name: | Sperm Freeze, Sperm Fridge |
|--------------------|--------------------------------------|
| Common Name: | Cryopreservation/refrigeration media |
| Regulation Number: | 21 CFR 884.6180 |
| Regulation Name: | Reproductive media and supplements |
| Product Code: | MQL (Media, Reproductive) |
| Regulatory Class: | Class II |
#### 5. Predicate Device(s)
| Device Name: | Freezing Medium TEST Yolk Buffer (TYB) with Glycerol, Refrigeration Medium |
|----------------|----------------------------------------------------------------------------|
| | TEST Yolk Buffer (TYB) |
| 510(k) Number: | K991421 |
| Manufacturer: | Irvine Scientific |
This predicate has not been subject to a design-related recall.
{4}------------------------------------------------
#### 6. Device Description
Sperm Freeze and Sperm Fridge are preservation media intended for preserving sperm (short-term and long-term) for in vitro fertilization (IVF) and assisted reproduction technologies (ART). There are two methods for sperm preservation; freezing and refrigeration. Freezing is necessary for long-term sperm preservation, but may have disadvantages such as reduced survival rate and/or sperm motility after thawing. Refrigeration makes it possible to preserve sperm for several days without freezing and may be preferable for short-term preservation. Both products are intended to be used exclusively in a laboratory environment by trained professionals.
Sperm Freeze is a is a Tris-buffered solution containing the cryoprotectants glycerol and trehalose that is used for long-term sperm cryopreservation procedures. Sperm Fridge is a buffered solution (TES and Tris) that includes a nutrient source for sperm (dextrose) that is used for short-term sperm preservation procedures. Sperm Freeze and Sperm Fridge are offered in two volumes (0.5 mL and 10 mL, and 0.5 mL is sold in a pack of 5).
Device specifications for Sperm Freeze and Sperm Fridge are detailed in Tables 1 and 2 below.
#### 7. Indications for Use
Sperm Freeze is intended for the cryopreservation of human semen samples prior to use in assisted reproductive technology procedures.
Sperm Fridge is intended to protect human semen samples during refrigerated storage prior to use in assisted reproductive technology procedures.
#### 8. Substantial Equivalence Discussion
A detailed comparison of the intended use and technological features of the subject and predicate device are described in the tables below:
#### Table 1: Comparison of Characteristics, Sperm Fridge
| Attribute | K191485<br>Subject Device:<br>Sperm Fridge | K991421<br>Predicate Device:<br>Refrigeration Medium- TEST Yolk<br>Buffer (TYB) | Comparison |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Kitazato | Irvine Scientific | Different |
| Product Code | MQL | MQL | Same |
| Indications for use | Sperm Fridge is intended to protect<br>human semen samples during<br>refrigerated storage prior to use in<br>assisted reproductive technology<br>procedures. | Refrigeration Medium -Test Yolk Buffer<br>and Freezing Medium - Test Yolk<br>Buffer with Glycerol are intended for<br>use in assisted reproductive<br>technology procedures that involve the<br>manipulation and storage of semen<br>samples prior to use in in vitro<br>fertilization and other similar<br>treatments. Refrigeration Medium -<br>Test Yolk Buffer is specifically designed<br>for to protect semen samples during<br>refrigerated storage, while Freezing | Different |
{5}------------------------------------------------
| | | Medium - Test Yolk Buffer with<br>Glycerol is intended to be used as a<br>cryopreservative of semen samples. | |
|---------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-----------|
| Rx/OTC | Rx | Rx | Same |
| Ingredients | TES; Tris; Dextrose; Gentamicin<br>Sulfate; Recombinant Human<br>Albumin | TES; Tris; Dextrose; Egg Yolk; Penicillin-<br>G; Streptomycin Sulfate | Different |
| pH | 7.2–7.6 | 7.2 ± 0.2 | Different |
