SAGE 1-Step GM-CSF with HSA, SAGE 1-Step GM-CSF with SPS
Applicant
Origio A/S
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Oct 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
The product is intended for in vitro culture of human embryos following fertilization until Day 5/6 of development. The media can also be used for embryo transfer.
Device Story
SAGE 1-Step™ GM-CSF (with HSA or SPS) is an embryo culture medium used in assisted reproduction. It contains physiological salts, energy substrates, amino acids, antibiotics, protein supplements, and 2 ng/ml of granulocyte-macrophage colony-stimulating factor (GM-CSF). The medium supports embryo development in vitro from fertilization to the blastocyst stage (Day 5/6). It is supplied in 3 ml glass vials, aseptically filtered, and intended for use by embryologists in clinical laboratory settings. The addition of GM-CSF is the primary technological difference from the predicate. The device is used to maintain optimal environmental conditions for embryo growth, potentially improving developmental outcomes. Healthcare providers use the medium to facilitate the culture process prior to embryo transfer.
Clinical Evidence
Bench testing only. Performance validated via pH (7.2-7.4), osmolality (257-273 mOsm/kg), endotoxin (<0.15 EU/ml), and sterility (USP <71>). Mouse embryo assay (MEA) showed ≥80% blastocyst development at 96h. GM-CSF concentration (ELISA) and potency (TF-1 cell assay) met 80-125% recovery/potency specifications. Real-time shelf-life testing confirmed specifications are met throughout the 16-week shelf-life.
Technological Characteristics
Embryo culture medium containing buffering agents, physiological salts, energy substrates, amino acids, antibiotics, protein supplements (HSA or SPS), and 2 ng/ml GM-CSF. Aseptically filtered (SAL 10^-3). Supplied in 3 ml glass vials. Shelf-life 16 weeks. Storage 2-8°C. Testing per USP <71> (sterility), USP <85> (endotoxin), and ISO 13408-2-2003 (aseptic processing).
Indications for Use
Indicated for in vitro culture of human embryos following fertilization until Day 5/6 of development and for embryo transfer procedures.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
SAGE 1-Step™ with HSA and SAGE 1-Step™ with SPS (K133707)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 13, 2017
ORIGIO a/s Tove Kjaer Director, Corporate Regulatory Affairs Knardrupvej 2 Malov, 2760 Denmark
Re: K172197 Trade/Device Name: SAGE 1-Step™ GM-CSF with HSA, SAGE 1-Step™ GM-CSF with SPS Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: Class II Product Code: MOL Dated: July 24, 2017 Received: July 26, 2017
Dear Tove Kjaer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K172197
Device Name
SAGE 1-Step™ GM-CSF with HSA SAGE 1-Step™ GM-CSF with SPS
Indications for Use (Describe)
The product is intended for in vitro culture of human embryos following fertilization until Day 5/6 of development. The media can also be used for embryo transfer
Type of Use (Select one or both, as applicable)
■ Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "origio" in a stylized, lowercase font. The letters are a deep purple color. The 'g' is designed with a unique, almost circular shape, and the overall design has a modern and minimalist aesthetic.
# a CooperSurgical Company
K172197 Page 1 of 4
## 510(k) SUMMARY K172197
- Submitter Information L
Submitter:
ORIGIO a/s Knardrupvej 2, 2760 Måløv, Denmark
Contact person:
Tove Kjær Director Corporate Regulatory Affairs ORIGIO a/s Phone: +45 4679 0220 Fax: +45 4679 0300 Email: tkjaer@origio.com
- II. Date prepared: Oct. 13 2017
## III. Device Information
| Device Name: | SAGE 1-Step™ GM-CSF with HSA<br>SAGE 1-Step™ GM-CSF with SPS |
|------------------------|--------------------------------------------------------------|
| Common name: | Embryo Culture Media |
| Classification name: | Reproductive media and supplements |
| Classification Number: | 21 CFR 884.6180 |
| Product Code: | MQL (Media, Reproductive) |
| Regulatory Class: | II |
#### IV. Predicate Device
ORIGIO a/s - SAGE 1-Step™ with HSA and SAGE 1-Step™ with SPS (K133707)
These predicate devices have not been subject to any design-related recalls.
# V. Indication for use
The product is intended for in vitro culture of human embryos following fertilization until Day 5/6 of development. The media can also be used for embryo transfer.
{4}------------------------------------------------
# VI. Device Description
SAGE 1-Step™ GM-CSF with HSA and SAGE 1-Step™ GM-CSF with SPS are assisted reproduction embryo culture media products consisting agents, physiological salts, energy substrates, amino acids, antibiotics, protein supplements (HSA or SPS), and granulocyte-macrophage colony-stimulating factor (GM-CSF, 2 ng/ml).