| Endotoxin | ≤ 0.25 EU/mL | ≤ 3 EU/mL | Different |
| Sterility | No growth | No growth | Same |
| Sperm Fridge-<br>survival | ≥ 80% (45 min) | N/A | Different |
| Storage Conditions | -20°C | -10°C | Different |
| Shelf-life | 1 year | 2 years | Different |
| Sterilization | Aseptic Filtration | Aseptic Filtration | Same |
### Table 2: Comparison of Characteristics, Sperm Freeze
| Attribute | K191485<br>Subject Device:<br>Sperm Freeze | K991421<br>Predicate Device:<br>Freezing Medium-TEST Yolk<br>Buffer(TYB) with Glycerol | Comparison |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Kitazato | Irvine Scientific | Different |
| Product Code | MQL | MQL | Same |
| Indications for Use | Sperm Freeze is intended for the<br>cryopreservation of human semen<br>samples prior to use in assisted<br>reproductive technology<br>procedures. | Refrigeration Medium -Test Yolk Buffer<br>and Freezing Medium - Test Yolk<br>Buffer with Glycerol are intended for<br>use in assisted reproductive<br>technology procedures that involve the<br>manipulation and storage of semen<br>samples prior to use in in vitro<br>fertilization and other similar<br>treatments. Refrigeration Medium -<br>Test Yolk Buffer is specifically designed<br>for to protect semen samples during<br>refrigerated storage, while Freezing<br>Medium - Test Yolk Buffer with<br>Glycerol is intended to be used as a<br>cryopreservative of semen samples. | Different |
| Rx/OTC | Rx | Rx | Same |
| Ingredients | Water; Calcium DL-Lactate;<br>Disodium Hydrogenphosphate;<br>Gentamicin; Glycerol; Glycine;<br>Lauryl Alcohol; Magnesium<br>Sulfate; Potassium Chloride;<br>Recombinant Human Albumin; | Water; TES; Tris; Sodium Citrate;<br>Fructose; Glycerol; Egg Yolk;<br>Penicillin-G; Streptomycin Sulfate | Different |
{6}------------------------------------------------
| | Sodium Bicarbonate; Sodium<br>Chloride; Sodium Phosphate;<br>Tocopherol; Trehalose; Tris | | |
|-------------------------|------------------------------------------------------------------------------------------|--------------------|-----------|
| pH | 7.2–7.6 | 7.0-7.4 | Different |
| Endotoxin | $≤$ 0.25 EU/mL | $≤$ 3 EU/mL | Different |
| Sterility | No growth | No growth | Same |
| Sperm Cryo-<br>survival | $≥$ 80% (45 min) | N/A | Different |
| Storage Conditions | -20 °C | -10 °C | Different |
| Shelf-Life | 1 year | 2 years | Different |
| Sterilization | Aseptic Filtration | Aseptic Filtration | Same |
The subject and predicate device have similar indications for use statements and have the same intended use - to cryopreserve or refrigerate human semen samples prior to use in assisted reproductive technology procedures. The subject and predicate device have different technological characteristics, including differences in formulation, pH, endotoxin, storage conditions, and shelf-life. These differences do not raise different questions of safety and effectiveness as compared to the predicate device.
#### 9. Summary of Non-Clinical Performance Testing
To demonstrate safety and effectiveness of Sperm Freeze and Sperm Fridge and to show substantial equivalence to the predicate device, Kitazato completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the devices are met.
The device passed all testing in accordance with internal requirements, national standards, and international standards shown below:
- Appearance
- Endotoxin per USP <85>
- Osmolality
- pH per USP <791>
- Sperm cryo-survival
- . Sperm refrigeration survival
- Sterility per USP <71> ●
Shelf-life performance testing was conducted at Time 0 and beyond the shelf-life (>12 months) to ensure that the product specifications listed above were met.
#### 10. Conclusions
The results of the performance testing described above demonstrate that Sperm Freeze and Sperm Fridge are as safe and effective as the predicate device and supports a determination of substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.