These devices are aseptically filtered (sterility assurance level of 10 %), and supplied in 3 ml glass vials. They are tested for pH, osmolality, embryotoxicity, endotoxin, sterility, and GM-CSF concentration and potency before lot release. These devices have a shelf-life of 16 weeks, and can be used for up to seven days after vial opening when stored at 2-8°C.
## VII. Comparison of Intended Use and Technological Characteristics of Subject and Predicate Devices
| Parameter | Subject device-<br>SAGE 1-StepTM<br>GM-CSF (with HAS<br>or SPS) (K172197) | Predicate Device-<br>SAGE 1-StepTM (with<br>HAS or SPS)<br>(K133707) | Comparison |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The product is<br>intended for in vitro<br>culture of human<br>embryos following<br>fertilization until Day<br>5/6 of development.<br>The media can also<br>be used for embryo<br>transfer. | This product is<br>intended for in vitro<br>fertilization and<br>culture of human<br>gametes and<br>embryos from<br>fertilization until Day<br>5/6 of development.<br>The medium can<br>also be used for<br>embryo transfer. | Different- The<br>subject and<br>predicate devices<br>are both indicated<br>for culturing embryos<br>following fertilization<br>until the blastocyst<br>stage (Day 5/6) of<br>development, and<br>for use in embryo<br>transfer procedures.<br>The predicate is also<br>indicated for use<br>during fertilization,<br>which is different<br>than the subject<br>device. This<br>difference does not<br>represent a new<br>intended use, but<br>rather a more limited<br>use of the subject<br>device. |
| Formulation | Buffering agents,<br>physiological salts | Buffering agents,<br>physiological salts | Different -Base<br>formulations of the |
{5}------------------------------------------------
| | energy substrates,<br>amino acids,<br>antibiotics, protein<br>supplements (HSA<br>or SPS), and GM-<br>CSF | energy substrates,<br>amino acids,<br>antibiotics, and<br>proteins<br>supplements (HSA<br>or SPS) | subject and<br>predicate devices<br>are similar; however,<br>the subject device<br>also includes GM-<br>CSF. This<br>difference does not<br>raise different<br>questions of safety<br>and effectiveness<br>(e.g.,<br>biocompatibility,<br>embryotoxicity, etc.). |
|-------------------|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| pH | 7.2-7.4 | 7.2-7.4 | Same |
| Osmolality | 257-273 | 257-273 | Same |
| MEA | 1-Cell MEA: ≥80%<br>blastocysts at 96h | 1-Cell MEA: ≥80%<br>blastocysts at 96h | Same |
| Endotoxin (EU/ml) | <0.15 | <0.15 | Same |
| Sterility | Aeptic Filteration<br>SAL 10⁻³, no growth | Aeptic Filteration<br>SAL 10⁻³, no growth | Same |
| Shelf-Life | 16 weeks | 26 weeks | Different - The<br>shelf-life of the<br>subject and<br>predicate devices<br>are different. This<br>difference does not<br>raise different<br>questions of safety<br>and effectiveness. |
As noted in the table above, the devices have the same intended use and are technologically comparable. Differences in technological characteristics noted above do not raise different questions of safety or effectiveness.
# VIII. Summary of Non-Clinical Performance Testing
The following studies have been performed to support substantial equivalence to the predicate devices:
- pH 7.2-7.4 ●
- . Osmolality - 257-273 mOsm/kg
- Aseptic processing validation testing that meet the requirements in ISO 13408-2-2003
{6}------------------------------------------------
- Sterility testing per USP <71> No microbial growth ●
- . Endotoxin testing per USP <85> - <0.15 EU/ml (LAL)
- Mouse embryo assay (MEA) 1-Cell MEA: ≥80% blastocysts at 96h ●
One-cell mouse embryos were exposed to subject devices and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the blastocyst stage at 96 hours were assessed in comparison with the control group.
- GM-CSF concentration (by ELISA) 80-120% recovery .
- GM-CSF potency (TF-1 cell assay) - 80-125% potency
- Shelf-life testing (real-time) was conducted to ensure that the following product specifications are met at time zero and the end of shelf-life.
- pH O
- Osmolality O
- 1-Cell MEA O
- Endotoxin o
- Sterility O
- GM-CSF concentration O
- o GM-CSF potency
#### IX. Conclusion
The results of the performance testing conducted on the subject devices demonstrates that they are as safe and effective as the predicate devices and support substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.